Browse hierarchy Radiology (RA) Subpart B — Diagnostic Devices 21 CFR 892.1550 Product Code IYN K052331 — ACUSON CYPRESS ULTRASOUND SYSTEM
ACUSON CYPRESS ULTRASOUND SYSTEM
K052331 · Siemens Medical Solutions USA, Inc. · IYN · Nov 10, 2005 · Radiology
Device Facts
Record ID K052331
Device Name ACUSON CYPRESS ULTRASOUND SYSTEM
Applicant Siemens Medical Solutions USA, Inc.
Product Code IYN · Radiology
Decision Date Nov 10, 2005
Decision SESE
Submission Type Traditional
Regulation 21 CFR 892.1550
Device Class Class 2
Attributes Pediatric
Intended Use
The Cypress platform is intended for use in the following applications: General Imaging and Cardiology for Fetal, Abdominal, Intraoperative (cardiac), Pediatrics, Neonatal Cephalic, Cardiac (adult, pediatric), Trans-esophageal, Peripheral Vessel, Intra-luminal and Intra-cardiac applications, and intended uses as defined in the FDA guidance document. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
Device Story
Cypress Ultrasound System is a multi-purpose diagnostic ultrasound platform. It transmits ultrasonic energy into the body via a transducer; reflected echoes are processed to generate 2D images of anatomical structures and fluid flow. Doppler principles are utilized to display blood flow as a spectrum or color-coded 2D images. The system includes analysis packages for anatomical measurements and clinical diagnosis. It is intended for use in clinical settings by healthcare professionals. The device benefits patients by providing non-invasive diagnostic visualization and hemodynamic assessment. The system operates according to the ALARA (as low as reasonably achievable) principle for acoustic output.
Clinical Evidence
Bench testing only. The submission relies on the substantial equivalence of the device's technological characteristics and intended use to previously cleared predicate devices. No clinical data was required or provided.
Technological Characteristics
Multi-purpose diagnostic ultrasound system. Transducer types include linear array, phased sector array, and intracardiac catheter. Modes: A, B, M, PWD, CWD, Color Doppler, Amplitude Doppler, Harmonic imaging, Contrast agent imaging. Connectivity: Standalone. Safety standards: CSA C22.2 No. 601-1, EN 60601-1, EN 60601-1-2, AIUM/NEMA UD-2/UD-3, ISO 10993 (biocompatibility).
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in fetal, abdominal, intraoperative (cardiac), pediatric, neonatal cephalic, adult/pediatric cardiac, trans-esophageal, and peripheral vessel applications. Includes intra-luminal and intra-cardiac visualization (right heart only) via AcuNav catheter. Contraindications: None stated.
Regulatory Classification
Identification An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
ACUSON Cypress™ Ultrasound System (K042055 )
ACUSON Cypress™ Ultrasound System (K021497 )
Cypress™ Ultrasound System (K010950 )
Lynx Ultrasound System (K991872 )
Ecton Lynx Ultrasound System (K982800 )
ACUSON Sequoia™ Diagnostic Ultrasound System (K973767 )
Related Devices
K042055 — CYPRESS ULTRASOUND SYSTEM · Siemens Medical Solutions USA, Inc. · Aug 16, 2004
K042044 — ACUSON CV70 CARDIOVASCULAR SYSTEM · Siemens Medical Solutions USA, Inc. · Aug 13, 2004
K052410 — ACUSON SEQUOIA ULTRASOUND SYSTEM · Siemens Medical Solutions USA, Inc. · Sep 22, 2005
K041319 — SEQUOIA DIAGNOSTIC ULTRASOUND SYSTEM · Siemens Medical Solutions USA, Inc. · Jun 7, 2004
K080935 — SONIX MDP ULTRASOUND SCANNER · Ultrasonix Medical Corporation · Aug 6, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
ACUSON Cypress™ Ultrasound System Special 510(k) Submission
2331
# SECTION 11
# 510(k) Summary of Safety and Effectiveness
- Siemens Medical Solutions USA, Inc., Ultrasound Division Sponsor: 1230 Shorebird Way P.O. Box 7393 Mountain View, California 94039-7393
- Iskra Mraković Contact Person: Regulatory Affairs (650) 694-5004 Telephone: (650) 943-7053 Fax:
- August 25, 2005 Submission Date:
- Device Name: Cypress Ultrasound System
- Common Name: Diagnostic Ultrasound System with Accessories
Classification:
Regulatory Class: II Review Category: Tier II
### 21 CFR 892.1550
| | FR # | Product Code |
|------------------------------------------|----------|--------------|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
| Diagnostic Intravascular Catheter | 870.1200 | 90-DQO |
### Predicate Devices:
- # K042055 (August 16, 2004) cleared as ACUSON Cypress™ Ultrasound System. .
- # K021497 (July 9, 2002) cleared as ACUSON Cypress™ Ultrasound System. .
- # K010950 (June 27, 2001) cleared as Cypress™ Ultrasound System. .
- # K991872 (June 16, 1999) cleared as Lynx Ultrasound System. .
- # K982800 (September 22, 1998) cleared as Ecton Lynx Ultrasound System. .
- # K973767 (December 23, 1997) cleared as ACUSON Sequoia™ Diagnostic . Ultrasound System.
{1}------------------------------------------------
### Device Description:
R89257
2/3
The Cypress system is a multi-purpose diagnostic ultrasound system with accessories and proprietary software, and is substantially equivalent to our current product that is already cleared for USA distribution under the following 510(k) Premarket Notification number(s):
- # K042055 (August 16, 2004) cleared as Cypress™ Ultrasound System .
- # K021497 (July 9, 2002) cleared as Cypress™ Ultrasound System .
- # K010950 (June 27, 2001) cleared as Cypress™ Ultrasound System .
- # K991872 (June 16, 1999) cleared as Lynx Ultrasound System .
- # K982800 (September 22, 1998) cleared as Ecton Lynx Ultrasound System .
The Cypress Ultrasound System has been designed to conform to the following product safety standards:
- CSA C22.2 No. 601-1, Safety Requirements for Medical Equipment ●
- AIUM/NEMA UD-2, 1998, Acoustic Output Measurement Standard for . Diagnostic Ultrasound
- AIUM/NEMA UD-3, 1998, Standard for Real Time Display of Thermal and . Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
- 93/42/EEC Medical Device Directive .
- Safety and EMC Requirements for Medical Equipment ●
- EN 60601-1 .
- EN 60601-1-1, Safety Requirements for Medical Equipment .
- EN 60601-1-2 .
- EN 60601-1-2-37 ●
- ISO 10993 Biocompatibility .
The system's acoustic output is in accordance with ALARA principle (as low as reasonably achievable)
### Intended Use:
The Cypress platform is intended for use in the following applications:
General Imaging and Cardiology for Fetal, Abdominal, Intraoperative (cardiac), Pediatrics, Neonatal Cephalic, Cardiac (adult, pediatric), Trans-esophageal, Peripheral Vessel, Intra-luminal and Intra-cardiac applications, and intended uses as defined in the FDA guidance document.
The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
{2}------------------------------------------------
# Technological Comparison to Predicate Device:
K052331
3/3
The Cypress is substantially equivalent in its technologies and functionality to the Cypress Ultrasound System that is already cleared under 510(k) premarket notification number K042055.
The Cypress functions in the same manner as other diagnostic ultrasound systems, in that they transmit ultrasonic energy into the body via a transducer. In the body, acoustic impedance of different tissues reflect different amounts of ultrasound energy back to the transducer, where post processing of received echoes is performed to generate twodimensional on-screen images of anatomic structures and fluid flow within the body. Doppler principles are used to process reflected ultrasound energy to display moving blood as a spectrum, or as color-coded two-dimensional images. All predicate device(s) listed above, allow for specialized measurements of structures and flow, and provide various calculations' functions.
Remainder of the page left blank intentionally.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and head. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV - 9 2005
Siemens Medical Solutions USA, Inc. c/o Ms. Iskra Mrakovic Manager, Regulatory Affairs Ultrasound Division 1230 Shorebird Way Mountain View CA 94043
### Re: K052331
Trade Name: Acuson Cypress Ultrasound System Regulation Number: 21 CFR 892.1550, 21 CFR 892.1560 and 21 CFR 892.1570 Regulation Name: Ultrasonic Pulsed Doppler Imaging System, Ultrasonic Pulsed Echo Imaging System and Diagnostic Ultrasonic Transducer Regulatory Class: II (Two) Product Code: 90 IYN, IYO, and ITX Dated: August 25, 2005 Received: August 26, 2005
Dear Ms. Mrakovic:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Iskra Mrakovic
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Brimmer for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# SECTION 6
# Diagnostic Ultrasound Indications for Use Form
510(k) Number (if known): Ko52331
Device Name:
# ACUSON Cypress™ ultrasound system
Intended Use:
Diagnostic imaging or fluid flow analysis of the human body as follows:
| | | | | | | | Mode of Operation | | | | |
|----------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | P | P | P | P | P | P | | | Note 3 | |
| Abdominal | | P | P | P | P | P | P | | | Note 3 | |
| Intraoperative<br>(Note 1) | | P | P | P | P | P | P | | | Note 3 | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Pediatric | | P | P | P | P | P | P | | | Note 3 | |
| Small Organ | | | | | | | | | | | |
| Neonatal Cephalic | | P | P | P | P | P | P | | | Note 3 | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | P | P | P | P | P | P | | | Note 3,4 | |
| Transesophageal | | P | P | P | P | P | P | | | Note 3,4 | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral vessel | | P | P | P | P | P | P | | | Note 3 | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other (Note 2) | | P | P | P | P | P | P | | | | |
Previously cleared under K042593, K042055, K021497, K010950, K991872, and K982800.
Note 1 For example: cardiac
Intra-Luminal, Intra-Cardiac Note 2
Note 3 Harmonic imaging
Note 4 Contrast agent imaging
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
B.Vammumton
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K052331
{6}------------------------------------------------
### Diagnostic Ultrasound Indications for Use Form
510(k) Number (if known): Ko52331
Device Name:
7L3 Linear Array Transducer for use with: ACUSON Cypress ultrasound system
Intended Use:
Diagnostic imaging or fluid flow analysis of the human body as follows:
| | | | | | | Mode of Operation | | | | |
|----------------------------------|---|---|---|-----|-----|-------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | | Note 3 |
| Abdominal | | P | P | P | | P | P | | | Note 3 |
| Intraoperative<br>(Note 1) | | P | P | P | | P | P | | | Note 3 |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | P | P | P | | P | P | | | Note 3 |
| Small Organ | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | P | P | P | | P | P | | | Note 3,4 |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | | P | P | | | Note 3 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
Previously cleared under # K021497.
Note 1 For example: cardiac
Note 3 Harmonic imaging
Note 4 Contrast agent imaging
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Blumman
(Division Sign-Off ardiovi 510/k)
{7}------------------------------------------------
### Diagnostic Ultrasound Indications for Use Form
510(k) Number (if known): Kos 2331
Device Name:
led Uses:
3V2c Phased Sector Array Transducer for use with ACUSON Cypress ultrasound system
Intended Use:
Diagnostic imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | | | | | | | | | | |
|----------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | P | P | P | | | Note 3 |
| Abdominal | | P | P | P | P | P | P | | | Note 3 |
| Intraoperative<br>(Note 1) | | P | P | P | P | P | P | | | Note 3 |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | P | P | P | P | P | P | | | Note 3 |
| Small Organ | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | P | P | P | P | P | P | | | Note 3,4 |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | P | P | P | | | Note 3 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
P = previously cleared under # K982800 and #K991872.
Note 1 For example: cardiac
Note 3 Harmonic imaging
Note 4 Contrast agent imaging
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Blummenstor
n Sign-Om islon of Cardlovascular Devices
{8}------------------------------------------------
-
## Diagnostic Ultrasound Indications for Use Form
# 510(k) Number (if known): Ko52331
Device Name:
7 V3c Phased Sector Array Transducer for use with; ACUSON Cypress ultrasound system
Intended Use:
Diagnostic imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | P | P | P | | | Note 3 |
| Abdominal | | P | P | P | P | P | P | | | Note 3 |
| Intraoperative<br>(Note 1) | | P | P | P | P | P | P | | | Note 3 |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | P | P | P | P | P | P | | | Note 3 |
| Small Organ | | | | | | | | | | |
| Neonatal Cephalic | | P | P | P | P | P | P | | | Note 3 |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | P | P | P | P | P | P | | | Note 3,4 |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | P | P | P | | | Note 3 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
Previously cleared under #K982800 and #K991872
Note 1 For example: cardiac
Note 3 Harmonic imaging
Note 4 Contrast agent imaging
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
B.Bmmmmma
(Division Sign-Off)
{9}------------------------------------------------
### Diagnostic Ultrasound Indications for Use Form
# 510(k) Number (if known): Kos 2331
Device Name:
AcuNav Ultrasound Catheter for use with; ACUSON Cypress ultrasound system
Intended Use:
For intracardiac and intra-luminal visualization of cardiac and (great vessel anatomy and physiology, and visualization of other devices in the heart - use in right heart only.
| | | Mode of Operation | | | | | | | | | |
|----------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other (Note 2) | | P | P | P | P | P | P | | | | |
Previously cleared under # K010950, #K033650, and #K042593.
Note 2 Intra-Luminal, Intra-Cardiac
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Blummo
Division Slan-O Division of Cardiovascular Devices
{10}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
510(k) Number (if known): Ko5233/
Device Name:
Aux CW Transducer for use with: ACUSON Cypress ultrasound system
Intended Use:
Diagnostic imaging or fluid flow analysis of the human body as follows:
| | | Mode of Operation | | | | | | | | |
|----------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | P | | | | | |
| Abdominal | | | | | P | | | | | |
| Intraoperative | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | P | | | | | |
| Small Organ | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | P | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | | | | P | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
Previously cleared under # K021497.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Bhimmer
(Division Sign-Off) Division of Cardlovascular Devices 510(k) Num
{11}------------------------------------------------
# Diagnostic Ultrasound Indications for Use Form
# 510(k) Number (if known): Ko5 2331
Device Name:
V5Ms Phased Sector Array TEE Transducer for use with; ACUSON Cypress ultrasound system
Intended Use:
Diagnostic imaging or fluid flow analysis of the human body as follows:
| | | Mode of Operation | | | | | | | | | |
|----------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intraoperative<br>(Note 1) | | P | P | P | P | P | P | | | Note 3 | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | P | P | P | P | P | P | | | Note 3,4 | |
| Transesophageal | | P | P | P | P | P | P | | | Note 3,4 | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
Previously cleared under #K973767 (on ACUSON Sequoia); TEE transducer indications cleared for use on Cypress via #K982800 and subsequent 510(k)s.
Note 1 For example: cardiac
Harmonic imaging Note 3
Note 4 Contrast agent imaging
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign Off
{12}------------------------------------------------
ACUSON Cypress™ Ultrasound System Special 510(k) Submission
### Diagnostic Ultrasound Indications for Use Form
# 510(k) Number (if known): Ko5 2331
Device Name:
5.0 MHz Biplane TEE Transducer for use with: ACUSON Cypress ultrasound system
Intended Use:
Diagnostic imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>(Note 1) | | P | P | P | P | P | P | | | Note 3 |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | P | P | P | P | P | P | | | Note 3,4 |
| Transesophageal | | P | P | P | P | P | P | | | Note 3,4 |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
Previously cleared under # K982800 and #K991872.
Note I For example: cardiac
Note 3 Harmonic imaging
Contrast agent imaging Note 4
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
K052331
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number
{13}------------------------------------------------
### Diagnostic Ultrasound Indications for Use Form
# 510(k) Number (if known): Ko52331
Device Name:
5.0 MHz Monoplane TEE Transducer for use with: ACUSON Cypress ultrasound system
Intended Use:
Diagnostic imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>(Note 1) | | P | P | P | P | P | P | | | Note 3 |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | P | P | P | P | P | P | | | Note 3,4 |
| Transesophageal | | P | P | P | P | P | P | | | Note 3,4 |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
Previously cleared under #K982800 and #K991872.
Note l For example: cardiac
Note 3 Harmonic imaging
Note 4 Contrast agent imaging
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Blammura
vision Sign-Off Ivision of Cardlovascular Devices 510(k) Numbe
{14}------------------------------------------------
### Diagnostic Ultrasound Indications for Use Form
# 510(k) Number (if known): Ko5 233/
Device Name:
## 4C1 Curvilinear Array Transducer for use with: ACUSON Cypress ultrasound system
Intended Use:
Diagnostic imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | | | Note 3 |
| Abdominal | | N | N | N | | N | N | | | Note 3 |
| Intraoperative<br>(Note 1) | | N | N | N | | N | N | | | Note 3 |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | | N | N | | | Note 3 |
| Small Organ | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | N | N | N | | N | N | | | Note 3,4 |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | N | N | N | | N | N | | | Note 3 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = new indication
Note 1 For example: cardiac
Note 3 Harmonic imaging
Note 4 Contrast agent imaging
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
> > Prescription Use (Per 21 CFR 801.109)
Blumman
Division Slan-Off Division of Cardiovascular Devices 510(k) Number
Indications for Use Form