ACUSON S1000/ S2000/ S3000 Diagnostic Ultrasound Systems

K142876 · Siemens Medical Solutions USA, Inc. · IYN · Oct 23, 2014 · Radiology

Device Facts

Record IDK142876
Device NameACUSON S1000/ S2000/ S3000 Diagnostic Ultrasound Systems
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeIYN · Radiology
Decision DateOct 23, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric, 3rd-Party Reviewed

Intended Use

The ultrasound imaging systems are intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides the ability to measure anatomical structures {fetal, abdominal, intraoperative, pediatric, small organ, neonatal cephalic, adult cephalic, cardiac, transesophageal, transrectal, transvaginal, peripheral vessel, musculo-skeletal (conventional), musculo-skeletal (superficial) and neonatal cardiac} and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging". The Acuson Acunav Ultrasound Catheter is intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients.

Device Story

Diagnostic ultrasound systems (Acuson S1000, S2000, S3000) acquire harmonic ultrasound echo data via various transducers; display in B-Mode, PW/CW Doppler, Color Doppler, Amplitude Doppler, or 3D/4D imaging. Used in clinical settings by physicians for anatomical measurement and diagnostic visualization. Includes specialized software packages like Arterial Health Package (AHP) for Intima Media Thickness measurement and various auto-measurement/segmentation tools (eSie Calc, syngo Auto OB). Output provides clinicians with visual data and quantitative measurements to support diagnosis and patient communication. Benefits include non-invasive visualization of internal anatomy, cardiac/vascular structures, and device placement within the heart.

Clinical Evidence

No clinical data required; substantial equivalence established through bench testing, acoustic output measurements, and adherence to recognized safety standards (IEC 60601-2-37, ISO 10993-1, etc.).

Technological Characteristics

Mobile, software-controlled diagnostic ultrasound systems. Supports B-mode, M-mode, PW/CW Doppler, Color Doppler, and 3D/4D imaging. Features include spatial compounding, tissue harmonic imaging, and elasticity imaging. Patient contact materials tested to ISO 10993-1. Connectivity includes wireless capability. 21-inch flat panel display.

Indications for Use

Indicated for fetal, abdominal, intraoperative, pediatric, small parts, transcranial, OB/GYN, cardiac, pelvic, neonatal/adult cephalic, vascular, musculoskeletal, superficial musculoskeletal, and peripheral vascular applications in adult and pediatric patients.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of three human profiles facing right, stacked on top of each other. The profiles are connected by flowing lines that resemble ribbons or waves. The graphic is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 23, 2014 Siemens Medical Solutions USA, Inc. % Mr. Mark Job Responsible Third Party Official 1394 25th Street, NW BUFFALO MN 55313 Re: K142876 Trade/Device Name: Acuson S1000/ S2000/ S3000 Diagnostic Ultrasound Systems Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX, OBJ Dated: September 30, 2014 Received: October 1, 2014 Dear Mr. Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. This determination of substantial equivalence applies to the following transducers intended for use with the Acuson S1000, S2000, S3000 Diagnostic Ultrasound Systems, as described in your premarket notification: Transducer Model Number | 12L4 | CW2 Probe | CW5 Probe | |---------------------------------|---------------------|---------------------------------| | EC9-4 Curved Array | MC9-4 Curved Array | 9L4 Linear Array | | 14L5 Multi-D Array | 4P1 Phased Array | 6C2 Curved Array | | 4C1 Curved Array | 6C1HD Curved Array | 8C3HD Curved Array | | 4V1 Phased Array | 10V4 Phased Array | 14L5 SP Linear Array | | 9EVF4 Curved Array | V5Ms Multiplane TEE | 18L6 HD Linear Array | | 8V3 Phased Array | 4V1c Phased Array | 6L3 | | EV8C4 | V7M TEE | 7CF2 Curved array mechanical 3D | | 7CF1 Curved array mechanical 3D | | AcuNav 8F Ultrasound Catheter | | AcuNav 10F Ultrasound Catheter | | | {1}------------------------------------------------ If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Sm.m.f) for Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### 1.3 INDICATIONS FOR USE 510(k) Number (if known): K142876 Device Name: S1000, S2000, S3000 Diagnostic Ultrasound Systems # Indications for Use: The ultrasound imaging systems are intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides the ability to measure anatomical structures {fetal, abdominal, intraoperative, pediatric, small organ, neonatal cephalic, adult cephalic, cardiac, transesophageal, transrectal, transvaginal, peripheral vessel, musculo-skeletal (conventional), musculo-skeletal (superficial) and neonatal cardiac} and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging". The Acuson Acunav Ultrasound Catheter is intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients. Prescription Use (Part 21CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {3}------------------------------------------------ Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) ## Diagnostic Ultrasound Indications for Use Form 510 (k) Number (if known): Device Name: Intended Use: ACUSON S1000, S2000, S3000 Ultrasound System Ultrasound imaging or fluid flow analysis of the human body as follows: | | | Mode of Operation | | | | | | | | | | |----------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|----------------------------------------|--| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | | | Ophthalmic | P | P | P | P | P | P | P | | BMDC | | | | Fetal | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11, 13 | | | Abdominal | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11, 13, 16, 18 | | | Intraoperative<br>(Note 9) | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11, 14 | | | Intraoperative<br>Neurological | P | P | P | P | P | P | P | | BMDC | | | | Pediatric | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11 | | | Small Organ<br>(Note 1) | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11,14, 16, 18 | | | Neonatal Cephalic | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10 | | | Adult Cephalic | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10 | | | Cardiac | P | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,6,7,8,10,15 | | | Trans-esophageal | P | P | P | P | P | P | P | | BMDC | Note 4 | | | Transrectal | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11,14 | | | Transvaginal | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11 | | | Transurethral | P | P | P | P | P | P | P | | BMDC | | | | Intravascular | P | P | P | P | P | P | P | | BMDC | | | | Peripheral vessel | P | P | P | P | P | P | P | | BMDC | Note2,3,4,5,6,7,8,10,<br>11,14,15 | | | Laparoscopic | P | P | P | P | P | P | P | | BMDC | | | | Musculo-skeletal<br>Conventional | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11,14, 18 | | | Musculo-skeletal<br>Superficial | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11,14, 18 | | | Other (specify) | P | P | P | P | P | P | P | | BMDC | Note 3,4,6, 10 | | N = new indication; P = previously cleared by FDA K140959 Additional Commnets: P. Tautenhahn, S. Reith, S. Schmitt - Note 1 i.e. breast, testes, thyroid, penis, prostate, etc. - Note 3 SieClear multi-view spatial compounding - Note 5 3-Scape real-time 3D imaging - Note 7 B&W SieScape panoramic imaging - Note 9 For example: vascular, abdominal - Note 11 Advanced Sieclear spatial compounding - Note 14 eSie™ Touch elasticity imaging / FTI - Note 16 Custom Tissue Imaging Note 18 VTI Note 4 Tissue Equalization Technology Note 6 Cadence contrast agent imaging Note 8 Power SieScape panoramic imaging Note 10 Clarify VE vascular enhancement technology Note 2 Ensemble tissue harmonic imaging - Note 13 STIC - Note 15 AHP - Note 17 eSie Fusion (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {4}------------------------------------------------ Concurrence of Center for Devices and Radiological Health (CDRH) #### Diagnostic Ultrasound Indications for Use Form | 510 (k) Number (if known): | _________________________ | |----------------------------|----------------------------------------------------------------------------| | Device Name: | 12L4 Transducer for use with ACUSON S1000, S2000 and<br>S3000 | | Intended Use: | Ultrasound imaging or fluid flow analysis of the human body as<br>follows: | | | Mode of Operation | | | Mode of Operation | | | | | | | | | | |----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|------------------------------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | | Ophthalmic | | | | | | | | | | | | Fetal | | | | | | | | | | | | Abdominal | | | | | | | | | | | | Intraoperative<br>Abdominal | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | Pediatric | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11, 14, 16 | | Small Organ<br>(Note 1) | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11, 14, 16 | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Trans-esophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral vessel | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5, 7,8,10,<br>11, 14 | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11, 14 | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other (specify) | | | | | | | | | | | N = new indication; P = previously cleared by FDA K140959 Additional Comments: - Note 1 i.e.: breast, testes, thyroid, penis, prostate, etc. - SieClear multi-view spatial compounding Note 3 - Note 5 3-Scape real-time 3D imaging - Note 7 B&W SieScape panoramic imaging Note 10 Clarify VE vascular enhancement technology compounding/DTCE Note 14 eSie™ Touch elasticity imaging - Note 2 Ensemble tissue harmonic imaging Tissue Equalization Technology Note 4 Note 6 Cadence contrast agent imaging - Note 8 Power SieScape panoramic imaging - Note 11 Advanced Sieclear spatial Custom Tissue Imaging /FTI Note 16 # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {5}------------------------------------------------ 510 (k) Number (if known): Device Name: #### CW2 Probe For Use On ACUSON S1000, S2000, S3000 Ultrasound System Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | |----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | | Ophthalmic | | | | | | | | | | | | Fetal | | | | | P | | | | | | | Abdominal | | | | | P | | | | | | | Intraoperative<br>(Note 9) | | | | | P | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | Pediatric | | | | | P | | | | | | | Small Organ<br>(Note 1) | | | | | P | | | | | | | Neonatal Cephalic | | | | | P | | | | | | | Adult Cephalic | | | | | P | | | | | | | Cardiac | | | | | P | | | | | | | Trans-esophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral vessel | | | | | P | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | P | | | | | | | Musculo-skeletal<br>Superficial | | | | | P | | | | | | | Other (specify) | | | | | | | | | | | N = new indication; P = previously cleared by FDA K140959 Additional Comments: For example: breast, testes, thyroid, penis, prostate, etc. Note 1 Note 9 For example: vascular, abdominal (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) {6}------------------------------------------------ 510 (k) Number (if known): Device Name: #### CW5 Probe For Use On ACUSON S1000, S2000, S3000 Ultrasound System Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | | | Mode of Operation | | | | | | | | | |----------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | | Ophthalmic | | | | | | | | | | | | Fetal | | | | | P | | | | | | | Abdominal | | | | | P | | | | | | | Intraoperative<br>(Note 9) | | | | | P | | | | | | | Intraoperative<br>Neurological | | | | | P | | | | | | | Pediatric | | | | | P | | | | | | | Small Organ<br>(Note 1) | | | | | P | | | | | | | Neonatal Cephalic | | | | | P | | | | | | | Adult Cephalic | | | | | P | | | | | | | Cardiac | | | | | P | | | | | | | Trans-esophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral vessel | | | | | P | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | P | | | | | | | Musculo-skeletal<br>Superficial | | | | | P | | | | | | | Other (specify) | | | | | | | | | | | N = new indication; P = previously cleared by FDA K140959 Additional Comments: Note 1 For example: breast, testes, thyroid, penis, prostate, etc. Note 9 For example: vascular, abdominal (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {7}------------------------------------------------ | 510 (k) Number (if | | |--------------------|--| | known): | | | Device Name: | | #### EC9-4 Curved Array Transducer For Use On ACUSON S1000, S2000, S3000 Ultrasound System Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | |----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|-----------------------------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | | Ophthalmic | | | | | | | | | | | | Fetal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10, 11 | | Abdominal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,6,7,8,10,<br>11 | | Intraoperative | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ<br>(Note 1) | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11,14 | | Neonatal Cephalic | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10, 11 | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Trans-esophageal | | | | | | | | | | | | Transrectal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5, 6, 7,8,10,<br>11,14 | | Transvaginal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10, 11 | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral vessel | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other (specify) | | | | | | | | | | | N = new indication; P = previously cleared by FDA K140959 Additional Comments: Note 1 i.e .: breast, testes, thyroid, penis, prostate, etc. Note 3 SieClear multi-view spatial compounding Note 5 3-Scape real-time 3D imaging Note 7 B&W SieScape panoramic imaging Note 11 Advanced Sieclear spatial compounding Note 14 eSie™ Touch elasticity imaging / FTI Note 4 Tissue Equalization Technology Note 6 Cadence contrast agent imaging Note 8 Power SieScape panoramic imaging Note 10 Clarify VE vascular enhancement technology #### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {8}------------------------------------------------ 510 (k) Number (if known): Device Name: #### MC9-4 Curved Array Transducer For Use On ACUSON S1000, S2000, S3000 Ultrasound Systems Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | |----------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|-----------------------------------| | Ophthalmic | | | | | | | | | | | | Fetal | P | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10, 11 | | Abdominal | P | P | P | P | | P | P | | BMDC | Note 2,3,4,5,6,,7,8,10,<br>11, | | Intraoperative<br>Note 9 | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ<br>(Note 1) | P | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11,14 | | Neonatal Cephalic | P | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10 | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Trans-esophageal | | | | | | | | | | | | Transrectal | P | P | P | P | | P | P | | BMDC | Note 2,3,4,5, 6, 7,8,10,<br>11,14 | | Transvaginal | P | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10, 11 | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral vessel | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other (specify) | | | | | | | | | | | N = new indication; P = previously cleared by FDA K140959 Additional Comments: Note 1 i.e.: breast, testes, thyroid, penis, prostate, etc Note 2 Ensemble tissue harmonic imaging Note 3 SieClear multi-view spatial compounding - Note 5 3-Scape real-time 3D imaging Note 7 B&W SieScape panoramic imaging Note 9 Abdomen and Vascular Note 4 Tissue Equalization Technology Note 6 Cadence contrast agent imaging Note 8 Power SieScape panoramic imaging Note 10 Clarify VE vascular enhancement technology Note 11 Advanced Sieclear spatial compounding Note 14 eSie™ Touch elasticity imaging / FTI (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {9}------------------------------------------------ | 510 (k) Number (if known): | | |----------------------------|--| | Device Name: | | 9L4 Linear Array Transducer For Use On ACUSON S1000, S2000, S3000 Ultrasound Systems Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | Mode of Operation | | | | | | | | | | |---------------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------------------------------| | | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | | Ophthalmic | | | | | | | | | | | | Fetal | P | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11 | | Abdominal<br>Intraoperative<br>Note 9 | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | Pediatric | P | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11 | | Small Organ<br>(Note 1) | P | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,6,7,8,10,<br>11,14, 16, 18 | | Neonatal Cephalic | P | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11 | | Adult Cephalic | P | P | P | P | | P | P | | BMDC | | | Cardiac | P | P | P | P | | P | P | | BMDC | Note 15 | | Trans-esophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral vessel | P | P | P | P | | P | P | | BMDC | Note 2,3,4,5,6,<br>7,8,10, 11, 14,15 | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | P | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,6,7,8,10, 11,<br>14 | | Musculo-skeletal<br>Superficial | P | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,6,7,8,10, 11,<br>14 | | Other (specify) | | | | | | | | | | | N = new indication; P = previously cleared by FDA K140959 Additional Comments: - Note 1 i.e.: breast, testes, thyroid, penis, prostate, etc. Note 2 Ensemble tissue harmonic imaging - Note 3 SieClear multi-view spatial compounding - Note 5 3-Scape real-time 3D imaging - Note 7 B&W SieScape panoramic imaging - Note 9 Abdomen and Vascular - Note 11 Advanced Sieclear spatial compounding - Note 15 AHP - Note 18 VTI (Virtual Touch Imaging) - - Note 4 Tissue Equalization Technology - Note 6 Cadence contrast agent imaging - Note 8 Power SieScape panoramic imaging - Note 10Clarify VE vascular enhancement technology - Note 14 eSie™ Touch elasticity imaging / FTI Note 16 Custom Tissue Imaging (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {10}------------------------------------------------ 510 (k) Number (if known): #### Device Name: #### 14L5 Multi-D Array Transducer For Use On ACUSON S1000, S2000, S3000 Ultrasound Systems Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | |----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|------------------------------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | | Ophthalmic | | | | | | | | | | | | Fetal | | | | | | | | | | | | Abdominal | | | | | | | | | | | | Intraoperative<br>Note 9 | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ<br>(Note 1) | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11, 14, 16 | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Trans-esophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral vessel | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,6,<br>7,8,10, 11, 14 | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11, 14 | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other (specify) | | | | | | | | | | | N = new indication: P = previously cleared by FDA K140959 Additional Comments: Note 1 i.e.: breast, testes, thyroid, penis, prostate, etc. Note 2 Ensemble tissue harmonic imaging Note 3 SieClear multi-view spatial compounding Note 5 3-Scape real-time 3D imaging Note 7 B&W SieScape panoramic imaging Note 9 Abdomen and Vascular technology Note 11 Advanced Sieclear spatial compounding Note 16 Custom Tissue Imaging te 2 Ensemble tissue harmonic imaging Note 4 Tissue Equalization Technology Note 6 Cadence contrast agent imaging Power SieScape panoramic imaging Note 8 Clarify VE vascular enhancement Note 10 Note 14. eSio™ Touch elasticity imaging (ETI) Note 14 eSie™ Touch elasticity imaging / FTI Note 18 Virtual Touch Imaging # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {11}------------------------------------------------ 510 (k) Number (if known): #### Device Name: #### 4P1 Phased Array Transducer For Use On ACUSON S1000, S2000, S3000 Ultrasound Systems Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | |----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|-----------------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | | Ophthalmic | | | | | | | | | | | | Fetal | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10 | | Abdominal | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10 | | Intraoperative<br>Note 9 | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10 | | Cardiac | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,6,7,8,10 | | Trans-esophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral vessel | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other (specify) | | | | | | | | | | | N = new indication; P = previously cleared by FDA K140959 Additional Comments: Note 2 Ensemble tissue harmonic imaging Note 3 SieClear multi-view spatial compounding - Note 4 Tissue Equalization Technology Note 5 3-Scape real-time 3D imaging Note 6 Cadence contrast agent imaging Note 7 B&W SieScape panoramic imaging Note 8 Power SieScape panoramic imaging Note 9 Abdomen and Vascular Note 10Clarify VE vascular enhancement technology # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {12}------------------------------------------------ | 510 (k) Number (if known): | |----------------------------| | Device Name: | #### 6C2 Curved Array Transducer For Use On ACUSON S1000, S2000, S3000 Ultrasound Systems Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | | | Mode of Operation | | | | | | | | | |----------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|----------------------------------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | | Ophthalmic | | | | | | | | | | | | Fetal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11 | | Abdominal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11, 14, 16, 17 | | Intraoperative<br>Note 9 | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | Pediatric | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11 | | Small Organ | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Trans-esophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral vessel | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11 | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other (specify) | | | | | | | | | | | N = new indication; P = previously cleared by FDA K140959 Additional Comments: Note 2 Ensemble tissue harmonic imaging Note 11 Advanced Sieclear spatial compounding Note 4 Tissue Equalization Technology Note 7 B&W SieScape panoramic imaging Note 9 Abdomen and Vascular - Note 3 SieClear multi-view spatial compounding Note 5 3-Scape real-time 3D imaging Power SieScape panoramic imaging Note 8 Note 10 Clarify VE vascular enhancement technology Note 14 eSie™ Touch elasticity imaging / FTI Note 16 Custom Tissue Imaging Note 17 eSie Fusion (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {13}------------------------------------------------ | 510 (k) Number (if known): | |----------------------------| | Device Name: | #### 4C1 Curved Array Transducer For Use On ACUSON S1000, S2000, S3000 Ultrasound Systems Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | |----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|----------------------------------------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | | Ophthalmic | | | | | | | | | | | | Fetal | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10, 11 | | Abdominal | P | P | P | P | P | P | P | | BMDC | Note2,3,4,5,6,7,8, 10,<br>11, 14, 16, 17, 18 | | Intraoperative<br>Note 9 | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ | P | P | P | P | P | P | P | | BMDC | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | P | P | P | P | P | P | P | | BMDC | | | Trans-esophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral vessel | P | P | P | P | P | P | P | | BMDC | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other (specify) | | | | | | | | | | | N = new indication; P = previously cleared by FDA K140959 Additional Comments: Note 2 Ensemble tissue harmonic imaging - Note 4 Tissue Equalization Technology - Note 6 Cadence contrast agent imaging Note 8 Power SieScape panoramic imaging - Note 10Clarify VE vascular enhancement technology compounding - Note 14 eSie™ Touch elasticity imaging / FTI - Note 17 eSie Fusion - Note 3 SieClear multi-view spatial compounding - Note 5 3-Scape real-time 3D imaging - Note 7 B&W SieScape panoramic imaging Note 9 Abdomen and Vascular Note 11 Advanced Sieclear spatial Note 16 Custom Tissue Imaging Note 18 VTI # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {14}------------------------------------------------ 510 (k) Number (if known): #### 6C1HD Curved Array Transducer For Use On ACUSON S2000, S3000 Ultrasound Systems Intended Use: Device Name: Ultrasound imaging or fluid flow analysis of the human body as follows: | | | Mode of Operation | | | | | | | | | |----------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|-------------------------------------------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | | Ophthalmic | | | | | | | | | | | | Fetal | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11 | | Abdominal | P | P | P | P | P | P | P | | BMDC | Note2,3,4,5,6,7,8,<br>10, 11, 14, 16, 17,<br>18 | | Intraoperative<br>Note 9 | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ | P | P | P | P | P | P | P | | BMDC | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | P | P | P | P | P | P | P | | BMDC | | | Trans-esophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral vessel | P | P | P | P | P | P | P | | BMDC | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other (specify) | | | | | | | | | | | N = new indication; P = previously cleared by FDA K140959 Additional Comments: Note 2 Ensemble tissue harmonic imaging Note 4 Tissue Equalization Technology Note 6 Cadence contrast agent imaging Note 8 Power SieScape panoramic imaging Note 10Clarify VE vascular enhancement technology compounding Note 14 eSie™ Touch elasticity imaging / FTI Note 17 eSie Fusion Note 3 SieClear multi-view spatial compounding Note 5 3-Scape real-time 3D imaging Note 7 B&W SieScape panoramic imaging Note 9 Abdomen and Vascular Note 11 Advanced Sieclear spatial Note 16Custom Tissue Imaging Note 18 VTI (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {15}------------------------------------------------ | 510 (k) Number (if known): | |----------------------------| | Device Name: | #### 8C3HD Curved Array Transducer For Use On ACUSON S2000, S3000 Ultrasound Systems Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | |----------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|------------------------------------| | Ophthalmic | | | | | | | | | | | | Fetal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11 | | Abdominal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11, 14, 16 | | Intraoperative<br>Note 9 | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | Pediatric | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11 | | Small Organ | | P | P | P | | P | P | | BMDC | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Trans-esophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | |…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...