DIAGNOSTIC ULTRASOUND SYSTEM

K123249 · Edan Instruments, Inc. · IYN · Feb 8, 2013 · Radiology

Device Facts

Record IDK123249
Device NameDIAGNOSTIC ULTRASOUND SYSTEM
ApplicantEdan Instruments, Inc.
Product CodeIYN · Radiology
Decision DateFeb 8, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric, 3rd-Party Reviewed

Intended Use

The diagnostic ultrasound system (U50) is applicable for adults, pregnant women, pediatric patients' ultrasound evaluation in hospitals and clinics. It is intended for use in abdominal, obstetrics, gynecology, pediatric, small parts, urology, peripheral vascular, musculoskeletal (conventional and superficial), transvaginal and cardiac clinical applications, by or on the order of a physician or similarly qualified health care professional.

Device Story

Portable diagnostic ultrasound system (U50) for clinical imaging; utilizes transducers (linear, convex, micro-convex) with 2.5-11 MHz frequency range. Inputs: ultrasonic signals; processing: Phased Inversion Harmonic Compound Imaging (eHCI), Multi-Beam-Forming (mBeam), Speckle Resistance Imaging (eSRI), and Spatial Compounding Imaging. Outputs: B-mode, M-mode, PW-mode, Color-mode, and PDI/DPDI mode images on 12.1-inch LCD. Operated by physicians or qualified professionals in hospitals/clinics. Provides real-time visualization for diagnostic assessment and biopsy guidance. Benefits: enables non-invasive diagnostic imaging and fluid flow analysis across multiple clinical specialties.

Clinical Evidence

No clinical data required. Safety and effectiveness established via bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), acoustic output testing per FDA 2008 guidelines, and biocompatibility (ISO 10993-1, -5, -10).

Technological Characteristics

Portable ultrasound system with 12.1-inch LCD. Features: eHCI, mBeam, eSRI, and Spatial Compounding. Transducers: linear, convex, micro-convex (2.5-11 MHz). Biocompatible needle-guide brackets. Standards: IEC 60601-1, IEC 60601-1-2, ISO 10993. Modes: B, M, PW, Color, PDI/DPDI.

Indications for Use

Indicated for ultrasound evaluation of adults, pregnant women, and pediatric patients in clinical settings. Applications include abdominal, obstetrics, gynecology, pediatric, small parts (thyroid, testes, breast), urology, peripheral vascular, musculoskeletal (conventional/superficial), transvaginal, and cardiac imaging.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K123249 #### Section 5 - 510(k) Summary FEB 8 2013 ## Prepared in accordance with the requirements of 21 CFR Part 807.92 | <b>1. Submitter:</b> | Edan Instruments, Inc.<br>3/F-B, Nanshan Medical Equipments Park, Nanhai Rd 1019#, Shekou,<br>Nanshan Shenzhen, 518067 P.R. China<br>Tel.: (0755) 26856469<br>Fax: (0755) 26882223 | |-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <b>Contact Person:</b> | Randy Jiang | | <b>Prepare date:</b> | Sep 25, 2012 | | <b>2. Device name<br/>and classification:</b> | Device Name:<br>Diagnostic Ultrasound System, Model U50<br>Classification Name:<br>892.1550 System, Imaging, Pulsed Doppler, Ultrasonic<br>Product code: IYN<br>892.1560 Ultrasonic, Pulsed echo, Imaging<br>Product code: IYO<br>892.1570 Transducer, Ultrasonic, Diagnostic<br>Product code: ITX<br>Regulatory Class: Class II | 3. Predicate Device: DUS 60 Digital Ultrasonic Diagnostic Imaging System / K110999 / Edan Instruments, Inc. DC-6 Diagnostic Ultrasound System / K072164 / Shenzhen Mindray Bio-medical Electronics Co., Ltd. M5 Diagnostic Ultrasound System / K102991 / Shenzhen Mindray Bio-medical Electronics Co., Ltd. GE LOGIQ E9 Diagnostic Ultrasound System/ K082185/ General Electric Co. {1}------------------------------------------------ | 4. Device<br>Description: | The U50 is a portable Diagnostic Ultrasound System, which applies<br>advanced technologies such as Phased Inversion Harmonic Compound<br>Imaging (eHCI), Multi-Beam-Forming (mBeam), Speckle Resistance<br>Imaging (eSRI), and Spatial Compounding Imaging, etc. Various image<br>parameter adjustments, 12.1 inch LCD and diverse probes are<br>configured to provide clear and stable images. | |---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Its function is to acquire and display Ultrasound images in B-mode,<br>M-mode, PW-mode, Color-mode, PDI/DPDI mode. This system<br>provides a series of probes that include linear array, convex array,<br>micro-convex array with a frequency range of approximately 2.5 MHz<br>to 11 MHz. | | 5. Intended Use: | The diagnostic ultrasound system (U50) is applicable for adults,<br>pregnant women, pediatric patients' ultrasound evaluation in hospitals<br>and clinics. It is intended for use in abdominal, obstetrics, gynecology, | pediatric, small parts, urology, peripheral vascular, musculoskeletal (conventional and superficial), transvaginal and cardiac clinical applications, by or on the order of a physician or similarly qualified health care professional. ### 6. Effectiveness and Safety Considerations: #### Clinical test: Clinical testing is not required. #### Non-clinical test: The following safety standards are conducted on the subject device: (1) IEC 60601-1 Electrical Safety (2) IEC 60601-1-2 Electromagnetic Compatibility (3) Acoustic output testing as per the guideline "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers" dated September 9, 2008. (4) ISO 10993-1, ISO 10993-5 and ISO 10993-10 #### 7.Comparison to the predicate device Comparison to the predicate devices, the subject device has same intended use, similar product design, same performance effectiveness, performance safety and the same needle-guide bracket material, property, and sterilization methods as the predicate devices. The differences above between the subject device and predicate device do not affect the basic design principle, usage, effectiveness and safety of the subject device. And no question is raised regarding to effectiveness and safety. ### 8. Substantially Equivalent Determination Verification and validation testing was conducted on the U50 Diagnostic Ultrasonic System. 5-2 {2}------------------------------------------------ This premarket notification submission demonstrates that U50 Diagnostic Ultrasonic System is substantially equivalent to the predicate devices. {3}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 ### February 8, 2013 Edan Instruments, Inc. c/o Mr. Ned Devine Responsible Third Party Official Underwriters Laboratories 333 Pfingsten Rd. NORTHBROOK IL 60062 Re: K123249 Trade/Device Name: U50 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: January 17, 2013 Received: January 30, 2013 Dear Mr. Devine: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. This determination of substantial equivalence applies to the following transducers intended for use with the U50 Diagnostic Ultrasound System, as described in your premarket notification: | Transducer Model Number<br>E | | |------------------------------|-----------------------------------------------------------------------------------| | | --------------------------------------------------------------------------------- | | C352UB | E612UB | C422UB | |---------|---------|--------| | L1042UB | C612UB | L552UB | | L742UB | C6152UB | | {4}------------------------------------------------ Page 2 - Mr. Devine If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, #### Sean M. Boyd -S for Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Diagnostic Ultrasound System 510K Submission ## Indications for Use 510(k) Number (if known): Device Name: U50 Diagnostic Ultrasound System #### Intended Use: The diagnostic ultrasound system (U50) is applicable for adults, pregnant women, pediatric patients' ultrasound evaluation in hospitals and clinics. It is intended for use in abdominal, obstetrics, gynecology, pediatric, small parts, urology, peripheral vasculoskeletal (conventional and superficial), transvaginal and cardiac clinical applications, by or on the order of a physician or similarly qualified health care professional. Prescription Use (21 CFR Part 801 Subpart D) · Or Over the Counter Use (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IFNEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) (Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health 510(k) (Division Sign Off) Division of Radlotogical Health Office of In Vitre Diagnostics and Re 510(k) K123219 {6}------------------------------------------------ ### U50 Diagnostic Ultrasound System Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Clinical Application | | Mode of Operation | | | | | | |-----------------------------|---------------------------------|---|-------------------|----|----|-------|--------------------|-----------------| | General | Specific | B | M | PW | CW | Color | Combined (Specify) | Other (Specify) | | | | | | | | | | | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal / Obstetrics | N | N | N | | N | N | Note 1,2 | | | Abdominal | N | N | N | | N | N | Note 1,2 | | | Intra-operative (Specify) | | | | | | | | | | Intra-operative (Neuro logical) | | | | | | | | | | Laparoscopic | | | | | | | | | | Pediatric | N | N | N | | N | N | Note 1,2 | | | Small Organ (Specify) * | N | N | N | | N | N | Note 1,2 | | Fetal<br>Imaging<br>& Other | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | | | | | | | | | | Trans-vaginal | N | N | N | | N | N | Note 1,2 | | | Trans-urethral | | | | | | | | | | Musculo-skeletal(Conventional) | N | N | N | | N | N | Note 1,2 | | | Musculo-skeletal (Superficial) | N | N | N | | N | N | Note 1,2 | | | Intravascular | | | | | | | | | | Other (Specify) ** | N | N | N | | N | N | Note 1,2 | | Cardiac | Adult Cardiac | N | N | N | | N | N | Note 1,2 | | | Pediatric Cardiac | N | N | N | | N | N | Note 1,2 | | | Intravascular(Cardiac) | | | | | | | | | | Trans-esoph.(Cardiac) | | | | | | | | | | Intra- cardiac | | | | | | | | | Peripheral<br>vascular | Peripheral vascular | N | N | N | | N | N | Note 1,2 | | | Other (Specify) | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix PDI= Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+PDVDPDI, B+Color+PW, B+PDV/DPDI +PW * Small Organ includes Thyroid, Testes, Breast ** Other use includes Urology, Kidney, Gynecology Note 1: Biopsy Guidance Note 2: Harmonic Imaging, This feature does not use contrast agent. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) (Division Sign Off) Division of Radiological Health Office of *In Vitro* Diagnostic and Radiological Health 510(k) K123249 {7}------------------------------------------------ ## U50 with C352UB Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Clinical Application | | Mode of Operation | | | | | | | |-----------------------------|----------------------|---------------------------------|-------------------|---|----|----|-------|-----------------------|--------------------| | | General | Specific | B | M | PW | CW | Color | Combined<br>(Specify) | Other<br>(Specify) | | Ophthalmic | Ophthalmic | Ophthalmic | | | | | | | | | Fetal<br>Imaging<br>& Other | | Fetal / Obstetrics | N | N | N | | N | N | Note 1,2 | | | | Abdominal | N | N | N | | N | N | Note 1,2 | | | | Intra-operative (Specify) | | | | | | | | | | | Intra-operative (Neuro logical) | | | | | | | | | | | Laparoscopic | | | | | | | | | | | Pediatric | | | | | | | | | | | Small Organ (Specify) * | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | Trans-rectal | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | Trans-urethral | | | | | | | | | | | Musculo-skeletal(Conventional) | | | | | | | | | | | Musculo-skeletal (Superficial) | | | | | | | | | | | Intravascular | | | | | | | | | | | Other (Specify) ** | N | N | N | | N | N | Note 1,2 | | Cardiac | | Adult Cardiac | | | | | | | | | Cardiac | | Pediatric Cardiac | | | | | | | | | | | Intravascular(Cardiac) | | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | | Intra- cardiac | | | | | | | | | Peripheral<br>vascular | | Peripheral vascular | | | | | | | | | vascular | | Other (Specify) | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix PDI=Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+Color+PW, B+Color+PW, B+PDV/DPDI +PW * Small Organ includes Thyroid, Testes, Breast ** Other use includes Urology, Kidney, Gynecology Note 1: Biopsy Guidance Note 2: Harmonic Imaging, This feature does not use contrast agent. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) 11 (División Sign Off) Division of Radiological Health Office of *In Vitro* Diagnostic and Radiological Health 510(k) K123249 {8}------------------------------------------------ ### U50 with L1042UB Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Clinical Application | | Mode of Operation | | | | | | | |-----------------------------|---------------------------------|---|-------------------|----|----|-------|-----------------------|--------------------|--| | General | Specific | B | M | PW | CW | Color | Combined<br>(Specify) | Other<br>(Specify) | | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal<br>Imaging<br>& Other | Fetal / Obstetrics | | | | | | | | | | | Abdominal | | | | | | | | | | | Intra-operative (Specify) | | | | | | | | | | | Intra-operative (Neuro logical) | | | | | | | | | | | Laparoscopic | | | | | | | | | | | Pediatric | | | | | | | | | | | Small Organ (Specify) * | N | N | N | | N | N | Note 1,2 | | | | Neonatal Cephalic | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | Trans-rectal | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | Trans-urethral | | | | | | | | | | | Musculo-skeletal(Conventional) | N | N | N | | N | N | Note 1,2 | | | | Musculo-skeletal (Superficial) | N | N | N | | N | N | Note 1,2 | | | | Intravascular | | | | | | | | | | | Other (Specify) ** | | | | | | | | | | Cardiac | Adult Cardiac | | | | | | | | | | | Pediatric Cardiac | | | | | | | | | | | Intravascular(Cardiac) | | | | | | | | | | | Trans-esoph.(Cardiac) | | | | | | | | | | | Intra- cardiac | | | | | | | | | | Peripheral<br>vascular | Peripheral vascular | N | N | N | | N | N | Note 1,2 | | | | Other (Specify) | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix PDI= Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+PDVDPDI, B+Color+PW, B+PDI/DPDI +PW * Small Organ includes Thyroid, Testes, Breast ** Other use includes Urology, Kidney, Gynecology_ ___________________________________________________________________________________________________________________________ Note 1: Biopsy Guidance Note 2: Harmonic Imaging, This feature does not use contrast agent. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) \$ (División Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health 510(k)________________________________________________________________________________________________________________________________________________________________________ {9}------------------------------------------------ ## U50 with L742UB Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Clinical Application | Mode of Operation | | | | | | | |------------|---------------------------------|-------------------|---|----|----|-------|-----------------------|--------------------| | General | Specific | B | M | PW | CW | Color | Combined<br>(Specify) | Other<br>(Specify) | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal / Obstetrics | | | | | | | | | | Abdominal | | | | | | | | | | Intra-operative (Specify) | | | | | | | | | | Intra-operative (Neuro logical) | | | | | | | | | | Laparoscopic | | | | | | | | | | Pediatric | | | | | | | | | | Small Organ (Specify) * | N | N | N | N | N | N | Note 1,2 | | Fetal | Neonatal Cephalic | | | | | | | | | Imaging | Adult Cephalic | | | | | | | | | & Other | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Musculo-skeletal(Conventional) | N | N | N | N | N | N | Note 1,2 | | | Musculo-skeletal (Superficial) | N | N | N | N | N | N | Note 1,2 | | | Intravascular | | | | | | | | | | Other (Specify) ** | | | | | | | | | | Adult Cardiac | | | | | | | | | | Pediatric Cardiac | | | | | | | | | Cardiac | Intravascular(Cardiac) | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | Intra- cardiac | | | | | | | | | Peripheral | Peripheral vascular | N | N | N | N | N | N | Note 1,2 | | vascular | Other (Specify) | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix PDI= Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+PDVDPDI, B+Color+PW, B+PDVDPDI +PW * Small Organ includes Thyroid, Testes, Breast ** Other use includes Urology, Kidney, Gynecology Note 1: Biopsy Guidance - Note 2: Harmonic Imaging, This feature does not use contrast agent. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) (Division Sign Off) Division of Radiological Health Office of *In Vitro* Diagnostic and Radiological Health 510(k) K123249 {10}------------------------------------------------ ## U50 with E612UB Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation | | | | | | | |-----------------------------|---------------------------------|---------------------|---|----|----|-------|--------------------|-----------------| | | | B | M | PW | CW | Color | Combined (Specify) | Other (Specify) | | General | Specific | | | | | | | | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal / Obstetrics | | | | | | | | | | Abdominal | | | | | | | | | | Intra-operative (Specify) | | | | | | | | | | Intra-operative (Neuro logical) | | | | | | | | | | Laparoscopic | | | | | | | | | | Pediatric | | | | | | | | | | Small Organ (Specify) * | | | | | | | | | Fetal<br>Imaging<br>& Other | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | | | | | | | | | | Trans-vaginal | N | N | N | | N | N | Note 1,2 | | | Trans-urethral | | | | | | | | | | Musculo-skeletal(Conventional) | | | | | | | | | | Musculo-skeletal (Superficial) | | | | | | | | | | Intravascular | | | | | | | | | | Other (Specify) ** | | | | | | | | | | Adult Cardiac | | | | | | | | | | Pediatric Cardiac | | | | | | | | | Cardiac | Intravascular(Cardiac) | | | | | | | | | | Trans-esoph.(Cardiac) | | | | | | | | | | Intra- cardiac | | | | | | | | | | Peripheral vascular | Peripheral vascular | | | | | | | | vascular | Other (Specify) | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix PDI= Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+Color+PW, B+Color+PW, B+PDVDPDI +PW * Small Organ includes Thyroid, Testes, Breast ** Other use includes Urology, Kidney, Gynecology Note 1: Biopsy Guidance Note 2: Harmonic Imaging, This feature does not use contrast agent. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) (Division Sign Off) Division of Radiological Health Office of *In Vitro* Diagnostic and Radiological Health 510(k) K123249 {11}------------------------------------------------ ### U50 with C612UB Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Clinical Application | | Mode of Operation | | | | | | | |--|-----------------------------|---------------------------------|-------------------|---|----|----|-------|-----------------------|--------------------| | | General | Specific | B | M | PW | CW | Color | Combined<br>(Specify) | Other<br>(Specify) | | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal<br>Imaging<br>& Other | Fetal / Obstetrics | | | | | | | | | | | Abdominal | | | | | | | | | | | Intra-operative (Specify) | | | | | | | | | | | Intra-operative (Neuro logical) | | | | | | | | | | | Laparoscopic | | | | | | | | | | | Pediatric | N | N | N | | N | N | Note 1,2 | | | | Small Organ (Specify) * | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | Trans-rectal | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | Trans-urethral | | | | | | | | | | | Musculo-skeletal(Conventional) | | | | | | | | | | | Musculo-skeletal (Superficial) | | | | | | | | | | | Intravascular | | | | | | | | | | | Other (Specify) ** | | | | | | | | | | Cardiac | Adult Cardiac | | | | | | | | | | | Pediatric Cardiac | N | N | N | | N | N | Note 1,2 | | | | Intravascular(Cardiac) | | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | | Intra- cardiac | | | | | | | | | | Peripheral<br>vascular | Peripheral vascular | | | | | | | | | | | Other (Specify) | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix PDI= Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+PDI/DPDI, B+Color+PW, B+PDI/DPDI +PW * Small Organ includes Thyroid, Testes, Breast ** Other use includes Urology, Kidney, Gynecology - Note 1: Biopsy Guidance Note 2: Harmonic Imaging, This feature does not use contrast agent. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Sz. D (Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health K/23249 510(k)________________________________________________________________________________________________________________________________________________________________________ 6-7 {12}------------------------------------------------ ## U50 with C6152UB Transducer · Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Clinical Application | Mode of Operation | | | | | | | |-----------------------------|---------------------------------|-------------------|---|----|----|-------|-----------------------|--------------------| | General | Specific | B | M | PW | CW | Color | Combined<br>(Specify) | Other<br>(Specify) | | | | | | | | | | | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal / Obstetrics | | | | | | | | | | Abdominal | | | | | | | | | | Intra-operative (Specify) | | | | | | | | | | Intra-operative (Neuro logical) | | | | | | | | | | Laparoscopic | | | | | | | | | | Pediatric | N | N | N | | N | N | Note 1,2 | | | Small Organ (Specify) * | | | | | | | | | Fetal<br>Imaging<br>& Other | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Musculo-skeletal(Conventional) | | | | | | | | | | Musculo-skeletal (Superficial) | | | | | | | | | | Intravascular | | | | | | | | | | Other (Specify) ** | | | | | | | | | | Adult Cardiac | | | | | | | | | | Pediatric Cardiac | N | N | N | | N | N | Note 1,2 | | Cardiac | Intravascular(Cardiac) | | | | | | | | | | Trans-esoph.(Cardiac) | | | | | | | | | | Intra- cardiac | | | | | | | | | Peripheral | Peripheral vascular | | | | | | | | | vascular | Other (Specify) | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix PDI=Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+PDV/DPDI, B+Color+PW, B+PDV/DPDI +PW * Small Organ includes Thyroid, Testes, Breast ** Other use includes Urology, Kidney, Gynecology Note 1: Biopsy Guidance Note 2: Harmonic Imaging, This feature does not use contrast agent. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) (Division Sign Off) Division of Radiological Health Office of *In Vitro* Diagnostic and Radiological Health 510(k) K123249 {13}------------------------------------------------ ## U50 with C422UB Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Clinical Application | | Mode of Operation | | | | | | | |-----------------------------|---------------------------------|---|-------------------|----|----|-------|-----------------------|--------------------|--| | General | Specific | B | M | PW | CW | Color | Combined<br>(Specify) | Other<br>(Specify) | | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal<br>Imaging<br>& Other | Fetal / Obstetrics | | | | | | | | | | | Abdominal | N | N | N | | N | N | Note 1,2 | | | | Intra-operative (Specify) | | | | | | | | | | | Intra-operative (Neuro logical) | | | | | | | | | | | Laparoscopic | | | | | | | | | | | Pediatric | | | | | | | | | | | Small Organ (Specify) * | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | Trans-rectal | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | Trans-urethral | | | | | | | | | | | Musculo-skeletal(Conventional) | | | | | | | | | | | Musculo-skeletal (Superficial) | | | | | | | | | | | Intravascular | | | | | | | | | | | Other (Specify) ** | | | | | | | | | | Cardiac | Adult Cardiac | N | N | N | | N | N | Note 1,2 | | | | Pediatric Cardiac | | | | | | | | | | | Intravascular(Cardiac) | | | | | | | | | | | Trans-esoph.(Cardiac) | | | | | | | | | | | Intra- cardiac | | | | | | | | | | Peripheral<br>vascular | Peripheral vascular | | | | | | | | | | | Other (Specify) | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix PDI=Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+PDI/DPDI, B+Color+PW, B+PDI/DPDI +PW * Small Organ includes Thyroid, Testes, Breast ** Other use includes Urology, Kidney, Gynecology Note 1: Biopsy Guidance Note 2: Harmonic Imaging, This feature does not use contrast agent. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) (Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health K123249 510(k) 6-9 {14}------------------------------------------------ ## U50 with L552UB Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Clinical Application | | Mode of Operation | | | | | | | |-----------------------------|----------------------|---------------------------------|-------------------|---|----|----|-------|--------------------|-----------------| | | General | Specific | B | M | PW | CW | Color | Combined (Specify) | Other (Specify) | | Ophthalmic | Ophthalmic | Ophthalmic | | | | | | | | | Fetal<br>Imaging<br>& Other | | Fetal / Obstetrics | | | | | | | | | | | Abdominal | | | | | | | | | | | Intra-operative (Specify) | | | | | | | | | | | Intra-operative (Neuro logical) | | | | | | | | | | | Laparoscopic | | | | | | | | | | | Pediatric | N | N | N | | N | N | Note 1,2 | | | | Small Organ (Specify) * | N | N | N | | N | N | Note 1,2 | | | | Neonatal Cephalic | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | Trans-rectal | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | Trans-urethral | | | | | | | | | | | Musculo-skeletal(Conventional) | N | N | N | | N | N | Note 1,2 | | | | Musculo-skeletal (Superficial) | N | N | N | | N | N | Note 1,2 | | | | Intravascular | | | | | | | | | | | Other (Specify) ** | | | | | | | | | Cardiac | | Adult Cardiac | | | | | | | | | | | Pediatric Cardiac | | | | | | | | | | | Intravascular(Cardiac) | | | | | | | | | | | Trans-esoph.(Cardiac) | | | | | | | | | | | Intra- cardiac | | | | | | | | | Peripheral<br>vascular | | Peripheral vascular | N | N | N | | N | N | Note 1,2 | | | | Other (Specify) | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix PDI= Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+PDVDPDI, B+Color+PW, B+PDVDPDI +PW * Small Organ includes Thyroid, Testes, Breast ** Other use includes Urology, Kidney, Gynecology Note 1: Biopsy Guidance Note 2: Harmonic Imaging, This feature does not use contrast agent. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) (Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health K123249 510(k)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...