The Teratech Corporation Terason™ t3200 is a general purpose Ultrasound. System intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Fetal, Abdominal, Intra-operative (abdominal, thoracic and PV); Pediatrics, Small Organ (Breast, testes, thyroid); Neonatal and Adult Cephalic; Trans-rectal and Trans-vaginal; Musculo-skeletal (Conventional and Superficial); Cardiac (Adult & Pediatric); Peripheral Vascular.
Device Story
Terason™ t3200 is a laptop-based diagnostic ultrasound system; consists of a Windows-based laptop connected to a custom engine for acoustic array control and echo processing. Inputs: ultrasound signals from transducers (6C1, 16HL7, 15L4). Processing: custom beamformer chip filters return signals for wider bandwidth and improved resolution. Outputs: real-time ultrasound images and flow analysis displayed on the laptop screen. Used in clinical settings by physicians for diagnostic imaging and guidance (e.g., needle/catheter placement, biopsy, follicle monitoring). Benefits: improved workflow, ease of use, and image resolution compared to the predicate Echo/t3000 system.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1 (safety), IEC 62366 (usability), IEC 60601-1-2 (EMC), IEC 60601-2-37 (ultrasonic safety), NEMA UD 3 (acoustic output), and ISO 10993 (biocompatibility) was verified. No clinical data provided.
Technological Characteristics
Laptop-based system with custom external engine. Transducers: 6C1, 16HL7, 15L4 with new acoustic arrays. Connectivity: Windows-based software. Safety standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37. Biocompatibility: ISO 10993. Imaging modes: B, M, PWD, Color Doppler, Power Doppler, Harmonic Imaging.
Indications for Use
Indicated for use by qualified physicians for ultrasound imaging or fluid flow analysis in fetal, abdominal, intra-operative, pediatric, small organ (breast, testes, thyroid), neonatal/adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal, cardiac, and peripheral vascular applications.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
{1}------------------------------------------------
KI10020
### 510(k) Summary
# Teratech Corporation
# Terason™ t3200 Ultrasound System
JAN 2 0 2011
### 1. Sponsor:
Teratech Corporation 77-79 Terrace Hall Ave. Burlington, MA 01803
Contact Person: Charles F. Hottinger, Ph.D, RAC, Regulatory Affairs Consultant Telephone: 206-780-7945
Date Prepared: November 1, 2010
### 2. Device Name
Proprietary Name: Terason™ t3200 Ultrasound System Common / Usual Name: Diagnostic Ultrasound System Diagnostic Ultrasound Transducer Classification Name:
(21 CFR 892.1570, 90-ITX) Ultrasonic Pulsed Echo Imaging System (21 CFR 892.1560, 90-IYO) Diagnostic Ultrasonic Transducer (21 CFR 892.1570, 90-ITX)
### 3. Predicate Device
Terason™ Echo/t3000 Ultrasound System (K080234)
### 4. Intended Use
The Teratech Corporation Terason™ t3200 is a general purpose Ultrasound. System intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Fetal, Abdominal, Intra-operative (abdominal, thoracic and PV); Pediatrics, Small Organ (Breast, testes, thyroid); Neonatal and Adult
{2}------------------------------------------------
Cephalic; Trans-rectal and Trans-vaginal; Musculo-skeletal (Conventional and Superficial); Cardiac (Adult & Pediatric); Peripheral Vascular.
# 5. Device Description
The Terason™ t3200 is a modified version of the Echo/t3000 Ultrasound System. The modifications include improved transducer acoustic arrays, an improvement in the ultrasound beam forming engine to improve the bandwidth of the receive signal processing capability and improvements in the software for workflow and ease of use of the system.
# 6. Technology Characteristics
The design and construction of the Terason™ t3200 is similar to the Terason™ Echo/t3000 Ultrasound system. These systems utilize a laptop computer running Windows to run the ultrasound application and a custom designed engine for control of the acoustic array and processing of the return echoes. The engine is housed in a compartment that is attached to the bottom of the laptop.
The differences between the Terason™ t3200 and the Terason Echo/t3000 Ultrasound System (the predicate device) include the following:
- The engine has a slight modification in the custom beamformer chip (as . compared to the Terason Echo/t3000 system) that provides for improved filtering of the return signal for wider bandwidth and better resolution across the whole image field.
- · The ultrasound application software has been modified to improve the user workflow and ease of use. The screen layout has been modified, and the user controls have been changed for better efficiency for the targeted exam types.
- . Transducers have been improved by utilizing new acoustic arrays.
15L4: Same housing construction as the 12L5 and 7L3 used on the previously cleared Echo/t3000 ultrasound system. The acoustic array is new with improved sensitivity and wider bandwidth than the 12L5.
6C1: Same material construction as used on the 15L4. The acoustic array is new with improved sensitivity and wider bandwidth
{3}------------------------------------------------
than the previously cleared 5C2 used on the Echo/t3000 ultrasound system.
16HL7: Same identical transducer as used on the previously cleared Echo/t3000 ultrasound system (labeled 12HL7 on the Echo / t3000).
## B1. Non Clinical Tests
The Terason™ t3200 system has been tested for compliance to the following standards (with the corresponding report referenced for each standard).
- IEC 60601-1 Medical Electrical Equipment Part 1: General . Requirements for Safety.
- o Intertek Test Record Number 100116897BOX-003
- IEC 62366, Medical Devices: Application of usability engineering to . medical devices.
- o Intertek Project: 100116897BOX-005
- IEC60601-1-6, Medical Electrical Equipment Part 1-6: General . requirements for safety- Collateral standard: Usability
- o Intertek Project: 100116897BOX-006
- IEC 60601-1-2:2007, Medical Electrical Equipment Part 1-2; General . requirements for basic safety and essential performance - Collateral Standard: Electromagnetic compatibility - Requirements and tests
- o IEC60601-1-2 Intertek Test Record Number, 100116897BOX-013
- o IEC60601-1-2 Intertek Test Record Number, 100116897BOX-010
- IEC 60601-2-37 / EN60601-2-37 Medical Electrical Equipment Part 2: . Particular requirements for the safety of ultrasonic medical diagnostic and monitoring equipment.
- o Transducer Model 6C1: Intertek Report Number IEC60601-2-37 t3200 6C1: 100116897BOX-008
{4}------------------------------------------------
- Transducer Model 15L4: Intertek Report Number IEC60601-2 o 37 t3200 15L4: 100116897BOX-007
- o Transducer Model 16HL7: Intertek Report Number IEC60601-2-37 t3200 16HL7: 100116897BOX-009
- . NEMA UD 3 Acoustic Output Display Terason t3200 Ultrasound System User Guide (16-3033)
- Biocompatibility Tests, ISO 10993 Part 5 and Part 10 .
- o Biocompatibility reports for all transducers (Attachments M-O of original submission)
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with outstretched wings, rendered in a stylized, minimalist design.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Teratech Corporation % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
JAN 2 0 201
Re: K110020
Trade/Device Name: TERASON™ t3200 Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYO, ITX, and IYN Dated: December 30, 2010 Received: January 3, 2011
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the TERASON™ t3200 Ultrasound System, as described in your premarket notification:
### Transducer Model Number
6C1 16HL7 1514
{6}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Paul Hardy at (301) 796-6542.
Sincerely Yours.
Mary S. Parker
Mary Pastel, ScD. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
#### Enclosure(s)
{7}------------------------------------------------
# Indications for Use Form
. "
JAN 2 0 2011
510(k) Number (if known): K110020 Device Name: _ Terason Ultrasound Systems Indications for Use:
The subject-modified device is a general purpose Ultrasound System intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow anaiysis of the human body. Specific clinical applications and exam types include: Fetal; Abdominal; Intra-operative (abdominal, thoracic and PV); Laparosopic; Pediatric; Small Organ (breast, testes, thyroid); Neonatal and Adult Cephalic; Trans-rectal and Trans-vaginal; Musculo-skeletal Conventional and Superficial; Cardiac (adult & pediatric); Peripheral Vascular.
Prescription Use Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mary S. Peake
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) 510020
Page 1 of _1_
Daniel A. Blumenthal Professor
{8}------------------------------------------------
#### System: TERASON™ t3200 Ultrasound System______________________________________________________________________________________________________________________________________________
| Clinical Application | | Mode of Operation | | | | | | Otherc |
|-----------------------------|--------------------------------------------------|-------------------|---|-----|-----|----------------|-----------------|--------|
| General<br>(Track I Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppa | Comb.<br>Modesb | |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetalf | N | N | N | | N | N | N |
| | Abdominalb: | N | N | N | | N | N | N |
| | Intra-operative (Spec.)d,e | N | N | N | | N | N | N |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatricc: | | | | | | | |
| | Small Organ (Thyroid,<br>Breast, Testes, etc.)d: | N | N | N | | N | N | N |
| | Neonatal Cephalicc: | | | | | | | |
| | Adult Cephalicc: | | | | | | | |
| | Trans-rectalf: | | | | | | | |
| | Trans-vaginal6: | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.)c: | N | N | N | | N | N | N |
| | Musculo-skel. (Superfic)c: | N | N | N | | N | N | N |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | N | N | N | | N | N | N |
| | Cardiac Pediatric | N | N | N | | N | N | N |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vesselb: | N | N | N | | N | N | N |
| | Other (Specify) | | | | | | | |
Transducer: (see comments)
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
a Includes Color Dopp. (CD), Directional Power Dopp. (DPD), and (non-directional) Power Dopp. (PD).
6 B+M; B+PWD; B+CD; B+OPD; B+PD
° Harmonic Imaging (HI)
0 Incl. ultrasound guidance for placement of needles,
catheters
° Abdominal organs and peripheral vessel.
I Incl. ultrasound guidance for piacement of needles,
catheters, cryosurgery, and brachytherapy
9 Incl. ultrasound guidance of transvaginal biopsy, infertility
monitoring of follicle development.
4 incl. guidance of amniocentesis, infertility monitoring of
follicle development.
' Incl. stress echo.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
N= new indication
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mary Slatt
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k).
Page 1 of __
{9}------------------------------------------------
#### _TERASON™ t3200 Ultrasound System_____________________________________________________________________________________________________________________________________________ System:
#### Transducer: 6C1
10
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation | | | | | | Otherc |
|-----------------------------|--------------------------------------------------|---|-------------------|-----|-----|----------------|-----------------|---|--------|
| General<br>(Track Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppa | Comb.<br>Modesb | | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetaln | N | N | N | | N | N | N | |
| | Abdominalc: | N | N | N | | N | N | N | |
| | Intra-operative (Spec.)d,e | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| Fetai<br>Imaging<br>& Other | Pediatricc: | N | N | N | | N | N | N | |
| | Small Organ (Thyroid,<br>Breast, Testes, etc.)d: | | | | | | | | |
| | Neonatal Cephalicc: | | | | | | | | |
| | Adult Cephalicc: | | | | | | | | |
| | Trans-rectalf: | | | | | | | | |
| | Trans-vaginalc: | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skel. (Convent.)c: | N | N | N | | N | N | N | |
| | Musculo-skel. (Superfic)c: | N | N | N | | N | N | N | |
| | Intra-luminal | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| | Cardiac Adult | N | N | N | | N | N | N | |
| Cardiac | Cardiac Pediatric | N | N | N | | N | N | N | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vesselc: | N | N | N | | N | N | N | |
| | Other (Specify) | | | | | | | | |
Includes Color Dopp. (CD), Directional Power Dopp. N= new indication
(DPD), and (non-directional) Power Dopp. (PD).
0 B+M: B+PWD; B+CD; B+DPD; B+PD
° Harmonic Imaging (HI)
0 Incl. ultrasound guidance for placement of needles,
catheters
° Abdominal organs and peripheral vessel.
I Incl. ultrasound guidance for placement of needles,
catheters, cryosurgery, and brachytherapy
6 Incl. ultrasound guidance of transvaginal biopsy, infertility
monitoring of follicle development.
" Incl. guidance of amniocentesis, infertility monitoring of
follicle development.
Incl. stress echo.
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mary Statts
Division Sign-Office of In Vitro Diagnostic Device Evaluation and Safety 510(k
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
{10}------------------------------------------------
#### TERASON™ t3200 Ultrasound System System:
16HL7 Transducer:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|------------------------------------------------------|--------------------------------------------------|-------------------|---|-----|-------------------|----------------|-----------------|--------|
| General | Specific | B | M | PWD | CWD | Color<br>Doppa | Comb.<br>Modesb | Otherc |
| (Track I Only) | (Tracks I & III) | | | | | | | |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetaln | | | | | | | |
| | Abdominalc: | | | | | | | |
| | Intra-operative (Spec.)a,e | N | N | N | | N | N | N |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatricc: | | | | | | | |
| | Small Organ (Thyroid,<br>Breast. Testes. etc.)d: | N | N | N | | N | N | N |
| | Neonatal Cephalicc: | | | | | | | |
| | Adult Cephalicc: | | | | | | | |
| | Trans-rectall: | | | | | | | |
| | Trans-vaginalg: | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.)o: | N | N | N | | N | N | N |
| | Musculo-skel. (Superfic)o: | N | N | N | | N | N | N |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vesselo: | N | N | N | | N | N | N |
| | Other (Specify) | | | | | | | |
| a Includes Color Dopp. (CD), Directional Power Dopp. | | | | | N= new indication | | | |
Includes Color Dopp. (CD), Directional Power Dopp.
(DPD), and (non-directional) Power Dopp. (PD). 6 B+M; B+PWD; B+CD; B+DPD; B+PD
° Harmonic Imaging (HI)
Harmonic Imaging (HI)
dial, without spectral equivalent
d incl. ultrasound guidance for placement of needles,
catheters
· Abdominal organs and peripheral vessel.
Incl. ultrasound guidance for placement of needles,
catheters, cryosurgery, and brachytherapy
9 Incl. ultrasound guidance of transvaginal biopsy, infertility
monitoring of follicie development.
" Incl. guidance of amniocentesis, infertility monitoring of
follicle development.
Incl. stress echo.
Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mary S. Pastet
Division Sign Office of in Vit Evaluation
510(k) K110020
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
{11}------------------------------------------------
#### TERASON™ t3200 Ultrasound System System:
#### Transducer: 15L4
.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|--------------------------------------------------|-------------------|---|-----|-----|----------------|-----------------|--------|
| General<br>(Track I Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppa | Comb.<br>Modesb | Otherc |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetaln | | | | | | | |
| | Abdominalo: | N | N | N | | N | N | N |
| | Intra-operative (Spec.)d,e | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatricc: | N | N | N | | N | N | N |
| | Small Organ (Thyroid,<br>Breast, Testes, etc.)d: | N | N | N | | N | N | N |
| | Neonatal Cephalicc: | | | | | | | |
| | Adult Cephalicc: | | | | | | | |
| | Trans-rectalf: | | | | | | | |
| | Trans-vaginalg: | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.)d: | N | N | N | | N | N | N |
| | Musculo-skel. (Superfic)d: | N | N | N | | N | N | N |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vesselo: | N | N | N | | N | N | N |
| | Other (Specify) | | | | | | | |
4 Includes Color Dopp. (CD), Directional Power Dopp. N= new indication ·
(DPD), and (non-directional) Power Dopp. (PD).
6 B+M; B+PWD; B+CD; B+DPD; B+PD
°Harmonic Imaging (HI)
4 incl. ultrasound guidance for piacement of needles,
catheters
్ Abdominal organs and peripheral vessel.
inci. ultrasound guidance for placement of needles,
catheters, cryosurgery, and brachytherapy
9 Incl. ultrasound guidance of transvaginal biopsy, infertility
monitoring of follicie development.
1 incl. guidance of amnìocentesis, infertility monitoring of
follicle development.
' Incl. stress echo.
Prescription Use _ × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mays. Posth
Division Sign-Q Office of In Vitro Diagnostic Device Evaluatio
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.