The Digi-One System is indicated for non-invasive evaluation of intracranial and extracranial vascular flow velocity irregularities in adults and in children. It is not intended for fetal use. It is not intended for neonatal use.
Device Story
Digi-One System is a portable digital Transcranial Doppler (TCD) system; connects to external Windows-based PC. Inputs: ultrasound signals from 1MHz, 2MHz, 4MHz, or 8MHz transducers. Transforms inputs via Doppler ultrasound processing to measure blood flow velocity in cerebral arteries and detect High Intensity Transient Signals (HITS). Outputs: M-mode display, spectral data, and flow velocity metrics. Used in clinical settings by healthcare providers. Output allows clinicians to diagnose circulation irregularities. Benefits: non-invasive assessment of cerebral vascular flow.
Clinical Evidence
Bench testing only. Testing included software validation per FDA guidelines, validation of output energies, parameter calculation accuracy, electrical/mechanical safety (IEC 60601-1), and electromagnetic compatibility (IEC 60601-1-2).
Indicated for non-invasive evaluation of intracranial and extracranial vascular flow velocity irregularities in adults and children. Not for fetal or neonatal use.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
K213009 — Transcranial Doppler Ultrasound System · Shenzhen Delica Medical Equipment Co., Ltd. · Jul 8, 2022
K191023 — Dolphin/IQ, Dolphin/4D and Dolphin/MAX · Viasonix , Ltd. · May 16, 2019
K170859 — Dolphin/IQ and Dolphin/4D · Viasonix , Ltd. · Nov 14, 2017
K173801 — Transcranial Doppler Ultrasound System · Shenzhen Delica Medical Equipment Co., Ltd. · Jul 11, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. The bottom half of the circle contains a stylized image of three human profiles facing to the right. The profiles are connected by a flowing line.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 23, 2015
Rimed Ltd. % Ms. Ahava Stein Regulatory Consultant 20 Hata'as St. (POB 124) Kfar Saba 44425 ISRAEL
Re: K143476
Trade/Device Name: Digi-One System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, ITX Dated: May 22, 2015 Received: May 28, 2015
Dear Ms. Stein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A. Ochs
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143476
Device Name Digi-One System
## Indications for Use (Describe)
The Digi-One System is indicated for non-invasive evaluation of intracranial vascular flow velocity irregularities in adult and in children. It is not intended for fetal use. It is not intended for neonatal use.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
### Indications for Use Form Fill out one form for each ultrasound system and each transducer. Digi-One System:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical<br>Application | A | B | C | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | X | X | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intra-operative<br>(Specify) | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Neonatal | | | | | | | | | | |
| Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | X | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Luminal | | | | | | | | | | |
| Peripheral | | | | | | | | | | |
| Vascular | | | | X | X | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-Skeletal<br>Conventional | | | | | | | | | | |
| Muscolo-Skeletal<br>Superficial | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | |
Mode of Oneratio
{4}------------------------------------------------
## Indications for Use Form Fill out one form for each ultrasound system and each transducer. 1 MHZ PW HAND-HELD TRANSDUCER
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | | | | | | | | | | |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|--------------------|
| Clinical<br>Application | A | B | C | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intra-operative<br>(Specify) | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | X | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Luminal | | | | | | | | | | |
| Peripheral<br>Vascular | | | | X | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-Skeletal<br>Conventional | | | | | | | | | | |
| Muscolo-Skeletal<br>Superficial | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | |
{5}------------------------------------------------
## Indications for Use Form Fill out one form for each ultrasound system and each transducer. 2 MHZ PW HAND-HELD TRANSDUCER
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | | | | | | Mode of Operation | | | | |
|----------------------------------|---|---|---|-----|-----|-------------------|---------------------------------|------------------------------|-----------------------|--------------------|
| Clinical<br>Application | A | B | C | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | X | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intra-operative<br>(Specify) | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Neonatal | | | | | | | | | | |
| Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | X | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Luminal | | | | | | | | | | |
| Peripheral<br>Vascular | | | | X | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-Skeletal<br>Conventional | | | | | | | | | | |
| Muscolo-Skeletal<br>Superficial | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | |
{6}------------------------------------------------
## Indications for Use Form Fill out one form for each ultrasound system and each transducer. 4 MHZ TRANSDUCER
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical<br>Application | A | B | C | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intra-operative<br>(Specify) | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Neonatal | | | | | | | | | | |
| Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Luminal | | | | | | | | | | |
| Peripheral<br>Vascular | | | | X | X | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-Skeletal<br>Conventional | | | | | | | | | | |
| Muscolo-Skeletal<br>Superficial | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | |
Mode of Oneration
{7}------------------------------------------------
## Indications for Use Form Fill out one form for each ultrasound system and each transducer. 8 MHZ TRANSDUCER
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | | | | | | | | | | |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|--------------------|
| Clinical<br>Application | A | B | C | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intra-operative<br>(Specify) | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Neonatal | | | | | | | | | | |
| Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Luminal | | | | | | | | | | |
| Peripheral<br>Vascular | | | | X | X | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-Skeletal<br>Conventional | | | | | | | | | | |
| Muscolo-Skeletal<br>Superficial | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | |
Mode of Operation
{8}------------------------------------------------
# SUMMARY OF SAFETY AND EFFECTIVENESS
K143476 (Premarket Notification [510(k)] Number)
### 1. Submitter Information
| <b>Manufacturer Name and Address</b> | Rimed Ltd. |
|--------------------------------------|-----------------------------------------------|
| | 25 Hacharoshet St., |
| | PO Box 2402 |
| | Industrial Park Raanana 4365613, |
| | Israel |
| <b>Official Correspondent</b> | Ahava Stein |
| | A. Stein - Regulatory Affairs Consulting Ltd. |
| | 20 Hata'as St. (Beit Hapaamon, Suite 102) |
| | Kfar Saba 44425, |
| | Israel |
2. **Date Prepared:** May 18, 2015
3. **Device Name** Digi-One System
**Proprietary Name:** Digi-One System
Common Name: Transcranial Doppler (TCD) Device
FDA Classification 21 CFR 892.1550; Ultrasonic Pulsed Doppler Imaging System Name:
FDA Classification: Class II, Product Code IYN; ITX
#### 4. Predicate Devices
The Digi-One System is substantially equivalent to the following devices:
| Manufacturer | Device | 510(k) | Date Cleared |
|--------------|------------------|---------|-----------------|
| Rimed Ltd. | Digi-Lite System | K062578 | August 23, 2006 |
### ട. Device Description
The Digi-One System is a small, lightweight, portable digital Transcranial Doppler (TCD) system with an advanced M-mode display which can be connected to any external Windows based PC.
{9}------------------------------------------------
It measures the blood flow velocity in the main cerebral arteries non-invasively for circulation diagnosis and HITS (High Intensity Transient Signals) detection using the same Rimed TCD software application as the predicate device.
The Digi-One System is a modification of the Digi-Lite System. The hardware was modified using current electronic components, and the software was modified to allow installation on a dedicated PC (such as a laptop device). Otherwise, the functionality of the system remains identical to that of the Digi-Lite system in all aspects, including acoustic power outputs, clinical parameter measurement accuracy, examination modes, available transducers, software menus and commands (with some minor user interface modifications). As the specifications remain identical, the safety and efficacy of the device are not affected when used as labeled. Therefore, the Digi-One system remains substantially equivalent to the predicate, un-modified Digi-Lite System.
| Technological<br>Characteristic | Digi-One System<br>Rimed Ltd. | Digi-Lite System<br>Rimed Ltd.<br>(K062578) |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code,<br>Class | IYN, ITX<br>Class II | IYN, ITX<br>Class II |
| Indications for<br>Use | The Digi-One System is<br>indicated for non-invasive<br>evaluation of intracranial and<br>extracranial vascular flow<br>velocity irregularities in adults<br>and in children. It is not<br>intended for fetal use. It is not<br>intended for neonatal use. | The Digi-Lite System is<br>indicated for non-invasive<br>evaluation of intracranial and<br>extracranial vascular flow<br>velocity irregularities in adults<br>and in children. It is not<br>intended for fetal use. It is not<br>intended for neonatal use. |
| Energy Used /<br>Delivered | Ultrasound energy | Ultrasound energy |
| Design:<br>(Modification<br>from predicate) | The Digi-One System consists<br>of a small unit which connects<br>to a PC and to which probes are<br>connected. | The Digi-Lite System consists<br>of a unit with a display to<br>which probes are connected. |
| - Mechanism of<br>Action | Doppler Ultrasound, with the<br>following modes:<br>Unilateral<br>Bilateral<br>Multichannel<br>Monitoring<br>Multidepth (up to 8 spectrums<br>at various depths)<br>M-Mode (64 gates) | Doppler Ultrasound, with the<br>following modes:<br>Unilateral<br>Bilateral<br>Multifrequency<br>Multichannel<br>Monitoring<br>Multidepth (up to 8 spectrums<br>at various depths)<br>M-Mode (64 gates) |
Predicate Comparison Table:
{10}------------------------------------------------
| Technological<br>Characteristic | Digi-One System<br>Rimed Ltd. | Digi-Lite System<br>Rimed Ltd.<br>(K062578) |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| - Accessories | Probe types:<br>- 1MHz PW, 16mm<br>- 2MHz PW, 14mm<br>- 4MHz PW/CW, 8mm<br>- 8MHz PW/CW, 5mm<br>Remote control<br>Monitoring probe holder<br>Color printer Footswitch | Probe types:<br>- 1MHz PW, 16mm<br>- 2MHz PW, 14mm<br>- 4MHz PW/CW, 8mm<br>- 8MHz PW/CW, 5mm<br>Remote control<br>Monitoring probe holder<br>Color printer Footswitch<br>CD ROM backup<br>DAT Tape |
| Performance | | |
| Sample volume | (for 1 MHz) (for 2 MHz)<br>15-92 mm 15-146 mm | (for 1 MHz) (for 2 MHz)<br>15-92 mm 15-146 mm |
| Depth | | |
| Power (%) | 0-100% at 7% steps | 0-100% at 7% steps |
| Acoustic output | 1MHz PW Ispta=34.15mW/cm2<br>2MHz PW Ispta=181.3mW/cm2<br>4MHz PW Ispta=232.2mW/cm2<br>4MHz CW Ispta=183mW/cm2<br>8MHz PW Ispta=149.8mW/cm2<br>8MHz CW Ispta=414mW/cm2 | 1MHz PW Ispta=34.15mW/cm2<br>2MHz PW Ispta=181.3mW/cm2<br>4MHz PW Ispta=232.2mW/cm2<br>4MHz CW Ispta=183mW/cm2<br>8MHz PW Ispta=149.8mW/cm2<br>8MHz CW Ispta=414mW/cm2 |
### Indications for Use 6.
The Digi-One System is indicated for non-invasive evaluation of intracranial and extracranial vascular flow velocity irregularities in adults and in children. It is not intended for fetal use. It is not intended for neonatal use.
### 7. Performance Standards
There are no performance standards under the Federal Food, Drug and Cosmetic Act, for the Digi-One System.
### 8. Performance Testing
The following Software and Performance tests were performed on the Digi-One System:
- Software Validation according to FDA Guidelines (tested modified ● software)
- Bench testing to validate output energies and parameter calculation accuracies
- Electrical and Mechanical Safety Standard (IEC 60601-1)
- Electromagnetic Compatibility Standard (IEC60601-1-2)
### 9. Technological Characteristics Compared to Predicate Device
The original Digi-Lite device underwent modification to electronics and software. The safety and effectiveness questions that were raised by these changes were
{11}------------------------------------------------
equivalence of acoustic power outputs, equivalence of measurement accuracy, and safe use under low-power modes. Performance testing of the modified device included measurements of output power and measurement accuracy. The testing results demonstrated that the modified device is substantially equivalent to the predicate, unmodified device. The software functions and testing modes did not change, except for some User Interface improvements; the system defaults for low-power modes were reviewed for equivalence to the predicate device, and were demonstrated to remain identical. Software testing was performed on the modified software. The technological characteristics, e.g., overall design, materials, mechanism of action, mode of operation, performance characteristics, etc., and the indications for use of the modified Digi-One System are substantially equivalent to the predicate device cited above. Based on the results of performance testing, the Digi-One device is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.