The Hitachi Aloka Medical, Ltd UST-5310 or UST-5311 or UST-5311-TIP transducers are intended for use with the PROSOUND ALPHA 6, PROSOUND ALPHA 7 and ARIETTA70 Diagnostic Ultrasound systems by trained personnel (doctor, sonographer, etc.) for the diagnostic ultrasound evaluation during intraoperative and Intra-operative (neurosurgery) procedures. This device is not indicated for Ophthalmic applications.
Device Story
Linear array ultrasound transducers (UST-5310, UST-5310-TIP, UST-5311, UST-5311-TIP) designed for intraoperative and neurosurgical imaging. Devices connect to PROSOUND ALPHA 6, PROSOUND ALPHA 7, and ARIETTA 70 ultrasound systems. Transducers capture acoustic signals, which are processed by the host ultrasound system to generate B-mode, M-mode, PWD, and Color Doppler images for real-time visualization of internal structures (e.g., liver, pancreas, gall bladder). Operated by physicians or sonographers in clinical/surgical settings. Output provides diagnostic visualization to assist clinical decision-making during procedures. Devices are sterile, single-use, and disposable.
Clinical Evidence
No clinical data required. Substantial equivalence supported by bench testing, including acoustic output measurements, biocompatibility testing per ISO 10993-1, electromagnetic compatibility, and electrical/mechanical safety evaluations.
Technological Characteristics
Linear array ultrasound transducers. Materials tested per ISO 10993-1. Operates with PROSOUND ALPHA 6, PROSOUND ALPHA 7, and ARIETTA 70 systems. Supports B-mode, M-mode, PWD, and Color Doppler. Sterile, single-use, disposable form factor. Connectivity via proprietary transducer interface to host ultrasound system.
Indications for Use
Indicated for diagnostic ultrasound evaluation during intraoperative and neurosurgical procedures. Intended for use by trained clinicians (doctors, sonographers). Not for ophthalmic use.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
Predicate Devices
UST-5310 or UST-5310-TIP / UST-5311 or UST-5311-TIP Intraoperative Ultrasound Transducer (K140854)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or a cape.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Hitachi Aloka Medical America, Inc. % Ms. Angela Van Arsdale RA/QA Manager 10 Fairfield Blvd. WALLINGFORD CT 06492
December 3, 2014
Re: K142618 Trade/Device Name: UST-5310 or UST-5310-TIP / UST-5311 or UST-5311-TIP Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: ITX Dated: September 12, 2014 Received: September 16, 2014
Dear Ms. Arsdale:
This letter corrects our substantially equivalent letter of October 16, 2014.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K142618
Device Name
UST-5310 or UST-5310-TIP / UST-5311 or UST-5311-TIP
# Indications for Use (Describe)
The Hitachi Aloka Medical, Ltd UST-5310 or UST-5311 or UST-5311-TIP transducers are intended for use with the PROSOUND ALPHA 6, PROSOUND ALPHA 7 and ARIETTA70 Diagnostic Ultrasound systems by trained personnel (doctor, sonographer, etc.) for the diagnostic ultrasound evaluation during intraoperative and Intra-operative (neurosurgery) procedures.
This device is not indicated for Ophthalmic applications.
Type of Use (Select one or both, as applicable)
x | Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
# DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM
Device Name: PROSOUND ALPHA 6, PROSOUND ALPHA 7 and ARIETTA 70 Transducer : UST -5310 or UST -5310-TIP Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | |
|---------------------------|--------------------------------|-------------------|---|-----|-----|------------------|------------------------|----------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Specify) | Other**<br>(Specify) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| Fetal Imaging<br>& Other | Fetal | | | | | | | | |
| | Abdominal | | | | | | | | |
| | Intra-operative (Specify)* | P | P | P | | P | P | P | |
| | Intra-operative (Neurosurgery) | P | P | P | | P | P | P | |
| | Laparoscopic ** | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (Specify)* | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | TEE (non-cardiac) | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | | |
| | Other: (Specify) * | | | | | | | | |
| | Other: Gynecological | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | |
| | Cardiac Adult, TEE | | | | | | | | |
| | Cardiac - Neonatal | | | | | | | | |
| | Cardiac - Pediatric | | | | | | | | |
| | Cardiac - Pediatric, TEE | | | | | | | | |
| Peripheral<br>Vessel | Peripheral Vascular | | | | | | | | |
| | Other (spec.) | | | | | | | | |
N = new indication. P = previously cleared by FDA via K140854
Combination of each operating mode includes:
* 1 Combination of each operating mode- B/M, B/PWD, M/CD, B/CD/PWD
**2 Includes: Mflow, B/Bflow, Power flow.
Intra-operative (Specify)* - (liver, pancreas, gall bladder..)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Prescription Use Only (Per 21 CFR 801.109)
> (Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)
Page 2 of 3
{4}------------------------------------------------
# DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM
Device Name: PROSOUND ALPHA 6, PROSOUND ALPHA 7 and ARIETTA 70 Transducer : UST -5311 or UST -5311-TIP Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|--------------------------------|-------------------|---|-----|-----|------------------|------------------------|----------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify)* | P | P | P | | P | P | P |
| | Intra-operative (Neurosurgery) | P | P | P | | P | P | P |
| | Laparoscopic** | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify)* | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | TEE (non-cardiac) | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Other: (Specify) * | | | | | | | |
| | Other: Gynecological | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Adult, TEE | | | | | | | |
| | Cardiac - Neonatal | | | | | | | |
| | Cardiac - Pediatric | | | | | | | |
| | Cardiac - Pediatric, TEE | | | | | | | |
| Peripheral<br>Vessel | Peripheral Vascular | | | | | | | |
| | Other (spec.) | | | | | | | |
N = new indication. P = previously cleared by FDA via K140854
Combination of each operating mode includes:
*1 Combination of each operating mode- B/M, B/PWD, M/CD, B/CD/PWD
**2 Includes: Mflow, B/Bflow, Power flow.
Intra-operative (Specify)* - (liver, pancreas, gall bladder..)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription Use Only (Per 21 CFR 801.109)
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)
Page 3 of 3
{5}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness in accordance with 21 CFR Part 807, Subpart E, Section 807.92.
#### 21 CFR 807.92, Subsection a
#### 1. Submitter's Information
Hitachi Aloka Medical America, Inc. 10 Fairfield Boulevard Wallingford, CT 06492-5903
Contact: Angela Van Arsdale RA/OA Manager Telephone: (203) 269-5088 Ext: 346 Fax Number: (203) 269-6075
Manufacturer: Hitachi Aloka Medical, Ltd. 6-22-1 Mure, Mitaka-Shi, 181-8622 Tokyo, Japan
Date Prepared: September 12, 2014
### 2. Device / Common / Classification Name / Classification / Product Code:
Device Proprietary Name - UST-5310 or UST-5310-TIP / UST-5311 or UST-5311-TIP Common name - Diagnostic Ultrasound Transducer Classification name - Diagnostic Ultrasonic Transducer Classification: Class II Product Code: 90-ITX 892.1570 Diagnostic Ultrasonic Transducer
### 3. Legally Marketed Predicate Device(s):
UST-5310 or UST-5310-TIP / UST-5311 or UST-5311-TIP Intraoperative Ultrasound Transducer for use with PROSOUND ALPHA 6 [K140854].
- 4. Device Description:
Linear Array transducer
- 5. Indication for Use:
The Hitachi Aloka Medical, Ltd UST-5310 or UST-5310-TIP / UST-5311-TIP transducers are intended for use with the PROSOUND ALPHA 6, PROSOUND ALPHA 7 and ARIETTA70 Diagnostic Ultrasound systems by trained personnel (doctor, sonographer, etc.) for the diagnostic ultrasound evaluation during intra- operative and Intraoperative (neurosurgery) procedures.
This device is not indicated for Ophthalmic applications.
- 6. Comparison to predicate device:
The Hitachi Aloka Medical, Ltd. UST-5310 or UST-5310-TIP / UST-5311-TIP Intraoperative transducers cleared via K140854 for use with the PROSOUND ALPHA 6 have not undergone any modifications, therefore there are no changes to either of the two transducers are being added for use with the PROSOUND ALPHA7 and ARIETTA70 Diagnostic Ultrasound systems. The subject systems: PROSOUND ALPHA7 & ARIETTA70 Diagnostic Ultrasound system and predicate system: PROSOUND ALPHA 6 (K093488 / K140854) are track 3 systems that incorporate the same fundamental and scientific technologies. There are no changes to the PROSOUND ALPHA7 and ARIETTA70 Diagnostic Ultrasound scanners with the exception of connection and recognition of the UST-5310 and UST-5311 transducers.
{6}------------------------------------------------
### 21 CFR Part 807.92, Section b
- 1. Non-clinical Testing
No new hazards were identified with the addition of the UST-5311 transducers to the PROSOUND ALPHA7 & ARIETTA70. The subject systems and the transducers have been evaluated for acoustic output, biocompatibility, electromagnetic compatibility, as well as electrical and mechanical safety, and have been found to conform to applicable medical device safety standards.
### 2. Clinical testing:
None required
- 3. Conclusions:
The Hitachi Aloka Medical, Ltd. Sterile Transducer is substantially equivalent in safety and effectiveness to the predicate device;
- . The subject and predicate device(s) are both indicated for diagnostic ultrasound imaging.
- . The subject and predicate device(s) have the same gray scale and Doppler capabilities.
- . The subject and predicate device(s) have the same essential technology for imaging, Doppler functions, and signal processing.
- 트 The subject and predicate device(s) have acoustic level below the Track 3 FDA limits.
- . The subject and predicate device(s) are manufactured in accordance to FDA 21 CFR 820 Quality System Regulations.
- 트 The subject and predicate device(s) are designed and manufactured to the same electrical and physical safety standards.
- 트 The subject and predicate device(s) are manufactured with materials that have been tested in accordance to ISO 10993-1; all biocompatibility testing has been conducted in accordance to each component material characterization, type of body contact, and duration contact risk profile.
- . The predicate transducers [UST-5310 & UST-5311] are sterile single use disposable devices.
### END OF SUMMARY
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.