K062635 · Clarimedex, Inc. · ILY · Dec 19, 2006 · Physical Medicine
Device Facts
Record ID
K062635
Device Name
MODEL 2200
Applicant
Clarimedex, Inc.
Product Code
ILY · Physical Medicine
Decision Date
Dec 19, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The CLARIMEDIX device is indicated for use to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting relaxation of muscle tissue; and to temporarily increase blood circulation where applied.
Device Story
Model 2200 System is a self-contained infrared lamp for topical heating; used in medical facilities by physical medicine healthcare professionals. Device features four treatment zones (popliteal, calf, foot, heel) utilizing multidiodes to emit infrared energy. System operates via 24V power supply; employs modulating waveform to elevate tissue temperature to <95°F. Healthcare providers apply device to patient; output provides temporary pain relief and increased local blood circulation. Device remains cool to touch during operation; includes firmware control for safety and performance management.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics, intended use, and compliance with electrical safety and EMC standards (IEC 60601-1, IEC 60601-1-2).
Technological Characteristics
Infrared lamp (21 CFR 890.5500); multidiodes; 24V power supply; modulating waveform; four treatment zones; dimensions 25.1 in x 10.0 in; compliant with IEC 60601-1 and IEC 60601-1-2; firmware-controlled; recommended use with biocompatible drape.
Indications for Use
Indicated for patients requiring temporary relief of minor muscle/joint pain, arthritis, muscle spasm, stiffness, and increased local blood circulation via topical infrared heating.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
K982546 — PHASER-2000 · Diomedics, Inc. · Sep 22, 1998
K033923 — TH0R DDII IR LAMP SYSTEM · Thor Intl. · Mar 17, 2004
K093204 — INNOVATIVE MED MULTI WAVE LIGHT THERAPY SYSTEM · Innovative Med, Inc. · Aug 23, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
K062635
:
### 510(k) Summary Prepared August 5, 2006
| Submitted by: | CLARIMEDIX, Inc.<br>1035 Pearl Street, Suite 400<br>Boulder, Colorado 80302 | | |
|-----------------|-----------------------------------------------------------------------------|-------------------------------------------------------------|--------------|
| Contact Person: | John Dunning<br>President<br>Telephone:<br>FAX:<br>e-mail: | (303) 905 6163<br>(303) 625 1030<br>jdunning@clarimedix.com | DEC 1 9 2006 |
| Product Name: | Model 2200 System | | |
| Common Name: | Infrared lamp | | |
| Classification: | ILY; Class II; CFR 21 890.5500 | | |
Predicate Devices:
| Device Name | Manufacturer | K Number |
|---------------------------------------------------------|----------------------------------------|----------|
| Warp 10 | Quantum Devices | K032229 |
| Spectropad<br>(a.k.a.<br>Restorative<br>Products, Inc.) | SMI (A.k.a. Anodyne<br>Therapy System) | K931261 |
{1}------------------------------------------------
#### Description of Device:
The device is indicated for use for the treatment of chronic pain by emitting energy in the IR spectrum for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffiess; promoting relaxation of muscle tissue; and to temporarily increase local blood circulation where applied
The system is a self-contained unit with four treatment areas.
- . Popliteal treatment zone
- Calf treatment zone .
- Foot treatment zone .
- . Heel treatment zone
#### Intended Use:
The device is indicated for use for the treatment of chronic pain by emitting energy in the IR spectrum for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting relaxation of muscle tissue; and to temporarily increase local blood circulation where applied
#### Comparison with Predicate Devices:
The submission device and the predicate devices have substantially equivalent intended use and technological specifications as shown in the tables below. The device is compliant with the standards for electrical safety and EMC, IEC 60601-1 IEC 60601-1-2 respectively.
{2}------------------------------------------------
| | Predicate Devices | | Subject Device |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Quantum Warp 10<br>K032229 | SMI Spectropad<br>K931261<br>(a.k.a. Anodyne Therapy<br>System) | CLARIMEDIX Model 2200 |
| FDA<br>classification | 21 CFR 890.5500 Infrared<br>Lamp (ILY)<br>a) Identification. An infrared<br>lamp is a device intended for<br>medical purposes that emits<br>energy at infrared frequencies<br>(approximately 700 nanometers<br>to 50,000 nanometers) to<br>provide topical heating<br>b) classification Class II | 21 CFR 890.5500 Infrared<br>Lamp (ILY)<br>a) Identification. An infrared<br>lamp is a device intended for<br>medical purposes that emits<br>energy at infrared frequencies<br>(approximately 700<br>nanometers to 50,000<br>nanometers) to provide topical<br>heating<br>b) classification Class II | 21 CFR 890.5500 Infrared<br>Lamp (ILY)<br>a) The CLARIMEDIX device<br>is indicated for use to emit<br>energy in the infrared<br>spectrum to provide topical<br>heating for the purpose of<br>elevating tissue temperature;<br>for the temporary relief of<br>minor muscle and joint pain,<br>arthritis and muscle spasm;<br>relieving stiffness; promoting<br>relaxation of muscle tissue;<br>and to temporarily increase<br>blood circulation where<br>applied.<br>b) classification Class II |
| Intended Use | Indicated for use for the<br>treatment of chronic pain by<br>emitting energy in the Near-IR<br>spectrum for the temporary<br>relief of minor muscle and joint<br>pain, arthritis and muscle<br>spasm; relieving stiffness;<br>promoting relaxation of muscle<br>tissue; and to temporarily<br>increase local blood circulation<br>where applied | Intended for relief of minor<br>muscle and joint pain and<br>improvement of superficial<br>circulation 1 | indicated for use for the<br>treatment of chronic pain by<br>emitting energy in the IR<br>spectrum for the temporary<br>relief of minor muscle and<br>joint pain, arthritis and muscle<br>spasm; relieving stiffness;<br>promoting relaxation of muscle<br>tissue; and to temporarily<br>increase local blood circulation<br>where applied |
| Intended<br>Users | Physical medicine health care<br>professionals | Physical medicine health care<br>professionals | Physical medicine health care<br>professionals |
| Intended Site<br>of Use | Medical health care facilities | Medical health care facilities | Medical health care facilities |
# Table 1 Comparison of Indications for Use
{3}------------------------------------------------
| | Predicate Devices | Predicate Devices | Subject Device |
|------------------------------------|-----------------------------------------------|-----------------------------------------------------------------|-----------------------------------------|
| | Quantum Warp 10<br>K032229 | SMI Spectropad<br>K931261<br>(a.k.a. Anodyne Therapy<br>System) | CLARIMEDIX Model 2200 |
| Method of<br>administration | Topical | Topical | Topical |
| Treatment areas | 10cm2 | Leg, foot, arm, neck | Lower leg, foot, adjustable<br>location |
| Number of<br>zones | Not specified in product<br>labeling | 4-8 therapy pads | Four zones |
| Temperature at<br>the skin surface | Not specified in labeling | Not specified in labeling | Less than 95° F |
| Energy source | Multidiodes dispersed over<br>treatment zones | 60 Infrared diodes per array | Multidiodes |
| Waveform | Constant | Not specified in labeling | Modulating |
| Power supply | 8 each 1.5 AA batteries<br>provides 50 doses | Not specified in labeling | 24V |
| Safety features | Remains cool to the touch | Not specified in labeling | Remains cool to touch |
| Biocompatibility | Not specified in labeling | Not specified in labeling; | Biocompatible drape<br>recommended |
| Configuration | Handheld | 4-8 pads applied | 4 zones applied |
| Length and<br>width | 5.55in x 2.65 in | Not specified in labeling | 25.1 in. x 10.0 in. |
| Portable | Yes | Yes | Yes |
| Electrical safety | Not specified in labeling | Not specified in labeling | Compliant with IEC 60601-1 |
| Firmware<br>control | Yes | Yes | Yes |
·
:
## Table 2 Comparison of Technological Characteristics
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
CLARIMEDEX, Inc. % Regulatory Technology Services, LLC Mr. Mark Job 1394 25TH Street, Northwest Buffalo, Minnesota 55313
DEC 1 9 2006
Re: K062635
Trade/Device Name: Model 2200 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: II Product Code: ILY Dated: December 6, 2006 Received: December 7, 2006
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Mr. Mark Job
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
For Pablo
Mark N. Melkerson
DEP D.R.SECTOR
Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known):
Device Name: Model 2200
Indications For Use:
The CLARIMEDIX device is indicated for use to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting relaxation of muscle tissue; and to temporarily increase blood circulation where applied.
X Prescription Use (Per 21CFR 801)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence Of CDRH, Office Of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
(K) Number 2062635
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.