← Product Code [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN) · K043196

# ALOKA SSD-ALPHA 10 ULTRASOUND SYSTEM (K043196)

_Aloka Co., Ltd. · IYN · Nov 30, 2004 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K043196

## Device Facts

- **Applicant:** Aloka Co., Ltd.
- **Product Code:** [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN.md)
- **Decision Date:** Nov 30, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric, 3rd-Party Reviewed

## Intended Use

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: [Table of clinical applications including Fetal, Abdominal, Intraoperative, Pediatric, Small Organ, Neonatal Cephalic, Cardiac, Transrectal, Transvaginal, Peripheral Vascular, and Musculo-skeletal Conventional, with modes including B, M, PWD, CWD, Color Doppler, and Amplitude Doppler].

## Device Story

The Aloka SSD-Alpha 10 is a diagnostic ultrasound system using piezo-electric transducers to transmit high-frequency sound waves into the body. The system detects reflected ultrasound energy based on tissue acoustic impedance to generate images; it also utilizes Doppler shift from moving tissues (blood) to display flow. Operated by clinicians in clinical settings, the system processes signals to provide real-time visualization of internal structures and blood flow. Healthcare providers use these images to aid in the diagnosis of disease or abnormalities. The device supports various modes including B-mode, M-mode, Pulsed Wave Doppler (PWD), Continuous Wave Doppler (CWD), Color Doppler, and Amplitude Doppler, as well as mixed-mode operations.

## Clinical Evidence

Bench testing only. The device complies with NEMA-UD2: 1992, AIUM 1994, IEC-60601-1, UL 2601-1, and ISO 10993-1 standards. Acoustic output levels are within FDA-allowed limits.

## Technological Characteristics

Piezo-electric transducer technology; supports B, M, PWD, CWD, Color Doppler, and Amplitude Doppler modes. Complies with IEC-60601-1, UL 2601-1, and ISO 10993-1 for electrical safety and biocompatibility. System console processes reflected ultrasound signals.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- Aloka SSD-5500 ([K032875](/device/K032875.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

043196

Date: 10/21/04

NOV 3 0 2004

### 510(k) Summary: Aloka Model SSD- Alpha 10

This summary statement complies with 21 CFR, section 807.92 as amended March 14, 1995.

This premarket notification has been submitted by Aloka Co., Ltd. and covers the Aloka SSD-Alpha 10 diagnostic ultrasound system and transducers. The address is:

10 Fairfield Boulevard Wallingford, CT 06492 (203) 269-5088

The contact person is: Richard J. Cehovsky, RA/QA Coordinator

The proprietary name is the Aloka SSD-Alpha 10 diagnostic ultrasound system and transducers. The common name for this type of device is a diagnostic ultrasound system and transducers.

The item in this submission is covered under the following classification:

| 90 IYN | Ultrasonic Pulsed Doppler Imaging System | 21 CFR 892.1550 |
|--------|------------------------------------------|-----------------|
| 90 ITX | Diagnostic Ultrasound Transducer         | 21 CFR 892.1570 |
| 90 IYO | Ultrasonic Pulsed Echo Imaging System.   | 21 CFR 892.1560 |

The above as stated in 21CFR, part 892.1570,1560 & 1550, has been classified as regulatory Class II.

The Aloka SSD- Alpha 10 and its transducers are substantially equivalent to its predicate; the Aloka SSD-5500 (K032875) and its transducers.

The Aloka SSD-Alpha 10 functions in the same manner as its predicate and other Aloka diagnostic ultrasound devices. High frequency sound waves are transmitted into the body by a piezo-electric transducer. In the body, differences in the acoustic impedance of different tissues reflect a certain amount of the ultrasound energy back to the transducer, where it is transmitted via the probe cable to the system console and processed into an image. The Aloka SSD-Alpha 10 transducers can also use the Doppler shift of sound reflected from moving tissues (blood) to detect and display flow.

The Aloka SSD-Alpha 10, like other Aloka marketed diagnostic ultrasound systems and transducers are indicated for imaging body structures to aid in the diagnosis of disease or abnormality.

The Aloka SSD-Alpha 10 diagnostic ultrasound system and transducers are similar in technological characteristics to its predicate system: SSD-5500 (K032875).

- The Aloka SSD-Alpha 10 is indicated for the same diagnostic ultrasound applications to Aloka's . ultrasound system: SSD-5500 (K032875).
- The Aloka SSD-Alpha 10 has the same gray-scale and Doppler abilities to Aloka's ultrasound . system as mentioned above.

{1}------------------------------------------------

# 510(k) Summary: Aloka Model SSD- Alpha 10

- . The SSD-Alpha 10 uses the same technologies for imaging, Doppler functions and signal processing as the following product currently marketed by Aloka : SSD-5500 (K032875).
- The SSD-Alpha 10 has the same method of use as the following product currently marketed by . Aloka: SSD-5500 (K032875).
- . The SSD-Alpha 10 acoustic power output levels are below the maximum levels allowed by the FDA.
- The SSD- Alpha 10 is subjected to the same Quality Assurance systems in development and . production as other products including the SSD-5500 (K032875) currently marketed by Aloka,
- . The patient contact materials used in the SSD-Alpha 10 have been evaluated for safety via the same standards and methods as the above mentioned product marketed by Aloka. These materials have been found to be safe for their intended uses,
- . The SSD-Alpha 10 complies with electrical and physical safety standards as other products currently marketed by Aloka such as the: SSD-5500 (K032875).
- Aloka Co., Ltd. Certifies that the SSD-Alpha 10 complies with NEMA-UD2: 1992, AIUM 1994 . "Acoustic Output Labeling Standard for Diagnostic Ultrasound Equipment", IEC-60601-1 (2001-09 Class A), UL 2601-1, 2nd edition (1997), Part 1, 2nd edition including Amendments 1&2 and ISO 10993-1:1997. All testing will be completed, prior to distribution, to meet the requirements of the standards listed above.

{2}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wing-like extensions, representing the department's mission of protecting the health of all Americans. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 3 0 2004

· Aloka Co., Ltd. % Mr. Daniel W. Lehtonen Staff Engineer - Medical Devices Intertek Testing Services NA, Inc. 70 Codman Hill Road BOXBOROUGH MA 01719

Re: K043196

Trade Name: Aloka Model SSD-Alpha 10 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulatory Name: Ultrasonic pulsed echo imaging system Regulatory Number: 21 CFR 892.1570 Regulatory Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: 90 IYN, IYO, and ITX Dated: November 17, 2004 Received: November 18, 2004

#### Dear Mr. Lehtonen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the Aloka Model SSD-Alpha 10 Diagnostic Ultrasound System, as described in your premarket notification:

#### Transducer Model Number

| ASU-1010 |
|----------|
| ASU-1012 |

UST-547 UST-675P

{3}------------------------------------------------

| UST-5293-5 | UST-9118  |
|------------|-----------|
| UST-5411   | UST-9120  |
| UST-5412   | UST-9128  |
| UST-5543   | UST-9130  |
| UST-5712   | UST-52101 |
| UST-5713T  | UST-52108 |
| UST-9115-5 |           |

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.

The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:

> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

[f you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled,

{4}------------------------------------------------

Page 2 - Mr. Lehtonen

《Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small ・ Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.

Sincerely yours,

Nancy C Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{5}------------------------------------------------

#### 4.3.1

# Diagnostic Ultrasound Indications for Use Form SSD-Alpha 10

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Modes of operation |   |   |     |     |                  |                      |                              |                       |                    |
|----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application             | A                  | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Opthalmic                        |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Fetal                            |                    | N | N | N   |     | N                | N                    |                              | See Below             |                    |
| Abdominal                        |                    | N | N | N   |     | N                | N                    |                              | See Below             |                    |
| Intraoperative (specify)         |                    | N | N | N   |     | N                | N                    |                              | See Below             |                    |
| Intraoperative Neurological      |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Pediatric                        |                    | N | N | N   |     | N                | N                    |                              | See Below             |                    |
| Small Organ (specify)            |                    | N | N | N   |     | N                | N                    |                              | See Below             |                    |
| Neonatal Cephalic                |                    | N | N | N   |     | N                | N                    |                              | See Below             |                    |
| Adult Cephalic                   |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Cardiac                          |                    | N | N | N   | N   | N                | N                    |                              | See Below             |                    |
| Transesophageal                  |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transrectal                      |                    | N | N | N   |     | N                | N                    |                              | See Below             |                    |
| Transvaginal                     |                    | N | N | N   |     | N                | N                    |                              | See Below             |                    |
| Transurethral                    |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Intravascular                    |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Peripheral Vascular              |                    | N | N | N   |     | N                | N                    |                              | See Below             |                    |
| Laparoscopic                     |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Conventional |                    | N | N | N   |     | N                | N                    |                              | See Below             |                    |
| Musculo-skeletal<br>Superficial  |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Other                            |                    |   |   |     |     |                  |                      |                              |                       |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix A

Additional Comments: Mixed mode operation includes BM, BPWD, MCD, BICD/PWD, BICD/PWD

Applications: Small Parts-breast/tested thyroid, Intra-operative- liver/ pascreas/ gall-bladder/ abdominal, gyncological, fetal, noonala, cardiac.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801 , 109

801.109)
Daniel A. Lynn

(Division Sign-Off) Division of Reproductive, Ab and Radiological Devices 510(k) Number _

{6}------------------------------------------------

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Modes of operation |   |   |     |     |                  |                      |                              |                       |                    |
|----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application             | A                  | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Opthalmic                        |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Fetal                            |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Abdominal                        |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Intraoperative (specify)         |                    | P | P | P   |     | P                | P                    |                              | See Below             |                    |
| Intraoperative Neurological      |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Pediatric                        |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Small Organ (specify)            |                    | P | P | P   |     | P                | P                    |                              | See Below             |                    |
| Neonatal Cephalic                |                    | P | P | P   |     | P                | P                    |                              | See Below             |                    |
| Adult Cephalic                   |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Cardiac                          |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transesophageal                  |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transrectal                      |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transvaginal                     |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transurethral                    |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Intravascular                    |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Peripheral Vascular              |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Laparoscopic                     |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Conventional |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Superficial  |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Other                            |                    |   |   |     |     |                  |                      |                              |                       |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments: Mixed mode operation includes BM, BPWD, M/CD, B/CD/PWD

Applications: Small Organ- breast,testes & thyroid, Intra-Operative- liver, pancrease & gall-bladder

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel A. Lennon

(Division Sign-Off) Division of Reproductive, Abdominal, ana Padiological Devices 5 I (K) Number __

{7}------------------------------------------------

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Modes of operation |   |   |     |     |                  |                      |                              |                       |                    |
|----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application             | A                  | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Opthalmic                        |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Fetal                            |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Abdominal                        |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Intraoperative (specify)         |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Intraoperative Neurological      |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Pediatric                        |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Small Organ (specify)            |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Neonatal Cephalic                |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Adult Cephalic                   |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Cardiac                          |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transesophageal                  |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transrectal                      |                    | P | P | P   |     | P                | P                    |                              | See Below             |                    |
| Transvaginal                     |                    | P | P | P   |     | P                | P                    |                              | See Below             |                    |
| Transurethral                    |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Intravascular                    |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Peripheral Vascular              |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Laparoscopic                     |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Conventional |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Superficial  |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Other                            |                    |   |   |     |     |                  |                      |                              |                       |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments: Mixed mode operation includes BM, BIPWD, MCD, B/CD/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David br. Lyman

(Division Sign-Off) Division of Reproductive, Abdo and Radiological Devices 510(k) Number __

{8}------------------------------------------------

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Modes of operation |   |   |     |     |                  |                      |                              |                       |                    |
|----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application             | A                  | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Opthalmic                        |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Fetal                            |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Abdominal                        |                    | N | N | N   |     | N                | N                    |                              | See Below             |                    |
| Intraoperative (specify)         |                    | N | N | N   |     | N                | N                    |                              | See Below             |                    |
| Intraoperative Neurological      |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Pediatric                        |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Small Organ (specify)            |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Neonatal Cephalic                |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Adult Cephalic                   |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Cardiac                          |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transesophageal                  |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transrectal                      |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transvaginal                     |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transurethral                    |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Intravascular                    |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Peripheral Vascular              |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Laparoscopic                     |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Conventional |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Superficial  |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Other                            |                    |   |   |     |     |                  |                      |                              |                       |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments: Mixed mode operation includes BM, BPWD, MCD, BCD/PWD

Applications: Abdominal, Gynecological, Fetal.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

01.109)
David A. Sypm

(Division Sign-Off) Division of Reproductive, Abdomina and Radiological Devices 510(k) Number __

{9}------------------------------------------------

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Modes of operation |   |   |     |     |                  |                      |                              |                       |                    |
|----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application             | A                  | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Opthalmic                        |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Fetal                            |                    | N | N | N   |     | N                | N                    |                              | See Below             |                    |
| Abdominal                        |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Intraoperative (specify)         |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Intraoperative Neurological      |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Pediatric                        |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Small Organ (specify)            |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Neonatal Cephalic                |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Adult Cephalic                   |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Cardiac                          |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transesophageal                  |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transrectal                      |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transvaginal                     |                    | N | N | N   |     | N                | N                    |                              | See Below             |                    |
| Transurethral                    |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Intravascular                    |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Peripheral Vascular              |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Laparoscopic                     |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Conventional |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Superficial  |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Other                            |                    |   |   |     |     |                  |                      |                              |                       |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments: Mixed mode operation includes BM, B/WD, M/CD, B/CD/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 8017109)

David A. Segerson

(Division Sign-Off) Division of Reproductive, Ab and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

{10}------------------------------------------------

.

#### Diagnostic Ultrasound Indications for Use Form UST- 5293-5

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Modes of operation |   |   |     |     |                  |                      |                              |                       |                    |  |
|----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical Application             | A                  | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |  |
| Opthalmic                        |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Fetal                            |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Abdominal                        |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Intraoperative (specify)         |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Intraoperative Neurological      |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Pediatric                        |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Small Organ (specify)            |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Neonatal Cephalic                |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Adult Cephalic                   |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Cardiac                          |                    | N | N | N   | N   | N                | N                    |                              | See Below             |                    |  |
| Transesophageal                  |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Transrectal                      |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Transvaginal                     |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Transurethral                    |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Intravascular                    |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Peripheral Vascular              |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Laparoscopic                     |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Musculo-skeletal<br>Conventional |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Musculo-skeletal<br>Superficial  |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Other                            |                    |   |   |     |     |                  |                      |                              |                       |                    |  |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments: Mixed mode operation includes BM, BPWD, MCD, BICDPWD, BCD/CWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David An. Bergman

(Division Sign-Off) Division of Reproductive, Abdomina and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

{11}------------------------------------------------

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Modes of operation |   |   |     |     |                  |                      |                              |                       |                    |
|----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application             | A                  | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Opthalmic                        |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Fetal                            |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Abdominal                        |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Intraoperative (specify)         |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Intraoperative Neurological      |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Pediatric                        |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Small Organ (specify)            |                    | N | N | N   |     | N                | N                    |                              | See Below             |                    |
| Neonatal Cephalic                |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Adult Cephalic                   |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Cardiac                          |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transesophageal                  |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transrectal                      |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transvaginal                     |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transurethral                    |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Intravascular                    |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Peripheral Vascular              |                    | N | N | N   |     | N                | N                    |                              | See Below             |                    |
| Laparoscopic                     |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Conventional |                    | N | N | N   |     | N                | N                    |                              | See Below             |                    |
| Musculo-skeletal<br>Superficial  |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Other                            |                    |   |   |     |     |                  |                      |                              |                       |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments: Mixed mode operation includes B/M, B/PWD, M/CD, B/CD/PWD

Application: Small Parts: breasts, testes & thyroid

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

801.109)

Division Sign Off)

(Division Sign-Off) (Division of Reproductive, Abdomina and Radiological Devices 510/k) Number ________________________________________________________________________________________________________________________________________________________________

{12}------------------------------------------------

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Modes of operation |   |   |     |     |                  |                      |                              |                       |                    |  |
|----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical Application             | A                  | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |  |
| Opthalmic                        |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Fetal                            |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Abdominal                        |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Intraoperative (specify)         |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Intraoperative Neurological      |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Pediatric                        |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Small Organ (specify)            |                    | N | N | N   |     | N                | N                    |                              | See Below             |                    |  |
| Neonatal Cephalic                |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Adult Cephalic                   |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Cardiac                          |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Transesophageal                  |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Transrectal                      |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Transvaginal                     |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Transurethral                    |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Intravascular                    |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Peripheral Vascular              |                    | N | N | N   |     | N                | N                    |                              | See Below             |                    |  |
| Laparoscopic                     |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Musculo-skeletal<br>Conventional |                    | N | N | N   |     | N                | N                    |                              | See Below             |                    |  |
| Musculo-skeletal<br>Superficial  |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Other                            |                    |   |   |     |     |                  |                      |                              |                       |                    |  |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments: Mixed mode operation includes BM, B/WD, M/CD, B/CD/PWD

Applications: Small Parts- breast, testes and thyroid.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel G. Kenyon

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

{13}------------------------------------------------

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Modes of operation |   |   |     |     |                  |                      |                              |                       |                    |
|----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application             | A                  | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Opthalmic                        |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Fetal                            |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Abdominal                        |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Intraoperative (specify)         |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Intraoperative Neurological      |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Pediatric                        |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Small Organ (specify)            |                    | P | P | P   |     | P                | P                    |                              | See Below             |                    |
| Neonatal Cephalic                |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Adult Cephalic                   |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Cardiac                          |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transesophageal                  |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transrectal                      |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transvaginal                     |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transurethral                    |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Intravascular                    |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Peripheral Vascular              |                    | P | P | P   |     | P                | P                    |                              | See Below             |                    |
| Laparoscopic                     |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Conventional |                    | P | P | P   |     | P                | P                    |                              | See Below             |                    |
| Musculo-skeletal<br>Superficial  |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Other                            |                    |   |   |     |     |                  |                      |                              |                       |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments: Mixed mode operation includes B/M, B/PWD, M/CD, B/CD/PWD

Applications: Small Parts- breast, testes and thyroid.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

CFR 801.109)

(Division Sign-Off)

Division of Reproductive, Abdominal, diciogical Devices ar : ్నా x: vumber ____________________________________________________________________________________________________________________________________________________________________

{14}------------------------------------------------

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Modes of operation |   |   |     |     |                  |                      |                              |                       |                    |  |
|----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical Application             | A                  | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |  |
| Opthalmic                        |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Fetal                            |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Abdominal                        |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Intraoperative (specify)         |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Intraoperative Neurological      |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Pediatric                        |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Small Organ (specify)            |                    | P | P | P   |     | P                | P                    |                              | See Below             |                    |  |
| Neonatal Cephalic                |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Adult Cephalic                   |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Cardiac                          |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Transesophageal                  |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Transrectal                      |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Transvaginal                     |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Transurethral                    |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Intravascular                    |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Peripheral Vascular              |                    | P | P | P   |     | P                | P                    |                              | See Below             |                    |  |
| Laparoscopic                     |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Musculo-skeletal<br>Conventional |                    | P | P | P   |     | P                | P                    |                              | See Below             |                    |  |
| Musculo-skeletal<br>Superficial  |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Other                            |                    |   |   |     |     |                  |                      |                              |                       |                    |  |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments: Mixed mode operation includes BM, B/WD, M/CD, B/CD/PWD

Applications: Small Parts- breast, testes and thyroid.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

David R. Ingram
(Division Sign-Off)

Division Sign-Off)
A
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K043196

28

{15}------------------------------------------------

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Modes of operation |   |   |     |     |                  |                      |                              |                       |                    |  |
|----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical Application             | A                  | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |  |
| Opthalmic                        |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Fetal                            |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Abdominal                        |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Intraoperative (specify)         |                    | P | P | P   |     | P                | P                    |                              | See Below             |                    |  |
| Intraoperative Neurological      |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Pediatric                        |                    | P | P | P   |     | P                | P                    |                              | See Below             |                    |  |
| Small Organ (specify)            |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Neonatal Cephalic                |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Adult Cephalic                   |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Cardiac                          |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Transesophageal                  |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Transrectal                      |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Transvaginal                     |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Transurethral                    |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Intravascular                    |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Peripheral Vascular              |                    | P | P | P   |     | P                | P                    |                              | See Below             |                    |  |
| Laparoscopic                     |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Musculo-skeletal<br>Conventional |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Musculo-skeletal<br>Superficial  |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Other                            |                    |   |   |     |     |                  |                      |                              |                       |                    |  |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments: Mixed mode operation includes BM, B/WD, M/CD, B/CD/PWD

Applications: Small Parts- breast, testes and thyroid, Intra-Operative- liver, pancreas & gall-bladder

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

David A. Sayre

(Division Sign-Off) Division of Reproductive, Abdomi anni finincical Devices 5 Van Number _________________________________________________________________________________________________________________________________________________________________

{16}------------------------------------------------

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Modes of operation |   |   |     |     |                  |                      |                              |                       |                    |
|----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application             | A                  | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Opthalmic                        |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Fetal                            |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Abdominal                        |                    | P | P | P   |     | P                | P                    |                              | See Below             |                    |
| Intraoperative (specify)         |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Intraoperative Neurological      |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Pediatric                        |                    | P | P | P   |     | P                | P                    |                              | See Below             |                    |
| Small Organ (specify)            |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Neonatal Cephalic                |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Adult Cephalic                   |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Cardiac                          |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transesophageal                  |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transrectal                      |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transvaginal                     |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Transurethral                    |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Intravascular                    |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Peripheral Vascular              |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Laparoscopic                     |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Conventional |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Superficial  |                    |   |   |     |     |                  |                      |                              |                       |                    |
| Other                            |                    |   |   |     |     |                  |                      |                              |                       |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments: Mixed mode operation includes BM, BRWD, MICD, BCDPWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

David A. Ingram

(Division Sign-Off) Division of Reproductive, Abdomina and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

{17}------------------------------------------------

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Modes of operation |   |   |     |     |                  |                      |                              |                       |                    |  |
|----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical Application             | A                  | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |  |
| Opthalmic                        |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Fetal                            |                    | P | P | P   |     | P                | P                    |                              | See Below             |                    |  |
| Abdominal                        |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Intraoperative (specify)         |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Intraoperative Neurological      |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Pediatric                        |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Small Organ (specify)            |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Neonatal Cephalic                |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Adult Cephalic                   |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Cardiac                          |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Transesophageal                  |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Transrectal                      |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Transvaginal                     |                    | P | P | P   |     | P                | P                    |                              | See Below             |                    |  |
| Transurethral                    |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Intravascular                    |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Peripheral Vascular              |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Laparoscopic                     |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Musculo-skeletal<br>Conventional |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Musculo-skeletal<br>Superficial  |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Other                            |                    |   |   |     |     |                  |                      |                              |                       |                    |  |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments: Mixed mode operation includes BM, BRWD, MCD, B/CD/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel A. Seymour

(Division Sign-Off) Division of Reproductive, Abdomina and Radiological Devices ", "The Number________________________________________________________________________________________________________________________________________________________________

{18}------------------------------------------------

.

# Diagnostic Ultrasound Indications for Use Form UST- 9120

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Modes of operation |   |   |     |     |                  |                      |                              |                       |                    |  |
|----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical Application             | A                  | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |  |
| Opthalmic                        |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Fetal                            |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Abdominal                        |                    |   |   |     |     |                  |                      |                              |                       |                    |  |
| Intraoperative (specify)         |                    | P | P | P   |     | P                | P                    |                              | See Below             |                    |  |
| Intraoperative Neurological      |                    |   |   |…

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K043196](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K043196)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
