IBI INQUIRY CLEARICE INTRACARDIAC ECHOCARDIOGRAPHY ULTRASOUND IMAGING CATHETER
Applicant
Irvine Biomedical,Inc.(Ibi)
Product Code
DQO · Cardiovascular
Decision Date
Mar 20, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1200
Device Class
Class 2
Indications for Use
The IBI Inquiry™ ClearICE™ Intracardiac Echocardiography Catheter is intended for intra-cardiac and intra-luminal visualization of cardiac anatomy and physiology as well as visualization of other devices in the heart.
Device Story
The IBI Inquiry™ ClearICE™ is a single-use, disposable, steerable intracardiac echocardiography (ICE) catheter. It features a phased-array ultrasonic imaging transducer at the distal tip. The catheter is 9 Fr (2.97 mm) in diameter with an insertable length of 90-110 cm. It provides bi-directional and quad-directional steering via a push/pull handle mechanism and rotating knob. The device connects to a GE Vivid i Diagnostic Ultrasound System using a reusable GE Connector Cable. Used in clinical settings, the catheter allows physicians to visualize cardiac anatomy, physiology, and other interventional devices within the heart. The ultrasound images are displayed on the GE system, assisting the clinician in real-time procedural guidance and decision-making during cardiac interventions.
Clinical Evidence
Bench testing only. The device was validated in compliance with ISO 13485, ISO 14971, ISO 10993 (biological evaluation), and various IEC/EN safety standards for medical electrical equipment. No clinical study data was provided.
Technological Characteristics
Phased-array ultrasonic imaging transducer; 9 Fr diameter; 90-110 cm length; bi-directional/quad-directional steering mechanism. Materials evaluated per ISO 10993-1. Sterilization via Ethylene Oxide (ANSI/AAMI/ISO 11135). Connectivity via GE Connector Cable to GE Vivid i Ultrasound System. Complies with IEC 60601-1, IEC 60601-2-2, and ISO 10555-1.
Indications for Use
Indicated for intra-cardiac and intra-luminal visualization of cardiac anatomy, physiology, and other devices in the heart. No specific age or gender restrictions provided.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
K101239 — VIEWFLEX PLUS ICE CATHETER · St Jude Medical · Jun 2, 2010
K212959 — ACUSON AcuNav Volume Intracardiac Echocardiography(ICE) Catheter · Siemens Medical Solutions USA, Inc. · Oct 15, 2021
K240050 — SOUNDSTAR CRYSTAL Ultrasound Catheter · Biosense Webster, Inc. · Jun 12, 2024
K121381 — VIEWFLEX XTRA ICE CATHETER · St Jude Medical · Jun 7, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
K 070011
# SECTION 14
## 510(k) Summary of Safety and Effectiveness
MAR 2 0 2007
- Sponsor: Irvine Biomedical Inc, 2375 Morse Avenue Irvine, CA 92614, USA
Contact Person: Rohit Patel Sr. Regulatory Affairs Specialist Telephone: (949) 271-1127 Fax: (949) 851-3062 E-mail: Rpatel(@sjm.com
Submission date: December 29, 2006
Device Name: IBI Inquiry™ ClearICE™ Intracardiac Echocardiography Catheter
Common Name: Intracardiac/Intravascular Ultrasound Catheter
Classification: Regulatory Class: II Regulatory Category: Tier II
| | FR Number | Product Code |
|----------------------------------|-----------|--------------|
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
| Diagnostic Ultrasound Catheter | 870.1200 | 74-DQO |
## Predicate Devices:
- #K042593 (January 5, 2005) cleared as ACUSON AcuNav 8Fr Ultrasound Catheter.
- #K033650 (February 24, 2004) cleared as ACUSON AcuNav 90/10 Diagnostic . Ultrasound Catheter.
- #K992632 (November 8, 2001) cleared as ACUSON AcuNav Diagnostic ● Ultrasound Catheter.
### Device Description :
The IBI Inquiry™ ClearICE™ Intracardiac Echocardiography Catheter is comprised of a single-use, disposable ultrasonic phased-array imaging transducer. A Inquiry ClearICE catheter is steerable: catheter, 9 Fr (2.97 mm) in diameter, 90 cm to 110 cm insertable length and available in various curve configuration.
The distal portion of the catheter can be deflected in four directions in two orthogonal planes: left-right (in a plane perpendicular to imaging plane) and anterior-posterior (in a plane coincident with the image plane). Bi-directional steering controlled by push/pull
SECTION 14
{1}------------------------------------------------
handle mechanism and quad-directional steering controlled by a push/pull mechanism and rotating knob.
The non-sterile, reusable GE Connector Cable is use to connect IBI InquiryTM ClearICE™ Intracardiac Echocardiography Catheter to the GE Vivid i Ultrasound system. The IBI Inquiry™ ClearICE™ Intracardiac Echocardiography Catheter is used with commercially available GE Connector Cable and Vivid- i Diagnostic Ultrasound System.
The IBI Inquiry™ ClearICE™ Intracardiac Echocardiography Catheter is comprised of three major components: (1) the catheter itself; (2) the steering mechanism; and (3) the reusable GE Connector Cable & GE Vivi i Ultrasound System.
The IBI Inquiry™ ClearICE™ Intracardiac Echocardiography Catheter is substantially equivalent to ACUSON AvuNav ultrasound catheters that are already cleared for USA distribution under the following 510(k) PreMarket Notification numbers:
- #K042593 (January 5, 2005) cleared as ACUSON AcuNav 8Fr Ultrasound . Catheter.
- #K033650 (February 24, 2004) cleared as ACUSON AcuNav 90/10 . Diagnostic Ultrasound Catheter.
- #K992632 (November 8, 2001) cleared as ACUSON AcuNav Diagnostic . Ultrasound Catheter.
The IBI Inquiry™ ClearICE Intracardiac Echocardiography Catheter designed and validated in compliance with the following standards:
- ISO 13485:2003: Medical devices -- Quality management systems --. Requirements for regulatory purposes
- 21 CFR Section 820: FDA Quality System Regulation .
- . Guidance Document -- 9/30/1997: Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers
- ISO 14971:2001: Medical devices Application of risk management to medical . devices
- . ISO 10555-1:1997: Sterile, single-use Intravascular catheter - Part 1: General requirements
- . IEC/EN 60601-1-1:2001: Medical electrical equivment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-2-2:1998: Medical Electrical equipment Part 2-2: Particular . requirements for the safety of high frequency surgical equipment
- BS EN 60601-1:1990, BS 5724: Part1: 1989: Medical electrical equipment Part . 1 : General requirements for safety
- . ISTA 2A: Performance Test for Individual Packaged Products 150 lb or Less
- ANSI / AAMI / ISO 11135:1994: Medical devices validation and routine ● control of Ethylene Oxide Sterilization
- . BS EN550:1994: Sterilization of Medical Devices - Validation and Routine Control of Ethylene Oxide sterilization
{2}------------------------------------------------
- BS EN556:2001: Sterilization of Medical Devices Requirements for Terminally . Sterilized Device to be labeled "Sterile"
- AAMI TIR28:2001: Product adoption and process equivalency for ethylene oxide . sterilization
- ISO 10993-7:1996: Biological evaluation of medical devices Part 7: Ethylene . oxide sterilization residuals
- ISO 10993-1:2003: Biological evaluation of medical devices Part 1: Evaluation . and testing
# Indication for Use:
The IBI Inquiry™ ClearICE™ Intracardiac Echocardiography Catheter is intended for intra-cardiac and intra-luminal visualization of cardiac anatomy and physiology as well as visualization of other devices in the heart.
# Technological Comparison to Predicate Device:
The IBI's Inquiry™ Intracardiac Echocardiography Catheter is substantially equivalent in its technologies and functionality to the ACUSON AcuNav ultrasound catheters that are already cleared for USA distribution under the following 510(k) PreMarket Notification numbers #K042593 (January 5, 2005), #K033650 (February 24, 2004), and #K992632 (November 8, 2001).
The IBI's Inquiry™ Intracardiac Echocardiography Catheter and ACUSON AcuNav ultrasound catheters are ultrasound-tipped catheter devices for ultrasound imaging.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The caduceus is depicted with intertwined snakes and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Rohit Patel Sr. Regulatory Affairs Specialist Irvine Biomedical, Inc. 2375 Morse Avenue IRVINE CA 92614
MAR 2 0 2007
Re: K070011
Trade Name: IBI Inquiry" ClearICE™ Intracardiac Echocardiography Catheter Regulation Number: 21 CFR §892.1570 Regulation Name: Diagnostic ultrasonic transducer Product Code: ITX Regulation Number: 21 CFR §870.1200 Regulation Name: Diagnostic intravascular catheter Produce Code: DOO Regulatory Class: II Dated: December 29, 2006 Received: January 3, 2007
Dear Mr. Patel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the IBI Inquiry" ClearICE" Intracardiac Echocardiography Catheter, as decriaed in your premarket notification:
Transducer Model Number - Inquiry - ClearICE™
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Rohit Patel
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this letter, please contact Andrew Kang, M.D. at (240) 276-1373.
Sincerely yours,
David A. Segerson
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{5}------------------------------------------------
# SECTION 13
#### Indications for Use
510(k) Number (if known): _K070011
Device Name: IBI Inquiry™ ClearICE™ Intracardiac Echocardiography Catheter
Indications for Use:
The IBI Inquiry™ ClearICE™ Intracardiac Echocardiography Catheter is intended for intra-cardiac and intra-luminal visualization of cardiac anatomy and physiology as well as visualization of other devices in the heart.
1 Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Thind A. Segman
Division Sign-Off) Division of Reproductive, Abdominal, a Radiological Devices 510(k) Number
Page / of /
{6}------------------------------------------------
# Diagnostic Ultrasound Indications for Use Form
510(k) Number (if known): K070011
Device Name: Inquiry™ ClearlCE™ Intracardiac Echocardiography Catheter
Indications for Use: The Inquiry™ ClearlCE™ Intracardiac Echocardiography Catheter is intended for intra-cardiac and intra-luminal visualization of cardiac anatomy and physiology as well as visualization of other devices in the heart.
Transducer: Inquiry™ ClearICE™ Intracardiac Echocardiography Catheter
| Clinical Application<br>Anatomy/Region of Interest | A | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other | |
|----------------------------------------------------|---|---|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|--|
| Ophthalmic | | | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | | | |
| Abdominal[1] | | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | | |
| Small Organ (specify)[2] | | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | | |
| Cardiac[3] | | N | N | N | N | N | N | N | N | N | | | |
| Transesophageal | | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | | |
| Intra-luminar | | N | N | N | N | N | N | N | N | N | | | |
| Peripheral Vascular | | | | | | | | | | | | | |
| Musculo-skeletal (Conventional) | | | | | | | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | | | | | | | |
| Other (intra-cardiac) | | N | N | N | N | N | N | N | N | N | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E Notes: [*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Amplitude/PWD.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
> Concurrence of CDRH, Office of Device Evaluation (ODE) morn
(Divin .. Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Page ( of 1
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.