← Product Code [EHD](/submissions/DE/subpart-b%E2%80%94diagnostic-devices/EHD) · K232068

# HyperLight Portable X-ray Unit (K232068)

_Changzhou Sifary Medical Technology Co., Ltd. · EHD · Sep 1, 2023 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/EHD/K232068

## Device Facts

- **Applicant:** Changzhou Sifary Medical Technology Co., Ltd.
- **Product Code:** [EHD](/submissions/DE/subpart-b%E2%80%94diagnostic-devices/EHD.md)
- **Decision Date:** Sep 1, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.1800
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Pediatric

## Indications for Use

Intended as extraoral x-ray sources to be used with intraoral image receptors for diagnostic imaging by dentists or dental technicians.

## Device Story

Handheld diagnostic X-ray system; generates ionizing radiation for dental radiographic examination of teeth, jaw, and oral structures. Operates by exposing intra-oral receptors (analog film, digital phosphorous plates, or digital sensors) to X-rays. Used in general dentistry, particularly for special needs or nursing home patients who cannot be easily moved. Operated by dentists or dental technicians. Features internal timer to control exposure duration; user interface includes LCD display with up-down buttons for exposure time selection, plus preset times based on patient size and tooth selection. Device includes Pb-filled acrylic scatter shield for operator protection. Output is diagnostic X-ray image captured on external receptor; assists clinicians in disease diagnosis. Benefits include portability and utility in non-standard clinical settings.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), software validation (IEC 62304), biocompatibility, and usability testing. Device performance verified against recognized standards including IEC 60601-1-3, IEC 60601-2-65, and IEC 61223-3-4.

## Technological Characteristics

Handheld X-ray unit; 65kVp, 2.5mA; constant potential (DC) waveform. Total filtration 1.8mmAl. Source-to-skin distance 20cm. Includes 13mm thick Pb-filled acrylic scatter shield. User interface: LCD with preset exposure times. Standards: IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-65, 62304, 62366, ISO 14971, IEC 61223-3-4. Software class: Moderate level of concern.

## Regulatory Identification

An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.

## Predicate Devices

- MobileX Portable X-ray System ([K180561](/device/K180561.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the logo for the Department of Health & Human Services - USA. To the right of that is the logo for the U.S. Food & Drug Administration. The FDA logo is in blue and white.

September 1, 2023

Changzhou Sifary Medical Technology Co., Ltd. % Mr. Lee W. Strong Regulatory Dept Manager 510k FDA Inc. 156 E. Granada Blvd. ORMOND BEACH FL 32176

Re: K232068

Trade/Device Name: HyperLight Portable X-ray Unit Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: Class II Product Code: EHD Dated: July 7, 2023 Received: July 12, 2023

#### Dear Mr. Strong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K232068

Device Name HyperLight Portable X-ray Unit

#### Indications for Use (Describe)

The HyperLight Portable X-ray Unit is a diagnostic X-ray system which is intended to be used by trained dentists and dental technicians as an extra-oral X-ray source for producing diagnostic x-ray images using intra-oral receptors. Its use is intended for both adults and pediatric subjects.

| Type of Use (Select one or both, as applicable)                                                               |                                                                                                              |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span>                      Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span>                      Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a stylized black and white drawing of a bird of prey, possibly an eagle or falcon. The bird is depicted in profile, facing right, with its head and beak clearly visible. The wing is prominently displayed, featuring several distinct feathers that create a sense of depth and texture. The overall design is sleek and modern, suggesting a logo or emblem.

# 510k FDA Consulting

Medical Device Clearances

156 East Granada Blvd.

Ormond Beach, FL 32176

386-506-8711

# 510(k) Summary - K232068

#### Submitter/Applicant

Changzhou Sifary Medical Technology Co., Ltd No.99 Qingyang Road, Xuejia County, Xinbei District Changzhou, Jiangsu, China 213000

Phone: + 86-186-02591004 Contact: Tao Huan, Head of Firm (simon@sifary.com)

Date Prepared: July 10, 2023

#### Preparer/Consultant

510K FDA Inc. 156 East Granada Blvd. Ormond Beach, FL 32176

Phone:

386-506-8711 Fax: (386) 675-4621 Lee Strong, Regulatory Dept. Mgr (lee@510kfda.com) Primary Contact: Secondary Contacts: Claude Berthoin, CEO (claude@denterpriseintl.com).

#### Device Classification

| Trade/Model Names:    | HyperLight Portable X-ray Unit    |
|-----------------------|-----------------------------------|
| Common Name:          | Portable X-ray System             |
| Regulation Name:      | Extra-oral Source X-ray System    |
| Regulation Number:    | 21 CFR 872.1800                   |
| Primary Product Code: | EHD                               |
| Classification Name:  | Unit, X-ray, Extraoral with Timer |
| Regulatory Class:     | II                                |
| 510k Review Panel:    | Dental                            |

{4}------------------------------------------------

### Predicate Device

The subject device claims equivalence to the following legally marketed predicate:

| 510(k) Number:        | K180561                                      |
|-----------------------|----------------------------------------------|
| Applicant:            | MobileX Portable X-ray System (Model T-100). |
| Date Cleared:         | April 4, 2018                                |
| Regulation Name:      | Extra-oral Source X-ray System               |
| Regulation Number:    | 21 CFR 872.1800                              |
| Primary Product Code: | EHD                                          |
| Classification Name:  | Unit, X-ray, Extraoral with Timer            |
| Regulatory Class:     | II                                           |
| 510k Review Panel:    | Dental                                       |

#### Indications for Use

The device is a diagnostic X-ray system which is intended to be used by trained dentists and dental technicians as an extra-oral X-ray source for producing diagnostic x-ray images using intra-oral receptors. Its use is intended for both adults and pediatric subjects.

#### Intended Use

Intended as extraoral x-ray sources to be used with intraoral image receptors for diagnostic imaging by dentists or dental technicians.

#### Device Description

HyperLight Portable X-ray Unit of Changzhou Sifary Medical Technology Co., Ltd., is a handheld x-ray device. The technology of portable x-ray devices was originally developed in the 1950s and was originally designed to be used in situations where there was no access to fixed x-ray units. With new technological possiblities arising, the use of portable handheld devices are becoming more mainstream.

The subject device is designed for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structure by exposing an x-ray image receptor to ionizing radiation. The x-ray source, a tube, is located inside the handheld device. All three conventional types of intraoral receptors can be used with this device-analog x-ray film, digital phosphorous plates, and digital x-ray sensors. The intraoral x-ray detectors are not part of the subject device. The functions of the HyperLight Portable device are supported by software. The software package is of Moderate level of concern and it's not based on the predicate system.

This device is used in general dentistry and is supplied with an internal timer to control the duration of the x-ray source to the patient. The handheld x-ray device is a choice model to assist doctors with special need patients, nursing home patients in the office that cannot be easily moved, as well as other special situations. The choice of an x-ray generator is a matter of functional utility in the dental operatory and personal preference by the medical professional.

{5}------------------------------------------------

## Comparison of Technological Characteristics with Predicate

| Device<br>Characteristic            | Subject Device<br>HyperLight Portable<br>X-ray Unit                                                                                                                                                                                                                                                       | Predicate Device Mobile-<br>X Portable X-ray System<br>(K180561)                                                                                                                                                                                                                                           | Comparison                                                                 |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| 510(k) Owner                        | Changzhou Sifary Medical<br>Technology Co., Ltd.<br>(China)                                                                                                                                                                                                                                               | Denterprise International,<br>Inc.<br>(USA)                                                                                                                                                                                                                                                                | NA                                                                         |
| Classification &<br>Product Code    | 872.1800; EHD                                                                                                                                                                                                                                                                                             | 872.1800; EHD                                                                                                                                                                                                                                                                                              | Similar                                                                    |
| Intended Use                        | Intended as extraoral x-ray<br>sources to be used with<br>intraoral image receptors for<br>diagnostic imaging by<br>dentists or dental<br>technicians.                                                                                                                                                    | Intended as extraoral x-ray<br>sources to be used with<br>intraoral image receptors for<br>diagnostic imaging by<br>dentists or dental<br>technicians.                                                                                                                                                     | Similar                                                                    |
| Indication for Use                  | The device is a diagnostic<br>X-ray system which is<br>intended to be used by<br>trained dentists and dental<br>technicians as an extra-oral<br>X-ray source for producing<br>diagnostic x-ray images<br>using intra-oral receptors.<br>Its use is intended for both<br>adults and pediatric<br>subjects. | The device is a diagnostic<br>X-ray system, which is<br>intended to be used by<br>trained dentists and dental<br>technicians as an extra-oral<br>X-ray source for producing<br>diagnostic x-ray images<br>using intra-oral receptors.<br>Its use is intended for both<br>adults and pediatric<br>subjects. | Similar                                                                    |
| Size                                | 11.85" x 4.58" x 9.97"                                                                                                                                                                                                                                                                                    | 6.5" x 6.0" x 10.5"                                                                                                                                                                                                                                                                                        | Difference of<br>design, size                                              |
| Source to Skin<br>Distance          | 20 cm                                                                                                                                                                                                                                                                                                     | 20.5 cm                                                                                                                                                                                                                                                                                                    | Difference .5 cm                                                           |
| Cone diameter                       | 5.7 cm                                                                                                                                                                                                                                                                                                    | 6.0 cm                                                                                                                                                                                                                                                                                                     | Difference .3 cm                                                           |
| Device<br>Characteristic            | Subject Device<br>HyperLight Portable<br>X-ray Unit                                                                                                                                                                                                                                                       | Predicate Device Mobile-<br>X Portable X-ray System<br>(K180561)                                                                                                                                                                                                                                           | Comparison                                                                 |
| User interface                      | Up-down buttons for<br>exposure time selection<br>with timer display.<br>Additionally, several user-<br>selectable preset times with<br>patient size, and tooth<br>selection icons on an LCD<br>display.                                                                                                  | Up-down buttons for<br>exposure time selection with<br>timer display. Additionally,<br>several user-selectable<br>preset times with patient<br>size, image-receptor type,<br>and tooth selection icons on<br>an LCD display.                                                                               | Difference;<br>subject device<br>does not set<br>image-receptor<br>type    |
| Backscatter<br>radiation protection | 159.5mm dia. 13mm thick<br>Pb-filled acrylic plastic<br>scatter shield                                                                                                                                                                                                                                    | 153mm dia. 12mm thick<br>Pb-filled acrylic plastic<br>scatter shield                                                                                                                                                                                                                                       | 6.5 mm<br>difference in<br>diameter and 1<br>mm difference in<br>thickness |
| Exposure switch                     | Exposure trigger at the<br>lower front area of the main<br>body or remote switch.                                                                                                                                                                                                                         | Exposure button at front<br>cover on right hand side or<br>remote switch.                                                                                                                                                                                                                                  | Different<br>location                                                      |
| Electrical<br>Information           |                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                            |                                                                            |
| Exposure time                       | 0.02 ~ 2.0 seconds<br>in .01-.40s increments (20<br>steps)                                                                                                                                                                                                                                                | 0.01 ~ 1.3 seconds in 0.01<br>or 0.05 increments                                                                                                                                                                                                                                                           | Difference;<br>subject device<br>has higher<br>exposure time<br>           |
| Time accuracy                       | ±5% or ±20ms, whichever          is greater                                                                                                                                                                                                                                                               | ± (10% +1 ms)                                                                                                                                                                                                                                                                                              | Slight difference                                                          |
| mA                                  | 2.5mA                                                                                                                                                                                                                                                                                                     | 2mA                                                                                                                                                                                                                                                                                                        | Subject has<br>greater mA                                                  |
| kVp                                 | 65kVp                                                                                                                                                                                                                                                                                                     | 70kVp                                                                                                                                                                                                                                                                                                      | Subject has<br>lesser kVp                                                  |
| Waveform                            | Constant Potential (DC)                                                                                                                                                                                                                                                                                   | Constant Potential (DC)                                                                                                                                                                                                                                                                                    | Similar                                                                    |
| Total Filtration                    | 1.8mmAl                                                                                                                                                                                                                                                                                                   | 1.5mmAl                                                                                                                                                                                                                                                                                                    | Subject has<br>greater Total<br>Filtration                                 |
| Device<br>Characteristic            | Subject Device<br>HyperLight Portable<br>X-ray Unit                                                                                                                                                                                                                                                       | Predicate Device Mobile-<br>X Portable X-ray System<br>(K180561)                                                                                                                                                                                                                                           | Comparison                                                                 |
| Performance<br>standards            | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-3<br>IEC 60601-1-6<br>IEC 60601-2-65<br>IEC 62304<br>IEC 62366<br>ISO 14971<br>IEC 61223-3-4                                                                                                                                                                  | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-3<br>IEC 60601-1-6<br>IEC 60601-2-65<br>IEC 62304<br>IEC 62366<br>ISO 14971                                                                                                                                                                                    | Difference;<br>subject device<br>tested to<br>IEC61223-3-4                 |

The following table compares technological and other characteristics of the subject and predicate device.

{6}------------------------------------------------

{7}------------------------------------------------

The above comparison shows the subject and predicate devices have substantially similar technological characteristics. Differences show up in the shape, size, design of the device and those are in cm and mm measurements of slight difference. The exposure time is slightly less with the subject device matched to the predicate. The differences of the device are minor and do not raise new issues of safety and effectiveness.

### Non-Clinical Performance Data

The following performance was completed on the subject device in support of the substantial equivalence determination of the predicate device. Clinical data was not needed to support substantial equivalence.

- Electrical Safety and EMC
- Software Validation ●
- Biocompatibility ●
- Usability
- Clinical Comparison ●
- Risk Assessment ●
- . All tests were performed in accordance with ISO standards and tests are recognized by FDA.
- None of the standards were adapted for application to the device under . review.
- . There were no requirements of any standard that were not applicable to the device.

{8}------------------------------------------------

- No deviations from the standards were applied. ●
- No differences exist between the tested device and the device to be marketed.
- . Conformity with all standards was determined by the device manufacturer, Changzhou Sifary Medical Technology Co., Ltd
- Electrical tests performed by TÜV Rheinland (Shanghai) Co., Ltd.

#### Specific Guidance Document

There are three FDA Specific Guidance Documents associated with the device:

- Radiation Safety Consideration for X-ray Equipment Designed for Hand-Held . Use. This manufacturer utilized this guidance to develop this device to ensure the safety of this device for both the operators and the patients.
- . Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. Details of this guidance are provided within the Software Validation Report.
- Pediatric Information for X-ray Imaging Device Premarket Notifications. The . submission includes pediatric patients in the Indications for Use and labelling.

### Labels

The labels on the device show that this device conforms to the following:

21 CFR 1020 Subchapter J: Performance Standards for Ionizing Radiation Emitting Products,

21 CFR 1020.30: Diagnostic x-ray systems and their major components,

21 CFR 1020.31: Radiographic Equipment

#### Substantial Equivalence

The above comparison chart shows the subject and predicate devices are substantially equivalent in technological characteristics.

Both devices have:

- The same function and used in the same environment.
- The same indications for use and the same intended use. ●
- The same manufacturing process and technological characteristics. ●
- Both devices have completed the ISO standardized testing and have passed ● and the tests are in the comparison chart shown above.

{9}------------------------------------------------

#### Conclusion

The subject and predicate device have the same indications for use, the same intended use and the same technological characteristics. The HyperLight Portable X-ray Unit performs the same identical functions, in the same environment as the predicate device. HyperLight Portable X-ray Unit uses the same technology as the predicate device, based on well-known technology. HyperLight Portable X-ray Unit is as safe and effective as the predicate device. We believe the subject device does not introduce any new safety concerns and is substantially equivalent to the predicate device. In conclusion, the subject device, HyperLight Portable X-ray Unit, is at least as safe and effective as the predicate device and warrants a finding of substantial equivalence to the legally marketed device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/EHD/K232068](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/EHD/K232068)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
