← Product Code [ITX](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/ITX) · K101008

# DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM, CONVEX ARRAY TRANSDUCER, LINEAR ARRAY (K101008)

_Shantou Institute of Ultrasonic Instruments (Siui · ITX · Jun 8, 2010 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/ITX/K101008

## Device Facts

- **Applicant:** Shantou Institute of Ultrasonic Instruments (Siui
- **Product Code:** [ITX](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/ITX.md)
- **Decision Date:** Jun 8, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1570
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

Diagnostic ultrasonic imaging for abdominal, pediatric, small organ, musculo-skeletal, cardiac, peripheral vascular applications in B, M, PWD, Color Doppler and 3D imaging modes.

## Device Story

Apogee 1100 is a digital color Doppler ultrasound imaging system. It utilizes various transducers (convex, linear, phased arrays) to acquire ultrasonic signals from patient anatomy. The system processes these signals to generate B-mode, M-mode, PWD, Color Doppler, and 3D images. Operated by clinicians in clinical settings, the device provides real-time visual output on a display. Healthcare providers interpret these images to assess anatomical structures and blood flow, aiding in diagnosis and clinical decision-making across abdominal, cardiac, vascular, and small organ applications. The system benefits patients by providing non-invasive diagnostic imaging.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Digital color Doppler ultrasound system. Includes convex (C3L60C, C5L40C), linear (L8L38C), and phased (P3F14C) array transducers. Supports B, M, PWD, Color Doppler, and 3D imaging modes. System is a standalone diagnostic imaging unit.

## Regulatory Identification

A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird-like figure with three stylized wing segments.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Shantou Institute if Ultrasonic Instruments Co., Ltd. (SIUI) % Mr. Bob Leiker ORS Representative Quality and Regulatory Services, Inc. 7263 Cronin Circle DUBLIN CA 95648

JUN = 8 2010

Re: K101008

Trade/Device Name: Apogee 1100 Digital Color Doppler Ultrasound Imaging System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYO, ITX, and IYN Dated: April 6, 2010 Received: April 12, 2010

Dear Mr. Leiker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the Apogee 1100 Digital Color Doppler Ultrasound Imaging System, as described in your premarket notification:

Transducer Model Number

Convex Array C3L60C Linear Array L8L38C Convex Array C5L40C Phased Array P3F14C

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

If you have any questions regarding the content of this letter, please contact Paul Hardy at (301) 796-6542.

Sincerely yours,

Sincerely yours,

Donald Sabi

Donald St. Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure(s)

{2}------------------------------------------------

# Indications for Use Statement

#### 510(k) Number (if known):

Device Name:

Apogee 1100 Digital Color Doppler Ultrasound Imaging System with

Convex Array Transducer C3L60C

Linear Array Transducer L8L38C

Convex Array Transducer C5L40C

Phased Array Transducer P3F14C

#### Indications for Use:

Diagnostic ultrasonic imaging for abdominal, pediatric, small organ, musculo-skeletal, cardiac, peripheral vascular applications in B, M, PWD, Color Doppler and 3D imaging modes.

#### Prescription Use (Part 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

# Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

(Division Sign-Off)

Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety

510K K161008

#### Indications for Use Statement

{3}------------------------------------------------

# 3.1 System Indications for Use Form

#### System: Apogce 1100

| Clinical Application        |                                    |   | Mode of Operation |     |     |                  |                       |                     |
|-----------------------------|------------------------------------|---|-------------------|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only)   | Specific<br>(Tracks 1 & 3)         | B | M                 | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic                  | Ophthalmic                         |   |                   |     |     |                  |                       |                     |
|                             | Fetal                              | N | N                 |     |     |                  |                       | N                   |
|                             | Abdominal                          | N | N                 | N   |     | N                |                       | N                   |
|                             | Intra-operative (Specify)          |   |                   |     |     |                  |                       |                     |
|                             | Intra-operative (Neuro)            |   |                   |     |     |                  |                       |                     |
|                             | Laparoscopic                       |   |                   |     |     |                  |                       |                     |
| Fetal<br>Imaging<br>& Other | Pediatric                          | N | N                 | N   |     | N                |                       |                     |
|                             | Small Organ (Specify)              | N | N                 | N   |     | N                |                       |                     |
|                             | Neonatal Cephalic                  |   |                   |     |     |                  |                       |                     |
|                             | Adult Cephalic                     |   |                   |     |     |                  |                       |                     |
|                             | Trans-rectal                       |   |                   |     |     |                  |                       |                     |
|                             | Trans-vaginal                      |   |                   |     |     |                  |                       |                     |
|                             | Trans-urethral                     |   |                   |     |     |                  |                       |                     |
|                             | Trans-esoph. (non-Card.)           |   |                   |     |     |                  |                       |                     |
|                             | Musculo-skeletal<br>(Conventional) | N | N                 | N   |     | N                |                       |                     |
|                             | Musculo-skeletal<br>(Superficial)  | N | N                 | N   |     | N                |                       |                     |
|                             | Intravascular                      |   |                   |     |     |                  |                       |                     |
|                             | Other (Specify)                    | N | N                 | N   |     | N                |                       | N                   |
|                             | Cardiac Adult                      | N | N                 | N   |     | N                |                       |                     |
| Cardiac                     | Cardiac Pediatric                  | N | N                 | N   |     | N                |                       |                     |
|                             | Intravascular (Cardiac)            |   |                   |     |     |                  |                       |                     |
|                             | Trans-esoph. (Cardiac)             |   |                   |     |     |                  |                       |                     |
|                             | Intra-cardiac                      |   |                   |     |     |                  |                       |                     |
|                             | Other (Specify)                    |   |                   |     |     |                  |                       |                     |
| Peripheral<br>Vessel        | Peripheral vessel                  | N | N                 | N   |     | N                |                       |                     |
|                             | Other (Specify)                    |   |                   |     |     |                  |                       |                     |

N = new indication; P = previously cleared by FDA; E = added under this appendix * Other modes of operation include: 3-D Imaging;

Additional Comments: Other uses include: Prostate, Kidney, Uterus, Ovary Small organs include: Thyroid, Testes, Breast

Prescription Use (Per 21 CFR 801.109)

(Division Sign-Off)

Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

510K K101008

TAB 3

Indications For Use

... ...

{4}------------------------------------------------

### 3.2 Transducer Indications for Use Form Transducer: Convex Array C3L60C

| Clinical Application        |                                    | Mode of Operation |   |     |     |                  |                       |                     |
|-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only)   | Specific<br>(Tracks 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
|                             | Ophthalmic                         |                   |   |     |     |                  |                       |                     |
|                             | Fetal                              | N                 | N |     |     |                  |                       |                     |
|                             | Abdominal                          | N                 | N | N   |     | N                |                       |                     |
|                             | Intra-operative (Specify)          |                   |   |     |     |                  |                       |                     |
|                             | Intra-operative (Neuro)            |                   |   |     |     |                  |                       |                     |
|                             | Laparoscopic                       |                   |   |     |     |                  |                       |                     |
| Fetal<br>Imaging<br>& Other | Pediatric                          |                   |   |     |     |                  |                       |                     |
|                             | Small Organ (Specify)              |                   |   |     |     |                  |                       |                     |
|                             | Neonatal Cephalic                  |                   |   |     |     |                  |                       |                     |
|                             | Adult Cephalic                     |                   |   |     |     |                  |                       |                     |
|                             | Trans-rectal                       |                   |   |     |     |                  |                       |                     |
|                             | Trans-vaginal                      |                   |   |     |     |                  |                       |                     |
|                             | Trans-urethral                     |                   |   |     |     |                  |                       |                     |
|                             | Trans-esoph. (non-Card.)           |                   |   |     |     |                  |                       |                     |
|                             | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                       |                     |
|                             | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                       |                     |
|                             | Intravascular                      |                   |   |     |     |                  |                       |                     |
|                             | Other (Specify)                    | N                 | N | N   |     | N                |                       |                     |
| Cardiac                     | Cardiac Adult                      |                   |   |     |     |                  |                       |                     |
|                             | Cardiac Pediatric                  |                   |   |     |     |                  |                       |                     |
|                             | Intravascular (Cardiac)            |                   |   |     |     |                  |                       |                     |
|                             | Trans-esoph. (Cardiac)             |                   |   |     |     |                  |                       |                     |
|                             | Intra-cardiac                      |                   |   |     |     |                  |                       |                     |
|                             | Other (Specify)                    |                   |   |     |     |                  |                       |                     |
| Peripheral<br>Vessel        | Peripheral vessel                  |                   |   |     |     |                  |                       |                     |
|                             | Other (Specify)                    |                   |   |     |     |                  |                       |                     |

N = new indication; P = previously cleared by FDA; E = added under this appendix Additional Comments: Other uses include: Prostate, Kidney, Uterus, Ovary

Prescription Use (Per 21 CFR 801.109)

(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

510K K101008

TAB 3

{5}------------------------------------------------

## 3:3 Transducer Indications for Use Form Transducer: Linear Array L8L38C

| Clinical Application        |                                    |   | Mode of Operation |     |     |                  |                       |                     |
|-----------------------------|------------------------------------|---|-------------------|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only)   | Specific<br>(Tracks 1 & 3)         | B | M                 | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic                  | Ophthalmic                         |   |                   |     |     |                  |                       |                     |
|                             | Fetal                              |   |                   |     |     |                  |                       |                     |
|                             | Abdominal                          |   |                   |     |     |                  |                       |                     |
|                             | Intra-operative (Specify)          |   |                   |     |     |                  |                       |                     |
|                             | Intra-operative (Neuro)            |   |                   |     |     |                  |                       |                     |
|                             | Laparoscopic                       |   |                   |     |     |                  |                       |                     |
| Fetal<br>Imaging<br>& Other | Pediatric                          | N | N                 | N   |     | N                |                       |                     |
|                             | Small Organ (Specify)              | N | N                 | N   |     | N                |                       |                     |
|                             | Neonatal Cephalic                  |   |                   |     |     |                  |                       |                     |
|                             | Adult Cephalic                     |   |                   |     |     |                  |                       |                     |
|                             | Trans-rectal                       |   |                   |     |     |                  |                       |                     |
|                             | Trans-vaginal                      |   |                   |     |     |                  |                       |                     |
|                             | Trans-urethral                     |   |                   |     |     |                  |                       |                     |
|                             | Trans-esoph. (non-Card.)           |   |                   |     |     |                  |                       |                     |
|                             | Musculo-skeletal<br>(Conventional) | N | N                 | N   |     | N                |                       |                     |
|                             | Musculo-skeletal<br>(Superficial)  | N | N                 | N   |     | N                |                       |                     |
|                             | Intravascular                      |   |                   |     |     |                  |                       |                     |
|                             | Other (Specify)                    |   |                   |     |     |                  |                       |                     |
|                             | Cardiac Adult                      |   |                   |     |     |                  |                       |                     |
| Cardiac                     | Cardiac Pediatric                  |   |                   |     |     |                  |                       |                     |
|                             | Intravascular (Cardiac)            |   |                   |     |     |                  |                       |                     |
|                             | Trans-esoph. (Cardiac)             |   |                   |     |     |                  |                       |                     |
|                             | Intra-cardiac                      |   |                   |     |     |                  |                       |                     |
|                             | Other (Specify)                    |   |                   |     |     |                  |                       |                     |
| Peripheral<br>Vessel        | Peripheral vessel                  | N | N                 | N   |     | N                |                       |                     |
|                             | Other (Specify)                    |   |                   |     |     |                  |                       |                     |

N = new indication: P = previously cleared by FDA: E = added under this appendix Additional Comments: Small organs include: Thyroid, Testes, Breast

510

Prescription Use (Per 21 CFR 801.109)

(Division Sign-Off) Division of Radiological Devices Vitro Diagnostic Device Evaluation and Safety Office of In

Indications For Use

{6}------------------------------------------------

## 3.4 Transducer Indications for Use Form Transducer: Convex Array C5L40C

| Clinical Application        |                                    | Mode of Operation |   |     |     |                  |                       |                     |
|-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only)   | Specific<br>(Tracks 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic                  | Ophthalmic                         |                   |   |     |     |                  |                       |                     |
|                             | Fetal                              |                   | N |     |     |                  |                       | N                   |
|                             | Abdominal                          |                   | N |     |     |                  |                       | N                   |
|                             | Intra-operative (Specify)          |                   |   |     |     |                  |                       |                     |
|                             | Intra-operative (Neuro)            |                   |   |     |     |                  |                       |                     |
|                             | Laparoscopic                       |                   |   |     |     |                  |                       |                     |
| Fetal<br>Imaging<br>& Other | Pediatric                          |                   |   |     |     |                  |                       |                     |
|                             | Small Organ (Specify)              |                   |   |     |     |                  |                       |                     |
|                             | Neonatal Cephalic                  |                   |   |     |     |                  |                       |                     |
|                             | Adult Cephalic                     |                   |   |     |     |                  |                       |                     |
|                             | Trans-rectal                       |                   |   |     |     |                  |                       |                     |
|                             | Trans-vaginal                      |                   |   |     |     |                  |                       |                     |
|                             | Trans-urethral                     |                   |   |     |     |                  |                       |                     |
|                             | Trans-esoph. (non-Card.)           |                   |   |     |     |                  |                       |                     |
|                             | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                       |                     |
|                             | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                       |                     |
|                             | Intravascular                      |                   |   |     |     |                  |                       |                     |
|                             | Other (Specify)                    |                   | N |     |     |                  |                       | N                   |
|                             | Cardiac Adult                      |                   |   |     |     |                  |                       |                     |
| Cardiac                     | Cardiac Pediatric                  |                   |   |     |     |                  |                       |                     |
|                             | Intravascular (Cardiac)            |                   |   |     |     |                  |                       |                     |
|                             | Trans-esoph. (Cardiac)             |                   |   |     |     |                  |                       |                     |
|                             | Intra-cardiac                      |                   |   |     |     |                  |                       |                     |
|                             | Other (Specify)                    |                   |   |     |     |                  |                       |                     |
| Peripheral<br>Vessel        | Peripheral vessel                  |                   |   |     |     |                  |                       |                     |
|                             | Other (Specify)                    |                   |   |     |     |                  |                       |                     |

N = new indication; P = previously cleared by FDA; E = added under this appendix * Other modes include: 3-D Imaging;

Additional Comments: Other uses include: Prostate, Kidney, Uterus, Ovary

Prescription Use (Per 21 CFR 801.109)

801.109)

(Division Sign-Off) Division of Radiological Devices Office of In Vitro Division of Nadiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety

510K K101008

Indications For Use

{7}------------------------------------------------

## 3.5 Transducer Indications for Use Form Transducer: Phased Array P3F14C

| Clinical Application        |                                    | Mode of Operation |   |     |     |                  |                       |                     |
|-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only)   | Specific<br>(Tracks 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic                  | Ophthalmic                         |                   |   |     |     |                  |                       |                     |
| Fetal<br>Imaging<br>& Other | Fetal                              |                   |   |     |     |                  |                       |                     |
|                             | Abdominal                          |                   |   |     |     |                  |                       |                     |
|                             | Intra-operative (Specify)          |                   |   |     |     |                  |                       |                     |
|                             | Intra-operative (Neuro)            |                   |   |     |     |                  |                       |                     |
|                             | Laparoscopic                       |                   |   |     |     |                  |                       |                     |
|                             | Pediatric                          |                   |   |     |     |                  |                       |                     |
|                             | Small Organ (Specify)              |                   |   |     |     |                  |                       |                     |
|                             | Neonatal Cephalic                  |                   |   |     |     |                  |                       |                     |
|                             | Adult Cephalic                     |                   |   |     |     |                  |                       |                     |
|                             | Trans-rectal                       |                   |   |     |     |                  |                       |                     |
|                             | Trans-vaginal                      |                   |   |     |     |                  |                       |                     |
|                             | Trans-urethral                     |                   |   |     |     |                  |                       |                     |
|                             | Trans-esoph. (non-Card.)           |                   |   |     |     |                  |                       |                     |
|                             | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                       |                     |
|                             | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                       |                     |
|                             | Intravascular                      |                   |   |     |     |                  |                       |                     |
|                             | Other (Specify)                    |                   |   |     |     |                  |                       |                     |
| Cardiac                     | Cardiac Adult                      | N                 | N | N   |     | N                |                       |                     |
|                             | Cardiac Pediatric                  | N                 | N | N   |     | N                |                       |                     |
|                             | Intravascular (Cardiac)            |                   |   |     |     |                  |                       |                     |
|                             | Trans-esoph. (Cardiac)             |                   |   |     |     |                  |                       |                     |
|                             | Intra-cardiac                      |                   |   |     |     |                  |                       |                     |
|                             | Other (Specify)                    |                   |   |     |     |                  |                       |                     |
| Peripheral<br>Vessel        | Peripheral vessel                  |                   |   |     |     |                  |                       |                     |
|                             | Other (Specify)                    |                   |   |     |     |                  |                       |                     |

N = new indication; P = previously cleared by FDA; E = added under this appendix

Prescription Use (Per 21 CFR 801.109)

(Division Sign-Off)

(Division of Radiological Devices
Division of Radiological Devices Eurolustio Office of In Vitro Diagnostic Device Evaluation and Safety

510K K101608

Indications For Use

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/ITX/K101008](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/ITX/K101008)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
