The Biomove 5000 is indicated for: - 1. Stroke Rehabilitation by Muscle Re-education - 2. Prevention of retardation of disuse atrophy - 3. Increasing local blood circulation - 4. Muscle re-education - 5. Maintaining or increasing range of motion
Device Story
Biomove 5000 is a powered EMG-triggered neuromuscular electrical stimulator; used for rehabilitation of paralyzed muscles, particularly post-stroke. Device inputs EMG signals from patient muscles; transforms signals via electrical stimulation to induce muscle contraction; outputs electrical pulses to re-educate muscles. Used in clinical or rehabilitation settings; operated by healthcare professionals. Provides feedback-driven stimulation to assist in muscle re-education and atrophy prevention; benefits patients by improving motor function and circulation.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and performance standards compliance.
Technological Characteristics
Powered EMG-triggered neuromuscular electrical stimulator. Includes patient cable and re-usable electrodes (Biotrodes). Complies with 21 CFR 898 for electrode lead wires and patient cables, IEC 60601-1, and IEC 60601-2-10. Class II device.
Indications for Use
Indicated for stroke rehabilitation, muscle re-education, prevention of disuse atrophy, increasing local blood circulation, and maintaining or increasing range of motion in patients with paralyzed muscles.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
K972997 — AM800 AUTOMOVE EMG TRIGGERED ELECTRICAL MUSCLE STIMULATION · Dan Med, Inc. · Nov 10, 1997
K023036 — CARE ETS · Care Rehab and Orthopaedic · Jun 19, 2003
K032955 — AUTOMOVE, MODEL AM800 · Danmeter A/S · Dec 15, 2003
K012885 — NEUROMOVE NM900 · Dan Med, Inc. · Nov 8, 2001
Submission Summary (Full Text)
{0}------------------------------------------------
K0860787
APR 1 7 2008
## 510(K) SUMMARY
## Biomove 5000 System
510(k) Number K_
A. Applicant's Name: Curatronic Ltd.
4 Atsmon Street. Hashmonaim 73127, Israel Tel.: +972-8-9761441 Fax .: +972-8-9762020 e-mail: philipson@curatronic.com
Ofer Hornick M.D. B. Contact Person: A. Stein - Regulatory Affairs Consulting 20 Ha'Ta'as St. Kfar Saba 44425 Israel Tel. + 972-9-7670002 Fax. +972-9-7668534 February 2008 C. Date Prepared:
Biomove 5000 System D. Trade Name:
Name: stimulator, muscle, powered E. Classification: Product Code: IPF Regulation No: 890.5850 Class': II Panel: Physical Medicine
{1}------------------------------------------------
- F. Predicate Devices: The Biomove 5000 is substantially equivalent to the Biomove 3000 device manufactured by the same company and cleared under number K042650 in terms of intended use, indications for use, technological characteristics, performance and user interface.
Both predicate devices are Class II medical devices.
A discussion of substantial equivalence is provided in this submission.
- G. Device Description: The Biomove 5000 is a powered EMG triggered neuromuscular electrical stimulator device used as a training system for rehabilitation of paralyzed muscles, mainly after stroke.
The following accessories are supplied with the system: Patient Cable, re-usable electrodes (Biotrodes).
H. Intended Use / Indication for Use: The Biomove 5000 is indicated for:
- 1. Stroke Rehabilitation by Muscle Re-education
- 2. Prevention of retardation of disuse atrophy
- 3. Increasing local blood circulation
- 4. Muscle re-education
- 5. Maintaining or increasing range of motion
- I. Performance Standards: The Biomove 5000 complies with U.S. Federal Performance Standard as set forth in 21 CFR 898 for electrode lead wires and Patient Cables.
The device complies with the following recognized standards:
{2}------------------------------------------------
- IEC 60601-1(1988), including amendments #1(1991), . #2(1995)
- IEC 60601-2-10 (1987) �
The device complies with the Guidance Document for Powered Muscle Stimulator 510(k)s; Final (1999).
- J. Technological Characteristics: The technological characteristics, e.g., overall design, materials, mechanism of action, mode of operation, performance characteristics, etc., and the intended use of the Biomove 5000 device are substantially equivalent to the predicate device cited above.
- K. Substantial Equivalence: The Biomove 5000 System is substantially equivalent to its predicate device as cited above and raises no new safety and/or effectiveness issues.
100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circle around the eagle. The eagle is black, and the text is also black.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 1 7 2008
Curatronic, Limited % Ms. Ahava Stein Regulatory Affairs Consulting 20 Hata'as Street Kfar Saba 44425 Israel
Re: K080787
> Trade/Device Name: Biomove 5000 system Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator. Regulatory Class: Class II Product Code: IPF Dated: March 16, 2008 Received: March 20, 2008
Dear Ms. Stein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
## Page 2 - Ms. Ahava Stein
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours.
Mark N Millman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known): _ K 080787
Device Name: Biomove 5000 System.
Indications for Use:
The Biomove 5000 is indicated for:
- 1. Stroke Rehabilitation by Muscle Re-education
- 2. Prevention of retardation of disuse atrophy
- 3. Increasing local blood circulation
- 4. Muscle re-education
- 5. Maintaining or increasing range of motion
Prescription Use V AND/OR (Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
Over-The-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
> Concurrence of CDRH, Office of Device Evaluation (ODE) AP 111
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
**510(k) Number** K090787
<sup>136</sup>
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.