ANDELA DYNAMIC COOLING DEVICE
Device Facts
| Record ID | K951033 |
|---|---|
| Device Name | ANDELA DYNAMIC COOLING DEVICE |
| Applicant | Candela Laser Corp. |
| Product Code | IMD · Physical Medicine |
| Decision Date | Aug 15, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 890.5710 |
| Device Class | Class 1 |
| Attributes | Therapeutic |
Indications for Use
The intended use of the Candela Dynamic Cooling Device is to minimize thermal injury to non-vascular skin structures during laser therapy of benign cutaneous vascular lesions, and to reduce pain associated with laser treatment.
Device Story
Device consists of CFC R-12 refrigerant source, electronically controlled solenoid delivery valve, and electronic timing circuitry. Integrated with SPTL laser triggerswitch system; delivers pulsed spray of CFC R-12 immediately preceding laser pulse. Evaporative cooling minimizes thermal injury to non-vascular skin structures and reduces pain. Used in clinical dermatology settings by physicians. Output is controlled refrigerant spray; assists clinical decision-making by enabling safer laser application; benefits patient via reduced pain and thermal damage.
Clinical Evidence
Bench testing only. Performance assessed via dimensional testing, functional properties, intended use verification, ease of use, and durability testing. No clinical data provided.
Technological Characteristics
Components: CFC R-12 refrigerant, solenoid delivery valve, electronic timing circuitry. Connectivity: In-line integration with SPTL laser triggerswitch. Principle: Evaporative cooling via pulsed refrigerant spray.
Indications for Use
Indicated for patients undergoing laser therapy for benign cutaneous vascular lesions to minimize thermal injury to non-vascular skin structures and reduce treatment-associated pain.
Regulatory Classification
Identification
A hot or cold disposable pack is a device intended for medical purposes that consists of a sealed plastic bag incorporating chemicals that, upon activation, provides hot or cold therapy for body surfaces.
Predicate Devices
- Cool Laser Optics Device
- Cold Packs
Related Devices
- K001589 — CANDELA DYNAMIC COOLING DEVICE · Candela Corp. · Aug 18, 2000
- K014253 — DERMACHILLER 4 · Paradigm Trex, LLC · Mar 26, 2002
- K060395 — CRYO, MODEL V6.0 · Zimmer Elektromedizin GmbH · Feb 24, 2006
- K090618 — CYNOSURE SMARTCOOL FAMILY COLD AIR DEVICE: SMARTCOOL, SMARTCOOL 6 AND CRYO C · Cynosure, Inc. · Apr 14, 2009