ULTRASOUND SCANNER, TYPE 2400

K024236 · B-K Medical A/S · IYN · Jan 17, 2003 · Radiology

Device Facts

Record IDK024236
Device NameULTRASOUND SCANNER, TYPE 2400
ApplicantB-K Medical A/S
Product CodeIYN · Radiology
Decision DateJan 17, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric

Intended Use

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Abdominal, Cardiac, Fetal Doppler, Intraoperative, Neurosurgery, Obstetrics, Pediatrics, Transrectal, Small Parts (organs), Transvaginal, Peripheral vascular, Musculo-skeletal.

Device Story

Ultrasound Scanner Type 2400 is a diagnostic ultrasound system supporting B-mode, M-mode, PWD, and CFM (Color/Amplitude Doppler) imaging. It utilizes linear, convex, and mechanical sector transducers to acquire ultrasound signals. The system features a digital beamformer and modified data acquisition techniques to process inputs. It supports optional ECG signal superimposition. The device performs geometric measurements and calculations for vascular, urology, cardiology, and OB/GYN applications; it also guides biopsy and puncture needles. Operated by clinicians in clinical settings, the system displays images and measurements to assist in diagnostic decision-making. Patient benefits include non-invasive visualization of internal structures and blood flow.

Clinical Evidence

Bench testing only. No clinical data provided. Safety and performance were verified through acoustic output measurements compliant with NEMA 1997 standards and IEC 60601-1 electrical/thermal safety testing. Acoustic output levels are maintained at Ispta ≤ 720 mW/cm² and MI ≤ 1.9.

Technological Characteristics

System utilizes digital beamforming, linear/convex/mechanical sector transducers, and supports B, M, PWD, and CFM modes. Patient contact materials comply with ISO 10993-1. Energy source is electrical; system is a Track 3 diagnostic ultrasound device. Connectivity includes optional ECG input. Software performs geometric measurements and calculations. Sterilization is not specified for the console; transducers are reusable.

Indications for Use

Indicated for diagnostic ultrasound imaging and fluid flow analysis in patients requiring abdominal, cardiac, fetal, intraoperative, neurosurgical, obstetric, pediatric, transrectal, small organ, transvaginal, peripheral vascular, or musculoskeletal assessment.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K.024236 # 510(k) Summary: This summary is provided as part of this Premarket Notification in compliance with 21CRF, Section 807.92. JAN 1 7 2003 Submitters name: B-K Medical A/S Address: Mileparken 34, DK2730 Gentofte, Denmark Phone: +45 44528100 Fax: +45 44528199 Contact person: Villy Braender, Quality Assurance Mnager Date prepared: 12 October, 2002 Trade name: Ultrasound Scanner Type 2400 Common name: Diagnostic Ultrasound System Classification names: Ultrasonic Pulsed Echo Imaging System (90 IYO, CFR 892.1560) Ultrasonic Pulsed Doppler Imaging System (90 IYN, CFR 892.1560) Diagnostic Ultrasonic Transducer (90 ITX, CFR 892.1570) Identification of predicate, legally marketed device: B-K Medical A/S 2102 Ultrasound Scanner K000567(MAR9 2000), K003986 (AUG21 2001), K011417 (JUN8 2001) ## Device description: 2400 supports the following scanning modes and combinations thereof: B-mode M-mode PWD mode and CFM mode. An optional ECG signal can be superimposed the ultrasound information in all modes and mode combinations. The system can perform simple geometric measurements, and perform calculations in the areas of Vascular, Urology, Cardiology and OB/GYN applications. The system can guide biopsy- and puncture needles. # Transducers Transducers are linear and convex arrays and mechanical sector. The patient contact materials comply with ISO10993-1 All transducers used together with 2400 are Track 3 transducers. ### Acoustic output The system controlling the Acoustic Output in 2400 is the same as the system in 2102. The system will assure that the acoustic output always will stay below the pre-amendments upper limits i.e. Ispta ≤ 720 mW/cm² and MI ≤ 1.9 (Track 3, non ophthalmic). The Thermal Index values are maximum 6.0, i.e. TI ≤ 6.0 #### Clinical measurement accuracy. Clinical measurements and calculations are described and accuracies are provided in the User Guide. Thermal, mechanical and electrical safety. The scanner 2400 has been tested by a recognized, certified body according to IEC 60601-1 . {1}------------------------------------------------ The Acoustic Output Reporting is made according to the standards required by "Information for Manufacturers Seeking Clearance of Diagnostic Ultrasound Systems and Transducers, FDA, CDRH, September 30, 1997" The acoustic output is measured and calculated according to: "Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment" (NEMA 1997). # Intended use. 2400 intended uses are contained within 2102-intended uses: | | Predicate device: | Submitted device: | |---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Ultrasound scanner Type 2102 | Ultrasound scanner Type | | | K000567(MAR9 2000), K003986 (AUG21 | 2400 | | | 2001), K011417 (JUN8 2001) | | | Modes of operation | B, M, PWD, CFM 1) and combinations.<br>Tissue harmonic imaging | B, M, PWD,CFM 1) and<br>combinations. Tissue<br>harmonic imaging | | Intended use(clinical<br>application) | Abdominal<br>Cardiac<br>Fetal Doppler<br>Intraoperative<br>Neurosurgery<br>Obstetrics<br>Pediatrics<br>Transrectal<br>Small Parts (organs)<br>Transvaginal<br>Peripheral vascular<br>Musculo-skeletal | Abdominal<br>Cardiac<br>Fetal Doppler<br>Intraoperative<br>Neurosurgery<br>Obstetrics<br>Pediatrics<br>Transrectal<br>Small Parts (organs)<br>Transvaginal<br>Peripheral vascular<br>Musculo-skeletal | | Features | ECG (not monitoring) | ECG (not monitoring) | 1) CFM= Color Flow Mapping=Color Doppler and Amplitude Doppler. # Technological characteristics compared to the predicate device. The predicate device has the same major technological characteristics as the subject device described above. Minor differences consist: Digital beamformer, modified data acquisition technique, modified user interface and mechanical outline. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem that resembles a stylized eagle or bird in flight, composed of three curved lines. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # JAN 1 7 2003 Mr. Villy Brænder Official Correspondent B-K Medical A/S Mileparken 34 DK-2730 Herlev DENMARK Re: K024236 Trade Name: Ultrasound Scanner Type 2400 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: 90 IYN, IYO, and ITX Dated: December 19, 2002 Received: December 23, 2002 #### Dear Mr. Brænder: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. This determination of substantial equivalence applies to the following transducers intended for use with the Ultrasound Scanner Type 2400, as described in your premarket notification: # Transducer Model Number | 1850 | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 8661 | | ANALA ------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>8662 | {3}------------------------------------------------ # 8803 8811 If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded. The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to: > Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850 This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. {4}------------------------------------------------ Page 3 - Mr. Brænder Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212. Sincerely yours, Qurid d. Eignon Nancy C. Brogdor Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure(s) {5}------------------------------------------------ K0242367p1 #### Diagnostic Ultrasound Indications for Use Form # System: 2400 #### Fill out one form for each ultrasound system and each transducer. Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | |----------------------------------|-------------------|---|---|-----|-------------------------------|------------------|----------------------|------------------------------|-------------------------|--------------------| | Clinical Application | A | B | M | PWD | Tissue<br>harmonic<br>imaging | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify 1) | Other<br>(specify) | | Ophthalmic | | | | | | | | | | | | Fetal | | × | × | × | × | × | × | | × | | | Abdominal | | × | × | × | × | × | × | | × | | | Intraoperative (specify) | | × | × | × | × | × | × | | × | | | Intraoperative Neurological | | × | × | × | × | × | × | | × | | | Pediatric | | × | × | × | × | × | × | | × | | | Small Organ (specify) | | × | × | × | × | × | × | | × | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | × | × | × | × | × | × | | × | | | Transesophageal | | | | | | | | | | | | Transrectal | | × | × | × | × | × | × | | × | | | Transvaginal | | × | × | × | × | × | × | | × | | | Transurethral | | × | × | × | × | × | × | | × | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | × | × | × | × | × | × | | × | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | × | × | × | × | × | × | | × | | | Musculo-skeletal Superficial | | × | × | × | × | × | × | | × | | | Other (specify) | | | | | | | | | | | N= new indication; P= previously cleared by FDA; E= added under Appendix E 1) B+M, B+D, B+C, B+D+C. B mode includes Tissue Harmonic Imaging. Additional Comments:_ D is PWD, C is Color Doppler. > (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) David A. Ingram Division Sign Off 510(k) {6}------------------------------------------------ K024i236 7 p.2 #### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM System: 2400 Transducer: 1850 (with interchangeable probes 8539,6004,6005) Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation | | | | | | | |---------------------------|------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------| | General<br>(Track 1 Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) | | Ophthalmic | Ophthalmic | | | | | | | | | Fetal Imaging<br>& Other | Fetal | | | | | | | | | | Abdominal | | | | | | | | | | Intra-operative (Specify) | P | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | Laparoscopic | | | | | | | | | | Pediatric | | | | | | | | | | Small Organ (Specify) | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | P | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | Musculo-skel. (Conventional) | | | | | | | | | | Musculo-skel. (Superficial) | | | | | | | | | | Intra-luminal | | | | | | | | | | Other (Specify) | | | | | | | | | Cardiac | Cardiac Adult | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | Other (Specify) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | Other (Specify) | | | | | | | | N= new indication; P= previously cleared by FDA(K002085); E= added under Appendix E *Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging Additional Comments: Intraoperative: Rectum, Urethra, Urinary bladder, (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation David H. Stamm (Division Sign-Off) Division of Reproductive, Abdominal and Radiological Device 510(k) Number . {7}------------------------------------------------ K024236 | System: | 2400 | |-------------|------| | Transducer: | 8661 | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation | | | | | | | |---------------------------|------------------------------|-------------------|---|-----|-----|------------------|-----------------------|------------------------| | General<br>(Track I Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Amplitude<br>e Doppler | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal | P | | P | | P | P 1) | P | | | Abdominal | | | | | | | | | | Intra-operative (Specify) | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | Laparoscopic | | | | | | | | | Fetal Imaging<br>& Other | Pediatric | | | | | | | | | | Small Organ (Specify) | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | P | | P | | P | P 1) | P | | | Trans-vaginal | P | | P | | P | P 1) | P | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | Musculo-skel. (Conventional) | | | | | | | | | | Musculo-skel. (Superficial) | | | | | | | | | | Intra-luminal | | | | | | | | | | Other (Specify) | | | | | | | | | Cardiac | Cardiac Adult | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | Other (Specify) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | Other (Specify) | | | | | | | | N= new indication; P= previously cleared by FDA; E= added under Appendix E * Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging Additional Comments: l)}Mode combinations: B+M, B+D, B+C, B+D+C. (D is PWD, C is Color Flow mapping Doppler including Amplitude(power)Doppler) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation David A. Bergman (Division Sign-Division of Reproductive and Radiologica 510(k) Numbe {8}------------------------------------------------ KO2H236 | System: | 2400 | |-------------|------| | Transducer: | 8662 | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation | | | | | | | |---------------------------|------------------------------|-------------------|---|-----|-----|------------------|-------------------------|------------------------| | General<br>(Track 1 Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify 1) | Amplitude<br>e Doppler | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal | | | | | | | | | | Abdominal | | | | | | | | | | Intra-operative (Specify 2) | E | E | E | | E | E | E | | | Intra-operative (Neuro) | E | E | E | | E | E | E | | | Laparoscopic | | | | | | | | | Fetal Imaging<br>& Other | Pediatric | E | E | E | | E | E | E | | | Small Organ (Specify) | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | Musculo-skel. (Conventional) | | | | | | | | | | Musculo-skel. (Superficial) | | | | | | | | | | Intra-luminal | | | | | | | | | | Other (Specify) | | | | | | | | | | Cardiac Adult | | | | | | | | | Cardiac | Cardiac Pediatric | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | Other (Specify) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | E | E | E | | E | E | E | | | Other (Specify) | | | | | | | | N= new indication; P= previously cleared by FDA; E= added under Appendix E * Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Tissue Motion Doppler, Color Velocity Imaging 1) Mode combinations: B+M, B+D, B+C, B+D+C. (D is PWD, C is Color Flow Additional Comments: mapping Doppler including Amplitude(power)Doppler) 2)Intraoperative: Gall bladder (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation David the hyrum (Division Sign-Off) Division of Reproductive, Al and Radiological Devic 510(k) Number {9}------------------------------------------------ K024236 | System: | 2400 | |-------------|------| | Transducer: | 8803 | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation | | | | | | | |---------------------------|------------------------------|-------------------|---|-----|-------------------------------|------------------|-----------------------|----------------------| | General<br>(Track I Only) | Specific<br>(Tracks I & III) | B | M | PWD | Tissue<br>harmonic<br>imaging | Color<br>Doppler | Combined<br>(Specify) | Amplitude<br>Doppler | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal | E | | E | N | E | E 1) | E | | | Abdominal | E | | E | N | E | E 1) | E | | | Intra-operative (Specify) | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | Laparoscopic | | | | | | | | | Fetal Imaging<br>& Other | Pediatric | E | | E | N | E | E 1) | E | | | Small Organ (Specify) | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | Musculo-skel. (Conventional) | | | | | | | | | | Musculo-skel. (Superficial) | | | | | | | | | | Intra-luminal | | | | | | | | | | Other (Specify) | | | | | | | | | | Cardiac Adult | E | | E | N | E | E 1) | E | | Cardiac | Cardiac Pediatric | E | | E | N | E | E 1) | E | | | Trans-esoph. (Cardiac) | | | | | | | | | | Other (Specify) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | Other (Specify) | | | | | | | | N= new indication; P= previously cleared by FDA; E= added under Appendix E * Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging Additional Comments: 1) ) Mode combinations: B+M, B+D, B+C, B+D+C. (D is PWD, C is Color Flow mapping Doppler including Amplitude(power)Doppler) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation David A. Ingram (Division Sign-Off) Division of Reproductive, Abdo and Radiological Devic 510(k) Number {10}------------------------------------------------ K024236 "1 | System: | 2400 | |-------------|------| | Transducer: | 8811 | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Clinical Application | | Mode of Operation | | | | | | | |--------------------------|------------------------------|------------------------------|-------------------|---|-----|-------------------------------|------------------|-------------------------|----------------------| | | General<br>(Track I Only) | Specific<br>(Tracks I & III) | B | M | PWD | Tissue<br>harmonic<br>imaging | Color<br>Doppler | Combined<br>(Specify 1) | Amplitude<br>Doppler | | Ophthalmic | Ophthalmic | | | | | | | | | | | Fetal | | | | | | | | | | | Abdominal | | | | | | | | | | | Intra-operative (Specify 2) | E | E | E | E | E | E | E | | | | Intra-operative (Neuro) | | | | | | | | | | | Laparoscopic | | | | | | | | | | Fetal Imaging<br>& Other | Pediatric | E | E | E | E | E | E | E | | | | Small Organ (Specify 3) | E | E | E | E | E | E | E | | | | Neonatal Cephalic | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | Trans-rectal | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | Trans-urethral | | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | | Musculo-skel. (Conventional) | E | E | E | E | E | E | E | | | | Musculo-skel. (Superficial) | E | E | E | E | E | E | E | | | | Intra-luminal | | | | | | | | | | | Other (Specify) | | | | | | | | | | Cardiac | Cardiac Adult | | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | | Other (Specify) | | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | E | E | E | E | E | E | E | | | | Other (Specify) | | | | | | | | | N= new indication; P= previously cleared by FDA; E= added under Appendix E *Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging Additional Comments: 2)Intraoperative: Breast, liver, pancreas, biliary system 3)Small Organ: Breast, testis, penis, thyroid, parathyroid, salivary glands, lymph nodes 1 ) mode combinations: B, B+M, B+C, B+D+C. (D is PWD, C is Color Flow mapping Doppler including Amplitude(power)Doppler) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation David A. Bergman Division of Renroductive. Abde and Radiological Device 510(k) Number
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...