Browse hierarchy Radiology (RA) Subpart B — Diagnostic Devices 21 CFR 892.1550 Product Code IYN K061643 — ECHO VIEW/SHIMASONIC DIAGNOSTIC ULTRASOUND SYSTEM, MODEL SDU-1200PRO
ECHO VIEW/SHIMASONIC DIAGNOSTIC ULTRASOUND SYSTEM, MODEL SDU-1200PRO
K061643 · Shimadzu Medical Systems · IYN · Jul 27, 2006 · Radiology
Device Facts
Record ID K061643
Device Name ECHO VIEW/SHIMASONIC DIAGNOSTIC ULTRASOUND SYSTEM, MODEL SDU-1200PRO
Applicant Shimadzu Medical Systems
Product Code IYN · Radiology
Decision Date Jul 27, 2006
Decision SESE
Submission Type Traditional
Regulation 21 CFR 892.1550
Device Class Class 2
Attributes Pediatric
Intended Use
The SDU-1200Pro is intended for the following applications: Fetal, Abdominal, Pediatric, Small Organs (Specify), Neonatal Cephalic, Adult Cephalic, Cardiac, Transvaginal, Peripheral Vascular, Muscular, Musculo-skeletal Superficial and Musculo-skeletal Conventional.
Device Story
SDU-1200Pro is a mobile diagnostic ultrasound system; utilizes flat linear array, convex linear, and sector probes (1.5-15 MHz). Operates in B, M, Pulsed Doppler, Continuous Doppler, and Color modes; supports combined modes (e.g., B/M, B/PWD, CFM/PWD). Used in clinical settings by healthcare professionals for diagnostic imaging of various anatomical structures. System processes acoustic echoes to generate real-time images and Doppler data; assists clinicians in visualizing internal structures and blood flow to support diagnostic decision-making. Benefits include non-invasive assessment of patient anatomy and hemodynamics.
Clinical Evidence
Bench testing only. The device complies with IEC 60601-1, UL60601-1, and AIUM/NEMA UD2/UD3 standards for acoustic output measurement and safety. No clinical data was required for this 510(k) clearance.
Technological Characteristics
Mobile ultrasound system; frequency range 1.5-15 MHz; supports linear, convex, and sector transducers. Modes: B, M, PWD, CWD, Color Doppler. Complies with IEC 60601-1, UL60601-1, AIUM NEMA UD2, and AIUM NEMA UD3. Connectivity/software details not specified beyond standard diagnostic ultrasound operation.
Indications for Use
Indicated for diagnostic ultrasound imaging or Doppler analysis of the human body, including fetal, abdominal, pediatric, small organ (thyroid, testicles, breast), neonatal/adult cephalic, cardiac, transrectal, transvaginal, peripheral vascular, and musculoskeletal applications.
Regulatory Classification
Identification An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Related Devices
K071289 — ECHOVIEW / SHIMASONIC DIAGNOSTIC ULTRASOUND SYSTEM, MODEL SDU-1200PRO · Shimadzu Medical Systems · Jan 9, 2008
K071287 — ECHOVIEW / SHIMASONIC DIAGNOSTIC ULTRASOUND DEVICE, MODEL SDU-2200PRO · Shimadzu Medical Systems · Jan 9, 2008
K061637 — ECHOVIEW/ SHIMASONIC DIAGNOSTIC ULTRASOUND SYSTEM, MODEL SDU-2200PRO · Shimadzu Medical Systems · Jul 25, 2006
K071291 — ECHOVIEW / SHIMASONIC DIAGNOSTIC ULTRASOUND SYSTEM, MODEL SDU-1100 · Shimadzu Medical Systems · Jan 9, 2008
K050510 — ECHOVIEW/SHIMASONIC DIAGNOSTIC ULTRASOUND SYSTEM, MODEL SDU-1100 · Shimadzu Medical Systems · Apr 1, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
K061643
### 510(k) Summary
### 1.0 SUBMITTER INFORMATION
1.1 Submitter: SHIMADZU MEDICAL SYSTEMS 20101 South Vermont Ave. Torrance, CA 90502-1328 PH: 310-217-8855 FX: 310-217-8869
1.2 Contact: Randal Walker
MAR. 1 7.2006 1.3 Date:
### 2.0 DEVICE NAME
| 2.1 Proprietary Name: | SDU-1200Pro |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2.2 Common Name: | Ultrasound Imaging System |
| 2.3 Classification: | Ultrasonic Pulsed Doppler Imaging System<br>FR # 892.1550, Product Code 90-IYN<br>Ultrasonic Pulsed Echo Imaging System<br>FR # 892.1560, Product Code 90-IYO<br>Diagnostic Ultrasound Transducer<br>FR # 892.1570, Product Code 90-ITX |
| 2.4 Predicate Device: | Shimadzu SDU-2200 (K003514, Feb./12/01) |
# 3.0 DEVICE DESCRIPTION
The SDU-1200Pro is a mobile diagnostic ultrasound system. This system has flat linear array, convex linear and sector probe with a frequency range of approximately 1.5 to 15 MHz. It has B mode, M mode, Pulsed Doppler mode, Continuous Doppler mode, Color mode, or in a combination of modes.
### 4.0 INTENDED USE
The SDU-1200Pro is intended for the following applications:
{1}------------------------------------------------
Fetal, Abdominal, Pediatric, Small Organs (Specify), Neonatal Cephalic, Adult Cephalic, Cardiac, Transvaginal, Peripheral Vascular, Muscular, Musculo-skeletal Superficial and Musculo-skeletal Conventional.
# 5.0 SAFETY CONSIDERATIONS
SDU-1200Pro has been designed to meet the following voluntary and measurement standards:
- IEC 60601-1 Safety of Medical Electric Equipment .
- UL60601-1:2003 Medical Electrical Equipment Part I : General Requirements . for Safety
- AIUM NEMA UD2 Acoustic Output Measurement Standard for Diagnostic . Ultrasound Equipment
- Acoustic Output Measurement and Labeling Standard for Diagnostic . Ultrasound Equipment Revision 1 (AIUM 1998)
- AIUM NEMA UD3 Standard for Real-time Display of Thermal and Mechanical . Acoustic Output Indices on Diagnostic Ultrasound Equipment
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an emblem that resembles an eagle or bird-like figure with stylized wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 27 2006
Mr. Randal Walker National Service Manager Shimadzu Medical Systems 20101 South Vermont Avenue TORRANCE CA 90502-1328
Re: K061643
Trade Name: SDU-1200Pro Diagnostic Ultrasound System Regulation Number: 21 CFR 89.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: March 17, 2006 Received: June 13, 2006
Dear Mr. Walker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the SDU-1200Pro Diagnostic Ultrasound System, as described in your premarket notification:
Image /page/2/Picture/11 description: The image is a circular logo with the letters "FDA" in the center. The letters are stylized and bold. Above the letters "FDA" are the letters "Est. 1906". Below the letters "FDA" is the word "Centennial". The logo is surrounded by a dotted border.
Jing Puttic . Health
{3}------------------------------------------------
## Transducer Model Number
| L040-075U | VA13R-050U | UB10R-065U |
|------------|--------------|------------|
| L040-120U | VA20R-035U | EC11R-055U |
| L040-120HU | VA40R-035U | S011-050U |
| L070-075U | VA40R-035HU | S017-035U |
| L072-050U | VA57R-0375WU | S020-025U |
| VA11R-055U | VA57R-035HU | |
| VA13R-035U | TV11R-055U | |
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this letter, please contact Andrew Kang, M.D. at (301) 594-1212.
Sincerely yours.
Nancy C Bugdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{5}------------------------------------------------
Ultrasound Device Indications Statement Page _1 of 20
510(k) Number (if known) :
Device Name : Diagnostic Ultrasound System SDU-1200Pro, system
Fill out one form for each ultrasound system or transducer.
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify)** | Tissue<br>Harmonic<br>Imaging | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-------------------------|-------------------------------|--------------------|
| Ophthalmic | | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | N | N | N | |
| Abdominal | | N | N | N | | N | N | N | N | N | |
| Intra-operative<br>(Specify) | | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ<br>(Specify) * | | N | N | N | | N | N | N | N | N | |
| Neonatal<br>Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | N | N | N | | N | N | N | N | N | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | N | N | N | | N | N | N | N | N | |
| Transvaginal | | N | N | N | | N | N | N | N | N | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | N | N | N | N | N | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | N | N | N | | N | N | N | N | N | |
| Musculo-skeletal<br>Superficial | | N | N | N | | N | N | N | N | N | |
| Other (Specify) | | | | | | | | | | | |
Mode of Operation
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Other Indications or Modes:
* Thyroid, Testicles, Breast
** B/M, B/PWD, CFM(B)/PWD, CFM(B)/CFM(M)
Prescription Use
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K061643
{6}------------------------------------------------
Ultrasound Device Indications Statement Page 2 _of_20
510(k) Number (if known) : ___________________________________________________________________________________________________________________________________________________
Fill out one form for each ultrasound system or transducer.
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify)** | Tissue<br>Harmoni<br>c<br>Imaging | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-------------------------|-----------------------------------|--------------------|
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intra-operative<br>(Specify) | | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ<br>(Specify) * | | N | N | N | | N | N | N | N | N | |
| Neonatal | | | | | | | | | | | |
| Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | N | N | N | N | N | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | N | N | N | | N | N | N | N | N | |
| Musculo-skeletal<br>Superficial | | N | N | N | | N | N | N | N | N | |
| Other (Specify) | | | | | | | | | | | |
Mode of Operation
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Other Indications or Modes:
* Thyroid, Testicles, Breast
** B/M, B/PWD, CFM(B)/PWD, CFM(B)/CFM(M)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
cscription Use
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K061643
{7}------------------------------------------------
Ultrasound Device Indications Statement Page _3_ of _ 20
510(k) Number (if known) : _ 上 ひん 1 6 43
Device Name : Diagnostic Ultrasound System SDU-1200Pro, L040-120U
Fill out one form for each ultrasound system or transducer.
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) ** | Tissue<br>Harmonic<br>Imaging | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|--------------------------|-------------------------------|--------------------|
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intra-operative<br>(Specify) | | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ<br>(Specify) * | N | N | N | N | | N | N | N | N | N | |
| Neonatal | | | | | | | | | | | |
| Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | N | N | N | N | N | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | N | N | N | | N | N | N | N | N | |
| Musculo-skeletal<br>Superficial | | N | N | N | | N | N | N | N | N | |
| Others (Specify) | | | | | | | | | | | |
Mode of Operation
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Other Indications or Modes:
* Thyroid, Testicles, Breast
** B/M, B/PWD, CFM(B)/PWD, CFM(B)/CFM(M)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Y. Marquez
(Division Sign-Off)
Division of Reproductive, Abdominal, and Padiological Devices 5 13(x) Number _______________________________________________________________________________________________________________________________________________________________
{8}------------------------------------------------
Ultrasound Device Indications Statement Page _4 of_20
510(k) Number (if known) : _ 1 0 6 l 6 4 3 Device Name : Diagnostic Ultrasound System SDU-1200Pro, L040-120HU
Fill out one form for each ultrasound system or transducer.
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
| Mode of Operation | | | | | | | | | | | |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|--------------------------|-------------------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) ** | Tissue<br>Harmonic<br>Imaging | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intra-operative<br>(Specify) | | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ<br>(Specify) * | | N | N | N | | N | N | N | N | N | |
| Neonatal | | | | | | | | | | | |
| Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | N | N | N | N | N | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | N | N | N | | N | N | N | N | N | |
| Musculo-skeletal<br>Superficial | | N | N | N | | N | N | N | N | N | |
| Others (Specify) | | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Other Indications or Modes:
* Thyroid, Testicles, Breast
** B/M, B/PWD, CFM(B)/PWD, CFM(B)/CFM(M)
prescription Use
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number
K061643
{9}------------------------------------------------
Ultrasound Device Indications Statement Page_5 of 20
6061643 510(k) Number (if known) : Device Name : Diagnostic Ultrasound System SDU-1200Pro, L070-075U
Fill out one form for each ultrasound system or transducer.
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) ** | Tissue<br>Harmonic<br>Imaging | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|--------------------------|-------------------------------|--------------------|
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intra-operative<br>(Specify) | | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ<br>(Specify) * | | N | N | N | | N | N | N | N | N | |
| Neonatal | | | | | | | | | | | |
| Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | N | N | N | N | N | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | N | N | N | | N | N | N | N | N | |
| Musculo-skeletal<br>Superficial | | N | N | N | | N | N | N | N | N | |
| Others (Specify) | | | | | | | | | | | |
Mode of Operation
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Other Indications or Modes:
* Thyroid, Testicles, Breast
** B/M, B/PWD, CFM(B)/PWD, CFM(B)/CFM(M)
**REVISED FINAL**
V. Larry C. Bearden
(Division Sign-Off) Division of Reproductive, Abdon and Radiological Devices 51(Xk) Number
{10}------------------------------------------------
Ultrasound Device Indications Statement Page_6 of 20
K061643 510(k) Number (if known) : Device Name : Diagnostic Ultrasound System SDU-1200Pro, L072-050U
Fill out one form for each ultrasound system or transducer.
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) ** | Tissue<br>Harmonic<br>Imaging | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|--------------------------|-------------------------------|--------------------|
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intra-operative<br>(Specify) | | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ<br>(Specify) * | N | N | N | N | N | N | N | N | N | N | |
| Neonatal<br>Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | N | N | N | N | N | N | N | N | N | N | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | N | N | N | N | N | N | N | N | N | N | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Others (Specify) | | | | | | | | | | | |
Mode of Operation
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Other Indications or Modes:
* Thyroid, Testicles, Breast
** B/M, B/PWD, CFM(B)/PWD, CFM(B)/CFM(M)
Janay Prather
(Division Sign-Off) Division of Reproductive, Abdominal, and Padiological Devices 51 (1) Number
{11}------------------------------------------------
Ultrasound Device Indications Statement Page _7 _of 20
510(k) Number (if known) : ___________________________________________________________________________________________________________________________________________________
Fill out one form for each ultrasound system or transducer.
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) ** | Tissue<br>Harmonic<br>Imaging | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|--------------------------|-------------------------------|--------------------|
| Ophthalmic | | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | N | N | N | |
| Abdominal | | N | N | N | | N | N | N | N | N | |
| Intra-operative<br>(Specify) | | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | | |
| Pediatric | | N | N | N | | N | N | N | N | N | |
| Small Organ<br>(Specify) * | | | | | | | | | | | |
| Neonatal<br>Cephalic | | N | N | N | | N | N | N | N | N | |
| Adult Cephalic | | N | N | N | | N | N | N | N | N | |
| Cardiac | | N | N | N | | N | N | N | N | N | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Others (Specify) | | | | | | | | | | | |
Mode of Oneration
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Other Indications or Modes:
** B/M, B/PWD, CFM(B)/PWD,CFM(B)/CFM(M)
\$\checkmark\$
Trusion Sign-Off & Reproductive, A olonica Devi 11 81 17100
{12}------------------------------------------------
Ultrasound Device Indications Statement Page_8 20 of
12061643 510(k) Number (if known) : _ Device Name : Diagnostic Ultrasound System SDU-1200Pro, VA13K-035U
Fill out one form for each ultrasound system or transducer.
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) ** | Tissue<br>Harmonic<br>Imaging | Other<br>(Specify) |
|---------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|--------------------------|-------------------------------|--------------------|
| Ophthalmic | | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | N | N | N | |
| Abdominal | | N | N | N | | N | N | N | N | N | |
| Intra-operative<br>(Specify) | | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ<br>(Specify) * | | | | | | | | | | | |
| Neonatal | | | | | | | | | | | |
| Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | N | N | N | | N | N | N | N | N | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | | |
| Conventional | | | | | | | | | | | |
| Superficial | | | | | | | | | | | |
| Others (Specify) | | | | | | | | | | | |
Mode of Operation
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Other Indications or Modes:
** B/M, B/PWD, CFM(B)/PWD,CFM(B)/CFM(M)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
V
Yancy C. Morgan
on of Reproductive. Abdominal P ... Choose Device
: : : : : : :
{13}------------------------------------------------
Ultrasound Device Indications Statement Page 9 20 of
510(k) Number (if known) : Device Name : Diagnostic Ultrasound System SDU-1200Pro, VA13R-050U
Fill out one form for each ultrasound system or transducer.
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) ** | Tissue<br>Harmonic<br>Imaging | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|--------------------------|-------------------------------|--------------------|
| Ophthalmic | | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | N | N | N | |
| Abdominal | | N | N | N | | N | N | N | N | N | |
| Intra-operative<br>(Specify) | | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ<br>(Specify) * | | | | | | | | | | | |
| Neonatal<br>Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | N | N | N | | N | N | N | N | N | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Others (Specify) | | | | | | | | | | | |
Mode of Operation
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Other Indications or Modes:
** B/M, B/PWD, CFM(B)/PWD,CFM(B)/CFM(M)
7 Nancy C. Beaton
Sign-Off
f Reproductive, Abdominal,
Radiological Devices K061643
Number
{14}------------------------------------------------
Ultrasound Device Indications Statement Page_10 of 20
#### 12061643 510(k) Number (if known) : _
5 Fo(x) Namber (1) laromic Ultrasound System SDU-1200Pro, VA20R-035U
Fill out one form for each ultrasound system or transducer.
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify)** | Tissue<br>Harmonic<br>Imaging | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-------------------------|-------------------------------|--------------------|
| Ophthalmic | | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | N | N | N | |
| Abdominal | | N | N | N | | N | N | N | N | N | |
| Intra-operative<br>(Specify) | | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ<br>(Specify) * | | | | | | | | | | | |
| Neonatal | | | | | | | | | | | |
| Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | N | N | N | | N | N | N | N | N | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Others (Specify) | | | | | | | | | | | |
Mode of Operation
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Other Indications or Modes:
** B/M, B/PWD, CFM(B)/PWD,CFM(B)/CFM(M)
Nancy C. Ferguson
(Person Sign Off)
(Person's Legal Name - Terminal
Employee ID: 406643
{15}------------------------------------------------
Ultrasound Device Indications Statement Page _ 1 l of 20
KO61643 510(k) Number (if known) : _ Device Name : Diagnostic Ultrasound System SDU-1200Pro, VA40R-035U
Fill out one form for each ultrasound system or transducer.
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
| Mode of Operation | | | | | | | | | | | |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-------------------------|-------------------------------|--------------------|
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify)** | Tissue<br>Harmonic<br>Imaging | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | N | N | N | |
| Abdominal | | N | N | N | | N | N | N | N | N | |
| Intra-operative<br>(Specify) | | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ<br>(Specify) * | | | | | | | | | | | |
| Neonatal<br>Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Others (Specify) | | | | | | | | | | | |
Mode of Operation
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Other Indications or Modes:
** B/M, B/PWD, CFM(B)/PWD,CFM(B)/CFM(M)
J Dwayne Heaton
Division Sign-Off
Division Sign-Off)
L
Dusien af Reproductive, Abdominal,
٠٩٤٠٠
Ser
RO61643
{16}------------------------------------------------
Ultrasound Device Indications Statement Page of 20
# 510(k) Number (if known) :
Device Name : Diagnostic Ultrasound System SDU-1200Pro, VA40R-035HU
Fill out one form for each ultrasound system or transducer.
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) ** | Tissue<br>Harmonic<br>Imaging | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|--------------------------|-------------------------------|--------------------|
| Ophthalmic | | | | | | | | | | | |
| Fetal | N | N | N | N | | N | N | N | N | N | |
| Abdominal | N | N | N | N | | N | N | N | N | N | |
| Intra-operative<br>(Specify) | | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ<br>(Specify) * | | | | | | | | | | | |
| Neonatal | | | | | | | | | | | |
| Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Others (Specify) | | | | | | | | | | | |
Mode of Oneration
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Other Indications or Modes:
** B/M, B/PWD, CFM(B)/PWD,CFM(B)/CFM(M)
Lacey (radion
no sinn Sinn Mit of Reproductive, Abdomina . minomasi De SE . THEIR
{17}------------------------------------------------
Ultrasound Device Indications Statement Page 13 of 20
#### 1061 510(k) Number (if known) : Device Name : Diagnostic Ultrasound System SDU-1200Pro, VA57R-0375WU
ce Name : Diagnostic Ultrasound System DDC-1200110, 71497R 09179 Rev
Fill out one form for each ultrasound system or transducer.
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) ** | Tissue<br>Harmonic<br>Imaging | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|-------…