M9 DIAGNOSTIC ULTRASOUND SYSTEM, M9T DIAGNOSTIC ULTRASOUND SYSTEM, M9CV DIAGNOSTIC ULTRASOUND SYSTEM

K141010 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · IYN · May 14, 2014 · Radiology

Device Facts

Record IDK141010
Device NameM9 DIAGNOSTIC ULTRASOUND SYSTEM, M9T DIAGNOSTIC ULTRASOUND SYSTEM, M9CV DIAGNOSTIC ULTRASOUND SYSTEM
ApplicantShenzhen Mindray Bio-Medical Electronics Co., Ltd.
Product CodeIYN · Radiology
Decision DateMay 14, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric, 3rd-Party Reviewed

Intended Use

The M9/M9CV/M9T Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ(breast, thyroid, testes), neonatal and adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal(conventional, superficial), adult and pediatric cardiac, trans-esoph.(Cardiac), peripheral vessel , urology exams.

Device Story

Software-controlled diagnostic ultrasound system; acquires/displays ultrasound data in B-Mode, PW-Mode, Color-Mode, Power/Dirpower Mode, THI, iScape, Color M, Elastography, and LVO; supports combined modes (e.g., B/M, B/PW/Color). Uses linear and convex array probes (3.0-10.0 MHz). Operated by clinicians in clinical settings. Transforms received echoes into onscreen displays of anatomic structures and fluid flow. Provides specialized measurements and calculations. Benefits patients through non-invasive diagnostic imaging and flow analysis.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including acoustic output measurement (UD 2, UD 3), biocompatibility (ISO 10993-1), cleaning/disinfection effectiveness, and thermal/electrical/mechanical safety (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37).

Technological Characteristics

Software-controlled ultrasound system; linear/convex array probes (3.0-10.0 MHz). Materials: biocompatible per ISO 10993-1. Imaging modes: B, M, Color Doppler, PWD, CWD, Amplitude Doppler, Color M, Anatomical M, and combined modes. Special functions: iScape, TDI, TDI QA, LVO, Elastography, Stress Echo, Tissue Tracking QA, iWorks, iNeedle, IMT. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 14971, IEC 62366, IEC 62304.

Indications for Use

Indicated for adults, pregnant women, pediatric patients, and neonates for fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal/adult cephalic, trans-rectal, trans-vaginal, musculoskeletal (conventional, superficial), adult/pediatric cardiac, trans-esophageal (cardiac), peripheral vessel, and urology exams.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(K) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92(c). The assigned 510(k) number is: ### 1. Submitter: Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China Tel: +86 755 8188 5658 Fax: +86 755 2658 2680 ### Contact Person: Wu Zicui Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China Date Prepared: February 19, 2014 2. Device Name: M9/M9T/M9CV Diagnostic Ultrasound System ### Classification Regulatory Class: II Review Category: Tier II 21 CFR 892.1550 Ultrasonic Pulsed Doppler Imaging System (IYN) 21 CFR 892.1560 Ultrasonic Pulsed Echo Imaging System (IYO) 21 CFR 892.1570 Diagnostic Ultrasound Transducer (ITX) ### 3. Device Description: M9/M9T/M9CV is a software controlled, ultrasonic diagnostic system. Its function is to acquire and display ultrasound data in B-Mode, PW-Mode, PW-Mode, Color-Mode , Power/Dirpower Mode, THI, iScape mode, Color M, Elastography, LVO or the combined mode (i.e. B/M-Mode, B/PW-mode, B/PW/Color). This system is a Track 3 device that employs an array of probes that include linear array and convex array with a frequency range of approximately 3.0 MHz to 10.0 {1}------------------------------------------------ MHz. ### 4. Intended Use: The M9/M9CV/M9T Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ(breast, thyroid, testes), neonatal and adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal(conventional, superficial), adult and pediatric cardiac, trans-esoph.(Cardiac), peripheral vessel , urology exams. ### 5. Comparison with Predicate Devices: M9/M9T/M9CV Diagnostic Ultrasound System is comparable with and substantially equivalent to these predicate devices: | Predicate Device | Manufacturer | Model | 510(k) Number | |------------------|--------------|------------|---------------| | 1 | Mindray | DC-8 | K132341 | | 2 | Mindray | M7 | K131690 | | 3 | GE | Voluson E8 | K101236 | M9/M9T/M9CV has the same technological characteristics, is comparable in key safety and effectiveness features, and has the same intended uses and basic operating modes as the predicate devices. All systems transmit ultrasonic energy into patients and perform post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body. All systems allow for specialized measurements of structures and flow, and calculations. - Subject device M9/M9T/M9CV has the same intended uses as the predicated device ● DC-8(K132341) | Subject Device | Predicate device | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | M9/M9T/M9CV | DC-8 (K132341) | | The M9/M9CV/M9T Diagnostic Ultrasound<br>System is applicable for adults, pregnant<br>women, pediatric patients and neonates. It is<br>intended for use in fetal, abdominal, pediatric,<br>small organ(breast, thyroid, testes), neonatal<br>and adult cephalic, trans-rectal, trans-vaginal,<br>musculo-skeletal(conventional, superficial),<br>adult and pediatric cardiac,<br>trans-esoph.(Cardiac), peripheral vessel ,<br>urology exams. | The DC-8/DC-8 PRO/DC-8 CV/DC-8<br>EXP/DC-8S diagnostic ultrasound system is<br>applicable for adults, pregnant women, pediatric<br>patients and neonates. It is intended for use in<br>fetal, abdominal, pediatric, small organ (breast,<br>thyroid, testes), neonatal cephalic, adult<br>cephalic, trans-rectal, trans-vaginal,<br>musculo-skeletal (conventional, superficial),<br>cardiac adult, cardiac pediatric, peripheral<br>vessel, urology and transesophageal (Cardiac)<br>exams | {2}------------------------------------------------ - All of the patient contact material of the M9/M9T/M9CV are the same as that of the predicated device DC-8(K132341) and the predicated device M7(K131690) with the exception of the materials for probe SP5-1s. Materials for the SP5-1s are tested under ISO 10993-1 for its biocompatibility. - The acoustic power levels of M9/M9T/M9CV are below the limits of FDA, which is the same as the predicated device DC-8(K132341). - M9/M9T/M9CV is designed in compliance with the FDA recognized electrical and physical safety standard, which is the same as the predicated device DC-8(K132341). - M9/M9T/M9CV has the same imaging modes as the predicated device DC-8(K132341): B, M, Color Doppler, PWD, CWD, Amplitude Doppler, Color M, Anatomical M Mode and combined mode). - M9/M9T/M9CV has some special functions: iScape, TDI, TDI QA, LVO, Elastography, Stress Echo, Tissue Tracking QA, iWorks, iNeedle and IMT. All of them are identical as the predicated device DC-8(K132341). - M9/M9T/M9CV has the same capacity in term of making comments and body marks on the images, reporting patient exam results as the predicated device DC-8(K132341). - . M9/M9T/M9CV has similar probes as the predicated device DC-8(K132341) and M7(K131609): | Subject device | Predicated device | Predicated device | NOTE | |----------------|-------------------|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | M9/M9T/M9CV | DC-8(K132341) | M7(K131690) | | | C5-1s | C5-2E | / | They have same indications for use, mode of operation, array elements. Different design is intended to improve penetrability and spacial resolution. C5-1s probe has been tested under the IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37. Differences do not influence the safety and effectiveness of this probe. | | L12-4s | / | L12-4s | Same. | | P10-4s | P10-4E | / | Same (except for the shape of the socket which does not affect the performance of the probe). | | P7-3Ts | / | P7-3Ts | Same. | {3}------------------------------------------------ | L14-6Ns | | L14-6Ns | Same. | |---------|-----------------|---------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------| | V11-3Ws | V11-3WE | / | Same (except for the shape-<br>of the socket which does<br>not affect the performance<br>of the probe). | | C11-3s | C11-3E | / | Same (except for the shape<br>of the socket which does<br>not affect the performance<br>of the probe). | | CW2s | CW2s | / | Same. | | CW5s | CW5s | / | Same. | | SP5-1s | P4-2NE<br>P4-2E | / | SP5-1s has the same<br>indications for use and<br>performance specification<br>with P4-2NE, except that<br>indications for use Neonatal<br>Cephalic is substantially | . M9/M9T/M9CV has the same measurement and calculation functions as the predicated device DC-8(K132341) except for the cardiac measurement and calculation MV ALL which is equivalent to predicated device Voluson E8(101236). ### 6. Non-clinical Tests: M9/M9T/M9CV Diagnostic Ultrasound System has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety, and has been designed to conform with applicable medical safety standards. This device has been tested and evaluated under the following standards: - UD 2 Acoustic Output Measurement Standard for Diagnostic Ultrasound . Equipment. - UD 3 Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment - 통 AAMI / ANS1 ES60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance. - 를 IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests (Edition 3) - IEC 60601-2-37 Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment - ISO14971 Medical devices Application of risk management to medical devices B - ISO 10993-1 Biological evaluation of medical devices -- Part 1: Evaluation and ## 0.5 {4}------------------------------------------------ K141010 Page 5 of 5 testing within a risk management process - 에 IEC 62366 Medical devices - Application of usability engineering to medical devices - l Medical device software - Software life cycle processes IEC 62304 These non-clinical tests relied on in this premarket notification submission can support the determination of substantial equivalence of the subject device. ## 7. Clinical Studies Not applicable. The subject of this submission, M9/M9T/M9CV Diagnostic Ultrasound System, does not require clinical studies to support substantial equivalence. ### Conclusion: Intended uses and other key features are consistent with traditional clinical practices, FDA guidelines and established methods of patient examination. The design, development and quality process of the manufacturer confirms with 21 CFR 820, ISO 9001 and ISO 13485 quality systems. The device conforms to applicable medical device safety standards. Therefore, the M9/M9T/M9CV Diagnostic Ultrasound System is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three parallel lines that curve and converge, resembling a stylized human figure. The symbol is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement. The text is in uppercase letters and is evenly spaced around the symbol. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 14, 2014 Shenzhen Mindray Bio-Medical Electronics, Co. Ltd. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Scrvices LLC 1394 25" Street NW BUFFALO MN 55313 Re: K141010 Trade/Device Name: M9 Diagnostic Ultrasound System. M9T Diagnostic Ultrasound System, M9CV Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: 11 Product Code: IYN, IYO, ITX Dated: April 24, 2014 Received: April 25, 2014 Dear Mr. Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRFI does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. This determination of substantial equivalence applies to the following transducers intended for use with the M9 Diagnostic Ultrasound System, M9T Diagnostic Ultrasound System, M9C V Diagnostic Ultrasound Systems, as described in your premarket notification: Transducer Model Number | C11-3s | C5-1s | L12-4s | |---------|---------|--------| | P7-3Ts | SP5-1s | P10-4s | | L14-6Ns | V11-3Ws | CW2s | | CW5s | | | {6}------------------------------------------------ Page 2-Mr. Job If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/McdicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ![](https://i.imgur.com/99999999999999999999999999999999.png) for Janine Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page. 510(k) Number (if known) K141010 Device Name M9/M9T/M9CV Indications for Use (Describe) The M91M9T Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatic, small organ(breast, thyroid, testes), neonatal and adult cephalic, transvaginal, musculo-skeletal(conventional, superficial), adult and pediatic cardiac, trans-soph.(Cardiac), peripheral vessel, urology exams. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Page 1 of 2 008-I PSC Publishing Services (30); 441-4740 EF FORM FDA 3881 (1/14) {8}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of Information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." **008-2** {9}------------------------------------------------ #### Diagnostic Ultrasound Indications For Use Format M9/M9T Diagnostic Ultrasound System System: Transducer: N/A Intended Use: Diagnostic ultrasound imaging or Nuid Now analysis of the human body as follows: | Clinical Application | | Mode of Operation | | | | | | | | | | |----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|-------------------------------------------------------|-------------------|-----|-----|--------------------------|----------------------|-----------------------|-----------------------|-----------------|--| | General<br>(Track 1<br>Only) | Specific (Track 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>(specify) | Other (specify) | | | | Ophthalmic | Ophthalmic | | | | | | | | | | | | Fetal Imaging<br>& Other | Fetal | N | N | N | N | N | N | N | Note 1, 2,4,5 | | | | | Abdominal | N | N | N | N | N | N | N | Note 1, 2,4,5 | | | | | Intra-operative (Specify*) | | | | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | Pediatric | N | N | N | N | N | N | N | Note 1, 2,4,5 | | | | | Small Organ (Specify**) | N | N | N | N | N | N | N | Note 1,2,5,6 | | | | | Neonatal Cephalic | N | N | N | N | N | N | N | Note 1,2,4,5 | | | | | Adult Cephalic | N | N | N | N | N | N | N | Note 1, 2,4,5 | | | | | Trans-rectal | N | N | N | N | N | N | N | Note 1,2,4,5 | | | | | Trans-vaginal | N | N | N | N | N | N | N | Note 1, 2,4,5 | | | | | Trans-urethral | | | | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | | | | Musculo-skeletal<br>(Conventional) | N | N | N | N | N | N | N | Note 1,2,4,5 | | | | | Musculo-skeletal (Superficial) | N | N | N | N | N | N | N | Note 1,2,5 | | | | | Intravascular | | | | | | | | | | | | | Cardiac Adult | N | N | N | N | N | N | N | Note 1,2,3,4,5,7 | | | | | Cardiac Pediatric | N | N | N | N | N | N | N | Note 1,2,3,4,5 | | | | Cardiac | Intravascular (Cardiac) | | | | | | | | | | | | | Trans-esoph. (Cardiac) | N | N | N | N | N | N | N | Note 1, 3,4 | | | | | Intra-cardiac | | | | | | | | | | | | Peripheral<br>vessel | Peripheral vessel | N | N | N | N | N | N | N | Note 1,2,4,5 | | | | | Other (Specify***) | N | N | N | N | N | N | N | Note 1,2,4,5 | | | | | N=new indication: P=previously cleared by FDA E=added under Appendix E | | | | | | | | | | | | | Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+ B、Power + PW +B. | | | | | | | | | | | | | * Intraoperative includes abdominal, thoracic, and vascular. | | | | | | | | | | | | | ** Small organ-breast, thyroid, testes. | | | | | | | | | | | | | *** Other use includes Urology. | | | | | | | | | | | | | Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents. | | | | | | | | | | | | | Note 2: iScape | | | | | | | | | | | | | Note 3:TDI | | | | | | | | | | | | | | Note 4: Color M | | | | | | | | | | | | | Notes: Biopsy Guidance | | | | | | | | | | | | | Note6: Elastography | | | | | | | | | | | | Note7: Contrast imaging (Contrast agent for LVO) | | | | | | | | | | | | | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | | | | | | | | | | | | | | | | | | | | | | | | | | | Concurrence of CDRH, Office of Device Evaluation(ODE) | | | | | | | | | | | | Clinical Application | | | | | | Mode of Operation | | | | | | General<br>(Track 1<br>Only) | Specific (Track 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>(specify) | Other (specify) | | | | Ophthalmic | Ophthalmic | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | Abdominal | N | N | N | | N | N | N | Note 1,2,4,5 | | | | | Intra-operative (Specify*) | | | | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | Pediatric | N | N | N | | N | N | N | Note 1,2,4,5 | | | | | Small Organ (Specify**) | | | | | | | | | | | | Fetal Imaging<br>& Other | Neonatal Cephalic | N | N | N | | N | N | N | Note 1,2,4,5 | | | | | Adult Cephalic | | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | | | | Musculo-skeletal | | | | | | | | | | | | | (Conventional) | | | | | | | | | | | | | Musculo-skeletal (Superficial) | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Cardiac Adult | | | | | | | | | | | | | Cardiac Pediatric | N | N | N | | N | N | N | Note 1,2,4,5 | | | | Cardiac | Intravascular (Cardiac) | | | | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | | Peripheral<br>vessel | Peripheral vessel | N | N | N | | N | N | N | Note 1,2,4,5 | | | | | Other (Specify***) | | | | | | | | | | | | | N=new indication; P=previously cleared by FDA | | | | | E=added under Appendix E | | | | | | | | Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW +Color+ B、 | | | | | | | | Power + PW +B. | | | | | * Intraoperative includes abdominal, thoracic, and vascular etc. | | | | | | | | | | | | | ** Small organ-breast, thyroid, testes. | | | | | | | | | | | | | *** Other use includes Urology. | | | | | | | | | | | | | | | | | | | | | | | | | Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.<br>Note 2: iScape | | | | | | | | | | | | | Note 3:TDI | | | | | | | | | | | | | Note 4: Color M | | | | | | | | | | | | | Note 5: Biopsy Guidance | | | | | | | | | | | | | | Note6: Elastography | | | | | | | | | | | | | Note7: Contrast imaging (Contrast agent for LVO) | | | | | | | | | | | | | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | | | | | | | | | | | | | Concurrence of CDRH, Office of Device Evaluation(ODE) | | | | | | | | | | | | | | | | | | | | | | | | | | Clinical Application | | | | | | Mode of Operation | | | | | | General<br>(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>(specify) | Other (specify) | | | | Only) | | | | | | | | | | | | | Ophthalmic | Ophthalmic | | | | | | | | | | | | | Fetal | N | N | N | | N | N | N | Note 1.2.4.5 | | | | | Abdominal | N | N | N | | N | N | N | Note 1,2,4,5 | | | | | Intra-operative (Specify®) | | | | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | Pediatric | N | N | N | | N | N | N | Note 1,2,4,5 | | | | | Small Organ (Specify**) | | | | | | | | | | | | Fetal Imaging | Neonatal Cephalic | | | | | | | | | | | | & Other | Adult Cephalic | | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | | | | Musculo-skeletal | N | N | N | | N | N | N | Note 1,2,4,5 | | | | | (Conventional) | | | | | | | | | | | | | Musculo-skeletal (Superficial) | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Cardiac Adult | | | | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | | | Cardiac | Intravascular (Cardiac) | | | | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | | Peripheral | Peripheral vessel | N | N | ਮ | | N | N | N | Note 1,2,4,5 | | | | vesse. | Other (Specify *** ) | | | | | | | | | | | | | N=new indication; P=previously cleared by FDA | | | | | E=added under Appendix E | | | | | | | | Additional comments: Combined modes-B+M. PW+B、Color + B、PW +Color+B、PW +Color+B、PW+B. | | | | | | | | | | | | | * Intraoperative includes abdominal, thoracic, and vascular. | | | | | | | | | | | | | * * Small organ-breast, thyroid, testes. | | | | | | | | | | | | | *** Other use includes Urology. | | | | | | | | | | | | | Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents. | | | | | | | | | | | | Note 2: iScape | | | | | | | | | | | | | Note 3:TDI | | | | | | | | | | | | | Note 4: Color M | | | | | | | | | | | | | Note5: Biopsy Guidance | | | | | | | | | | | | | Note6: Elastography | | | | | | | | | | | | | | Note7: Contrast imaging (Contrast agent for LVO) | | | | | | | | | | | | | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | | | | | | | | | | | | | | | | | | | | | | | | | | Concurrence of CDRH, Office of Device Evaluation(ODE) | | | | | | | | | | | | | Clinical Application | Mode of Operation | | | | | | | | | | | General<br>(Track 1<br>Only) | Specific (Track 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>(specify) | Other (specify) | | | | Ophthalmic | Ophthalmic | | | | | | | | | | | | Fetal Imaging<br>& Other | Fetal | | | | | | | | | | | | | Abdominal | P | P | P | | P | P | P | Note 1,2,5 | | | | | Intra-operative (Specify*) | | | | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | Pediatric | P | P | P | | P | P | P | Note 1.2,5 | | | | | Small Organ (Specify**) | P | P | P | | P | P | P | Note 1,2,5,6 | | | | | Neonatal Cephalic | P | P | P | | P | P | P | Note 1,2,5 | | | | | Adult Cephalic | | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | | | | Musculo-skeletal<br>(Conventional) | P | P | P | | P | P | P | Note 1,2,5 |…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...