The device is air pressure massager intended to temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas.
Device Story
System comprises air pump, leg sleeves, and hoses; pump connects to sleeves via hoses. Each sleeve contains 4 compression chambers. Device operates by sequentially inflating and then deflating air chambers in a cycle to provide compression massage from limb end toward body center. Microprocessor-controlled system uses sensors to monitor pressure and timing, cycling airflow to achieve compression. Used in clinics, hospitals, athlete training, and home environments; operated by patients or clinicians. Output is mechanical compression of body parts. Benefits include temporary relief of minor muscle aches/pains and increased circulation.
Clinical Evidence
Bench testing only. Device tested for compliance with AAMI ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 10993-5 (cytotoxicity), and ISO 10993-10 (irritation/sensitization). Functional testing included minimum/maximum air pressure, maximum electric current, maximum airflow, noise level, and mode verification. Results compared favorably to predicate devices.
Technological Characteristics
Powered inflatable tube massager; 4 compression chambers per sleeve; Thermoplastic Urethane sleeve material; microprocessor-controlled compressor and valve system; 110V, 60Hz power source; pressure range 20-200 mmHg; cycle time 70s. Complies with AAMI ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 10993-5, and ISO 10993-10.
Indications for Use
Indicated for temporary relief of minor muscle aches and pains and temporary increase in circulation to treated areas in general population.
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
Predicate Devices
Pulse 2.0, Pulse Pro 2.0 of NormaTec Industries, LP (K183169)
K201982 — Air compression therapy system FO-3001; Air compression therapy system FO-3008 · Foshan Hongfeng Co., Ltd. · Mar 25, 2021
K233536 — Air Compression Recovery System · Shenzhen Dongjilian Medical Tech Co., Ltd. · Dec 16, 2023
K181409 — Pressure Therapy System PT1003 · Xiamen Senyang Co., Ltd. · Feb 19, 2019
K182668 — Rapid Reboot Compression Therapy System · Rapid Reboot Recovery Products, LLC · Dec 14, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 23, 2020
Xiamen Senyang Co., Ltd. % Raymond Luo Technical Manager Shanghai SUNGO Management Consulting Co., Ltd. 13th F, 1500# Century Avenue Shanghai, China 200122
Re: K192358
Trade/Device Name: Pressure Therapy System PT1003 Regulation Number: 21 CFR 890.5650 Regulation Name: Powered Inflatable Tube Massager Regulatory Class: Class II Product Code: IRP Dated: September 12, 2019 Received: February 26, 2020
Dear Raymond Luo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Vivek J. Pinto, PhD Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K192358
Device Name Pressure Therapy System PT1003
Indications for Use (Describe)
The device is air pressure massager intended to temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas.
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995,
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
## A. Applicant:
Name: XIAMEN SENYANG CO., LTD. Address: 4-5 FLOOR, XINGBEI INDUSTRY, NO 95-99, WEST 2 ROAD, JIU TIANHU, XINGLIN XIAMEN, 361000, P.R. China
Official Contact Person Information Shanghai SUNGO Management Consulting Co., Ltd. Name: Raymond Luo Tel: 0086-21-68828050 Mail: fda.sungo@gmail.com
## B. Subject device:
| Trade name: | Pressure Therapy System PT1003 |
|------------------------------|-----------------------------------|
| Common name: | Powered inflatable Tube Massage |
| Classification name: | Massager, Powered Inflatable Tube |
| Regulation Medical Specialty | Physical Medicine |
| Regulation Number | 890.5650 |
| Product Code | IRP |
| Classification | Class II |
### C. Predicate device:
01. Pulse 2.0, Pulse Pro 2.0 of NormaTec Industries, LP, K183169
02. Rapid Reboot Compression Therapy System, Rapid Reboot Recovery Products, LLC, K182668
## D. Reference device:
K181409, Pressure Therapy System PT1003, Rx Only
### E. Indications for Use:
The device is air pressure massager intended to temporarily relieve minor muscle aches and to temporarily increase circulation to the treated areas.
### F. Prescriptive
The Pressure Therapy System PT1003, and the predicate (K183169 and K182668) are OTC devices.
### G. Device Description:
The system consists of an air pump, leg sleeves and hoses working together as one unit. The air pump is connected to the dedicated sleeves via a series of hoses; one sleeve has 4 compression chambers. The compression massage direction is from limb end to body center by inflating the air chambers sequentially and then deflating as one cycle, the pressure can be adjusted to avoid any discomfort to the patient. The sleeve works under the action of sensor and microprocessor. This software is to control the timing and the pressure reflected by the sensor, it cycles the airflow to reach the function of cycling compression of body parts.
{4}------------------------------------------------
# H. Substantial Equivalence Table
| Device | Subject Device | Predicate Device 01 | Predicate Device 02 |
|----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | XIAMEN SENYANG CO., LTD. | NormaTec Industries, LP, | Rapid Reboot Recovery Products, LLC |
| Model Name | Pt 1003 | Pulse 2.0, Pulse Pro 2.0 | Not Publicly Available |
| Classification | Class II Device, IRP (21 CFR890.5650) | Class II Device, IRP (21 CFR890.5650) | Class II Device, IRP (21 CFR890.5650) |
| Prescriptive | No. OTC | No. OTC | No. OTC |
| Indications for<br>use | The device is air pressure massagers intended to<br>temporarily relieve minor muscle aches and/or<br>pains, and to temporarily increase circulation to<br>the treated areas. | The device is air pressure massagers intended to<br>temporarily relieve minor muscle aches and/or<br>pains, and to temporarily increase circulation to<br>the treated areas. | The Rapid Reboot Compression Therapy System<br>is indicated for the temporary relief of minor<br>muscle aches and pains and for temporary<br>increase in circulation to the treated areas in<br>people who are in good health. The Rapid<br>Reboot Compression Therapy System simulates<br>kneading and stroking of tissues by using an<br>inflatable garment. |
| Intended use<br>Environment | Clinics, hospital, athlete training, and<br>home environments | Clinics, hospital, athlete training, and<br>home environments | Clinics, hospital, athlete training, and home<br>environments |
| Standard | AAMI ANSI ES<br>60601-1:2005/(R)2012 and A1:2012<br>IEC 60601-1-11.<br>IEC 60601-1-2<br>ISO10993-5& ISO10993-10 | AAMI ANSI ES<br>60601-1:2005/(R)2012 and A1:2012<br>IEC 60601-1-11.<br>IEC 60601-1-2<br>ISO10993-5& ISO10993-10 | IEC 60601-1:2014; IEC<br>60601-1-2:2014; EN ISO<br>10993-5:2009 & EN ISO<br>10993-10:2010 |
| Software<br>Micro-processor<br>Control | Microprocessor | Microprocessor | Microprocessor |
| Mode of<br>Compression | Sequential Gradient, Peristaltic and<br>Pulsing | Sequential Gradient, Peristaltic and<br>Pulsing | Sequential Gradient, Peristaltic and<br>Pulsing |
| Power Source | 110 V, 60Hz | 15 VDC via an IEC 60601-1<br>compliant power supply (100- 240 VAC | 110 V, 60Hz |
| | | | |
| Therapy Time | 0-30Min | input) Integrated rechargeable battery<br>Stays on until the user turns it off or can be set<br>up to turn off in a range of 10 minutes to<br>continuous. | User determines therapy time. Choose from 10,<br>20, or 30 minute session time, with option to add<br>additional 10 minutes to any therapy time. |
| Max Pressure<br>Min Pressure | 20-200mm Hg | 0-110mm Hg | 0-200 mmHg |
| Number of<br>Chambers | 4 Chambers for each unit | 5 or less | 4 Chambers for each unit |
| Compression<br>Applicator<br>Garments<br>Sleeve Material | Thermoplastic Urethane | 200 denier nylon with a polyurethane<br>laminate/extrusion | Nylon with a Polyurethane<br>laminate |
| Technology | Compressor and valve system that<br>sequentially inflates cells of appliance. | Compressor and valve system that<br>sequentially inflates cells of appliance. | Compressor and valve system which<br>sequentially inflates cells of appliance |
| Cycle time | 70s | Not Publicly Available | 1 min 20s |
| Patient Contact | Non-conductive appliances | Non-conductive appliances | Non-conductive appliances |
| Size | Size 285*175*130mm | Size: 4.4" x 3.8" x 8.1 " | Size: 10" x 6.5" x 5" |
| Preprogrammed<br>modes | Model Adjustable: The number of zones that are<br>enabled in the attachment can be changed<br>between one and five zones. | Model Adjustable: The number of zones that are<br>enabled in the attachment can be changed<br>between one and five zones. | 2 modes: "A" mode inflates and deflates<br>chambers from bottom up (distal to proximal<br>chambers), one at a time. "B" mode also inflates<br>chambers from bottom up, but maintains<br>pressure in lower chambers as works its way to<br>top. Then all chambers release pressure at same<br>time once all chambers have sequentially<br>inflated |
{5}------------------------------------------------
{6}------------------------------------------------
#### Determination of Substantial Equivalence I.
The following factors of Pressure Therapy System PT1003 are substantially equivalent to currently marketed and cleared devices (K183169).
- A Indications for Use: The Pressure Therapy System PT1003 indications for use are identical to the predicate, NormaTec Pulse 2.0, Pulse Pro 2.0, 510(k) K183169.
- A Prescriptive: The Pressure Therapy System PT1003, and the predicate (K183169) are OTC devices.
- > Performance and Specifications: The Pressure Therapy System PT1003 have equivalent specifications of performance when compared to the predicate, NormaTec Pulse 2.0, Pulse Pro 2.0, 510(k) K183169.
- > Compliance with Standards: Both subject device and predicate device comply with the same standards expect the special standard related to the blue tooth function.
- > Intended Use Environment: Clinics, hospital, athlete training, and home environments, which are identical to the predicate.
- > Materials for patient contact are Non-conductive appliances.
The following factors of Pressure Therapy System PT1003 are different from currently marketed and cleared devices (K183169).
- A Pressure range of the predicate device is 0-110mm Hg while current device is 20-200mm Hg. To ensure the pressure is safe, we identified another predicate device. K182668, which is also for OTC use and the pressure scope is 0 to 200mmHg.
- A Design, Technology, and Principle of Operation: The proposed device has not got Bluetooth capability but this will not bring any new risk to the subjected device.
### J. Performance characteristic
The Pt1003 pressure therapy system has been tested and met the following standards: AAMI ANSI ES 60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
IEC 60601-1-2 Edition 3: 2007-03. Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
IEC 60601-1-11 Edition 2.0 2015-01, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
ISO10993-5, Biological evaluations of medical devices -- Part 5: Tests for In Vitro cytotoxicity
ISO 10993-10, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
Function test
- Minimum and maximum air pressure test
- Test the maximum electric current value
- Test the maximum airflow value
- Test the noise level
- Test the modes A and B.
The test was done on PT1003 following the test procedure defined. When we compare the test data with the predicate device, we can found the test result is almost same.
{7}------------------------------------------------
#### K. Conclusion
The Pt1003 Pressure Therapy System has substantially equivalent intended use as the cleared NormaTec Pulse 2.0, Pulse Pro 2.0, (K183169) and has substantially equivalent technological and performance characteristics. The pressure scope is safe by compare with the predicate device Rapid Reboot Recovery Products, LLC, K182668. After analyzing laboratory testing to applicable standards, it is concluded that Pt1003 Pressure Therapy System is as safe and effective as the predicate device, has few technological differences, but there are no new indications for use and does not raise any new safety and/or effectiveness concerns. Consequently, it is substantially equivalent to the predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.