Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal (Obstetrics); Abdominal (including renal and GYN/pelvic); Pediatric; Small Organ (breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional and Superficial; Transrectal (TR); Transvaginal (TV).
Device Story
Mobile ultrasound console with keyboard, controls, LCD display, and electronic-array transducers. Performs real-time digital acquisition, processing, and display of ultrasound signals. Used in clinical settings by healthcare professionals for general radiology, cardiology, and vascular imaging. Provides B-mode, M-mode, PW/CW Doppler, Color/Power Doppler, and 3D/4D imaging. Output displayed on screen for diagnostic evaluation, biopsy guidance, and follicle monitoring. Benefits include non-invasive visualization of internal anatomy and blood flow to support clinical decision-making.
Clinical Evidence
No clinical studies were required to support substantial equivalence. The device was evaluated through bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, electromagnetic, and mechanical safety testing. Quality assurance measures included risk analysis, design reviews, and comprehensive system verification and validation.
Technological Characteristics
Mobile console with electronic-array transducers. Modalities: B, M, PW/CW Doppler, Color/Power Doppler, 3D/4D, Harmonic, Coded Pulse. Connectivity: Digital acquisition/display. Biocompatible patient-contact materials. Safety: Complies with applicable medical device safety standards for acoustic output, electrical, and mechanical safety.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in fetal, abdominal, pediatric, small organ, neonatal/adult cephalic, cardiac, peripheral vascular, musculoskeletal, transrectal, and transvaginal applications. Patient population includes pediatric, neonatal, and adult patients.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo of the General Electric (GE) company. The logo consists of the letters "GE" intertwined and enclosed within a circular border. The design has a vintage or slightly distressed appearance, with rough edges and a textured look, suggesting it might be an older version of the logo.
# APR 2 7 2012
#### GE Healthcare
510(k) Premarket Notification Submission
#### 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | March 7, 2012 |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare<br>9900 Innovation Dr<br>Wauwatosa, WI 53226 |
| Primary Contact Person: | Bryan Behn<br>Regulatory Affairs Manager<br>GE Healthcare<br>T:(414)721-4214<br>F:(414)918-8275 |
| Trade Name:<br>Device: | Voluson S6, Voluson S8 Ultrasound System |
| Common/Usual Name: | Voluson S6, Voluson S8 |
| Classification Names:<br>Product Code: | Class II<br>Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-IYN<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
K103649 Voluson S6, Voluson S8 Diagnostic Ultrasound System Predicate Device(s):
K113758, K112213 Voluson E6E8E8ExpertE10
K053435 Voluson i
K113690 LOGIQ e
Device Description:
The subject device consists of a mobile console with keyboard, specialized controls, a color video LCD display with electronicarray transducers. It has the same general appearance, dimensions and weight as the unmodified device, it is a Track 3 generalpurpose imaging and analysis system providing real-time digital acquisition, processing and display capability intended for general radiology imaging and evaluation with some cardiology and vascular applications.
The device is a general-purpose ultrasound system. Specific Intended Use: clinical applications and exam types include: Fetal (Obstetrics); Abdominal (including renal and GYN/pelvic); Pediatric; Small Organ (breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional and Superficial; Transrectal (TR); Transvaginal (TV).
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a circular logo with the letters 'GE' intertwined in the center. The letters are stylized with curved lines and a swirling design. The logo is surrounded by a decorative border with a repeating pattern. The image is in black and white.
510(k) Premarket Notification Submission
#### Technology:
Determination of Substantial Equivalence:
The Voluson S6, Voluson S8 employs the same fundamental scientific technology as its predicate devices
Summary of Non-Clinical Tests:
The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform to applicable medical device safety standards. Voluson S6, Voluson S8 and its applications comply with voluntary standards as detailed in Section 9, 11 and 17 of this premarket submission. The following quality assurance measures were applied to the development of the system:
- Risk Analysis .
- Requirements Reviews
- Design Reviews .
- Testing on unit level (Module verification) .
- Integration testing (System verification)
- Final Acceptance Testing (Validation)
- Performance testing (Verification)
- Safety testing (Verification)
Transducer materials and other patient contact materials are biocompatible.
#### Summary of Clinical Tests:
The subject of this premarket submission, Voluson S6, Voluson S8, did not require clinical studies to support substantial equivalence.
GE Healthcare considers the Voluson S6, Voluson S8 to be as Conclusion: safe, as effective, and performance is substantially equivalent to the predicate device(s).
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
APR 2 7 2012
Mr. Bryan Behn Regulatory Affairs Manager GE Healthcare, GE Medical Systems Ultrasound and Primary Care, LLC 9900 Innovation Drive WAUWATOSA WI 53226
Re: K120741
Trade/Device Name: Voluson S6. Voluson S8 Regulation Number: 21 CFR 892-1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: March 7, 2012 Received: March 12, 2012
Dear Mr. Behn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Voluson S6, Voluson S8, as described in your premarket notification:
Transducer Model Number
| RAB4-8-RS | AB2-7-RS | 3Sc-RS |
|-----------|------------|--------|
| 4C-RS | RIC5-9W-RS | 9L-RS |
| E8C-RS | RAB2-5-RS | P2D |
| C1-5-RS | 8C-RS | |
| 12L-RS | ML6-15-RS | |
{3}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Shahram Vaezy at (301) 796-6242.
Sincerely Yours,
Michael O'Keeffe, Jr.
Janine M. Morris Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
{4}------------------------------------------------
#### 510(k) Number (if known):
Device Name:
Voluson S6, Voluson S8
Indications for Use:
The device is a general-purpose ultrasound system. Specific clinical applications and exam types include: Fetal (Obstetrics); Abdominal (including renal and GYN/pelvic); Pediatric, Small Organ (breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional and Superficial; Transrectal (TR); Transvaginal (TV).
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use_NA (Part 21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Aunh D. Th
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) Number K120741
{5}------------------------------------------------
#### Indications for Use Forms
The following forms represent indications with clinical applications and exam types along with the modes of operation for the Voluson S6, Voluson S8 system. Combinations identified "P" for the system represents those previously cleared in K103649. laominod " 101 the of brow" for the system represents those that are new in this release. Combinations identified "P" for the transducers represents those previously cleared with this or another GE Ultrasound system. Combinations identified "N" for the transducers uns of allouise. On the viewiously cleared any GE Ultrasound system. Please see section 11 Table 11.2.1 for information on previous clearance information on these transducers.
---
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Devices Evaluation and Safety
510K.K120741
Confidential and Privileged. This document contidential and privileged trade secrets and other Connidential and Priviteged. This document connance connecting connection of thers not employed by General Electric Co. All rights reserved
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo, which consists of the letters 'G' and 'E' intertwined within a circular border. The logo is presented in black and white, with the letters and border appearing in black against a white background. The design is simple yet recognizable, representing the well-known multinational conglomerate.
#### Diagnostic Ultrasound Indications for Use Form GE Voluson S6/S8 Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|-------------------------------|-------------------|---|---------|---------|---------|---------|---------|----------|----------|-------|----------|
| Clinical Application | B | M | PW | CW | Color | Color M | Power | Combined | Harmonic | Coded | Other |
| Anatomy/Region of Interest | | | Doppler | Doppler | Doppler | Doppler | Doppler | Modes | Imaging | Pulse | [Notes] |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics 17 | P | P | P | N | . P | P | b | P | . P | P | [ 5,6,9] |
| Abdominal(!) | P | P | P | N | P | P | P | P | P | P | [ 5,6,9] |
| Pediatric | P | P | P | N | P | P | P | P | P | P | [ 5,6,9] |
| Small Organ®] | P | P | P | N | P | P | P | P | P | P | [ 2,6,9] |
| Neonatal Cephalic | P | P | P | N | P | P | P | P | P | P | [ર] |
| Adult Cephalic | P | P | P | N | P | P | P | P | P | P | |
| Cardiaclil | P | P | P | N | P | P | P | P | P | P | (ર) |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | P | [5,6,9] |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | [5,6,9] |
| Musculo-skeletal Superficial | p | P | P | | P | P | P · | P | P | P | [5,6,9] |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal[8] | P | P | P | | P | P | P | P | P | P | [5,6,9] |
| Transvaginal | P | P | P | | P | P | P | P | P | P | [5,6,9] |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA
Notes: [1] Abdominal includes renal, GYN/Pelvic
[1] I loominal inolodes renar, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients
[3] Cardiac is Adult and Pediatric.
[5] 3D/4D Imaging Mode.
[6] Includes imaging of guidance of biopsy (2D/3D/4D).
[7] Includes infertility monitoring of follicle development.
on and Selety
[8] Includes urology/prostate.
[9] Elastography imaging- Elasticity
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR, 801.109)
(Divsion Bigh-Om
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circular border. The logo is in black and white and appears to be a vintage or classic version of the GE logo.
โ
# Diagnostic Ultrasound Indications for Use Form GE Voluson S6/S8 with RAB4-8-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>(Notes) |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [5,6] |
| Abdominal[1] | P | P | P | | P | P | P | P | P | P | [5,6] |
| Pediatric | P | P | P | | P | P | P | P | P | P | [5,6] |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | [5,6] |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal [8] | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA
[1] Abdominal includes renal, GYN/Pelvic Notes:
[1] Addoninal includes frenci, OTTC. Salivary gland, lymph nodes, pediatric and neonatal patients
[3] Cardiac is Adult and Pediatric.
[5] 3D/4D Imaging Mode.
[6] Includes imaging of guidance of biopsy (2D/3D/4D).
Svaluation and Safety
[7] Includes infertility monitoring of follicle development.
[8] Includes urology/prostate.
[9] Elastography imaging- Elasticity
[9] Elastography imagings Elastony
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Radiologicel Devices
Office of In Vitro Di
EIOK
21
{8}------------------------------------------------
### Diagnostic Ultrasound Indications for Use Form
#### GE Voluson S6/S8 with 4C-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>(Notes) |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [6] |
| Abdominal[1] | P | P | P | | P | P | P | P | P | P | [6] |
| Pediatric | P | P | P | | P | P | P | P | P | P | [6] |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | P | [6] |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | [6] |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal[8] | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA
[1] Abdominal includes renal, GYN/Pelvic Notes:
[1] Adominial indudes rendry of Nr.) of the stires, the mary gland, lymph nodes, pediatric and neonatal patients [3] Cardiac is Adult and Pediatric.
[5] 3D/4D Imaging Mode.
[6] Includes imaging of guidance of biopsy (2D/3D/4D).
[7] Includes infertility monitoring of follicle development.
[8] Includes urology/prostate.
[9] Elastography imaging- Elasticity
[5] Clastography Imeging B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
(Division Sign-On)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
SIOK
{9}------------------------------------------------
510(k) Premarket Notification Submission
# Diagnostic Ultrasound Indications for Use Form GE Voluson S6/S8 with E8C-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [6] |
| Abdominal[1] | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic | P | P | P | | P | P | P | P | P | P | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal[8] | P | P | P | | P | P | P | P | P | P | [6] |
| Transvaginal | P | P | P | | P | P | P | P | P | P | [6] |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| I anarosconic | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA
[1] Abdominal includes renal, GYN/Pelvic Notes:
[1] Addonifial molodou Tonal, D Proid, salivary gland, lymph nodes, pediatric and neonatal patients
[3] Cardiac is Adult and Pediatric.
[5] 3D/4D Imaging Mode.
[6] Includes imaging of guidance of biopsy (2D/3D/4D).
[7] Includes infertility monitoring of follicle development.
[8] Includes urology/prostate.
[9] Elastography imaging- Elasticity
[0] Clactography ime are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
(Division Sign-
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K
{10}------------------------------------------------
# Diagnostic Ultrasound Indications for Use Form
#### GE Voluson S6/S8 with C1-5-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [6] | |
| Abdominal[1] | P | P | P | | P | P | P | P | P | P | [6] | |
| Pediatric | P | P | P | | P | P | P | P | P | P | [6] | |
| Small Organ[2] | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | [6] | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Other | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal[8] | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA
[1] Abdominal includes renal, GYN/Pelvic Notes:
[2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients
[3] Cardiac is Adult and Pediatric.
[5] 3D/4D Imaging Mode.
[6] Includes imaging of guidance of biopsy (2D/3D/4D).
[7] Includes infertility monitoring of follicle development.
[8] Includes urology/prostate.
[9] Elastography imaging- Elasticity
[0] Clastography imbging "Exection")
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
(Division Sign-Om
Division of Radiological Devices
Office of In Vitro Diagriostic Device Evaluation and Safety
Deynep Eyelust
510K
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the General Electric (GE) logo, which consists of the letters 'G' and 'E' intertwined within a circular border. The letters are stylized and connected, creating a distinctive monogram. The logo is presented in black and white, with the letters and border appearing in black against a white background.
# Diagnostic Ultrasound Indications for Use Form GE Voluson S6/S8 with 12L-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|-------------------------------|-------------------|---|---------|---------|---------|---------|---------|-------------------|---------|-------|---------|--|
| Clinical Application | B | M | PW | CW | Color | Color M | Power | Combined Harmonic | | Coded | Other | |
| Anatomy/Region of Interest | | | Doppler | Doppler | Doppler | Doppler | Doppler | Modes | Imaging | Pulse | (Notes) | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics 77 | | | | | | | | | | | | |
| Abdomina[1] | | | | | | | | | | | | |
| Pediatric | P | P | P | | P | P | P | P | P | P | [6,9] | |
| Small Organ[2] | P | P | P | | P | P | P | P | P | P | [6,9] | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiaclil | | | | | | | | | | | | |
| Peripheral Vascular | b | P | P | | P | P | P | P | P | P | (୧' à) | |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | (୧'୨) | |
| Musculo-skeletal Superficial | P | P | P | | P | P | P | P | P | P | [6,9] | |
| Other | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal[8] | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| 7 operaceanie | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA
[1] Abdominal includes renal, GYN/Pelvic Notes:
[1] Abdominal includes Terrar, C. Fron Civis
[2] Small organ includes breast, testes, thyroid, saliyary gland, lymph nodes, pediatric and neonatal patients
[3] Cardiac is Adult and Pediatric.
[5] 3D/4D Imaging Mode.
[6] Includes imaging of guidance of biopsy (2D/3D/4D).
[7] Includes infertility monitoring of follicle development.
[8] Includes urology/prostate.
[9] Elastography imaging- Elasticity
[0] Liactography images are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Rediological Devices Office of In Vitro Diagnostic Device Evaluation and Satety
510K
{12}------------------------------------------------
510(k) Premarket Notification Submission
#### Diagnostic Ultrasound Indications for Use Form GE Voluson S6/S8 with AB2-7-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [6] |
| Abdominal[1] | P | P | P | | P | P | P | P | P | P | [6] |
| Pediatric | P | P | P | | P | P | P | P | P | P | [6] |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | [6] |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type. Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal[8] | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA
[1] Abdominal includes renal, GYN/Pelvic Notes:
[1] I loonliner malasot renay, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients
[3] Cardiac is Adult and Pediatric.
[5] 3D/4D Imaging Mode.
[6] Includes imaging of guidance of biopsy (2D/3D/4D).
[7] Includes infertility monitoring of follicle development.
[8] Includes urology/prostate.
[9] Elastography imaging- Elasticity
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510K. 120741
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the General Electric (GE) logo, which consists of the letters 'G' and 'E' intertwined within a circular border. The letters are stylized with curved lines, giving them a flowing appearance. The logo is presented in black and white, with the letters and border appearing in black against a white background.
#### Diagnostic Ultrasound Indications for Use Form GE Voluson S6/S8 with RIC5-9W-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [ 5,6] | |
| Abdominal[1] | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | |
| Small Organ[2] | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Other | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal[8] | P | P | P | | P | P | P | P | P | P | [5,6] | |
| Transvaginal | P | P | P | | P | P | P | P | P | P | [5,6] | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA
[1] Abdominal includes renal, GYN/Pelvic Notes:
[2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients
[3] Cardiac is Adult and Pediatric.
[5] 3D/4D Imaging Mode.
[6] Includes imaging of guidance of biopsy (2D/3D/4D).
[7] Includes infertility monitoring of follicle development.
[8] Includes urology/prostate.
[9] Elastography imaging- Elasticity
[၁] Clastography imaging - Loonshi
[၁] Combined modes are Β/Μ, Β/Color M, B/PWD, B/Color/PWD, B/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21-CFR 801.109)
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K
27
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters are stylized and connected in a flowing, cursive-like manner. The logo is black and white.
T
510(k) Premarket Notification Submission
# Diagnostic Ultrasound Indications for Use Form GE Voluson S6/S8 with RAB2-5-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | (5,6) | |
| Abdominal[1] | P | P | P | | P | P | P | P | P | P | (5,6) | |
| Pediatric | | | | | | | | | | | | |
| Small Organ[2] | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac [3] | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | (5,6) | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Other | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal[8] | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA (K053435)
[1] Abdominal includes renal, GYN/Pelvic Notes:
[1] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients
[3] Cardiac is Adult and Pediatric.
[5] 3D/4D Imaging Mode.
[6] Includes imaging of guidance of biopsy (2D/3D/4D).
[7] Includes infertility monitoring of follicle development.
[8] Includes urology/prostate.
[9] Elastography imaging- Elasticity
[5] Clastogruphy Imaging "LiaoRing" LiaoRing", "LiaoRiPWD, B/Color/PWD, B/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
(Division Sign-Off)
on of Radiological Devices Office of In ice Evaluation and Safety
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The logo is black and white and has a vintage look.
### Diagnostic Ultrasound Indications for Use Form
#### GE Voluson S6/S8 with 8C-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | | P | N | P | P | P | N | |
| Abdominal[1] | P | P | P | | P | N | P | P | P | N | |
| Pediatric | P | P | P | | P | N | P | P | P | N | |
| Small Organ[2] | P | P | P | | P | N | P | P | P | N | (9) |
| Neonatal Cephalic | P | P | P | | P | N | P | P | P | N | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | P | P | P | | P | N | P | P | P | N | |
| Peripheral Vascular | P | P | P | | P | N | P | P | P | N | |
| Musculo-skeletal Conventional | P | P | P | | P | N | P | P | P | N | (9) |
| Musculo-skeletal Superficial | P | P | P | | P | N | P | P | P | N | (9) |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal[8] | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA(K113690)
[1] Abdominal includes renal, GYN/Pelvic Notes:
[2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients
[3] Cardiac is Adult and Pediatric.
[5] 3D/4D Imaging Mode.
[6] Includes imaging of guidance of biopsy (2D/3D/4D).
[7] Includes infertility monitoring of follicle development.
[8] Includes urology/prostate.
[9] Elastography imaging- Elasticity
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
ision Sign-On)
Division of Rediological Devices valuation and Safety Office of In, Vitro Diagnostic Dovit し
510K
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized manner. The letters are enclosed within a circular border that has a decorative, swirling pattern around the edges.
510(k) Premarket Notification Submission
# Diagnostic Ultrasound Indications for Use Form GE Voluson S6/S8 with ML6-15-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes) |
| Ophthalmic |…