← Product Code [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN) · K063518

# HI VISION 900 DIAGNOSTIC ULTRASOUND SCANNER (K063518)

_Hitachi Medical Systems America, Inc. · IYN · Dec 21, 2006 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K063518

## Device Facts

- **Applicant:** Hitachi Medical Systems America, Inc.
- **Product Code:** [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN.md)
- **Decision Date:** Dec 21, 2006
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Intended Use

Diagnostic ultrasound imaging or fluid flow analysis if the human body as follows:

## Device Story

The Hitachi HI VISION 900 is a diagnostic ultrasound system utilizing pulsed echo and pulsed Doppler imaging. It processes acoustic signals from various transducers to generate B-mode, M-mode, Pulsed Wave Doppler, Continuous Wave Doppler, Color Flow, Amplitude Doppler, Harmonic, and 3D images. Operated by clinicians in clinical settings, the system provides real-time visualization of internal structures and blood flow. Output is displayed for the clinician to guide diagnostic assessments, interventional procedures (e.g., biopsies), and surgical planning. The device aids in clinical decision-making by providing non-invasive anatomical and hemodynamic data, potentially improving diagnostic accuracy and procedural safety.

## Clinical Evidence

Bench testing only. The device complies with NEMA UD-3 (Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices) and the 510(k) Diagnostic Ultrasound Guidance (April 14, 1994). General safety is verified through compliance with IEC 60601-1.

## Technological Characteristics

Diagnostic ultrasound system supporting B-mode, M-mode, PWD, CWD, Color Doppler, Amplitude Doppler, Harmonic, and 3D imaging. Complies with NEMA UD-3 and IEC 60601-1. Supports multiple transducer types (e.g., EUP-B514, EUP-F531, EUP-850, EUP-C514, EUP-L53S, EUP-TC3, EUP-C532, EUP-L65, EUP-U533, EUP-CC531, EUP-053T, EUP-V53W, EUP-CV524, EUP-OL334, EUP-ES52M, EUP-R54AW).

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# Device Description

The Hitachi HI VISION 900 Diagnostic Ultrasound Scanner is a Track 3 Diagnostic Ultrasound Pulsed Doppler and Pulsed Echo Imaging System capable of the following operating functions:

- . B Mode
- M Mode .
- . Pulsed Wave Doppler
- . Continuous Wave Doppler

- . Color Flow
- . Amplitude Doppler
- Harmonic imaging .
- . Superficial musculoskeletal imaging
- . 3D imaging

# Safety

As a Track 3 ultrasound device, the Hitachi HI VISION 900 Diagnostic Ultrasound Scanner complies with the Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment (1992) - published by NEMA as UD-3.

With respect to limits on acoustic outputs, the Hitachi HI VISION 900 Diagnostic Ultrasound Scanner complies with the guideline limits set in the 510(k) Diagnostic Ultrasound Guidance - Revision: April 14, 1994.

With regard to general safety, the Hitachi HI VISION 900 Diagnostic Ultrasound Scanner is designed to comply with IEC 606601-1 (1998) Medical Electrical Equipment, Part 1 - General Requirements for Safety.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

1

DEC 2 1 2006

Mr. Douglas J. Thistlewaite Manager of Regulatory Affairs HITACHI Medical Systems America, Inc. 1959 Summit Commerce Park TWINSBURG OH 44087-2371

Re: K063518

Trade Name: HI VISION 900 Diagnostic Ultrasound Scanner Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: November 20, 2006 Received: November 21, 2006

Dear Mr. Thistlewaite:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the HI VISION 900 Diagnostic Ultrasound Scanner, as described in your premarket notification:

Image /page/1/Picture/11 description: The image shows a circular logo with the letters "FDA" in bold, stylized font at the center. Above the letters, the years "1906-2006" are printed. Below the letters, the word "Centennial" is written in cursive. The logo is surrounded by text that reads "A CENTURY OF SERVICE TO CONSUMER PROTECTION & HEALTH IN THE SCIENCE OF FOODS". The logo appears to be a commemorative emblem, possibly celebrating the 100th anniversary of an organization.

Protecting and Promoting Public Health

{2}------------------------------------------------

#### Transducer Model Number EUP-B514 EUP-F531 EUP-850 EUP-C514 EUP-L53S EUP-TC3 EUP-C532 EUP-L65 EUP-U533 EUP-CC531 EUP-053T EUP-V53W EUP-CV524 EUP-OL334 EUP-ES52M EUP-R54AW-19, -33

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.

The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:

> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

le

{3}------------------------------------------------

Page 2 - Mr. Thistlewaite

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

If you have any questions regarding the content of this letter, please contact Andrew Kang at (240) 276-3666.

Sincerely yours,

David A. Segura

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{4}------------------------------------------------

### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM 006 NOTSTA IH

| Clinical Application      |                              | Mode of Operation |    |     |     |                  |                      | Other**<br>(Spec.) |
|---------------------------|------------------------------|-------------------|----|-----|-----|------------------|----------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B                 | M  | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) |                    |
| Ophthalmic                | Ophthalmic                   | N                 | N  | N   | N   | N                | N                    | N                  |
|                           | Fetal                        | N                 | N  | N   | N   | N                | N                    | N                  |
|                           | Abdominal                    | Na                | Na | Na  | Na  | Na               | Na                   | Na                 |
|                           | Intra-operative (Spec.)      | Nb                | Nb | Nb  |     | Nb               | Nb                   | Nb                 |
|                           | Intra operative (Neuro.)     |                   |    |     |     |                  |                      |                    |
|                           | Laparoscopic                 | N                 | N  | N   |     | N                | N                    | N                  |
|                           | Pediatric                    | N                 | N  | N   | N   | N                | N                    | N                  |
|                           | Small Organ (Spec.)          | Nd                | Nd | Nd  |     | Nd               | Nd                   | Nd                 |
|                           | Neonatal Cephalic            | N                 | N  | N   |     | N                | N                    | N                  |
| Fotal Imaging<br>& Other  | Adult Cephalic               | N                 | N  | N   | N   | N                | N                    | N                  |
|                           | Trans rectal                 | Nh                | Nh | Nh  |     | Nh               | Nh                   | Nh                 |
|                           | Trans-vaginal                | Nf                | Nf | Nf  |     | Nf               | Nf                   | Nf                 |
|                           | Trans-urethral               |                   |    |     |     |                  |                      |                    |
|                           | Trans esoph. (non-Card.)     |                   |    |     |     |                  |                      |                    |
|                           | Musculo-skel. (Convent.)     | N                 | N  | N   |     | N                | N                    | N                  |
|                           | Musculo-skel. (Superfic.)    | N                 | N  | N   |     | N                | N                    | N                  |
|                           | Intra-luminal                |                   |    |     |     |                  |                      |                    |
|                           | Other (spec.)                |                   |    |     |     |                  |                      |                    |
| Cardiac                   | Cardiac Adult                | N                 | N  | N   | N   | N                | N                    | N                  |
|                           | Cardiac Pediatric            | N                 | N  | N   | N   | N                | N                    | N                  |
|                           | Trans esophageal (card.)     | Ng                | Ng | Ng  | N   | Ng               | Ng                   | Ng                 |
|                           | Other (spec.)                |                   |    |     |     |                  |                      |                    |
| Peripheral<br>Vessel      | Peripheral vessel            | N                 | N  | N   | N   | N                | N                    | N                  |
|                           | Other (spec.)                |                   |    |     |     |                  |                      |                    |

Intended use. Diagnostic ultrasound imaging or fluid flow analysis if the human body as follows:

N = new indication, P = previous indication

*Combination of each operating mode, B, M, PWD, CWD and Color Doppler.

** Amplitude Dopplor, Harmonic Imaging and 3D Imaging.

#### Additional Comments:

System:

Includes imaging for guidance of percutaneous biopay of abdominal organs and structures Subscript "a": (including amniocentesis)

| Subscript "b": | Includes imaging of organs and structures exposed during surgery<br>(excluding neurosurgery and laparoscopic procedures). |
|----------------|---------------------------------------------------------------------------------------------------------------------------|
| Subscript "c": | Includes thyroid, parathyroid, breast, scrotum, penis.                                                                    |
| Subscript "d": | Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy.                                 |
| Subscript "e": | Includes imaging for guidance of transrectal biopsy.                                                                      |
| Subscript "f": | Includes imaging for guidance of transvaginal biopsy.                                                                     |
| Subscript "g": | For pediatric patients.                                                                                                   |
| Subscript "h": | Includes imaging for guidance of transrectal biopsy.                                                                      |

An a paran A DEBOY

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices

510(k) Number: K063518

David R. Hyman

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

Prescription Use (Per 21 CFR 801.109)

647089808814

BRITEMA SMATEVS lesibeM idaesiH-mori

me 2p : 7 |

{5}------------------------------------------------

System: HI VISION 900

Transducer: EUP-B514

| Clinical Application      |                              | Mode of Operation |    |     |     |                  |                      |                    |
|---------------------------|------------------------------|-------------------|----|-----|-----|------------------|----------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B                 | M  | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) |
| Ophthalmic                | Ophthalmic                   |                   |    |     |     |                  |                      |                    |
|                           | Fetal                        | P                 | P  | P   |     | P                | P                    | P                  |
|                           | Abdominal                    | Pa                | Pa | P   |     | Pa               | Pa                   | Pa                 |
|                           | Intra-operative (Spec.)      |                   |    |     |     |                  |                      |                    |
|                           | Intra-operative (Neuro.)     |                   |    |     |     |                  |                      |                    |
|                           | Laparoscopic                 |                   |    |     |     |                  |                      |                    |
|                           | Pediatric                    |                   |    |     |     |                  |                      |                    |
|                           | Small Organ (Spec.)          |                   |    |     |     |                  |                      |                    |
|                           | Neonatal Cephalic            |                   |    |     |     |                  |                      |                    |
| Fetal Imaging<br>& Other  | Adult Cephalic               |                   |    |     |     |                  |                      |                    |
|                           | Trans-rectal                 |                   |    |     |     |                  |                      |                    |
|                           | Trans-vaginal                |                   |    |     |     |                  |                      |                    |
|                           | Trans-urethral               |                   |    |     |     |                  |                      |                    |
|                           | Trans-esoph. (non-Card.)     |                   |    |     |     |                  |                      |                    |
|                           | Musculo-skel. (Convent.)     |                   |    |     |     |                  |                      |                    |
|                           | Musculo-skel. (Superfic.)    |                   |    |     |     |                  |                      |                    |
|                           | Intra-luminal                |                   |    |     |     |                  |                      |                    |
|                           | Other (spec.)                |                   |    |     |     |                  |                      |                    |
| Cardiac                   | Cardiac Adult                |                   |    |     |     |                  |                      |                    |
|                           | Cardiac Pediatric            |                   |    |     |     |                  |                      |                    |
|                           | Trans-esophageal (card.)     |                   |    |     |     |                  |                      |                    |
|                           | Other (spec.)                |                   |    |     |     |                  |                      |                    |
| Peripheral<br>Vessel      | Peripheral vessel            |                   |    |     |     |                  |                      |                    |
|                           | Other (spec.)                |                   |    |     |     |                  |                      |                    |

Intended use: Diagnostic ultrasound imaging or fluid flow analysis if the human hody as follows

N = new indication, P = previous indication

*Combination of each operating mode, B, M, PWD and Color Doppler.

** Amplitude Doppler and Harmonic Imaging

Additional Comments:

| Subscript "a": | Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures<br>(including amniocentesis).     |
|----------------|---------------------------------------------------------------------------------------------------------------------------|
| Subscript "b": | Includes imaging of organs and structures exposed during surgery<br>(excluding neurosurgery and laparoscopic procedures). |
| Subscript "c": | Includes thyroid, parathyroid, breast, scrotum, penis.                                                                    |
| Subscript "d": | Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy.                                 |
| Subscript "e": | Includes imaging for guidance of transrectal biopsy.                                                                      |
| Subscript "f": | Includes imaging for guidance of transvaginal biopsy.                                                                     |
| Subscript "g": | For pediatric patients.                                                                                                   |
| Subscript "h": | Includes imaging for guidance of transrectal biopsy.                                                                      |

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Reproductive, Abdominal, ENT and Radiological Devices

510(k) Number: K063518

David A. Lygum

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _

Prescription Use (Per 21 CFR 801.109)

6720E960868889

ше Zy:Z |

{6}------------------------------------------------

### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM 06 NOTERS H

System: Transducer:

EUP-C514

| Clinical Application      |                              | Mode of Operation |    |     |     |                  |                      |                    |
|---------------------------|------------------------------|-------------------|----|-----|-----|------------------|----------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B                 | M  | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) |
| Ophthalmic                | Ophthalmic                   |                   |    |     |     |                  |                      |                    |
| Fetal Imaging<br>& Other  | Fetal                        | P                 | P  | P   |     | P                | P                    | P                  |
|                           | Abdominal                    | Pa                | Pa | Pa  |     | Pa               | Pa                   | Pa                 |
|                           | Intra-operative (Spec.)      |                   |    |     |     |                  |                      |                    |
|                           | Intra-operative (Neuro.)     |                   |    |     |     |                  |                      |                    |
|                           | Laparoscopic                 |                   |    |     |     |                  |                      |                    |
|                           | Pediatric                    | P                 | P  | P   |     | P                | P                    | P                  |
|                           | Small Organ (Spec.)          | Pd                | Pd | Pd  |     | Pd               | Pd                   | Pd                 |
|                           | Neonatal Cephalic            |                   |    |     |     |                  |                      |                    |
|                           | Adult Cephalic               |                   |    |     |     |                  |                      |                    |
|                           | Trans-rectal                 |                   |    |     |     |                  |                      |                    |
|                           | Trans-vaginal                |                   |    |     |     |                  |                      |                    |
|                           | Trans-urethral               |                   |    |     |     |                  |                      |                    |
|                           | Trans-esoph. (non-Card.)     |                   |    |     |     |                  |                      |                    |
|                           | Musculo-skel. (Convent.)     |                   |    |     |     |                  |                      |                    |
|                           | Musculo-skel. (Superfic.)    |                   |    |     |     |                  |                      |                    |
|                           | Intra-luminal                |                   |    |     |     |                  |                      |                    |
|                           | Other (spec.)                |                   |    |     |     |                  |                      |                    |
| Cardiac                   | Cardiac Adult                |                   |    |     |     |                  |                      |                    |
|                           | Cardiac Pediatric            |                   |    |     |     |                  |                      |                    |
|                           | Trans esophageal (card.)     |                   |    |     |     |                  |                      |                    |
|                           | Other (spec.)                |                   |    |     |     |                  |                      |                    |
| Peripheral<br>Vessel      | Peripheral vessel            |                   |    |     |     |                  |                      |                    |
|                           | Other (spec.)                |                   |    |     |     |                  |                      |                    |

#### Intended use. Diagnostic ultragound imaging or fluid flow analysis if the human body as follows

N = new indication, P = previous indication

* Combination of each operating mode, B, M, PWD and Color Doppler.

** Amplitude Doppler, Harmonic Imaging and 3D Imaging,

Additional Comments:

Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures Subscript "a": (including amniocentesis).

| Subscript "b": Includes imaging of organs and structures exposed during surgery<br>(excluding neurosurgery and laparoscopic procedures). |
|------------------------------------------------------------------------------------------------------------------------------------------|
| Subscript "c": Includes thyroid, parathyroid, breast, scrotum, penis.                                                                    |
| Subscript "d": Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy.                                 |
| Subscript "e": Includes imaging for guidance of transrectal biopsy.                                                                      |
| Subscript "f": Includes imaging for guidance of transvaginal biopsy.                                                                     |
| Subscript "g": For pediatric patients.                                                                                                   |
| Subscript "h": Includes imaging for guidance of transrectal biopsy.                                                                      |
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)                                                                 |

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-Off Division of Reproductive, Abdominal, ENT, and Radiological Devices

510(k) Number: K063518

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K06351

Prescription Use (Per 21 CFR 801,109)

44706880000114

180178mA 2metry2 leaibeM idastin-mori 90-61-200 me2☆:Z}

{7}------------------------------------------------

System -Transduce:

| HI VISION 900 |  |
|---------------|--|
| EUP-C532      |  |

| Clinical Application      |                              | Mode of Operation |    |     |     |                  |                      |                    |
|---------------------------|------------------------------|-------------------|----|-----|-----|------------------|----------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B                 | M  | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) |
| Ophthalmic                | Ophthalmic                   |                   |    |     |     |                  |                      |                    |
|                           | Fetal                        |                   |    |     |     |                  |                      |                    |
|                           | Abdominal                    | Pa                | Pa | Pa  |     | Pa               | Pa                   | Pa                 |
|                           | Intra-operative (Spec.)      | Pb                | Pb | Pb  |     | Pb               | Pb                   | Pb                 |
|                           | Intra-operative (Neuro.)     |                   |    |     |     |                  |                      |                    |
|                           | Laparoscopic                 |                   |    |     |     |                  |                      |                    |
|                           | Pediatric                    | P                 | P  | P   |     | P                | P                    | P                  |
|                           | Small Organ (Spec.)          | Pd                | Pd | Pd  |     | Pd               | Pd                   | Pd                 |
|                           | Neonatal Cephalic            | P                 | P  | P   |     | P                | P                    | P                  |
| Fetal Imaging<br>& Other  | Adult Cephalic               |                   |    |     |     |                  |                      |                    |
|                           | Trans-rectal                 |                   |    |     |     |                  |                      |                    |
|                           | Trans-vaginal                |                   |    |     |     |                  |                      |                    |
|                           | Trans-urethral               |                   |    |     |     |                  |                      |                    |
|                           | Trans-esoph. (non-Card.)     |                   |    |     |     |                  |                      |                    |
|                           | Musculo-skel. (Convent.)     |                   |    |     |     |                  |                      |                    |
|                           | Musculo-skel. (Superfic.)    |                   |    |     |     |                  |                      |                    |
|                           | Intra-luminal                |                   |    |     |     |                  |                      |                    |
|                           | Other (spec.)                |                   |    |     |     |                  |                      |                    |
|                           | Cardiac Adult                |                   |    |     |     |                  |                      |                    |
| Cardiac                   | Cardiac Pediatric            |                   |    |     |     |                  |                      |                    |
|                           | Trans esophageal (card.)     |                   |    |     |     |                  |                      |                    |
|                           | Other (spec.)                |                   |    |     |     |                  |                      |                    |
| Peripheral<br>Vessel      | Peripheral vessel            | P                 | P  | P   |     | P                | P                    | P                  |
|                           | Other (spec.)                |                   |    |     |     |                  |                      |                    |

Intended use. Diagnostic ultrasound imaging or fluid flow analysis if the human body as follows

N = new indication, P = previous indication

*Combination of each operating mode, B, M. PWD and Color Doppler.

** Amplitude Doppler and Harmonic Imaging

Additional Comments:

Subscript "Includes imaging for guidance of percutaneous biopoy of abdominal organs and structures

|                | (including amniocentesis).                                                                |
|----------------|-------------------------------------------------------------------------------------------|
| Subscript "b": | Includes imaging of organs and structures exposed during surgery                          |
|                | (excluding neurosurgery and laparoscopic procedures).                                     |
| Subscript "c": | Includes thyroid, parathyroid, breast, scrotum, penis.                                    |
| Subscript "d": | Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy. |
| Subscript "e": | Includes imaging for guidance of transrectal biopsy.                                      |
| Subscript "f": | Includes imaging for guidance of transvaginal biopsy.                                     |
| Subscript "g": | For pediatric patients.                                                                   |
| Subscript "h": | Includes imaging for guidance of transrectal biopsy.                                      |
|                | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)                  |

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Reproductive, Abdominal, ENT. and Radiological Devices

510(k) Number: K063518

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K063518

Prescription Use (Per 21 CFR 801.109)

1997 1977 1991 1991 18 189 1991 1991 19 90-61-200 12:488mm

{8}------------------------------------------------

System Transducer:

006 NOTSION 900 EUP-CC531

| Clinical Application      |                                                                                                                                                        | Mode of Operation |          |          |     |                  |                      |                    |  |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|----------|----------|-----|------------------|----------------------|--------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III)                                                                                                                           | B                 | M        | FWD      | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) |  |
| Ophthalmic                | Ophthalmic                                                                                                                                             |                   |          |          |     |                  |                      |                    |  |
| Fetal Imaging<br>& Other  | Fetal<br>Abdominal<br>Intra operative (Spec.)<br>Intra-operative (Neuro.)<br>Laparoscopic<br>Pediatric<br>Small Organ (Spec.)<br>Neonatal Cephalic     | P                 | P        | P        |     | P                | P                    | P                  |  |
|                           | Adult Cephalic<br>Trans-rectal<br>Trans-vaginal<br>Trans-urethral<br>Trans-esoph. (non-Card.)<br>Musculo-skel. (Convent.)<br>Musculo-skel. (Superfic.) | Pe<br>Pf          | Pe<br>Pf | Pe<br>Pf |     | Pe<br>Pf         | Pe<br>Pf             | Pe<br>Pf           |  |
|                           | Intra-luminal<br>Other (spec.)                                                                                                                         |                   |          |          |     |                  |                      |                    |  |
| Cardiac                   | Cardiac Adult<br>Cardiac Pediatric<br>Trans-esophageal (card.)<br>Other (spec.)                                                                        |                   |          |          |     |                  |                      |                    |  |
| Peripheral<br>Vessel      | Peripheral vessel<br>Other (spec.)                                                                                                                     |                   |          |          |     |                  |                      |                    |  |

Intended use: Diagnostic ultrasound inaging or fluid flow analysis if the human body as follows

N = new indication, P = previous indication

*Combination of each operating mode, B, M, PWD and Color Doppler.

** Amplitude Doppler and Harmonic Imaging

Additional Comments:

Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures Subscript "a": (including amniocentesis).

| Subscript "b": | Includes imaging of organs and structures exposed during surgery<br>(excluding neurosurgery and laparoscopic procedures). |
|----------------|---------------------------------------------------------------------------------------------------------------------------|
| Subscript "c": | Includes thyroid, parathyroid, breast, scrotum, penis.                                                                    |
| Subscript "d": | Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy.                                 |
| Subscript "e": | Includes imaging for guidance of transrectal biopsy.                                                                      |
| Subscript "f": | Includes imaging for guidance of transvaginal biopsy.                                                                     |
| Subscript "g": | For pediatric patients.                                                                                                   |
| Subscript "h": | Includes imaging for guidance of transrectal biopsy.                                                                      |

WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED (FLEASE DO MO) Concurrence of CDRH, Office of Device Evaluation (ODE)

> (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices

510(k) Number: K063518

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number 2063518

Prescription Use (Per 21 CFR 801.109)

ની

{9}------------------------------------------------

#### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM 006 NOTSION 900

System: Transducer:

CUP-CV524

| Clinical Application      |                              | Mode of Operation |   |     |     |                  |                     |                    |
|---------------------------|------------------------------|-------------------|---|-----|-----|------------------|---------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other**<br>(Spec.) |
| Ophthalmic                | Ophthalmic                   |                   |   |     |     |                  |                     |                    |
|                           | Fetal                        | P                 | P | P   |     | P                | P                   | P                  |
|                           | Abdominal                    | P                 | P | P   |     | P                | P                   | P                  |
|                           | Intra-operative (Spec.)      |                   |   |     |     |                  |                     |                    |
|                           | Intra-operative (Neuro.)     |                   |   |     |     |                  |                     |                    |
|                           | Laparoscopic                 |                   |   |     |     |                  |                     |                    |
|                           | Pediatric                    | P                 | P | P   |     | P                | P                   | P                  |
|                           | Small Organ (Spec.)          | P                 | P | P   |     | P                | P                   | P                  |
|                           | Neonatal Cephalic            |                   |   |     |     |                  |                     |                    |
| Fetal Imaging<br>& Other  | Adult Cephalic               |                   |   |     |     |                  |                     |                    |
|                           | Trans rectal                 |                   |   |     |     |                  |                     |                    |
|                           | Trans-vaginal                |                   |   |     |     |                  |                     |                    |
|                           | Trans-urethral               |                   |   |     |     |                  |                     |                    |
|                           | Trans-esoph. (non-Card.)     |                   |   |     |     |                  |                     |                    |
|                           | Musculo-skel. (Convent.)     |                   |   |     |     |                  |                     |                    |
|                           | Musculo-skel. (Superfic.)    |                   |   |     |     |                  |                     |                    |
|                           | Intra-luminal                |                   |   |     |     |                  |                     |                    |
|                           | Other (spec.)                |                   |   |     |     |                  |                     |                    |
|                           | Cardiac Adult                |                   |   |     |     |                  |                     |                    |
| Cardiac                   | Cardiac Pediatric            |                   |   |     |     |                  |                     |                    |
|                           | Trans esophageal (card.)     |                   |   |     |     |                  |                     |                    |
|                           | Other (spec.)                |                   |   |     |     |                  |                     |                    |
| Peripheral<br>Vessel      | Peripheral vessel            |                   |   |     |     |                  |                     |                    |
|                           | Other (spec.)                |                   |   |     |     |                  |                     |                    |

Intended use' Diagnostic ultrasound imaging or fluid flow analysis if the human body as follows

N = new indication, P = previous indication

* Combination of each operating mode, B, M, PWD and Color Doppler.

** Amplitude Doppler, Harmonic Imaging and 3D Imaging.

### Additional Comments:

| Subscript "a": | Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures<br>(including amniocentesis).     |
|----------------|---------------------------------------------------------------------------------------------------------------------------|
| Subscript "b": | Includes imaging of organs and structures exposed during surgery<br>(excluding neurosurgery and laparoscopic procedures). |
| Subscript "c": | Includes thyroid, parathyroid, breast, scrotum, penis.                                                                    |
| Subscript "d": | Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy.                                 |
| Subscript "e": | Includes imaging for guidance of transrectal biopsy.                                                                      |
| Subscript "f": | Includes imaging for guidance of transvaginal biopsy.                                                                     |
| Subscript "g": | For pediatric patients.                                                                                                   |
| Subscript "h": | Includes imaging for guidance of transrectal biopsy.                                                                      |

WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED PLEASE DO NOT

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off Division of Reproductive, Abdominal, ENT, and Radiological Devices

510(k) Number: K063518

David h. Segram

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

Prescription Use (Per 21 CFR 801.109)

1688-1 020/800 . 9 EBO-T - 677069606800814

so irema smetzy s les ibel interit-mori שפיל : קישורים 90-61-200

{10}------------------------------------------------

Transducer: EUP ES52M

System:

| Clinical Application      |                              | Mode of Operation |   |     |     |                  |                      |                    |  |
|---------------------------|------------------------------|-------------------|---|-----|-----|------------------|----------------------|--------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B                 | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) |  |
| Ophthalmic                | Ophthalmic                   |                   |   |     |     |                  |                      |                    |  |
|                           | Fetal                        |                   |   |     |     |                  |                      |                    |  |
|                           | Abdominal                    |                   |   |     |     |                  |                      |                    |  |
|                           | Intra operative (Spec.)      |                   |   |     |     |                  |                      |                    |  |
|                           | Intra-operative (Neuro.)     |                   |   |     |     |                  |                      |                    |  |
|                           | Laparoscopic                 |                   |   |     |     |                  |                      |                    |  |
|                           | Pediatric                    |                   |   |     |     |                  |                      |                    |  |
|                           | Small Organ (Spec.)          |                   |   |     |     |                  |                      |                    |  |
|                           | Neonatal Cephalic            |                   |   |     |     |                  |                      |                    |  |
| Fetal Imaging<br>& Other  | Adult Cephalic               |                   |   |     |     |                  |                      |                    |  |
|                           | Trans-rectal                 |                   |   |     |     |                  |                      |                    |  |
|                           | Trans-vaginal                |                   |   |     |     |                  |                      |                    |  |
|                           | Trans-urethral               |                   |   |     |     |                  |                      |                    |  |
|                           | Trans-esoph. (non-Card.)     |                   |   |     |     |                  |                      |                    |  |
|                           | Musculo-skel. (Convent.)     |                   |   |     |     |                  |                      |                    |  |
|                           | Musculo-skel. (Superfic.)    |                   |   |     |     |                  |                      |                    |  |
|                           | Intra-luminal                |                   |   |     |     |                  |                      |                    |  |
|                           | Other (spec.)                |                   |   |     |     |                  |                      |                    |  |
| Cardiac                   | Cardiac Adult                |                   |   |     |     |                  |                      |                    |  |
|                           | Cardiac Pediatric            |                   |   |     |     |                  |                      |                    |  |
|                           | Trans-esophageal (card.)     | P                 | P | P   | P   | P                | P                    | P                  |  |
|                           | Other (spec.)                |                   |   |     |     |                  |                      |                    |  |
| Peripheral<br>Vessel      | Peripheral vessel            |                   |   |     |     |                  |                      |                    |  |
|                           | Other (spec.)                |                   |   |     |     |                  |                      |                    |  |

N = new indication, 1' = previous indication

*Combination of each operating mode. B. M. PWD, CWD and Color Doppler.

** Amplitude Doppler

Additional Comments'

Subscript "g" Includes imaging for guidance of percutancous biopsy of abdominal organs and structures

|                | (including amniocentesis).                                                                |
|----------------|-------------------------------------------------------------------------------------------|
| Subscript "b": | Includes imaging of organs and structures exposed during surgery                          |
|                | (excluding neurosurgery and laparoscopic procedures).                                     |
| Subscript "c": | Includes thyroid, parathyroid, breast, scrotum, penis.                                    |
| Subscript "d": | Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy. |
| Subscript "e": | Includes imaging for guidance of transrectal biopsy.                                      |
| Subscript "f": | Includes imaging for guidance of transvaginal biopsy.                                     |
| Subscript "g": | For pediatric patients.                                                                   |
| Subscript "h": | Includes imaging for guidance of transrectal biopsy.                                      |
|                | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)                  |

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices

510(k) Number: K063518

David A. Ingram

---

(Division Sign-Off)

Division of Reproductive, Abdomi and Radiological Devices 510(k) Number _

Prescription Use (Per 21 CFR 801, 109)

{11}------------------------------------------------

# DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM 006 NOTSION 900

Transducer: EUP-F531

System -

|                           | Clinical Application         |    |    |     |     | Mode of Operation |                      |                    |
|---------------------------|------------------------------|----|----|-----|-----|-------------------|----------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B  | M  | PWD | CWD | Color<br>Doppler  | Combined*<br>(Spec.) | Other**<br>(Spec.) |
| Ophthalmic                | Ophthalmic                   |    |    |     |     |                   |                      |                    |
|                           | Fetal<br>Abdominal           | P  | P  | P   |     | P                 | P                    | P                  |
|                           | Intra-operative (Spec.)      | Pb | Pb | Pb  |     | Pb                | Pb                   | Pb                 |
|                           | Intra operative (Neuro.)     |    |    |     |     |                   |                      |                    |
|                           | Laparoscopic                 |    |    |     |     |                   |                      |                    |
|                           | Pediatric                    | P  | P  | P   |     | P                 | P                    | P                  |
|                           | Small Organ (Spec.)          | Pc | Pc | Pc  |     | Pc                | Pc                   | Pc                 |
|                           | Neonatal Cephalic            | P  | P  | P   |     | P                 | P                    | P                  |
| Fetal Imaging             | Adult Cephalic               |    |    |     |     |                   |                      |                    |
| & Other                   | Trans-rectal                 | P  | P  | P   |     | P                 | P                    | P                  |
|                           | Trans vaginal                | P  | P  | P   |     | P                 | P                    | P                  |
|                           | Trans-urethral               |    |    |     |     |                   |                      |                    |
|                           | Trans-esoph, (non Card.)     |    |    |     |     |                   |                      |                    |
|                           | Musculo-skel. (Convent.)     |    |    |     |     |                   |                      |                    |
|                           | Musculo skel. (Superfic.)    |    |    |     |     |                   |                      |                    |
|                           | Intra luminal                |    |    |     |     |                   |                      |                    |
|                           | Other (spec.)                |    |    |     |     |                   |                      |                    |
|                           | Cardiac Adult                |    |    |     |     |                   |                      |                    |
| Cardiac                   | Cardiac Pediatric            |    |    |     |     |                   |                      |                    |
|                           | Trans esophageal (card.)     |    |    |     |     |                   |                      |                    |
|                           | Other (spec.)                |    |    |     |     |                   |                      |                    |
| Peripheral                | Peripheral vessel            | P  | P  | P   |     | P                 | P                    | P                  |
| Vessel                    | Other (spec.)                |    |    |     |     |                   |                      |                    |

Intended use: Diagnostic ultrasound imaging or fluid flow enalysis if the human body as follows:

N = new indication, P = previous indication

* Combination of each operating mode, B, M, PWD and Color Doppler.

** Amplitude Doppler and Harmonic Imaging.

Additional Comments: Stibscript "4": ۲۰۰۵۱۰۰۵

| Subscript "a": | Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures<br>(including amniocentesis).     |
|----------------|---------------------------------------------------------------------------------------------------------------------------|
| Subscript "b": | Includes imaging of organs and structures exposed during surgery<br>(excluding neurosurgery and laparoscopic procedures). |
| Subscript "c": | Includes thyroid, parathyroid, breast, scrotum, penis.                                                                    |
| Subscript "d": | Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy.                                 |
| Subscript "e": | Includes imaging for guidance of transrectal biopsy.                                                                      |
| Subscript "f": | Includes imaging for guidance of transvaginal biopsy.                                                                     |
| Subscript "g": | For pediatric patients.                                                                                                   |
| Subscript "h": | Includes imaging for guidance of transrectal biopsy.                                                                      |

S LINE-CONTINUE ON ANOTHER PAGE IF NEEDED ALLE, BALLOW

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices

510(k) Number:

David A. Bergmann

KO6351

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number .

Prescription Use (Per 21 CFR 801.109)

:

{12}------------------------------------------------

System' Transduce

| HI VISION 900 |
|---------------|
| EUP-L53S      |

|                           | Clinical Application         | Mode of Operation |    |     |     |                  |                      |                    |
|---------------------------|------------------------------|-------------------|----|-----|-----|------------------|----------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B                 | M  | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) |
|                           | Ophthalmic                   |                   |    |     |     |                  |                      |                    |
|                           | Fetal                        |                   |    |     |     |                  |                      |                    |
|                           | Abdominal                    | Pa                | Pa | Pa  |     | Pa               | Pa                   | Pa                 |
|                           | Intra-operative (Spec.)      |                   |    |     |     |                  |                      |                    |
|                           | Intra-operative (Neuro.)     |                   |    |     |     |                  |                      |                    |
|                           | Laparoscopic                 |                   |    |     |     |                  |                      |                    |
|                           | Pediatric                    | P                 | P  | P   |     | P                | P                    | P                  |
| Fetal Imaging<br>& Other  | Small Organ (Spec.)          | Pd                | Pd | Pd  |     | Pd               | Pd                   | Pd                 |
|                           | Neonatal Cephalic            |                   |    |     |     |                  |                      |                    |
|                           | Adult Cephalic               |                   |    |     |     |                  |                      |                    |
|                           | Trans-rectal                 |                   |    |     |     |                  |                      |                    |
|                           | Trans-vaginal                |                   |    |     |     |                  |                      |                    |
|                           | Trans-urethral               |                   |    |     |     |                  |                      |                    |
|                           | Trans-esoph. (non-Card.)     |                   |    |     |     |                  |                      |                    |
|                           | Musculo-skel (Convent.)      | P                 | P  | P   |     | P                | P                    | P                  |
|                           | Musculo-skel. (Superfic.)    | P                 | P  | P   |     | P                | P                    | P                  |
|                           | Intra-luminal                |                   |    |     |     |                  |                      |                    |
|                           | Other (spec.)                |                   |    |     |     |                  |                      |                    |
|                           | Cardiac Adult                |                   |    |     |     |                  |                      |                    |
| Cardiac                   | Cardiac Pediatric            |                   |    |     |     |                  |                      |                    |
|                           | Trans-esophageal (card.)     |                   |    |     |     |                  |                      |                    |
|                           | Other (spec.)                |                   |    |     |     |                  |                      |                    |
| Peripheral                | Peripheral vessel            | P                 | P  | P   |     | P                | P                    | P                  |
| Vessel                    | Other (spec.)                |                   |    |     |     |                  |                      |                    |

Intended use: Diagnostic ultrasound imaging or fluid flow analysis if the human body as follows

N = new indication, P = previous indication

* Combination of each operating mode, B, M, PWD and Color Doppler.

** Amplitude Doppler, Harmonic Imaging and 3D Imaging,

Additional Comments:

Subscript "A" Includes imaging for guidance of perculaneous biopsy of abdominal organs and structures (including amniocentesis).

| Subscript "b": | Includes imaging of organs and structures exposed during surgery                          |
|----------------|-------------------------------------------------------------------------------------------|
|                | (excluding neurosurgery and laparoscopic procedures).                                     |
| Subscript "c": | Includes thyroid, parathyroid, breast, scrotum, penis.                                    |
| Subscript "d": | Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy. |
| Subscript "e": | Includes imaging for guidance of transrectal biopsy.                                      |
| Subscript "f": | Includes imaging for guidance of transvaginal biopsy.                                     |
| Subscript "g": | For pediatric patients.                                                                   |
| Subscript "h": | Includes imaging for guidance of transrectal biopsy.                                      |

WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) (PLEASE DO NOT Concurrence of CDRH. Office of Device Evaluation (ODE)

> Division Sign-Off Division of Reproductive, Abdominal, ENT, and Radiological Devices

510(k) Number: K063518

David R. Lehman

(Division Sign-Off) Onliston of Reproductive, Abdom ann Radiological Devices 5114k) Number _

Preseription Use (Per 21 CFR 801.109)

। ୧୫---020/210.9 £80-1 BATOEGROEE I +

soilema smatzy2 feaibeM idastiH-mor7 12:44am 80-61-200

{13}------------------------------------------------

System: Transducer:

# EUP-LG5

| Clinical Application      |                              | Mode of Operation |    |     |     |                  |                      |                    |
|---------------------------|------------------------------|-------------------|----|-----|-----|------------------|----------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B                 | M  | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) |
| Ophthalmic                | Ophthalmic                   |                   |    |     |     |                  |                      |                    |
|                           | Fetal                        |                   |    |     |     |                  |                      |                    |
|                           | Abdominal                    | P                 | P  | P   |     | P                | P                    | P                  |
|                           | Intra-operative (Spec.)      |                   |    |     |     |                  |                      |                    |
|                           | Intra-operative (Neuro.)     |                   |    |     |     |                  |                      |                    |
|                           | Laparoscopic                 |                   |    |     |     |                  |                      |                    |
|                           | Pediatric                    | P                 | P  | P   |     | P                | P                    | P                  |
|                           | Small Organ (Spec.)          | Pc                | Pc | Pc  |     | Pc               | Pc                   | Pc                 |
|                           | Neonatal Cephalic            |                   |    |     |     |                  |                      |                    |
| Fetal Imaging<br>& Other  | Adult Cephalic               |                   |    |     |     |                  |                      |                    |
|                           | Trans-rectal                 |                   |    |     |     |                  |                      |                    |
|                           | Trans-vaginal                |                   |    |     |     |                  |                      |                    |
|                           | Trans-urethra)               |                   |    |     |     |                  |                      |                    |
|                           | Trans-esoph. (non-Card.)     |                   |    |     |     |                  |                      |                    |
|                           | Musculo-skel. (Convent.)     | P                 | P  | P   |     | P                | P                    | P                  |
|                           | Musculo-skel. (Superfic.)    | P                 | P  | P   |     | P                | P                    | P                  |
|                           | Intra-luminal                |                   |    |     |     |                  |                      |                    |
|                           | Other (spec.)                |                   |    |     |     |                  |                      |                    |
|                           | Cardiac Adult                |                   |    |     |     |                  |                      |                    |
| Cardiac                   | Cardiac Pediatric            |                   |    |     |     |                  |                      |                    |
|                           | Trans-esophageal (card.)     |                   |    |     |     |                  |                      |                    |
|                           | Other (spec.)                |                   |    |     |     |                  |                      |                    |
|                           |                              |                   |    |     |     |                  |                      |                    |
| Peripheral<br>Vessel      | Peripheral vessel            | P                 | P  | P   |     | P                | P                    | P                  |
|                           | Other (spec.)                |                   |    |     |     |                  |                      |                    |

Intended use. Dirasound imaging or fluid flow analysis if the human body as follows:

N = new indication, P = previous indication

*Combination of each operating mode, B, M, PWD and Color Doppler.

** Amplitude Doppler and Harmonic Imaging.

# Additional Comments:

| Subscript "a": | Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures<br>(including amniocentesis).     |
|----------------|---------------------------------------------------------------------------------------------------------------------------|
| Subscript "b": | Includes imaging of organs and structures exposed during surgery<br>(excluding neurosurgery and laparoscopic procedures). |
| Subscript "c": | Includes thyroid, parathyroid, breast, scrotum, penis.                                                                    |
| Subscript "d": | Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy.                                 |
| Subscript "e": | Includes imaging for guidance of transrectal biopsy.                                                                      |
| Subscript "f": | Includes imaging for guidance of transvaginal biopsy.                                                                     |

Subscript "g": For pediatric patients. Subscript "h": ipt "h": Includes imaging for guidance of transectal biopay.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEBDED

(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUED ON ANOTHER PAGE)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-Off Division of Reproductive, Abdominal, ENT, and Radiological Devices

510(k) Number: K063518

David R. Segmen

---

(Division Sign-Off)

Division of Reproductive, Abdomin and Radiological Devices 510(k) Number ___

Prescription Use (Per 21 CFR 801.109)

Ania Sales Saletses Tesibel infortill-mori

ાજે

{14}------------------------------------------------

System: 006 NOISION 900

Transducer: EUP-0531

|                           | Clinical Application                                 |    |    |     |     | Mode of Operation |                      |                    |
|---------------------------|------------------------------------------------------|----|----|-----|-----|-------------------|----------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)                         | B  | M  | PWD | CWD | Color<br>Doppler  | Combined*<br>(Spec.) | Other**<br>(Spec.) |
| Ophthalmic                | Ophthalmic                                           |    |    |     |     |                   |                      |                    |
|                           | Fetal<br>Abdominal                                   |    |    |     |     |                   |                      |                    |
| Fetal Imaging<br>& Other  | Intra-operative (Spec.)<br>Intra-operative (Neuro.)  | Pb | Pb | Pb  |     | Pb                | Pb                   | Pb                 |
|                           | Laparoscopic<br>Pediatric                            |    |    |     |     |                   |                      |                    |
|                           | Small Organ (Spec.)<br>Neonatal Cephalic             |    |    |     |     |                   |                      |                    |
|                           | Adult Cephalic<br>Trans-rectal                       |    |    |     |     |                   |                      |                    |
|                           | Trans-vaginal<br>Trans-urethral                      |    |    |     |     |                   |                      |                    |
|                           | Trans-esoph. (non-Card.)<br>Musculo-skel. (Convent.) |    |    |     |     |                   |                      |                    |
|                           | Musculo-skel. (Superfic.)<br>Intra-luminal           |    |    |     |     |                   |                      |                    |
|                           | Other (spec.)                                        |    |    |     |     |                   |                      |                    |
| Cardiac                   | Cardiac Adult<br>Cardiac Pediatric                   |    |    |     |     |                   |                      |                    |
|                           | Trans-esophageal (card.)<br>Other (spec.)            |    |    |     |     |                   |                      |                    |
| Peripheral<br>Vessel      | Peripheral vessel<br>Other (spec.)                   |    |    |     |     |                   |                      |                    |

| Intended use: Diagnostic ultrasound imaging or fluid flow analysis if the human body as follows: |  |
|--------------------------------------------------------------------------------------------------|--|
|--------------------------------------------------------------------------------------------------|--|

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

N = new indication, P = previous indication

* Combination of each operating mode, B, M. PWD and Color Doppler.

** Amplitude Doppler and Harmonic Imaging.

#### Additional Comments:

| Subscript "a": | Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures<br>(including amniocentesis).     |
|----------------|---------------------------------------------------------------------------------------------------------------------------|
| Subscript "b": | Includes imaging of organs and structures exposed during surgery<br>(excluding neurosurgery and laparoscopic procedures). |
| Subscript "c": | Includes thyroid, parathyroid, breast, scrotum, penis.                                                                    |
| Subscript "d": | Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy.                                 |
| Subscript "e": | Includes imaging for guidance of transrectal biopsy.                                                                      |
| Subscript "f": | Includes imaging for guidance of transvaginal biopsy.                                                                     |
| Subscript "g": | For pediatric patients.                                                                                                   |
| Subscript "h": | Includes imaging for guidance of transrectal biopsy.                                                                      |

WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED CLEASE DO NOT Concurrence of CDRH, Office of Device Evaluation (ODE)

> (Division Sign-Off) Division of Reproductive, Abdominal, ENT and Radiological Devices

510(k) Number: K063518

David Ch. Lyonn

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _

Prescription Use (Per 21 CFR 801.109)

| 88-3 020/410.9 180-1 647069808814

so inema smetzy si system idae substil-mori

14

Dac-18-06

12:44am

{15}------------------------------------------------

# DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM 006 NOTSION 900

System: Transducer: EUP-OL334

|                           | Clinical Application         |   |   |     |     |                  | Mode of Operation    |                     |  |  |
|---------------------------|------------------------------|---|---|-----|-----|------------------|----------------------|---------------------|--|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other **<br>(Spec.) |  |  |
| Ophthalmic                | Ophthalmic                   |   |   |     |     |                  |                      |                     |  |  |
|                           | Fetal                        |   |   |     |     |                  |                      |                     |  |  |
|                           | Abdominal                    |   |   |     |     |                  |                      |                     |  |  |
|                           | Intra-operative (Spec.)      |   |   |     |     |                  |                      |                     |  |  |
|                           | Intra-operative (Neuro.)     |   |   |     |     |                  |                      |                     |  |  |
|                           | Laparoscopic                 | P | P | P   |     | P                | P                    | P                   |  |  |
|                           | Pediatric                    |   |   |     |     |                  |                      |                     |  |  |
|                           | Small Organ (Spec.)          |   |   |     |     |                  |                      |                     |  |  |
|                           | Neonatal Cephalic            |   |   |     |     |                  |                      |                     |  |  |
| Fetal Imaging<br>& Other  | Adult Cephalic               |   |   |     |     |                  |                      |                     |  |  |
|                           | Trans-rectal                 |   |   |     |     |                  |                      |                     |  |  |
|                           | Trans-vaginal                |   |   |     |     |                  |                      |                     |  |  |
|                           | Trans-urethral               |   |   |     |     |                  |                      |                     |  |  |
|                           | Trans-esoph. (non-Card.)     |   |   |     |     |                  |                      |                     |  |  |
|                           | Musculo-skel. (Convent.)     |   |   |     |     |                  |                      |                     |  |  |
|                           | Musculo-skel. (Superfic.)    |   |   |     |     |                  |                      |                     |  |  |
|                           | Intra-luminal                |   |   |     |     |                  |                      |                     |  |  |
|                           | Other (spec.)                |   |   |     |     |                  |                      |                     |  |  |
| Cardiac                   | Cardiac Adult                |   |   |     |     |                  |                      |                     |  |  |
|                           | Cardiac Pediatric            |   |   |     |     |                  |                      |                     |  |  |
|                           | Trans-esophageal (card.)     |   |   |     |     |                  |                      |                     |  |  |
|                           | Other (spec.)                |   |   |     |     |                  |                      |                     |  |  |
| Peripheral<br>Vessel      | Peripheral vessel            |   |   |     |     |                  |                      |                     |  |  |
|                           | Other (spec.)                |   |   |     |     |                  |                      |                     |  |  |

# Intended use . Diagnostic ultrasound imaging or fluid flow analysis if the human body as follows:

N = new indication, P = previous indication

"Combination of each operating mode, B, M, PWD and Color Doppler.

**Amplitude Doppler

Additional Comments:

| Subscript "a": | Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures<br>(including amniocentesis).     |
|----------------|---------------------------------------------------------------------------------------------------------------------------|
| Subscript "b": | Includes imaging of organs and structures exposed…

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K063518](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K063518)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
