← Product Code [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN) · K121699

# ACUSON X300 DIAGNOSTIC ULTRASOUND SYSTEM SONOVISTA X300 DIAGNOSTIC ULTRASOUND SYSTEM (K121699)

_Siemens Medical Solutions USA, Inc. · IYN · Jun 22, 2012 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K121699

## Device Facts

- **Applicant:** Siemens Medical Solutions USA, Inc.
- **Product Code:** [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN.md)
- **Decision Date:** Jun 22, 2012
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric, 3rd-Party Reviewed

## Intended Use

The Siemens Acuson X300 ultrasound imaging system is intended for the following applications: General Radiology, Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Neonatal/Adult Cephalic, Cardiac, Transesophageal, Pelvic, Transcranial, OB/GYN, Urology, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes. The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging".

## Device Story

Multi-purpose mobile diagnostic ultrasound system; acquires harmonic ultrasound echo data; displays in B-Mode, Pulsed (PW) Doppler, Continuous (CW) Doppler, Color Doppler, Amplitude Doppler, Harmonic Imaging, 3D/4D imaging. Operated by clinicians in clinical settings. Provides anatomical measurements and analysis packages (e.g., Arterial Health Package for Intima Media Thickness). Output displayed on flat panel; assists physicians in clinical diagnosis and patient communication regarding cardiovascular health. Benefits include non-invasive assessment of anatomical structures and vascular health.

## Clinical Evidence

Bench testing only. Evaluated for acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, electromagnetic, and mechanical safety. Conforms to UL 60601-1, IEC 60601-2-37, AIUM/NEMA UD-2/UD-3, and ISO 10993-1 standards.

## Technological Characteristics

Mobile, software-controlled ultrasound system. Supports various transducers (phased, convex, linear, TEE, intracardiac). Modes: B, M, PW/CW Doppler, Color/Amplitude Doppler, 3D/4D. Connectivity: Flat panel display. Safety: Thermal/mechanical indices display. Standards: IEC 60601-1, IEC 60601-2-37, ISO 10993-1.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- ACUSON X300 Diagnostic Ultrasound System ([K090276](/device/K090276.md), [K093531](/device/K093531.md))
- ACUSON SC2000 Diagnostic Ultrasound System ([K113179](/device/K113179.md))
- ZONARE z.one Ultra Ultrasound System ([K101091](/device/K101091.md))
- GE LOGIQ i, LOGIQ e, Vivid e ([K113690](/device/K113690.md))
- GE LogiqBooK XP ([K014206](/device/K014206.md))
- Sonosite M-turbo ([K101757](/device/K101757.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

ACUSON X300™ Diagnostic Ultrasound System Special 510(k) Submission

# SECTION 12 510(k) Summary

JUN 2 2 2012

- 11.1. Sponsor: Siemens Medical Solutions USA, Inc., Ultrasound Division 685 East Middlefield Road Mountain View, California 94043
11.2. Contact Person: Shelly Pearce Title: Regulatory Affairs Specialist (650) 694 5988 Telephone: (650) 694 5580 Fax:

11.3. Submission Date: May 15, 2012

- 11.4. Device Name: ACUSON X300™ Diagnostic Ultrasound System SONOVISTA X300 Diagnostic Ultrasound System ACUSON X300 ™ Diagnostic Ultrasound System, premium edition (K090276, K093531)
#### 11:5. Common Name: Diagnostic Ultrasound System with Accessories

11.6. Classification: Regulatory Class: ll Review Category: Tier II Radiology Classification Panel:

| Ultrasonic Pulsed Doppler Imaging System | FR # 892.1550 | Product Code 90-IYN |
|------------------------------------------|---------------|---------------------|
| Ultrasonic Pulsed Echo Imaging System    | FR # 892.1560 | Product Code 90-IYO |
| Diagnostic Ultrasound Transducer         | FR # 892.1570 | Product Code 90-ITX |
| Diagnostic Intravascular Catheter        | FR # 870.1200 | Product Code 74-DQC |

A. Legally Marketed Predicate Devices

The Siemens ACUSON X300 Diagnostic Ultrasound System is a multi-purpose diagnostic ultrasound system with accessories and proprietary software, and is substantially equivalent to the current product that is already cleared for USA distribution under the following 510(k) PreMarket Notification number:

- K090276. K093531. ACUSON X300™ Diagnostic Ultrasound System, SONOVISTA . X300 Diagnostic Ultrasound System, ACUSON X300™ Diagnostic Ultrasound System, premium edition
- K113179, ACUSON SC2000™ Diagnostic Ultrasound System .
- K101091, ZONARE z.one Ultra Ultrasoud System .
- K113690, GE LOGIQ i, LOGIQ e, Vivid e .
- K014206, GE LogiqBooK XP .
- . K101757, Sonosite M-turbo

### B. Device Description:

The Siemens ACUSON X300 Diagnostic Ultrasound System is a multi-purpose mobile, software controlled, diagnostic ultrasound system with an on-screen display for thermal and mechanical

{1}------------------------------------------------

indices related to potential bio-effect mechanisms. Its function is to acquire primary or secondary harmonic ultrasound echo data and display it in B-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppier Mode, Amplitude Doppler Mode, a combination of modes, or Harmonic Imaging and 3D Imaging, or Harmonic Imaging and 4D imaging on a Flat Panel Display.

#### C. Intended Use

The Siemens Acuson X300 ultrasound imaging system is intended for the following applications: General Radiology, Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Neonatal/Adult Cephalic, Cardiac, Transesophageal, Pelvic, Transcranial, OB/GYN, Urology, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.

The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.

The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging".

#### D. Substantial Equivalence

The submission device is substantially equivalent to the predicate with regard to both intended use and technological characteristics.

#### E. Performance Data

The modifications to the Siemens ACUSON X300 Diagnostic Ultrasound System are verified and validated according to the company's design control process.

F. A brief discussion of nonclinical tests submitted, referenced, or relied on in

the 510(k) for a determination of substantial equivalence.

The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety and has been found to conform with applicable medical device safety standards. The system complies with the following voluntary standards:

UL 60601-1, Safety Requirements for Medical Equipment

IEC 60601-2-37 Diagnostic Ultrasound Safety Standards

CSA C22.2 No. 601-1, Safety Requirements for Medical Equipment

AIUM/NEMA UD-3, Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment

AIUM/NEMA UD-2, Acoustic Output Measurement Standard for Diagnostic Ultrasound

93/42/EEC Medical Devices Directive

Safety and EMC Requirements for Medical Equipment

EN/IEC 60601-1

{2}------------------------------------------------

EN/IEC 60601-1-1

EN/IEC 60601-1-2

IEC 1157 Declaration of Acoustic Power

ISO 10993-1 Biocompatibility

Cleared patient contact materials, electrical and mechanical safety are unchanged.

{3}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized emblem. The emblem consists of three abstract human profiles facing to the right, stacked on top of each other. The profiles are connected by a flowing line, creating a sense of unity and movement.

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

JUN 2 2 2012

Siemens Medical Solutions USA, Inc. Ultrasound Group % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

Re: K121699

K121077
Trade/Device Name: ACUSON X300™ Diagnostic Ultrasound System; SONOVISTA X 300 Diagnostic Ultrasound Systems; ACUSON X300™ Diagnostic Ultrasound System, premium edition

Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX, and DQO Dated: June 7, 2012 Received: June 8, 2012

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced we nave leviewed your beenou broke is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to the enclosure) to regally mankelou production Device Amendments, or to devices that have been May 26, 1970, the Chacement and of the Federal Food, Drug, and Cosmetic Act (Act). You reclassified in accordance with the provisions to the general controls provisions of the Act. The general may, therefore, thanket the device, basyon is month registration, listing of devices, good Controls provisions of the beling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with This determination of Salostantial System, SONOVISTA X 300 Diagnostic Ultrasound the ACUSON X 300 - Diagnostic Ultrasound System, premium edition, as described in your premarket notification:

### Transducer Model Number

P4-2 Phased Sector Array CH5-2 Convex Array VF10-5 Linear Array L9-5 Linear Array EC9-4 Convex Array EV9-4 Convex Array VF13-5 Linear Array P8-4 Phased Sector Array

BE 9-4 Convex Array CW2 Continuous Wave Doppler CW5 Continuous Wave Doppler AcuNav 8F Intracardiac AcuNav 10F Intracardiac V5Ms TEE 4V1c Phased Sector Array VF13-5SP Linear Array

{4}------------------------------------------------

C8-5 Tightly Curved Array 8L3 Linear "Regel" 10V4 Phased Array C7F2 Curved Array Mechanical 3D/4D EV9F4 Curved Array Mechanical 3D/4D L13F5 3D/4D Mechanical Wobbler Linear VF8-3 Linear Array P5-1 Phased Array C6-2 Convex Array SoundStar 10F P9-4 Phased Array

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Subled of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any I DA has made a deceminations administered by other Federal agencies. You must comply with all the I cocrar statutes and regulations as annot limited to: registration and listing (21 CFR Part 807); labeling (21 Act 3 requirements, morable wing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your premarket notification. This FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to 11 you desire specific advice for your er offices/CDRH/CDRHOffices/ucm11.5809.htm for the Center for Intip://www.ida.gov/Abouti Drivellers/Office of Compliance. Also, please note the regulation Devices and Radiological reallir o (SEPT Part 807.97). For questions regarding the reporting of referents in promote the MDR regulation (21 CFR Part 803), please go to regarding the reporting of daverse overse w/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

If you have any questions regarding the content of this letter, please contact Joshua Nipper at (301) 796-6524.

Sincerely Yours,

Janine M. Morris Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure(s)

{5}------------------------------------------------

## SECTION 6 Indications for Use Forms

### Indications for Use Statement

510(k) Number (if known):

Device Name:

ACUSON X300TM Diagnostic Ultrasound System SONOVISTA X300 Diagnostic Ultrasound System ACUSON X300™ Diagnostic Ultrasound System, premium edition

Indications For Use:

The Siemens Acuson X300 ultrasound imaging system is intended for the following applications: General Radiology, Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Neonatal/Adult Cephalic, Cardiac, Transesophageal, Pelvic, Transcranial, OB/GYN, Urology, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.

The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.

The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging".

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off Office of In Vitro Diagnostic Devices

510(k)K121699

Page 1 of

{6}------------------------------------------------

510(k) Number (if known):

Device Name: Intended Use:

ACUSON X300 Diagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows:

| Clinical Application        |                               | Mode of Operation |   |   |     |     |                  |                      |                              |                       |                          |
|-----------------------------|-------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------------|
| Other<br>(Track1<br>Only)   | Specific<br>(Tracks 1& 3)     | A                 | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify)       |
| Ophthalmic                  | Ophthalmic                    |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Fetal                         | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Abdominal                     | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Intra-operative<br>(Note 6)   | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Intra-operative<br>(Neuro)    | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Laparoscopic                  | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
| Fetal<br>Imaging<br>& Other | Pediatric                     | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | SmallOrgan<br>(Note 1)        | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Neonatal Cephalic             | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Adult Cephalic                | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2.3.4.5.7.8.9,11 |
|                             | Trans-rectal                  | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Trans-vaginal                 | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Trans-urethral                |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Trans-esoph.<br>(non-Card.)   | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Musculo-skel.<br>(Convent.)   | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Musculo-skel.<br>(Superfic)   | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Intra -vascular               | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Other<br>(Specify)            | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2.3.4.5.7.8.9.11 |
|                             | Cardiac Adult                 | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
| Cardiac                     | Cardiac Pediatric             | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Intra-vascular<br>(Cardiac)   | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Trans-esophageal<br>(Cardiac) | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2.3,4.5,7.8,9,11 |
|                             | Intra-Cardiac                 | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7.8,9,11 |
|                             | Other (Specify)               | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
| Peripheral                  | Peripheral vessel             | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
| Vessel                      | Other (Specify)               | P                 | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |

N = new indication; P = previously cleared by K080760; K090276

For example: breast, testes, thyroid, penis, prostate, etc. Note 1

Note 3 3D imaging

3D imaging
Power SieScape panoramic imaging Note 5

Tissue Equalization Technology Note 9

Dynamic TCE Note 11

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Note 2

Note 4

Note 6

Note 8

Note 10

Ensemble tissue harmonic imaging

B&W SieScape panoramic imaging

For example: abdominal, vascular

Intracardiac imaging

SieClear multi-view spatial compounding

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off - Office of In Vitro Diagnostic Devices
K121699

Page 2 of

Contrast agent imaging Note 7

{7}------------------------------------------------

### ACUSON X300™ Diagnostic Ultrasound System Special 510(k) Submission

#### Diagnostic Ultrasound Indications for Use Form

510(k) Number (if known):

Device Name: Intended Use:

P4-2 Phased Sector Array Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows:

| Clinical Application        | Mode of Operation             |   |   |   |     |     |                  |                      |                              |                       |                          |
|-----------------------------|-------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------------|
| Other<br>( Track 1<br>Only) | Specific<br>(Tracks11& 3)     | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify)       |
| Ophthalmic                  | Ophthalmic                    |   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Fetal                         |   | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Abdominal                     |   | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,5,6,7,8,9,11 |
|                             | Intra-operative<br>(Note 6)   |   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Intra-operative<br>(Neuro)    |   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Laparoscopic                  |   |   |   |     |     |                  |                      |                              |                       |                          |
| Fetal<br>Imaging<br>& Other | Pediatric                     |   | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | SmallOrgan<br>(Note 1)        |   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Neonatal Cephalic             |   | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Adult Cephalic                |   | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Trans-rectal                  |   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Trans-vaginal                 |   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Trans-urethral                |   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Trans-esoph.<br>(non-Card.)   |   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Musculo-skel.<br>(Convent.)   |   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Musculo-skel.<br>(Superfic)   |   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Intra-vascular                |   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Other<br>(Specify)            |   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Cardiac Adult                 |   | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2.3.4.5.7.8,9.11 |
| Cardiac                     | Cardiac Pediatric             |   | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Intra-vascular<br>(Cardiac)   |   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Trans-esophageal<br>(Cardiac) |   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Intra-cardiac                 |   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Other (Specify)               |   |   |   |     |     |                  |                      |                              |                       |                          |
| Peripheral                  | Peripheral vessel             |   | P | P | P   | P   | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
| Vessel                      | Other (Specify)               |   |   |   |     |     |                  |                      |                              |                       |                          |

N = new indication; P = previously cleared by K080760

| Note 1 | For example: breast, testes, thyroid, penis, prostate, etc. |
|--------|-------------------------------------------------------------|
|--------|-------------------------------------------------------------|

- Note 3 3D imaging
Power SieScape panoramic imaging Note 5 Contrast agent imaging Note 7

- Tissue Equalization Technology Note 9
- Note 11 Dynamic TCE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Note 2 Note 4

Note 6

Note 8

Note 10

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off - Office/of In Vitro Diagnostic Devices
KlAl699

Page 3 of

Ensemble tissue harmonic imaging

B&W SieScape panoramic imaging

For example: abdominal, vascular

Intracardiac imaging

{8}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form

510(k) Number (if known):

Device Name:

CH5-2 Convex Array Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows:

Intended Use:

Clinical Application Mode of Operation Color Other Color Amplitude Cambined Other Specific PWD CWD B M Velocity ( Track1 A Doppler Doppler (Specify) (Specify) (Tracks11& 3) Imaging Only) Ophthalmic Ophthalmic Note P P BMDC P P P Fetal 2.3.4.5.7,8.9.11 Note BMDC P P P P P Abdominal 2,3,4,5,7,8,9,11 Intra-operative (Note G) Intra-operative (Neuro) Laparoscopic Note P BMDC P P P P Pediatric Fetal 2,3,4,5,7,8,9,11 SmallOrgan Imaging & Other (Note 1) Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph. (non-Card.) Note Musculo-skel. P P P P P BMDC 2.3.4.5.7.8.9.11 . (Convent.) Musculo-skel. (Superfic) Intra-vascular Qther (Specify) Note P P BMDC P P P Cardiac Adult 2,3.4,5,7,8,9,11 Note P p P P BMDC P Cardiac Cardiac Pediatric 2,3,4,5,7,8,9,11 Intra-vascular (Carliac) Trans-esophageal (Cardiac) Intra-cardiac Other (Specify) Note P BMDC P P P Peripheral vessel P . 2,3,4,5,7,8,9,11 Peripheral Vessel Other (Specify)

N = new indication; P = previously cleared by K080760

For example: breast, testes, thyroid, penis, prostate, etc. Note 1

- Note 3 3D imaging
- Note 5 Power SieScape panoramic imaging
- Note 7 Contrast agent imaging
- Tissue Equalization Technology Note 9

Note 11 Dynamic TCE

(FLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off - Office of In Vitro Diagnostic Devices

510(k) K121699

Page 4 of

- Note 4 Note 6
Note 2

- For example: abdominal, vascular
- SieClear multi-view spatial compounding Note 8
- Note 10 Intracardiac imaging

Ensemble tissue harmonic imaging B&W SieScape panoramic imaging

{9}------------------------------------------------

510(k) Number (if known):

Device Name:

VF10-5 Linear Array Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems

Intended Use:

Diagnostic imaging or fluid flow analysis of the human body as follows:

| Clinical Application      | Mode of Operation             |   |   |   |     |     |                  |                      |                              |                       |                          |
|---------------------------|-------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------------|
| Other<br>(Track1<br>Only) | Specific<br>(Tracks1& 3)      | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify)       |
| Ophthalmic                | Ophthalmic                    |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Fetal                         |   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                           | Abdominal                     |   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                           | Intra-operative<br>(Note 6)   |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Intra-operative<br>(Neuro)    |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Laparoscopic                  |   |   |   |     |     |                  |                      |                              |                       |                          |
| Fetal                     | Pediatric                     |   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9.11 |
| Imaging<br>& Other        | SmallOrgan<br>(Note 1)        |   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                           | Neonatal Cephalic             |   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                           | Adult Cephalic                |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Trans-rectai                  |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Trans-vaginal                 |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Trans-urethral                |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Trans-esoph.<br>(non-Card.)   |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Musculo-skel.<br>(Convent.)   |   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                           | Musculo-skei.<br>(Superfic)   |   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                           | Intra-vascular                |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Other<br>(Specify)            |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Cardiac Adult                 |   |   |   |     |     |                  |                      |                              |                       |                          |
| Cardiac                   | Cardiac Pediatric             |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Intra-vascular<br>(Cardiac)   |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Trans-esophageal<br>(Cardiac) |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Intra-cardiac                 |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Other (Specify)               |   |   |   |     |     |                  |                      |                              |                       |                          |
| Peripheral                | Peripheral vessel             |   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9.11 |
| Vessel                    | Other (Specify)               |   |   |   |     |     |                  |                      |                              |                       |                          |

N = new indication; P = previously cleared by K080760

- For example: breast, testes, thyroid, penis, prostate, etc. Note 2 Note 1
- 3D imaaina Note 3
- Power SieScape panoramic imaging Note 5
- Contrast agent imaging Note 7
- Tissue Equalization Technology Note 9
- Note 11 Dynamic TCE

(FLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Note 4

Note 6

Note 8

Note 10

Division Sign-Off - Office of In-Vitro Diagnostic Devices

Kla1699
510(k)

Page 5 of

Ensemble tissue harmonic imaging

B&W SieScape panoramic imaging
For example: abdominal, vascular

Intracardiac imaging

{10}------------------------------------------------

#### Diagnostic Ultrasound Indications for Use Form

510(k) Number (if known):

Device Name:

L9-5 Linear Array Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows:

Intended Use:

Clinical Application Mode of Operation Other Color Specific Color Amplitude Combined Other A B M PWD CWD Velocity ( Track 1 (Tracks11& 3) Doppler Doppler (Specify) (Specify) Only) Imaging Ophthalmic Ophthalmic Note P P P P P BMDC Fetal 2,3,4,5,7,8,9,11 Note P P P P P BMDC Abdominal 2,3,4,5,7,8,9,11 Intra-operative (Note 6) Intra-operative (Neuro) Laparoscopic Note P P P BMDC Fetal Pediatric P P 2.3.4.5.7.8.9,11 · SmallOrgan Imaging Note P P P P P BMOC 2,3,4,5,7,8,9,11 & Other (Note 1) Note P Neonatal Cephalic P P P p BMDC 2,3,4,5,7,8,9,11 Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph. (non-Card.) Musculo-skel. Note P P P P P BMDC 2,3,4,5,7,8,9,11 (Convent.) Musculo-skel. Note P P P P BMDC P 2,3,4,5,7,8,9,11 (Superfic) Intra-vascular Other (Specify) Cardiac Adult Cardiac Cardiac Pediatric Intra-vascular (Cardiac) Trans-esophageal (Cardiac) Intra-Cardaic · Other (Specify) Note P P P BMDC Peripheral vessel P P Peripheral 2.3.4.5.7.8.9.11

Other (Specify) N = new indication; P = previously cleared by K080760

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

- 3D imaging Note 3
- Power SieScape panoramic imaging Note 5 Contrast agent imaging

Note 7

Vessel

Tissue Equalization Technology Note 9 Dynamic TCE Note 11

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Note 2

Note 4

Note 6

Note 8

Note 10

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off - Office of In Vitro Diagnostic Devices
510(k) K121699

Ensemble tissue harmonic imaging

B&W SieScape panoramic imaging

For example: abdominal, vascular

Intracardiac imaging

{11}------------------------------------------------

### ACUSON X300™ Diagnostic Ultrasound System Special 510(k) Submission

#### Diagnostic Ultrasound Indications for Use Form

510(k) Number (if known):

Device Name: Intended Use: EC9-4 Convex Array Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows:

Diagnostic imaging or fluid flow analysis o

| Clinical Application      | Mode of Operation             |   |   |   |     |     |                  |                      |                              |                       |                          |
|---------------------------|-------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------------|
| Other<br>(Track1<br>Only) | Specific<br>(Tracks 11& 3)    | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify)       |
| Ophthalmic                | Ophthalmic                    |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Fetal                         | P | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                           | Abdominal                     |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Intra-operative<br>(Note 6)   |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Intra-operative<br>(Neuro)    |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Laparoscopic                  |   |   |   |     |     |                  |                      |                              |                       |                          |
| Fetal                     | Pediatric                     |   |   |   |     |     |                  |                      |                              |                       |                          |
| Imaging<br>& Other        | SmallOrgan<br>(Note 1)        |   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                           | Neonatal Cephalic             |   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2.3.4.5,7,8,9,11 |
|                           | Adult Cephalic                |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Trans-rectal                  |   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                           | Trans-vaginal                 |   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8.9,11 |
|                           | Trans-urethral                |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Trans-esoph.<br>(non-Card.)   |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Musculo-skel.<br>(Convent.)   |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Musculo-skel.<br>(Superfic)   |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Intra-vascular                |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Other<br>(Specify)            |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Cardiac Adult                 |   |   |   |     |     |                  |                      |                              |                       |                          |
| Cardiac                   | Cardiac Pediatric             |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Intra-vascular<br>(Cardiac)   |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Trans-esophageal<br>(Cardiac) |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Intra-Cardiac                 |   |   |   |     |     |                  |                      |                              |                       |                          |
|                           | Other (Specify)               |   |   |   |     |     |                  |                      |                              |                       |                          |
| Peripheral                | Peripheral vessel             |   |   |   |     |     |                  |                      |                              |                       |                          |
| Vessel                    | Other (Specify)               |   |   |   |     |     |                  |                      |                              |                       |                          |

N = new indication; P = previously cleared by K080760

For example: breast, testes, thyroid, penis, prostate, etc. Note 1

- 3D imaging Note 3
- Power SieScape panoramic imaging Note 5
- Note 7 Contrast agent imaging
- Tissue Equalization Technology Note 9
- Note 11 Dynamic TCE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Note 2

Note 4

Note G

Note 8

Nota 10

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off - Office of In Vitro Diagnostic Devices

510(k) K121699

Page 7 of

Ensemble tissue harmonic imaging

B&W SieScape panoramic imaging

For example: abdominal, vascular

Intracardiac imaging

{12}------------------------------------------------

510(k) Number (if known):

Device Name:

EV9-4 Convex Array Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems

Intended Use:

Diagnostic imaging or fluid flow analysis of the human body as follows:

| Clinical Application       |                               | Mode of Operation |   |   |     |     |                  |                      |                              |                       |                          |
|----------------------------|-------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------------|
| Other<br>(Track 1<br>Only) | Specific<br>(Tracks11& 3)     | A                 | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify)       |
| Ophthalmic                 | Ophthalmic                    |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                            | Fetal                         |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                            | Abdominal                     |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                            | Intra-operative<br>(Note 6)   |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                            | Intra-operative<br>(Neuro)    |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                            | Laparoscopic                  |                   |   |   |     |     |                  |                      |                              |                       |                          |
| Fetal                      | Pediatric                     |                   |   |   |     |     |                  |                      |                              |                       |                          |
| Imaging<br>& Other         | SmallOrgan<br>(Note 1)        |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                            | Neonatal Cephalic             |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                            | Adult Cephalic                |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                            | Trans-rectal                  |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                            | Trans-vaginal                 |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                            | Trans-urethral                |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                            | Trans-esoph.<br>(non-Card.)   |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                            | Musculo-skel.<br>(Convent)    |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                            | Musculo-skel.<br>(Superfic)   |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                            | Intra-vascular                |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                            | Other<br>(Specify)            |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                            | Cardiac Adult                 |                   |   |   |     |     |                  |                      |                              |                       |                          |
| Cardiac                    | Cardiac Pediatric             |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                            | Intra-vascular<br>(Cardiac)   |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                            | Trans-esophageal<br>(Cardiac) |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                            | intra-Cardiac                 |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                            | Other (Specify)               |                   |   |   |     |     |                  |                      |                              |                       |                          |
| Peripheral                 | Peripheral vessel             |                   |   |   |     |     |                  |                      |                              |                       |                          |
| Vessel                     | Other (Specify)               |                   |   |   |     |     |                  |                      |                              |                       |                          |

N = new indication; P = previously cleared by K080760

For example: breast, testes, thyroid, penis, prostate, etc. Note 1

- 3D imaging Note 3
- Power SieScape panoramic imaging Note 5
- Contrast agent imaging
Tissue Equalization Technology Note 7
- Note 9
- Dynamic TCE Note 11
	- (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

oncurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Note 2

Note 4

Note 6

Note 8

Note 10

Division Sign-Off - Office of In Vitro Diagnostic Devices
510(k) K121699

Page 8 of

Ensemble tissue hamonic imaging

B&W SieScape panoramic imaging

For example: abdominal, vascular

Intracardiac imaging

{13}------------------------------------------------

510(k) Number (if known):

Device Name:

VF13-5 Linear Array Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems

Intended Use:

Diagnostic imaging or fluid flow analysis of the human body as follows:

| Clinical Application        |                               |  | Mode of Operation |   |   |     |     |                  |                      |                              |                       |                          |
|-----------------------------|-------------------------------|--|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------------|
| Other<br>(Track 1<br>Only)  | Specific<br>(Tracks1 & 3)     |  | A                 | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify)       |
| Ophthalmic                  | Ophthalmic                    |  |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Fetal                         |  |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Abdominal                     |  |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Intra-operative<br>(Note 6)   |  |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Intra-operative<br>(Neuro)    |  |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Laparoscopic                  |  |                   |   |   |     |     |                  |                      |                              |                       |                          |
| Fetal<br>Imaging<br>& Other | Pediatric                     |  |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | SmallOrgan<br>(Note 1)        |  |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Neonatal Cephalic             |  |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Adult Cephalic                |  |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Trans-rectal                  |  |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Trans-vaginal                 |  |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Trans-urethral                |  |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Trans-esoph,<br>(non-Card.)   |  |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Musculo-skel.<br>(Convent)    |  |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Musculo-skel.<br>(Superfic)   |  |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Intra-vascular                |  |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Other<br>(Specify)            |  |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Cardiac Adult                 |  |                   |   |   |     |     |                  |                      |                              |                       |                          |
| Cardiac                     | Cardiac Pediatric             |  |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Intra-vascular<br>(Cardiac)   |  |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Trans-esophageal<br>(Cardiac) |  |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Intra-cardiac                 |  |                   |   |   |     |     |                  |                      |                              |                       |                          |
|                             | Other (Specify)               |  |                   |   |   |     |     |                  |                      |                              |                       |                          |
| Peripheral<br>Vessel        | Peripheral vessel             |  | P                 | P | P | P   |     | P                | P                    |                              | BMDC                  | Note<br>2,3,4,5,7,8,9,11 |
|                             | Other (Specify)               |  |                   |   |   |     |     |                  |                      |                              |                       |                          |

N = new indication; P = previously cleared by K080760

For example: breast, testes, thyroid, penis, prostate, etc. Note 1

- Note 3 3D imaging
- Power SieScape panoramic imaging Note 5
- Contrast agent imaging Note 7

Tissue Equalization Technology Note 9

- Dynamic TCE Note 11
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Note 2

Note 4

Note 6

Note 8

Note 10

Ensemble tissue harmonic imaging

B&W SieScape panoramic imaging

For example: abdominal, vascular

Intracardiac imaging

Division Sign-Off - Office of in Vitro Diagnostic Devices

510(k) K121699

{14}------------------------------------------------

510(k) Number (if known):

Device Name:

P8-4 Phased Sector Array Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems

Intended Use:

Other

( Track 1

Only) Ophthalmic

Diagnostic imaging or fluid flow analysis of the human body as follows: Clinical Application Mode of Operation Calor Specific Color Amplitude Combined Other CWD Velocity B M PWD A (Specify) (Tracks11& 3) Doppler Doppler (Specify) Imaging Ophthalmic

|                    | Fetal                         | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 |
|--------------------|-------------------------------|---|---|---|---|---|---|------|--------------------------|
|                    | Abdominal                     | P | P | P | P | P | P | BMDC | Note<br>2,3,5,6,7,8,9,11 |
|                    | Intra-operative<br>(Note 6)   |   |   |   |   |   |   |      |                          |
|                    | Intra-operative<br>(Neuro)    |   |   |   |   |   |   |      |                          |
|                    | Laparoscopic                  |   |   |   |   |   |   |      |                          |
| Fetal              | Pediatric                     | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 |
| Imaging<br>& Other | SmallOrgan<br>(Note 1)        |   |   |   |   |   |   |      |                          |
|                    | Neonatal Cephalic             | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 |
|                    | Adult Cephalic                |   |   |   |   |   |   |      |                          |
|                    | Trans-rectal                  |   |   |   |   |   |   |      |                          |
|                    | Trans-vaginal                 |   |   |   |   |   |   |      |                          |
|                    | Trans-urethral                |   |   |   |   |   |   |      |                          |
|                    | Trans-esoph.<br>(non-Card.)   |   |   |   |   |   |   |      |                          |
|                    | Musculo-skel.<br>(Convent)    | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 |
|                    | Musculo-skel.<br>(Superfic)   |   |   |   |   |   |   |      |                          |
|                    | Intra-vascular                |   |   |   |   |   |   |      |                          |
|                    | Other<br>(Specify)            |   |   |   |   |   |   |      |                          |
|                    | Cardiac Adult                 | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 |
| Cardiac            | Cardiac Pediatric             | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 |
|                    | Intra-vascular<br>(Cardiac)   |   |   |   |   |   |   |      |                          |
|                    | Trans-esophageal<br>(Cardiac) |   |   |   |   |   |   |      |                          |
|                    | Intra-cardiac                 |   |   |   |   |   |   |      |                          |
|                    | Other (Specify)               |   |   |   |   |   |   |      |                          |
| Peripheral         | Peripheral vessel             | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 |
| Vessel             | Other (Specify)               |   |   |   |   |   |   |      |                          |

N = new indication; P = previously cleared by K080760

| Note 1 | For example: breast, testes, thyroid, penis, prostate, etc |
|--------|------------------------------------------------------------|
|--------|------------------------------------------------------------|

- Note 3 3D imaging
- Power SieScape panoramic imaging Note 5

Contrast agent imaging Note 7 Tissue Equalization Technology Note 9

- Dynamic TCE Note 11
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Note 2

Note 4

Note 6

Note 8

Note 10

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off - Office of In Vitro Diagnostic Devices
510(k) K121699

Page 10 of _

Ensemble tissue harmonic imaging

B&W SieScape panoramic imaging

SieClear multi-view spatial compounding

For example: abdominal, vascular

{15}------------------------------------------------

510(k) Number (if known):

BE 9-4 Convex Array Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems

Device Name: Intended Use:

Diagnostic imaging or fluid flow analysis of the human body as follows:

| Clinical Application       |                                                       | Mode of Operation |   |   |     |     |                  |                      |                              |                       |                    |
|----------------------------|-------------------------------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Other<br>(Track 1<br>Only) | Specific<br>(Tracks1& 3)                              | A                 | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic                 | Ophthalmic                                            |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                            | Fetal                                                 |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,9 |
|                            | Abdominal                                             |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                            | Intra-operative<br>(Note 6)                           |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                            | Intra-operative<br>(Neuro)                            |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                            | Laparoscopic                                          |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Fetal                      | Pediatric                                             |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Imaging<br>& Other         | SmallOrgan<br>(Note 1)                                |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,9 |
|                            | Neonatal Cephalic                                     |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,9 |
|                            | Adult Cephalic                                        |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                            | Trans-rectal                                          |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,9 |
|                            | Trans-vaginal                                         |                   | P | P | P   |     | P                | P                    |                              | BMDC                  | Note 2,3,4,5,7,8,9 |
|                            | Trans-urethral                                        |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                            | Trans-esoph.<br>(non-Card.)                           |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                            | Musculo-skel.<br>(Convent)                            |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                            | Musculo-skel.<br>(Superfic)                           |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                            | Intra-vascular                                        |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                            | Other<br>(Specify)                                    |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                            | Cardiac Adult                                         |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Cardiac                    | Cardiac Pediatric                                     |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                            | Intra-vascular<br>(Cardiac)                           |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                            | Trans-esophageal<br>(Cardiac)                         |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                            | Intra-cardiac                                         |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                            | Other (Specify)                                       |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Peripheral                 | Peripheral vessel                                     |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Vessel                     | Other (Specify)                                       |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                            | N = new indication; P = previously cleared by K080760 |                   |   |   |     |     |                  |                      |                              |                       |                    |

For example: breast, testes, thyroid, penis, prostate, etc. Note 1

- 3D imaging Note 3
- Power SieScape panoramic imaging Note 5

Note 7 Contrast agent imaging

Tissue Equalization Technology Note 9

- Note 11 Dynamic TCE
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Note 2

Note 4

Note 6

Note 8

Note 10

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off Office of In Vitro Olagnostic Devices

510(k) K121699

Page 11 of

Ensemble tissue harmonic imaging

B&W SieScape panoramic imaging
 For example: abdominal, vascular

Intracardiac imaging

{16}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form

510(k) Number (if known):

Device Name:

CW2 Continuous Wave Doppler Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems

Intended Use:

Diagnostic imaging or fluid flow analysis of the human body as follows:

| Clinical Application        |                               | Mode of Operation |   |   |     |     |                  |                      |                              |                       |                    |
|-----------------------------|-------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Other<br>( Track 1<br>Only) | Specific<br>(Tracks11& 3)     | A                 | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic                  | Ophthalmic                    |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                             | Fetal                         |                   |   |   |     | P   |                  |                      |                              |                       |                    |
|                             | Abdominal                     |                   |   |   |     | P   |                  |                      |                              |                       |                    |
|                             | Intra-operative<br>(Note 6)   |                   |   |   |     | P   |                  |                      |                              |                       |                    |
|                             | Intra-operative<br>(Neuro)    |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                             | Laparoscopic                  |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Fetal                       | Pediatric                     |                   |   |   |     | P   |                  |                      |                              |                       |                    |
| Imaging<br>& Other          | SmallOrgan<br>(Note 1)        |                   |   |   |     | P   |                  |                      |                              |                       |                    |
|                             | Neonatal Cephalic             |                   |   |   |     | P   |                  |                      |                              |                       |                    |
|                             | Adult Cephalic                |                   |   |   |     | P   |                  |                      |                              |                       |                    |
|                             | Trans-rectal                  |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                             | Trans-vaginal                 |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                             | Trans-urethral                |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                             | Trans-esoph.<br>(non-Card.)   |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                             | Musculo-skel.<br>(Convent.)   |                   |   |   |     | P   |                  |                      |                              |                       |                    |
|                             | Musculo-skel.<br>(Superfic)   |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                             | Intra-vascular                |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                             | Other<br>(Specify)            |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                             | Cardiac Adult                 |                   |   |   |     | P   |                  |                      |                              |                       |                    |
| Cardiac                     | Cardiac Pediatric             |                   |   |   |     | P   |                  |                      |                              |                       |                    |
|                             | Intra-vascular<br>(Cardiac)   |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                             | Trans-esophageal<br>(Cardiac) |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                             | Intra-cardiac                 |                   |   |   |     |     |                  |                      |                              |                       |                    |
|                             | Other (Specify)               |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Perpheral                   | Peripheral vessel             |                   |   |   |     | P   |                  |                      |                              |                       |                    |
| Vessel                      | Other (Specify)               |                   |   |   |     |     |                  |                      |…

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K121699](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K121699)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
