Ai Ray Dental X-Ray Device

K222569 · Guilin Woodpecker Medical Instrument Co., Ltd. · EHD · Nov 23, 2022 · Dental

Device Facts

Record IDK222569
Device NameAi Ray Dental X-Ray Device
ApplicantGuilin Woodpecker Medical Instrument Co., Ltd.
Product CodeEHD · Dental
Decision DateNov 23, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.1800
Device ClassClass 2
AttributesPediatric

Intended Use

Ai Ray Dental X-Ray Device is an extraoral diagnostic dental X-ray source to produce X-ray images using intraoral image receptors. It is indicated for use by a dental technician for both adult and pediatric patients.

Device Story

Portable, handheld dental X-ray source; supercapacitor-operated; rechargeable 10.8V DC Li-ion polymer battery. Used in dental offices, orthodontic offices, and hospital wards by dentists or dental technicians. Emits X-rays by supplying high voltage to an X-ray tube, freeing electrons from cathode to strike anode. Produces diagnostic quality X-ray images using intraoral film or digital receptors. Features jog dial for mode selection, preset exposure times based on patient size/tooth type, and Pb-filled acrylic backscatter shield. Output assists clinicians in diagnosing diseases of teeth, jaw, and oral structures. Benefits include mobility for environments where traditional X-ray systems are impractical.

Clinical Evidence

No clinical data was required to demonstrate substantial equivalence. Bench testing only.

Technological Characteristics

Handheld extraoral X-ray system; 70 kV fixed voltage; 3 mA fixed current; 0.04-1.0s exposure time. Powered by 10.8V DC Li-ion polymer battery. Includes Pb-filled acrylic backscatter shield. Standards: ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-65, IEC 61223-3-4, 21 CFR 1020.30/31.

Indications for Use

Indicated for use by dental technicians for adult and pediatric patients to produce diagnostic dental radiographs of teeth, jaw, and oral structures using intraoral image receptors.

Regulatory Classification

Identification

An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services logo. The logo on the right is the FDA logo, which stands for U.S. Food & Drug Administration. The FDA logo is in blue. November 23, 2022 Guilin Woodpecker Medical Instrument Co., Ltd. % Charles Mack Principal Engineer IRC 2950 E Lindrick Drive CHANDLER, AZ 85249 Re: K222569 Trade/Device Name: Ai Ray Dental X-Ray Device Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral Source X-Ray System Regulatory Class: Class II Product Code: EHD Dated: August 24, 2022 Received: August 24, 2022 Dear Charles Mack: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, 2022.11.23 *Lu Jiang* 09:33:38 -05'00' Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K222569 Device Name Ai Ray Dental X-Ray Device Indications for Use (Describe) Ai Ray Dental X-Ray Device is an extraoral diagnostic dental X-ray source to produce X-ray images using intraoral image receptors. It is indicated for use by a dental technician for both adult and pediatric patients. | Type of Use (Select one or both, as applicable) | |------------------------------------------------------------------------------------------------------------------------------| | <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Remediation Use (Part 21 CFR 221 Subpart B) </span> | | <span style="font-size: 10pt;"> <input type="checkbox"/> Source Term Control Use (21 CFR 1271 Subpart D) </span> | X Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ | K22269 | 510(k) SUMMARY | |----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Preparation Date: | November 21, 2022 | | Manufacturer's Name and Address: | Guilin Woodpecker Medical Instrument<br>Co Ltd<br>Information Industrial Park, Guilin<br>National High-Tech Zone, Guilin City,<br>Guangxi Province, China 541004 | | Corresponding Official: | Charles Mack | | Telephone Number: | 931-625-4938 | | Email Address: | charliemack@irc-us.com | | Trade Name: | Ai Ray Dental X-Ray Device | | Common Name(s): | unit, x-ray, extraoral with timer, | | Regulation Name(s): | Extraoral source x-ray system | | Regulation Number(s): | 21CFR872.1800 | | Product Code: | EHD | | Device Class: | Class II | | Predicate Device: | Vatech Co., Ltd K200182 | | Trade Name: | EzRay Air Portable (Model: VEX-P300) | | Common Name(s): | unit, x-ray, extraoral with timer, | | Regulation Name(s): | Extraoral source x-ray system | | Regulation Number(s): | 21CFR872.1800 | | Product Code: | EHD | | Device Class: | Class II | {4}------------------------------------------------ #### Device Description: Ai Ray Dental X-Ray Device is a super capacitor-operated, portable dental X-ray source designed for handheld operation. It is designed to produce diagnostic quality X-rays images utilizing either film or digital imaging techniques. The Portable Dental X-ray Device is designed for use in a dental office. It can also be used in other similar environments (orthodontic office, general practitioner's office, hospital ward, etc.) where appropriate safeguards are implemented. The device uses a rechargeable supercapacitor to allow for the use of the Portable Dental X-ray Device where transportation or use of other x-ray devices might be prohibitive due to the other device's size and/or lack of mobility. The proposed Ai Ray Dental X-Ray Device is similar in design to the predicate EzRay Air Portable, VEX-P300 (K200182). The proposed Ai Ray Device and the predicates is a portable dental X-ray system intended for use by a trained and qualified dentist or dental technician for both adult and pediatric subjects for producing diagnostic dental radiographs for treatment of diseases of the teeth, jaw, and other oral structures using intra-oral image receptors. The differences between the proposed Ai Ray Dental X-Ray Device and the predicate EzRay Air Portable, VEX-P300 (K200182) are as follows: 1. The opening key is positioned differently, EzRay Air Portable at the bottom and Ai Ray at the top. 2. The structural shapes of the two devices are different. #### Principles of Operation The equipment acquires images by emitting X-rays continuously on a human tooth. X-rays are emitted when high voltage is supplied to the X-ray tube assembly, which frees electrons from the cathode. They hit the anode to produce X-rays. {5}------------------------------------------------ | Characteristics | Subject Device | Predicate Device | Discussion | |----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------| | Device | Ai Ray Dental X-Ray Device | EzRay Air Portable (Model: VEX-P300) | - | | 510K Applicant | Guilin Woodpecker Medical Instrument Co., Ltd. | Vatech Co., Ltd | - | | 510(K) Number Regulation Number | Pending<br>CFR872.1800 | K200182<br>CFR872.1800 | Identical | | Product Code | EHD | EHD | Identical | | Classification Name | Extraoral source x-ray system | Extraoral source x-ray system | Identical | | OTC or Prescription | Prescription Use | Prescription Use | Identical | | Medical Specialty | Dental | Dental | Identical | | Indication for Use | Ai Ray Dental X-Ray Device is an extraoral diagnostic dental X-ray source to produce X-ray images using intraoral image receptors. It is indicated for use by a dentist or a dental technician for both adult and pediatric patients. | EzRay Air Portable (Model: VEX-P300) is an extraoral diagnostic dental X-ray source to produce X-ray images using intraoral image receptors. It is indicated for use by a dentist or a dental technician for both adult and pediatric patients. | Identical | | Principle of Use | X-ray tube | X-ray tube | Identical | | Mechanical | | | | | Size (L x W x H) | $114mm×363.8mm×2$<br>45.6mm | 280 x 165 x 296 mm | Different. Note 1 | | Source to Skin Distance | 200~230mm | 200 mm | Different. Note 2 | | X-ray Field Size | Ф5.6cm±0.1cm | 6cm round | Different. Note 3 | | Mechanical | | | | | Characteristics | Subject Device | Predicate Device | Discussion | | User Interface | Jog dial for operating mode selection. Additionally, several user-selectable preset times with patient size and tooth selection icons on a display module. | Jog dial for operating mode selection. Additionally, several user-selectable preset times with patient size and tooth selection icons on a display module. | Identical | | Backscatter Radiation Protection | 162 mm dia., Pb-filled acrylic plastic, Back Scattering shield | 165 mm dia., Pb-filled acrylic plastic, Back Scattering shield | Different. Note 4 | | Exposure Switch | Exposure button on the handset | Exposure button on the handset | Identical | | Tube Head Mounting | Handheld | Handheld | Identical | | Electrical | | | | | Energy Source | Rechargeable 10.8V DC Li-ion polymer battery pack (Nominal Capacity: 2450~2500mAh ) | Rechargeable 21.6 V DC Li-ion polymer battery pack (Nominal Capacity: 2,500 mAh) | Different. Note 5 | | Exposure Time | 0.04-1.0 seconds in 0.01 increments | 0.05 - 1.0 seconds in 0.01 increments | Different. Note 6 | | Tube Current (mA) | 3 mA fixed, error±20% | 2.5 mA fixed | Different. Note 7 | | Tube Voltage (kVp) | 70 kV fixed, error±10% | 60 or 65 kV fixed | Different. Note 8 | | Waveform | Constant Potential (DC) | Constant Potential (DC) | Identical | | Applied Standard | IEC 60601-1, IEC 60601-1-3, IEC 60601-2-65, IEC 60601-1-2, 21 CFR 1020.30, 1020.31 | IEC 60601-1, IEC 60601-1-3, IEC 60601-2-65, IEC 60601-1-2, 21 CFR1020.30, 1020.31 | Identical | # Comparison of Technological Characteristics with the Predicate Device {6}------------------------------------------------ {7}------------------------------------------------ #### Differences Discussion: Note1 The physical shape/size difference has no impact on safety and performance. Both devices conform to the same safety and performance standards. Note 2 The distance difference has no impact on safety and performance. Both devices conform to the same safety and performance standards. Note 3 The difference has no impact on safety and performance. Both devices conform to the same safety and performance standards. Note 4 The size of the backscattering shield is different due to different device shapes. The radiation leakage value of the device with a backscattering shield after testing meets the requirements of the FDA. Security and performance are not affected. Note 5 Although the battery specifications are different, they conform to the same requirement of IEC 62133 for the rechargeable battery. The differences do not raise issues of safety and effectiveness. Note 6 Both devices conform to the same requirement of federal standard requirements. The performance test demonstrated that the difference doesn't raise safety and effectiveness issues. Note 7 Although the X-ray tube's specifications are not identical and conform to the same performance standards IEC 60601-2-65, The performance test demonstrated the difference doesn't raise safety and effectiveness issues. Note 8 Although the X-ray tube's specifications are not identical, and they conform to the same performance standards IEC 60601-2-65. The performance test demonstrated that the difference doesn't raise safety and effectiveness issues. {8}------------------------------------------------ ## Performance Testing Performance testing was provided to support the substantial equivalence determination and to validate and verify that the Ai Ray Dental X-Ray Device met all requirements of related applicable standards. The results of these tests demonstrate compliance with the requirements of the consensus standards noted below. ## Non-Clinical Performance Testing Non-clinical performance testing was performed to verify substantial equivalence to predicate devices. ### Safety and EMC: - ANSI AAMI ES60601-1:2005 Medical electrical equipment Part 1: General । requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) - IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests #### Performance: - IEC 60601-1-3 Edition 2.1 2013-04 Medical electrical equipment Part 1-3: । General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment - IEC 60601-1-6 Edition 3.1 2013-10 Medical electrical equipment Part 1-6: -General requirements for basic safety and essential performance - Collateral standard - IEC 60601-2-65 Edition 1.1 2021-05 CONSOLIDATED VERSION Medical electrical equipment - Part 2-65: Particular requirements for the basic safety and essential performance of dental intra-oral-X-ray equipment - IEC 61223-3-4 First edition 2000-03 Evaluation and routine testing in medical imaging departments - Part 3-4: Acceptance tests - Imaging performance of dental X-ray - IEC 62304:2006/A1:2015 "Medical device software Software life cycle processes." {9}------------------------------------------------ - IEC 62366-1 Edition 1.0 2015-02 Medical devices Part 1: Application of usability engineering to medical devices [Including CORRIGENDUM 1 (2016)] - FDA Guidance for the Content of Premarket Submissions for Software -Contained in Medical Devices - FDA Guidance for Radiation Safety Considerations for X-Ray Equipment -Designed for Handheld Use - FDA Guidance: "Pediatric Information for X-ray Imaging Device Premarket -Notifications," dated November 28, 2017 - 21 CFR 1020.30 Diagnostic x-ray system and their major components - - Content of Premarket Submissions for Management of Cybersecurity in -Medical Devices ### Clinical Performance Data No clinical data was required to demonstrate substantial equivalence. ## Conclusions: The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The Ai Ray Dental X-Ray Device is substantially equivalent to the Vatech Co., Ltd, Inc. EzRay Air Portable (Model: VEX-P300) cleared under K200182 with respect to the indications for use, target populations, treatment method, and technological characteristics. -END-
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...