SONOACE X8 DIAGNOSTIC ULTRASOUND SYSTEM

K093714 · Medison Co., Ltd. · ITX · Dec 10, 2009 · Radiology

Device Facts

Record IDK093714
Device NameSONOACE X8 DIAGNOSTIC ULTRASOUND SYSTEM
ApplicantMedison Co., Ltd.
Product CodeITX · Radiology
Decision DateDec 10, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1570
Device ClassClass 2
AttributesPediatric, 3rd-Party Reviewed

Indications for Use

The SONOACE X8 system and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body. The clinical applications include: General, abdomen, obstetrics, gynecology, vascular, extremity, pediatric, cardiac, breast, urology, neonatal cephalic, trans-esophageal. and etc.

Device Story

The SONOACE X8 is a mobile, software-controlled diagnostic ultrasound system. It acquires ultrasound data via various transducers and displays it in 2D, M, Color Doppler, Power Doppler, PW/CW Spectral Doppler, and Tissue Doppler modes; 3D/4D imaging is supported via mechanical scan probes. The system automatically displays mechanical and thermal indices for real-time acoustic output monitoring. It provides measurement and analysis packages for clinical diagnostic use by healthcare professionals. The device is used in clinical settings to visualize anatomical structures and fluid flow, aiding in diagnosis and biopsy guidance. Output is viewed on an integrated LCD display, supporting clinical decision-making through quantitative and qualitative assessment of patient anatomy and hemodynamics.

Clinical Evidence

Bench testing only. The device complies with safety standards including UL 60601-1, CSA C22.2 No. 601.1, IEC 60601-2-37, NEMA UD-2, NEMA UD-3, IEC 61157, and ISO 10993-1 for biocompatibility.

Technological Characteristics

Mobile diagnostic ultrasound system with LCD display. Supports B, M, PW/CW Doppler, Color/Power Doppler, and 3D/4D imaging. Materials comply with ISO 10993-1. Safety standards: UL 60601-1, CSA C22.2 No. 601.1, IEC 60601-2-37, NEMA UD-2, NEMA UD-3, IEC 61157. Connectivity includes standard ultrasound transducer interfaces.

Indications for Use

Indicated for diagnostic ultrasound imaging and fluid flow analysis in adult, pediatric, and neonatal populations. Clinical applications include general, abdominal, obstetrical (fetal imaging, infertility monitoring), gynecological, vascular (peripheral vessel), extremity, pediatric, cardiac (adult/pediatric), breast, urological (trans-rectal, trans-vaginal), neonatal cephalic, and trans-esophageal imaging. Includes guidance for biopsy.

Regulatory Classification

Identification

A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K093714 510(k) Premarket Notification SONOACE X8 Diagnostic Ultrasound System DEC 1 0 2009 #### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92. #### 1. Submitter's Information: 21 CFR 807.92(a)(1) MEDISON CO., LTD. 1003, Daechi-dong, Gangnam-gu, Seoul 135-280, Korea Contact Person: - Mr. Kyung-Am, Shim Regulatory Affairs Manager | Telephone: | 82.2.2194.1381 | |------------|--------------------| | Facsimile: | 82.2.2194.1399 | | Email: | kashim@medison.com | Data Prepared: September 7, 2009 #### 2. Name of the device: Common/Usual Name: Diagnostic Ultrasound System and Accessories Proprietary Name: SONOACE X8 Diagnostic Ultrasound System | Classification Names: | FR Number | Product Code | |------------------------------------------|-----------|--------------| | Ultrasonic Pulsed Doppler Imaging System | 892.1550 | IYN | | Ultrasound Pulsed Echo Imaging System | 892.1560 | IYO | | Diagnostic Ultrasound Transducer | 892.1570 | ITX | #### 3. Identification of the predicate or legally marketed device: K063580, SONOACE X8 Diagnostic Ultrasound System K092159, ACCUVIX V20 Diagnostic Ultrasound System #### 4. Device Description: The SONOACE X8 is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode, M mode, Color Doppler mode, Power Doppler mode, PW Spectral Doppler, CW Spectral Doppler mode, and Tissue Doppler Image mode on the LCD display. It also provides the 3D/4D imaging mode using the 3D/4D probe in the Mechanical scan mode. The SONOACE X8 has real time acoustic output display with two basic indices, a {1}------------------------------------------------ mechanical index and a thermal index, which are both automatically displayed. The system also provides for the measurement of anatomical structures and for analysis packages that provide information used for clinical diagnostic purposes by competent health care professionals. The SONOACE X8 has been designed to meet the following product safety standards: - UL 60601-1, Safety requirements for Medical Equipment - CSA C22.2 No. 601.1, Safety requirements for Medical Equipment - IEC60601-2-37. Diagnostic Ultrasound Safety Standards - EN/IEC60601-1, Safety requirements for Medical Equipment - EN/IEC60601-1-2, EMC requirements for Medical Equipment - NEMA UD-2. Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment - NEMA UD-3, Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment - IEC 61157. Declaration of the acoustic output - ISO10993-1. Biocompatibility #### 5. Intended Uses: The SQNOACE X8 system and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body. The clinical applications include: General, abdomen, obstetrics, gynecology, vascular, extremity, pediatric, cardiac, breast, urology, neonatal cephalic, trans-esophageal. and etc. #### 6. Technological Characteristics: The SONOACE X8 is substantially equivalent to the SONOACE X8 Diagnostic Ultrasound System, cleared via K063580, and the ACCUVIX V20 Diagnostic Ultrasound System, cleared via K092159. All systems transmit ultrasonic energy into patients, then perform post processing of received echoes to generate on-screen display of anatomic structures and fluid flow within the body. All system allow for specialized measurements of structures and flow, and calculations. END of 510(K) Summary {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three arms or extensions, representing health and human services. The figure is positioned to the right of a circular text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". #### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002 Medison Co., Ltd. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313 DEC 1 0 2009 Re: K093714 Trade/Device Name: SONOACE X8 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system' Regulatory Class: II Product Code: IYO, IYN, and ITX Dated: December 1, 2009 Received: December 2, 2009 Dear Mr. Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. This determination of substantial equivalence applies to the following transducers intended for use with the SONOACE X8 Diagnostic Ultrasound System, as described in your premarket notification: | Transducer Model Number | | | | |-------------------------|------------|------------|--------| | 3D2-6ET | C2-5EP | NEV4-9ES | P2-4AA | | 3D4-8EK | C2-8 | HL5-12ED | P2-4AH | | 3D4-8ET | C3-7EP | L4-7EL | P3-5AC | | 3D4-9ES | C4-9/10ED | L5-12/50EP | P3-7AC | | 3D5-9EK | ER4-9/10ED | L5-12EC | CW 2.0 | | 3DC2-6 | EV4-9/10ED | L5-12EP | CW 4.0 | | C2-5EL | NER4-9ES | MPT4-7 | | {3}------------------------------------------------ #### Page 2 - Mr. Mark Job If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. If you have any questions regarding the content of this letter, please contact Paul Hardy at (301) 796-6542. Sincerely yours, Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure(s) {4}------------------------------------------------ ## Section 1.3 INDICATIONS FOR USE | DIAGNOSTIC ULTRASOUND INDICATIONS STATEMENT | | | | | | | | | |--------------------------------------------------------------------------------------------------|-----------------------------------------|---|---|-----|-----|-------------------|---------------------------------------------------|--------------------| | 510(k) No.: | K093714 | | | | | | | | | Device Name: | SONOACE X8 Diagnostic Ultrasound System | | | | | | | | | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | | | | Clinical Application | | | | | | Mode of Operation (*includes simultaneous B-mode) | | | General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal (See Note 3) | P | P | P | P | P | Note 1 | Notes 2, 7, 8 | | | Abdominal | P | P | P | P | P | Note 1 | Notes 2, 7, 8 | | | Intra-operative (See Note 6) | | | | | | | | | | Intra-operative (Neuro.) | | | | | | | | | Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | | | Pediatric | P | P | P | P | P | Note 1 | Note 2,4,5,6,7,8,9 | | | Small Organ (See Note 5) | P | P | P | P | P | Note 1 | Note 2,5,6,8,9 | | | Neonatal Cephalic | P | P | P | P | P | Note 1 | | | | Adult Cephalic | P | P | P | P | P | Note 1 | Note 4, 7 | | | Trans-rectal | P | P | P | P | P | Note 1 | Note 2, 3, 8 | | | Trans-vaginal | P | P | P | P | P | Note 1 | Note 2, 3, 8 | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Cardiac) | N | N | N | N | N | Note 1 | Note 4, 7 | | | Musculo-skel. (Convent.) | P | P | P | P | P | Note 1 | Note 2,5,6,8,9 | | | Musculo-skel. (Superfic.) | P | P | P | P | P | Note 1 | Note 2,5,6,8,9 | | | Intra-luminal | | | | | | | | | | Other (spec.) | | | | | | | | | | Cardiac Adult | P | P | P | P | P | Note 1 | Note 4, 7 | | Cardiac | Cardiac Pediatric | P | P | P | P | P | Note 1 | Note 4, 7 | | | Trans-esophageal (Cardiac) | N | N | N | N | N | Note 1 | Note 4, 7 | | | Other (spec.) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | P | P | P | P | P | Note 1 | Note 5, 6, 9 | | | Other (spec.) | | | | | | | | N= new indication; P= previously cleared by FDA K063580; E= added under Appendix E Additional Comments: Color Doppler includes Power (Amplitude) Doppler Note 1: B/M, B/PW Doppler, B/Color Doppler/PW Doppler/PW Doppler, B/Power Doppler/PW Doppler, B/Color Doppler/Color M Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, parathyroid, breast, scroturn and penis in adult, pediatric and neonatal patients Note 6: Abdominal organs and peripheral vessel Note 7: Tissue Harmonic Imaging (THI) Note 8: 3D imaging Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) | (Division Sign-Off) | |---------------------| |---------------------| Division of Reproductive, Abdominal, and Radiological Devices | 510(k) Number | K093714 | |---------------|---------| |---------------|---------| {5}------------------------------------------------ | 510(k) No.: | | | | | | | | | |---------------------------|---------------------------------------------------------------------------|---|---|-----|-----|-------------------|----------------------|-----------------------------------------------------------------------------------| | Device Name: | 3D2-6ET for use with SONOACE X8 | | | | | | | | | Intended Use: | | | | | | | | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows | | | Clinical Application<br>Mode of Operation (*includes simultaneous B-mode) | | | | | | | | | General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal (See Note 3) | P | P | P | | P | Note 1 | Notes 2, 7, 8 | | | Abdominal | P | P | P | | P | Note 1 | Notes 2, 7, 8 | | Fetal Imaging<br>& Other | Intra-operative (See Note 6) | | | | | | | | | | Intra-operative (Neuro.) | | | | | | | | | | Laparoscopic | | | | | | | | | | Pediatric | P | P | P | | P | Note 1 | Notes 2, 7, 8 | | | Small Organ (See Note 5) | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Cardiac) | | | | | | | | | | Musculo-skel. (Convent.) | | | | | | | | | | Musculo-skel. (Superfic.) | | | | | | | | | | Intra-luminal | | | | | | | | | | Other (spec.) | | | | | | | | | | Cardiac Adult | | | | | | | | | Cardiac | Cardiac Pediatric | | | | | | | | | | Trans-esophageal (Cardiac) | | | | | | | | | | Other (spec.) | | | | | | | | | | Peripheral vessel | | | | | | | | | Peripheral<br>Vessel | Other (spec.) | | | | | | | | N= new indication; P= previously cleared by FDA K063580; E= added under Appendix E Additional Comments: Color Doppler includes Power (Amplitude) Doppler Note 1: B/M, B/PW Doppler, B/Color Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler, B/Color Doppler/Color M Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 6: Abdominal organs and peripheral vessel Note 7: Tissue Harmonic Imaging (THI) Note 8: 3D imaging Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) *(Division Sign-Off)* Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number K93714 {6}------------------------------------------------ 510(k) No .: | 3D4-8EK for use with SONOACE X8<br>Device Name:<br>Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows | | | | | | | | | |----------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|---|---|-----|-----|-------------------|----------------------|------------------| | Clinical Application | | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | General<br>(Track I only) | Ophthalmic | | | | | | | | | Fetal Imaging<br>& Other | Fetal (See Note 3) | p | p | p | | p | Note 1 | Notes 2, 7, 8 | | | Abdominal | p | p | p | | p | Note 1 | Notes 2, 7, 8 | | | Intra-operative (See Note 6) | | | | | | | | | | Intra-operative (Neuro.) | | | | | | | | | | Laparoscopic | | | | | | | | | | Pediatric | p | p | p | | p | Note 1 | Notes 2, 7, 8 | | | Small Organ (See Note 5) | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Cardiac) | | | | | | | | | | Musculo-skel. (Convent.) | | | | | | | | | | Musculo-skel. (Superfic.) | | | | | | | | | | Intra-luminal | | | | | | | | | | Other (spec.) | | | | | | | | | Cardiac | Cardiac Adult | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | Trans-esophageal (Cardiac) | | | | | | | | | | Other (spec.) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | Other (spec.) | | | | | | | | N= new indication; P= previously cleared by FDA K063580; E= added under Appendix E Additional Comments: Color Doppler includes Power (Amplitude) Doppler Note 1: B/M, B/PW Doppler, B/Color Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler, B/Color Doppler/Color M Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients - Note 6: Abdominal organs and peripheral vessel Note 7: Tissue Hannonic Imaging (THI) Note 8: 3D imaging Note 9: Panoramic imaging #### Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Pey 21 CFR 801.109) (Division Sign-Off) Division of Reproductive, Abdominal. and Radiological Dev 510(k) Number {7}------------------------------------------------ | 510(k) No.: | | | | | | | | | |-----------------------------------------------------------------------------------------------------|------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------| | Device Name:<br>3D4-8ET for use with SONOACE X8 | | | | | | | | | | Intended Use:<br>Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | | | | Clinical Application | Mode of Operation (*includes simultaneous B-mode) | | | | | | | | General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | | Ophthalmic | | | | | | | | | | Fetal (See Note 3) | P | P | P | | P | Note 1 | Notes 2, 7, 8 | | | Abdominal | P | P | P | | P | Note 1 | Notes 2, 7, 8 | | | Intra-operative (See Note 6) | | | | | | | | | | Intra-operative (Neuro.) | | | | | | | | | Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | | | Pediatric | P | P | P | | P | Note 1 | Notes 2, 7, 8 | | | Small Organ (See Note 5) | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Cardiac) | | | | | | | | | | Musculo-skel. (Convent.) | | | | | | | | | | Musculo-skel. (Superfic.) | | | | | | | | | | Intra-luminal | | | | | | | | | | Other (spec.) | | | | | | | | | | Cardiac Adult | | | | | | | | | Cardiac | Cardiac Pediatric | | | | | | | | | | Trans-esophageal (Cardiac) | | | | | | | | | | Other (spec.) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | Other (spec.) | | | | | | | | N= new indication; P= previously cleared by FDA K063580; E= added under Appendix E Additional Comments: Color Doppler includes Power (Amplitude) Doppler Note 1: B/M, B/PW Doppler, B/Color Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler, B/Color Doppler/Color M Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 6: Abdominal organs and peripheral vessel Note 7: Tissue Harmonic Imaging (THI) Note 8: 3D imaging Note 9: Panoramic imaging #### Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) Heleum (Division Sign-Off) Division of Reproductive, Abdominal. and Radiological I 510(k) Numbe {8}------------------------------------------------ 510(k) No .: Device Name: 3D4-9ES for use with SONOACE X8 Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use: Clinical Application Mode of Operation (*includes simultaneous B-mode) General Specific B M PWD CWD Color Combined* Other (Track I only) (Tracks I & III) Doppler* (Spec.) (Spec.) Ophthalmic Ophthalmic Fetal (See Note 3) Abdominal Intra-operative (See Note 6) Intra-operative (Neuro.) Fetal Imaging Laparoscopic Pediatric & Other Small Organ (See Note 5) Neonatal Cephalic Adult Cephalic Trans-rectal P P P P Notes 2, 8 Note 1 Trans-vaginal P P Notes 2, 8 P P Note 1 Trans-urethral Trans-esoph. (non-Cardiac) Musculo-skel. (Convent.) Musculo-skel. (Superfic.) Intra-luminal Other (spec.) Cardiac Adult Cardiac Cardiac Pediatric Trans-esophageal (Cardiac) Other (spec.) Peripheral Peripheral vessel Vessel Other (spec.) N= new indication; P= previously cleared by FDA K063580; E= added under Appendix E Additional Comments: Color Doppler includes Power (Amplitude) Doppler Note 1: B/M, B/PW Doppler, B/Color Doppler/PW Doppler, B/Power Doppler/PW Doppler/PW Doppler. B/Color Doppler/Color M Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 6: Abdominal organs and peripheral vessel Note 7: Tissue Harmonic Imaging (THI) Note 8: 3D imaging Note 9: Panoramic imaging Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) Division Sign Off (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devic 510(k) Number {9}------------------------------------------------ | 510(k) No.: | | | | | | | | | |--------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|---|---|-----|-----|-------------------|----------------------|------------------| | Device Name:<br>3D5-9EK for use with SONOACE X8 | | | | | | | | | | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | | | | Clinical Application<br>Mode of Operation (*includes simultaneous B-mode) | | | | | | | | | General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal (See Note 3) | | | | | | | | | | Abdominal | | | | | | | | | | Intra-operative (See Note 6) | | | | | | | | | | Intra-operative (Neuro.) | | | | | | | | | Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | | | Pediatric | | | | | | | | | | Small Organ (See Note 5) | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | P | P | P | | P | Note 1 | Notes 2, 8 | | | Trans-vaginal | P | P | P | | P | Note 1 | Notes 2, 8 | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Cardiac) | | | | | | | | | | Musculo-skel. (Convent.) | | | | | | | | | | Musculo-skel. (Superfic.) | | | | | | | | | | Intra-luminal | | | | | | | | | | Other (spec.) | | | | | | | | | | Cardiac Adult | | | | | | | | | Cardiac | Cardiac Pediatric | | | | | | | | | | Trans-esophageal (Cardiac) | | | | | | | | | | Other (spec.) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | Other (spec.) | | | | | | | | N= new indication; P= previously cleared by FDA K063580; E= added under Appendix E Additional Comments: Color Doppler includes Power (Amplitude) Doppler Note 1: B/M, B/PW Doppler, B/Color Doppler, B/Color Doppler/PW Doppler/PW Doppler/PW Doppler, B/Color Doppler/Color M Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 6: Abdominal organs and peripheral vessel Note 7: Tissue Harmonic Imaging (THI) Note 8: 3D imaging Note 9: Panoramic imaging Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Rer 21 CFR 801.109) (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number {10}------------------------------------------------ | 510(k) No .: | | | | | | | | | | |-----------------------------------------------------------------------------------------------------|--------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|--| | Device Name: | 3DC2-6 for use with SONOACE X8 | | | | | | | | | | Intended Use:<br>Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | | | | | Clinical Application | Mode of Operation (*includes simultaneous B-mode) | | | | | | | | | General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | | Ophthalmic | Ophthalmic | | | | | | | | | | | Fetal (See Note 3) | N | N | N | | N | Note 1 | Notes 2, 7, 8 | | | | Abdominal | N | N | N | | N | Note l | Notes 2, 7, 8 | | | | Intra-operative (See Note 6) | | | | | | | | | | | Intra-operative (Neuro.) | | | | | | | | | | Fetal Imaging | Laparoscopic | | | | | | | | | | & Other | Pediatric | N | N | N | | N | Note 1 | Notes 2, 7, 8 | | | | Small Organ (See Note 5) | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | Trans-rectal | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | Trans-urethral | | | | | | | | | | | Trans-esoph. (non-Cardiac) | | | | | | | | | | | Musculo-skel. (Convent.) | | | | | | | | | | | Musculo-skel. (Superfic.) | | | | | | | | | | | Intra-luminal | | | | | | | | | | | Other (spec.) | | | | | | | | | | | Cardiac Adult | | | | | | | | | | Cardiac | Cardiac Pediatric | | | | | | | | | | | Trans-esophageal (Cardiac) | | | | | | | | | | | Other (spec.) | | | | | | | | | | Peripheral | Peripheral vessel | | | | | | | | | | Vessel | Other (spec.) | | | | | | | | | N= new indication; P= previously cleared; E= added under Appendix E Additional Comments: Color Doppler includes Power (Amplitude) Doppler Note 1: B/M, B/PW Doppler, B/Color Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler, B/Color Doppler/Color M Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 6: Abdominal organs and peripheral vessel Note 7: Tissue Harmonic Imaging (THI) Note 8: 3D imaging Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) Hes. Klem gn-C (fr) (Divisic Division of Reproductive, Abdominal and Radiological D 510(k) Number {11}------------------------------------------------ 510(k) No .: Device Name: C2-5EL for use with SONOACE X8 Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use: Clinical Application Mode of Operation (*includes simultaneous B-mode) General Specific B PWD CWD M Color Combined* Other (Track I only) (Tracks I & III) Doppler* (Spec.) (Spec.) Ophthalmic Ophthalmic Fetal (See Note 3) P P P P Note 1 Notes 2, 7, 8 Abdominal P Notes 2, 7, 8 P P P Note l Intra-operative (See Note 6) Intra-operative (Neuro.) Fetal Imaging Laparoscopic & Other Pediatric P P P P Note l Notes 2, 7, 8 Small Organ (See Note 5) Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph. (non-Cardiac) Musculo-skel. (Convent.) Musculo-skel. (Superfic.) Intra-luminal Other (spec.) Cardiac Adult Cardiac Cardiac Pediatric Trans-esophageal (Cardiac) Other (spec.) Peripheral Peripheral vessel Vessel Other (spec.) N= new indication; P= previously cleared by FDA K063580; E= added under Appendix E Additional Comments: Color Doppler includes Power (Amplitude) Doppler Note 1: B/M, B/PW Doppler, B/Color Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler, B/Color Doppler/Color M Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 6: Abdominal organs and peripheral vessel Note 7: Tissue Harmonic Imaging (THI) Note 8: 3D imaging | Concurrence of CDRH, Office of Device Evaluation (ODE) | |--------------------------------------------------------| | Prescription Use (Per 21 CFR 801.109) | Sign-0 Division of Reproductive, Abdominal, and Radiological D 510(k) Number {12}------------------------------------------------ 510(k) No .: Device Name: C2-5EP for use with SONOACE X8 Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use: Clinical Application Mode of Operation (*includes simultaneous B-mode) Specific General B M PWD CWD Color Combined* Other (Track I only) (Tracks I & III) Doppler3 (Spec.) (Spec.) Ophthalmic Ophthalmic Fetal (See Note 3) N Note 1 Notes 2, 7, 8 N N N Abdominal N N N N Note 1 Notes 2, 7, 8 Intra-operative (See Note 6) Intra-operative (Neuro.) Laparoscopic Fetal Imaging & Other Pediatric N Note 1 Notes 2, 7, 8 N N N Small Organ (See Note 5) Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph. (non-Cardiac) Musculo-skel. (Convent.) Musculo-skel. (Superfic.) Intra-luminal Other (spec.) Cardiac Adult Cardiac Cardiac Pediatric Trans-csophagcal (Cardiac) Other (spec.) Peripheral Peripheral vessel Vessel Other (spec.) N= new indication; P= previously cleared; E= added under Appendix E Additional Comments: Color Doppler includes Power (Amplitude) Doppler Note 1: B/M, B/PW Doppler, B/Color Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler, B/Color Doppler/Color M Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 6: Abdominal organs and peripheral vessel Note 7: Tissue Harmonic Imaging (THI) Note 8: 3D imaging | Concurrence of CDRH, Office of Device Evaluation (ODE) | | |--------------------------------------------------------|--| | Prescription Use (Per 21 CFR 801.109) | | (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological I 510/k) Number {13}------------------------------------------------ | 510(k) No.: | | | | | | | | | |-----------------------------------------------------------------------------------------------------|------------------------------|---|---|-----|-----|-------------------|---------------------------------------------------|------------------| | C2-8 for use with SONOACE X8<br>Device Name: | | | | | | | | | | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:<br>Intended Use: | | | | | | | | | | | Clinical Application | | | | | | Mode of Operation (*includes simultaneous B-mode) | | | General | Specific | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | (Track I only) | (Tracks I & III) | | | | | | | | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal (See Note 3) | N | N | N | | N | Note 1 | Notes 2, 7, 8 | | | Abdominal | N | N | N | | N | Note 1 | Notes 2, 7, 8 | | | Intra-operative (See Note 6) | | | | | | | | | | Intra-operative (Neuro.) | | | | | | | | | Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | | | Pediatric | N | N | N | | N | Note 1 | Notes 2, 7, 8 | | | Small Organ (See Note 5) | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Cardiac) | | | | | | | | | | Musculo-skel. (Convent.) | | | | | | | | | | Musculo-skel. (Superfic.) | | | | | | | | | | Intra-luminal | | | | | | | | | | Other (spec.) | | | | | | | | | | Cardiac Adult | | | | | | | | | Cardiac | Cardiac Pediatric | | | | | | | | | | Trans-esophageal (Cardiac) | | | | | | | | | | Other (spec.) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | Other (spec.) | | | | | | | | N= new indication; P= previously cleared; E= added under Appendix E Additional Comments: Color Doppler includes Power (Amplitude) Doppler Note 1: B/M, B/PW Doppler, B/Color Doppler/PW Doppler/PW Doppler, B/Power Doppler/PW Doppler, B/Color Doppler/Color M Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 6: Abdominal organs and peripheral vessel Note 7: Tissue Harmonic Imaging (THI) Note 8: 3D imaging | Concurrence of CDRH, Office of Device Evaluation (ODE) | |--------------------------------------------------------| | Prescription Use (Per 21 CFR 801.109) | Herlea (Division Sign-Off) Division of Reproductive, Abdominal and Radiological Dev 510(k) Number {14}------------------------------------------------ | 510(k) No.: | | | | | | | | | | |--------------------------------------------------------------------------------------------------|------------------------------|--|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------| | Device Name:<br>C3-7EP for use with SONOACE X8 | | | | | | | | | | | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | | | | General<br>(Track I only) | Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | | | | Specific<br>(Tracks I & III) | | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal Imaging<br>& Other | Fetal (See Note 3) | | P | P | P | | P | Note 1 | Notes 2, 7, 8 | | | Abdominal | | P | P | P | | P | Note 1 | Notes 2, 7, 8 | | | Intra-operative (See Note 6) | | | | | | | | | | | Intra-operative (Neuro.) | | | | | | | | | | | Laparoscopic | | | | | | | | | | | Pediatric | | P | P | P | | P | Note 1 | Notes 2, 7, 8 | | | Small Organ (See Note 5) | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | Trans-rectal | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | Trans-urethral | | | | | | | | | | | Trans-esoph. (non-Cardiac) | |…
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...