Transducer Probe Cover

K241615 · Hony Medical Co., Ltd. · ITX · Jul 3, 2024 · Radiology

Device Facts

Record IDK241615
Device NameTransducer Probe Cover
ApplicantHony Medical Co., Ltd.
Product CodeITX · Radiology
Decision DateJul 3, 2024
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1570
Device ClassClass 2

Indications for Use

Transducer Probe Cover placed over diagnostic ultrasound transducer/ probe scan head instruments. The cover allows use of the transducer in scanning and needle guided procedures for body surface, endocavity, and intra-operative diagnostic ultrasound, while helping to prevent transfer of microorganisms, body fluids, and particulate material to the patient and healthcare worker during reuse of the transducer. The cover also provides a means for maintenance of a sterile field. Transducer Probe Cover are furnished sterile; single use patient/procedure, disposable.

Device Story

Transparent polyurethane sleeve; tubular shape; heat-sealed distal end; applied over ultrasound transducer probe. Used in body surface, endocavity, and intra-operative diagnostic ultrasound. Operator (clinician) applies sterile coupling gel to transducer face or inside cover, inserts transducer, and unrolls cover over probe length; secures with bands. Acts as physical barrier preventing transfer of microorganisms, body fluids, and particulates between patient and probe. Facilitates easier cleaning/disinfection; maintains sterile field. Single-use, disposable; EO sterilized. No impact on ultrasound imaging performance.

Clinical Evidence

No clinical data. Bench testing only. Biocompatibility evaluation performed per ISO 10993-1, including ISO 10993-4 (Thrombogenicity and Complement Activation), ISO 10993-5, ISO 10993-10, and ISO 10993-11. Microbial barrier performance validated per ASTM F1671-13.

Technological Characteristics

Polyurethane film (0.05mm, 49 GSM); tubular form factor. EO sterilization (SAL 10^-6). Biocompatibility per ISO 10993-1/4/5/10/11. Microbial barrier per ASTM F1671-13. Acoustic impedance 1.60 x 10^6 Pa s/m; velocity 1594 m/s; attenuation 0.01 dB/(cm-MHz).

Indications for Use

Indicated for use as a protective barrier over diagnostic ultrasound transducers during body surface, endocavity, and intra-operative ultrasound scanning and needle-guided procedures to prevent cross-contamination and maintain a sterile field.

Regulatory Classification

Identification

A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ July 3, 2024 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. Hony Medical Co., Ltd. % Boyle Wang General Manager Shanghai Truthful Information Technology Co., Ltd. RM.1801, No. 161, East Lujiazui Rd. Pudong, Shanghai 200120 CHINA Re: K241615 Trade/Device Name: Transducer Probe Cover Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic Ultrasonic Transducer Regulatory Class: Class II Product Code: ITX Dated: June 5, 2024 Received: June 5, 2024 Dear Boyle Wang: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, #### Marjan Nabili -S for Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging And Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ # Indications for Use Submission Number (if known) K241615 Device Name Transducer Probe Cover #### Indications for Use (Describe) Transducer Probe Cover placed over diagnostic ultrasound transducer/ probe scan head instruments. The cover allows use of the transducer in scanning and needle quided procedures for body surface, endocavity, and intra-operative diagnostic ultrasound, while helping to prevent transfer of microorganisms, body fluids, and particulate material to the patient and healthcare worker during reuse of the transducer. The cover also provides a means for maintenance of a sterile field. Transducer Probe Cover are furnished sterile; single use patient/procedure, disposable. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary K241615 This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92. ## 1.0 Submitter's Information | Name: | Hony Medical Co., Ltd. | |----------|-----------------------------------------------------------------------| | Address: | No.12,9 Road, Changlong Ind.Zone, Taishan City,Guangdong China 529222 | | Tel: | +86 15916327827 | | Contact: | Zhu Huina | #### Designated Submission Correspondent | Contact: | Mr. Boyle Wang | |----------|------------------------------------------------------------------------| | Name: | Shanghai Truthful Information Technology Co., Ltd. | | Address: | Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai, 200120<br>China | | Tel: | +86-21-50313932 | | Email: | Info@truthful.com.cn | Date of Preparation: Jun.27,2024 ## 2.0 Device Information | Trade name: | Transducer Probe Cover | |----------------------|----------------------------------------------| | Common name: | Ultrasonic Diagnostic Transducer Probe Cover | | Classification name: | Transducer, Ultrasonic, Diagnostic. | | Model(s): | Various Dimension | | Production code: | ITX | | Regulation number: | 21 CFR 892.1570 | | Classification: | Class II | | Panel: | Radiology | #### 3.0 Predicate and Reference Device Information ## Predicate# | Manufacturer: | Hony Medical Co., Ltd. | |---------------|------------------------| | Trade Device: | Transducer Probe Cover | {4}------------------------------------------------ 510(k) number: K221278 #### Reference# | Manufacturer: | CIVCO MEDICAL INSTRUMENTS CO., INC. | |----------------|-------------------------------------| | Trade Device: | GENERAL PURPOSE TRANDUCER COVER | | 510(k) number: | K970513 | #### 4.0 Device Description The subject device is composed of transparent thin-walled polyurethane sleeve, rubber ring and (or) tape strip. A conventional 0.05mm thin, 49 GSM (Grams per Square Meter), transparent high strength polyurethane film tube shape, in various dimensions with heat sealed distal end to be applied over a transducer probe to provide a Transducer Cover that can be used to minimize contamination between patient and ultrasound probe during ultrasound scanning procedures for body surface, endocavity and intra-operative diagnostic ultrasound. This may help with easier cleaning and disinfection of the probe. Ultrasound imaging is not impaired by use of the cover as it is intended. Adequate coupling between the cover and the transducer is required. The Transducer Probe Cover is utilized by applying sterile transmission, coupling, or lubricating gel onto the transducer face or into closed end of cover, inserting ultrasound transducer into closed end of cover and unrolling cover over length of the transducer as desired, and securing open end of cover with bands as necessary the removal process is accomplished by pulling the cover off the transducer in a reverse method from the application. The subject device is furnished in sterile condition, for single use patient/procedure use, disposable. As the device is single use device, which is individually packaged sterile devices. The packaging is compatible with the product's EO sterilization method. The sterilization validation confirms the packaging is qualified bacterial film to maintain the sterilization condition of the device. #### 5.0 Indication for Use Statement Transducer Probe Cover placed over diagnostic ultrasound transducer/ probe scan head instruments. The cover allows use of the transducer in scanning and needle guided procedures for body surface, endocavity, and intra-operative diagnostic ultrasound, while helping to prevent transfer of microorganisms, body fluids, and particulate material to the patient and healthcare worker during reuse of the transducer. The cover also provides a means for maintenance of a sterile field. {5}------------------------------------------------ Transducer Probe Cover are furnished sterile; single use patient/procedure, disposable. ### 6.0 Summary of Non-Clinical Testing In this current submission, just expand the intended use scope of the subject device. As there were no materials, structure, dimensions or performance changes made to the subject device when compared to the primary predicate device, no additional non-clinical testing was considered necessary to support the substantial equivalence. #### Performance Bench Testing As there are no any technological characteristics changes as part of this submission from the previous clearance, no performance bench testing was included as part of this submission. There are no differences between the predicate Transducer Probe Cover and the proposed Transducer Probe Cover that could affect the shelf life or sterility; therefore, sterilization and shelf-life evaluations were not required. ## Biocompatibility Testing The proposed Transducer Probe Cover does not introduce any new direct or indirect patient contacting materials, as compared to the previous clearance predicate Transducer Probe Cover. Since the subject device expand the indication for use scope, it can be used in endocavity, and intra-operative diagnostic ultrasound, additionally biocompatibility evaluation Items including Thrombogenicity and Complement Activation Testing Per ISO 10993-4 were performed due to the contact level of tissue. And the biocompatibility test results reviewed by the FDA for the predicate Transducer Probe Cover (reference K221278) also valid, can be support the proposed Transducer Probe Cover. {6}------------------------------------------------ ## 7.0 Technological Characteristic Comparison Table | Item | Subject Device | Predicate Device | Reference Device | Conclusion | |---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) No. | K241615 | K221278 | K970513 | -- | | Product Code | ITX | ITX | ITX | Same | | Regulation No. | 21 CFR 892.1570 | 21 CFR 892.1570 | 21 CFR 892.1570 | Same | | Class | II | II | II | Same | | Intended<br>Use/Indication for<br>Use | Transducer Probe Cover placed over<br>diagnostic ultrasound transducer/<br>probe scan head instruments. The<br>cover allows use of the transducer in<br>scanning and needle guided<br>procedures for body surface,<br>endocavity and intra-operative<br>diagnostic ultrasound, while helping to<br>prevent transfer of microorganisms,<br>body fluids, and particulate material to<br>the patient and healthcare worker<br>during reuse of the transducer. The<br>cover also provides a means for<br>maintenance of a sterile field.<br>Transducer Probe Cover are furnished<br>sterile; single use patient/procedure,<br>disposable. | Transducer Probe Cover placed over<br>diagnostic ultrasound transducer/<br>probe scan head instruments. The<br>cover allows use of the transducer in<br>scanning and needle guided<br>procedures for external intact skin<br>diagnostic ultrasound, while helping to<br>prevent transfer of microorganisms,<br>body fluids, and particulate material to<br>the patient and healthcare worker<br>during reuse of the transducer. The<br>cover also provides a means for<br>maintenance of a sterile field.<br>Transducer Probe Cover are furnished<br>sterile; single use patient/procedure,<br>disposable. | Protective cover or sheath placed<br>over diagnostic ultrasound<br>transducer / probe scan head<br>instruments. The cover allows use of<br>the transducer in scanning and<br>needle guided procedures for body<br>surface, endocavity, and<br>intra-operative diagnostic ultrasound,<br>while helping to prevent transfer of<br>microorganisms, body fluids , and<br>particulate material to the patient and<br>healthcare worker during reuse of<br>the transducer (both sterile and<br>non-sterile covers) . The cover also<br>provides a means for maintenance of<br>a sterile field (sterile covers only) .<br>CIVCO Poly Uitrasound Transducer | Compare to the predicate<br>device, the subject device<br>expanded the scope of<br>the indication for use. The<br>subject device<br>additionally can be used<br>for endocavity, and<br>intra-operative diagnostic<br>ultrasound.<br>Thrombogenicity and<br>Complement Activation<br>Testing were performed<br>due to the contact level<br>of tissue. | | Materials &<br>Construction | Polyurethane, tubular, sealed | Polyurethane, tubular, sealed | Covers are furnished sterile &<br>non-sterile single use patient /<br>procedure, disposable.<br>Polyurethane and polyethylene<br>extruded thermoplastic film, in<br>one-piece, open on one end, closed<br>on other end. | Same | | Model | Various Size | Various Size | Various Size | Same | | Microbial Barrier | Meets requirements of ASTM<br>F1671-13 for prevention of<br>blood-borne pathogens | Meets requirements of ASTM<br>F1671-13 for prevention of<br>blood-borne pathogens | Not Publicly Available | Same | | Acoustic<br>Performance | Acoustic Impedance: 1.60 x 106 Pa<br>s/m<br>Acoustic Velocity: 1594 m/s<br>Acoustic Attenuation:<br>0.01dB/(cm-MHZ) | Acoustic Impedance: 1.60 x 106 Pa<br>s/m<br>Acoustic Velocity: 1594 m/s<br>Acoustic Attenuation:<br>0.01dB/(cm-MHZ) | Not Publicly Available<br>Not Publicly Available<br>Not Publicly Available | Same | | Sterile | EO sterilization, SAL 10-6 | EO sterilization, SAL 10-6 | Both in EO sterilization and<br>non-sterile | Same | | Disposable,<br>Single Use Only | Yes | Yes | Yes | Same | | Shelf Life | 3 years | 3 years | Not Publicly Available | Same | | Biocompatibility | Conform with ISO10993-1<br>(ISO10993-4, ISO10993-5,<br>ISO10993-10, ISO10993-11) | Conform with ISO10993-1<br>(ISO10993-4, ISO10993-5,<br>ISO10993-10, ISO10993-11) | Conform with ISO 10993 standards | Thrombogenicity,<br>Complement Activation<br>Testing and Platelet<br>Activation Test per ISO | | | | | 10993-4<br>were performed due to<br>the contact level of<br>tissue. | | # Table 2- Comparison of Technology Characteristics {7}------------------------------------------------ {8}------------------------------------------------ {9}------------------------------------------------ Except the expand the applicable scope of indication for use for the subject device, there is no any changes made on the technological characteristics of the subject device are identical to those of predicate device. The materials, dimensions, sterilization methods, processing and manufacturing are no any changes were made to the subject device. The subject device is substantially equivalent to the predicate devices with respect to technological characteristics, and performance characteristics. The differences between the applicable scope of the proposed and predicate devices do not alter the substantially equivalent since additional biocompatibility testing evaluation has been performed. ## 8.0 Conclusion The conclusions drawn from the comparison and analysis above demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed predicated device and raises no new questions of safety or effectiveness. The differences between both devices are insignificant in terms of safety and effectiveness.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...