GAUNYING INFRARED LAMP MODEL IL50

K101693 · Foshan Gaunying Electronics Co., Ltd. · ILY · Jan 5, 2011 · Physical Medicine

Device Facts

Record IDK101693
Device NameGAUNYING INFRARED LAMP MODEL IL50
ApplicantFoshan Gaunying Electronics Co., Ltd.
Product CodeILY · Physical Medicine
Decision DateJan 5, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Gaunying Infrared Lamp ( Model #: IL50) is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness, or muscle spasm; the temporary increase in local blood circulation.

Device Story

Gaunying Infrared Lamp (IL50) provides topical heating via infrared radiation; utilizes glass ceramic plate with illuminants to emit 600-2000 nm (near/short-wavelength) infrared energy. Device intended for OTC use; operated by lay users. User positions lamp at 30-40 cm distance from target area to elevate tissue temperature. Visible light emission serves as safety indicator to reduce burn risk compared to non-visible far-infrared devices. Output provides therapeutic heat for pain/stiffness relief and increased local circulation. Bench testing confirms temperature maintenance above 40°C without exceeding 45°C at skin surface. Usability testing validated that lay users can safely operate device following provided instructions.

Clinical Evidence

No clinical data. Bench testing performed comparing heat output of subject device to predicate at 30, 35, and 40 cm distances over 55 minutes. Subject device maintained temperatures >40°C and <45°C, demonstrating performance equivalence to predicate. Usability testing conducted with 10 participants (5 male, 5 female) confirmed lay users could safely operate device and understand instructions.

Technological Characteristics

Glass ceramic plate infrared emitter; 600-2000 nm wavelength range. Complies with IEC 60601-1 (safety) and IEC 60601-1-2 (EMC). Bench-tested for thermal output at 30-40 cm distances. Standalone device.

Indications for Use

Indicated for individuals requiring topical heating for temporary relief of minor muscle and joint pain, stiffness, muscle spasms, or to increase local blood circulation.

Regulatory Classification

Identification

An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.

Special Controls

*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K101693 # 510(K) SUMMARY JAN - 5 2011 This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 807.92 #### Submitter's Information 1.0 ### Establishment Registration Name: Foshan Gaunying Electronics Co., Ltd. 4F, #4 Industry Country, Cheng Nan Park of Foshan Hi-tech Industrial Development Zone Foshan, Guangdong, China 528000 # Contact Person (US Agent/Official Correspondent) of the Applicant: Mr. Guenter Ginsberg (President) Media Trade Corporation 11820 Red Hibiscus Drive Bonita Springs, FL 34135, USA Tel: 239 948 2001 Fax: 239 948 2002 Email: g.ginsberg@gmx.net ### . Device Information Type of 510(k) submission: Traditional Device Common Name: Infrared Lamp Infrared Lamp Trade Name: IL50 Model: lamp, infrared, therapeutic heating Classification name: Physical Medicine Review Panel: ILY Product Code: Class II Regulation Class: 21CFR 890.5500 Regulation Number: {1}------------------------------------------------ K101693 #### 2.0 Predicate Device Information | Sponsor: | Chongqing Xinfeng Medical Instrument Co., Ltd. | |----------------|--------------------------------------------------------------------------------| | Device: | Xinfeng Heating Lamp, including 3 different models CQ-<br>27, CQ-36 and CQ-55A | | 510(K) Number: | K043558 | #### 3.0 Device description Gaunying Infrared Lamp (Model#: IL50) has a high-quality glass ceramic plate (ceramic infrared). Glass ceramic plates are also used on hobs and when used in conjunction with illuminants guarantee intensive and safe infrared radiation. #### 4.0 Intended Use The Gaunying Infrared Lamp (Model #: IL50) is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness, or muscle spasm; the temporary increase in local blood circulation. #### Performance Summary 5.0 Testing of the Infrared Lamp (Model#: IL50) includes functional performance testing and electrical safety testing. The device is manufactured to comply with the following international standards: - IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements . for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995. (General) - IEC 60601-1-2. Medical Electrical Equipment Part 1-2: General . Requirements for Safety - Collateral standard: Electromagnetic Compatibility - Requirements and Tests (Edition 2:2001 with Amendment 1:2004; Edition 2.1 (Edition 2:2001 consolidated with Amendment 1:2004)). (General) - UL 60601-1, first edition, 2003, deviations to IEC 60601-1 . - Performance Bench Tests: . Two (2) devices were tested for heat output at distances of 30, 35 and 40cm from the infrared source. Temperatures were recorded every one (1) minute, up to 15 minutes, using a digital temperature sensor. The test results concluded that the Infrared Lamp (Model#: IL50) was able to maintain temperatures above 40° C (the desired effective temperature) at least 10 minutes of the 15 minutes cycle time. The tests also showed that the temperature at the sensor (skin) never exceeded a temperature of 45° C. Both devices used for this test tracked each near 100%. {2}------------------------------------------------ K101693 # Usability Test: A group of 10 people were selected, 5 females and 5 men with different ages and educational backgrounds. They were given the Operating Instructions (OI) to read and study before using the subject device. The test group had no problem operating the device, but had questions and clarifications about some statements in the OI. The OI was changed accordingly. A new study group of 10 people were tested and observed. Nobody had any difficulties understanding the OI and operating the device. It was concluded that the Operating Instructions are clear and have enough information and warnings that lay people should not have any difficulties understanding the instructions and operating the device safely and effectively. #### 6.0 Comparison to predicate device and conclusion Compared with predicate device Xinfeng CQ-27, CQ-36 and CQ-55A Heating Lamps made by Chongqing Xinfeng Medical Instrument Co., Ltd. (K043558), the device Infrared Lamp (Model#: IL50) has same intended use with the predicate device. Although, there is a little different technological characteristics; The infrared wavelength used in subject device and predicate device is deferent. The subject device uses 600 nm - 2000 nm infrared, which belongs to Near-infrared (0.75-1.4 µm) and Short-wavelength infrared (1.4-3 um). The predicate device use 5 to 25 microns, which belongs to Longwavelength infrared (8-15 µm) and far infrared (15-1,000 µm). The human visible light is in wavelength 400-700 nm. So the subject device heating can be seen when it is working. The predicate device heating is not visible when it is working. That means the possibility of burns risk for subject device is less than the predicate device. ## Comparison Temperature Test: A bench test was set up to compare the heat output of the predicate device with the subject device. Temperatures were recorded for both devices at distances of 30, 35, and 40 cm and at 1 minute intervals up to 55 minutes. The results show that the Gaunying Infrared Lamp (the subject) tracks the temperature readings of the Xingfing Heating Lamp (the predicate device) and is therefore equivalent in performance to the predicate device. Whatever Near-infrared, Short-wavelength infrared, Long-wavelength infrared and Far infrared, they all can heat and therapy disease. The different wavelength will not affect the effectiveness of device. The testing report shows that Infrared {3}------------------------------------------------ Gaunying Infrared Lamp IL50 510(K) Files: Section 5 K101693 Lamp (Model#: IL50) is as safe and effective as the predicate device. So, the difference does not raise new questions of safety and effectiveness. The Infrared Lamp (Model#: IL50), is substantially equivalent to the predicate device. #### 7.0 Submission date: June 16, 2010 {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird in flight, composed of three curved lines that suggest wings and a body. The bird is oriented diagonally, moving from the upper right to the lower left of the image. Encircling the bird is text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA". Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Foshan Gaunying Electronics Co., Ltd. % Media Trade Corporation Mr. Guenter Ginsberg 11820 Red Hibiscus Drive Bonita Springs, Florida 34135 JAN - 5 2011 Re: K101693 Trade/Device Name: Gaunying Infrared Lamp, Model IL50 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Class II Product Code: ILY Dated: December 10, 2010 Received: December 10, 2010 Dear Mr. Ginsberg: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {5}------------------------------------------------ Page 2 – Mr. Guenter Ginsberg or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Sincerely yours, Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ pg 1 of 1 ### Statement of Indications for Use 510(k) Number (if known): K101693 Device Name: Gaunying Infrared Lamp Model: IL50 JAN - 5 2011 Indications For Use: The Gaunying Infrared Lamp ( Model #: IL50) is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness, or muscle spasm; the temporary increase in local blood circulation. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use X (21 CFR 801 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Neil P. Olken Ew mkn Page 1 of of 1 lo Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K101693 Section 4, Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...