Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: [Table of clinical applications including Fetal, Abdominal, Pediatric, Small Organ (breast, testes & thyroid), Neonatal Cephalic, Adult Cephalic, Cardiac, Transvaginal, Peripheral Vascular, and Musculo-skeletal Superficial].
Device Story
Aloka SSD-Alpha 5 is a diagnostic ultrasound system using piezo-electric transducers to transmit high-frequency sound waves into the body. System detects reflected ultrasound energy based on tissue acoustic impedance differences; processes signals into images. Doppler shift technology used to detect and display blood flow. Operated by clinicians in clinical settings. Output displayed on console for diagnostic interpretation to aid in disease or abnormality identification. System supports B-mode, M-mode, Pulsed Wave Doppler (PWD), Continuous Wave Doppler (CWD), Color Doppler, and Amplitude Doppler modes, including mixed-mode operations.
Clinical Evidence
Bench testing only. No clinical data provided. Compliance with acoustic output labeling (NEMA-UD2: 1992, AIUM 1994), electrical safety (IEC-60601-1, UL 2601-1), and biocompatibility (ISO 10993-1) standards demonstrated.
Technological Characteristics
Diagnostic ultrasound system using piezo-electric transducers. Supports B, M, PWD, CWD, Color Doppler, and Amplitude Doppler modes. Complies with NEMA-UD2: 1992, AIUM 1994, IEC-60601-1 (2001-09 Class A), UL 2601-1 (1997), and ISO 10993-1:1997. Patient contact materials evaluated for safety. System console architecture with probe connectivity.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in patients requiring evaluation of abdominal, gynecological, fetal, small organ (breast, testes, thyroid), neonatal cephalic, adult cephalic, cardiac, peripheral vascular, and superficial musculo-skeletal structures.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
K043196 — ALOKA SSD-ALPHA 10 ULTRASOUND SYSTEM · Aloka Co., Ltd. · Nov 30, 2004
K032875 — ALOKA MODEL SSD-5500 V6.0 ULTRASOUND SYSTEM · Aloka Co. Ltd USA · Sep 30, 2003
K060059 — ALOKA SD-3500 ULTRASOUND SYSTEM · Aloka Co., Ltd. · Jan 13, 2006
K033311 — ALOKA SSD-5000 V5.0 ULTRASOUND SYSTEM · Aloka Co., Ltd. · Oct 29, 2003
K972465 — ALOKA SSD-1400 DIAGNOSTIC ULTRASOUND SYSTEM · Aloka Co., Ltd. · Aug 15, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
K04/1916
:
540(k)
Aloka Co., Ltd.
·
・
Model SSD- Alpha 5
้JUL 3 0 2004
## Section 4.2.13b
# 510(k) Summary of S&E
( Immediately follows this page )
12
:
{1}------------------------------------------------
Date: 6/8/04
#### 510(k) Summary: Aloka Model SSD- Alpha 5
This summary statement complies with 21 CFR, section 807.92 as amended March 14, 1995.
This premarket notification has been submitted by Aloka Co., Ltd. and covers the Aloka SSD-Alpha 5 Ver. 1.0 diagnostic ultrasound system and transducers. The address is:
> 10 Fairfield Boulevard Wallingford, CT 06492 (203) 269-5088
The contact person is: Richard J. Cehovsky, RA/QA Coordinator
The proprietary name is the Aloka SSD-Alpha 5 diagnostic ultrasound system and transducers. The common name for this type of device is a diagnostic ultrasound system and transducers.
The item in this submission is covered under the following classification:
| 90 IYN | Ultrasonic Pulsed Doppler Imaging System | 21 CFR 892.1550 |
|--------|------------------------------------------|-----------------|
| 90 ITX | Diagnostic Ultrasound Transducer | 21 CFR 892.1570 |
| 90 IYO | Ultrasonic Pulsed Echo Imaging System. | 21 CFR 892.1560 |
The above as stated in 21CFR, part 892.1570.1560 & 1550, has been classified as regulatory Class II.
The Aloka SSD- Alpha 5 and its transducers are substantially equivalent to its predicate; the Aloka SSD-5500 (K032875) and its transducers.
The Aloka SSD-Alpha 5 functions in the same manner as its predicate and other Aloka diagnostic ultrasound devices. High frequency sound waves are transmitted into the body by a piezo-electric transducer. In the body, differences in the acoustic impedance of different tissues reflect a certain amount of the ultrasound energy back to the transducer, where it is transmitted via the probe cable to the system console and processed into an image. The Aloka SSD-Alpha 5 transducers can also use the Doppler shift of sound reflected from moving tissues (blood) to detect and display flow.
The Aloka SSD-Alpha 5, like other Aloka marketed diagnostic ultrasound systems and transducers are indicated for imaging body structures to aid in the diagnosis of disease or abnormality.
The Aloka SSD-Alpha 5 diagnostic ultrasound system and transducers are similar in technological characteristics to its predicate system: SSD-5500 (K032875).
- The Aloka SSD-Alpha 5 is indicated for the same diagnostic ultrasound applications to Aloka's . ultrasound system: SSD-5500 (K032875).
- The Aloka SSD-Alpha 5 has the same gray-scale and Doppler abilities to Aloka's ultrasound . system as mentioned above.
{2}------------------------------------------------
#### 510(k) Summary: Aloka Model SSD- Alpha 5
- The SSD-Alpha 5 uses the same technologies for imaging, Doppler functions and signal processing . as the following product currently marketed by Aloka : SSD-5500 (K032875).
- The SSD-Alpha 5 has the same method of use as the following product currently marketed by . Aloka: SSD-5500 (K032875).
- The SSD-Alpha 5 acoustic power output levels are below the maximum levels allowed by the . FDA.
- The SSD- Alpha 5 is subjected to the same Quality Assurance systems in development and . production as other products including the SSD-5500 (K032875) currently marketed by Aloka.
- The patient contact materials used in the SSD-Alpha 5 have been evaluated for safety via the same . standards and methods as the above mentioned product marketed by Aloka. These materials have been found to be safe for their intended uses.
- The SSD-Alpha 5 complies with electrical and physical safety standards as other products . currently marketed by Aloka such as the: SSD-5500 (K032875).
- Aloka Co., Ltd. Certifies that the SSD-Alpha 5 complies with NEMA-UD2: 1992, AIUM 1994 . "Acoustic Output Labeling Standard for Diagnostic Ultrasound Equipment", IEC-60601-1 (2001-09 Class A), UL 2601-1, 200 edition (1997), Part 1, 200 edition including Amendments 1&2 and ISO10993-1:1997. All testing will be completed, prior to distribution, to meet the requirements of the standards listed above.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circle around the eagle. The logo is black and white and has a simple, clean design.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 3 0 2004
Aloka Co., Ltd. % Mr. Daniel W. Lehtonen Staff Engineer - Medical Devices Intertek Testing Services NA, Inc. 70 Codman Hill Road BOXBOROUGH MA 01719
Re: K041916
Trade Name: Aloka SSD-Alpha 5 Ultrasound System Regulation Number: 21 CFR §892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR §892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR §892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: 90 IYN, IYO, and ITX Dated: July 15, 2004 Received: July 16, 2004
Dear Mr. Lehtonen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Aloka SSD-Alpha 5 Ultrasound System, as described in your premarket notification:
#### Transducer Model Number
| UST-5410 | UST-9128 |
|----------|-----------|
| UST-9118 | UST-52101 |
| UST-9126 | UST-52108 |
{4}------------------------------------------------
## Page 2 - Mr. Lehtonen
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device it inay be subject to sach additional Regulations, Title 21, Parts 800 to 898. In addition, FDA can be found in the Court of receits concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I lease be advised that I Dr over device complies with other requirements of the Act that I DA has made a active regulations administered by other Federal agencies. You must or any I cachar statutes and regulations and limited to: registration and listing (21 Comply with an the Hot 3 requirements)01); good manufacturing practice requirements as set CFK i at 607), aboning (21 OF X 1 at 800); and if applicable, the electronic forul in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping I ms device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested information, merading accepter's September 30, 1997 "Information for Manufacturers m Appendix &, (encrease of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and rne special roperty marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the econotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
{5}------------------------------------------------
Page 3 - Mr. Lehtonen
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
Sincerely yours,
David h. byssom
for Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{6}------------------------------------------------
#### 4.3.1
## Diagnostic Ultrasound Indications for Use Form SSD-Alpha 5
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Modes of operation | | | | | | | | | | |
|----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | |
| Opthalmic | | | | | | | | | | | |
| Fetal | | N | N | N | N | N | N | | See Below | | |
| Abdominal | | N | N | N | N | N | N | | See Below | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Pediatric | | N | N | N | N | N | N | | See Below | | |
| Small Organ (specify) | | N | N | N | N | N | N | | See Below | | |
| Neonatal Cephalic | | N | N | N | N | N | N | | See Below | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | N | N | N | N | N | N | | See Below | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | N | N | N | N | N | N | | See Below | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | N | N | N | | See Below | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | N | N | N | N | N | N | | See Below | | |
| Other | | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix A
Additional Comments: Mixed mode operation includes BM, BPWD, MCD, B/CD/PWD, B/CWD, B/CD/CWD Applications: Small Parts-breast, testes & thyroid, abdominal, gynecological, fetal, neonatal, cardiac.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (C
Prescription Use (Per 21 CFR 801,109)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices K041916
510(k) Number
·
{7}------------------------------------------------
Intended Use Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Modes of operation | | | | | | | | | |
|----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Opthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | N | N | N | N | N | N | | See Below | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | N | N | N | | See Below | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | N | N | N | N | N | N | | See Below | |
| Other | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: Mixed mode operation includes BM, BPWD, MCD, BCDTWD, BCDCWD
Applications: breast, testes and thyroid.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Usc (Pcr 21 CFR 801 109)
FR 801.109)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K041916
{8}------------------------------------------------
Intended Use Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Modes of operation | | | | | | | | | |
|----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Opthalmic | | | | | | | | | | |
| Fetal | | P | P | P | P | P | P | | See Below | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | P | P | P | P | P | P | | See Below | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: Mixed mode operation includes BM, BPWD, MCD, B/CD/PWD, B/CD/CWD,
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)
Nancy C Brogdon
(Division Sign-O Division of Reproducti and Radiological Devices 510(k) Number
{9}------------------------------------------------
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Modes of operation | | | | | | | | | |
|----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | P | P | P | | See Below | |
| Abdominal | | P | P | P | P | P | P | | See Below | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: Mixed mode operation includes BM, BPWD, MCD, BCDTWD, BICWD, BICDCWD.
Applications: Abdominal, Gynecological, Fetal.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)
Nancy C. Swagdon
Division of Reproduct and Radiological Devic 510(k) Number
{10}------------------------------------------------
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows i
| | Modes of operation | | | | | | | | | | |
|----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | |
| Opthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | P | P | P | P | P | P | | See Below | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Pediatric | | P | P | P | P | P | P | | See Below | | |
| Small Organ (specify) | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascı·lar | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other | | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: Mixed mode operation includes BM, BPWD, MCD, B/CD/WD, B/CD/WD,
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Nancy Brogdon
(Division Sign Division of Reproductive, Abdomi and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________
{11}------------------------------------------------
Intended Use Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows i
| | Modes of operation | | | | | | | | | |
|----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Opthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | P | P | P | P | P | P | | See Below | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: Mixed mode operation includes BM, BPWD, MCD, B/CD/WWD, B/CWD, B/CWD, B/CWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)
Nancy C. Brogdon
(Division Sian-Of Division of Reproductive and Radiological Devices 510(k) Number _
{12}------------------------------------------------
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Modes of operation | | | | | | | | | | |
|---------------------------------------------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | |
| Opthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | | |
| Neonatal Cephalic | | P | P | P | P | P | P | | See Below | | |
| Adult Cephalic | | P | P | P | P | P | P | | See Below | | |
| Cardiac | | P | P | P | P | P | P | | See Below | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional<br>Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other | | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: Mixed mode operation includes BAM, BPWD, MCD, B/CD/TWD, B/CD/CWD
Application: Neonatal
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)
Nancy hroador
(Division Sign-Off) Division of Reproductive, I and Radiological Devices 510(k) Number ..
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.