K043055 · Escada International, Inc. · ILY · Jul 25, 2005 · Physical Medicine
Device Facts
Record ID
K043055
Device Name
TERRAQUANT MQ2000 LASER THERAPY DEVICE
Applicant
Escada International, Inc.
Product Code
ILY · Physical Medicine
Decision Date
Jul 25, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TerraQuant MQ2000 Laser Therapy Device is a non-invasive infrared lamp intended to provide topical heating. The Terraquant MQ2000 in indicated for temporary relief of minor muscle and joint pain, arthritis and muscle spasm, relieving stiffness, promoting relaxation of muscle tissue, and to temporarily increase local blood circulation where heat is indicated.
Device Story
The TerraQuant MQ2000 is a non-invasive, hand-held infrared lamp system for pain therapy. It comprises a desktop control unit, a hand-held emitter, and a 110V power adapter. The device emits laser, infrared, and visible red light (wavelengths 600-700 nm and 860-960 nm) at 60-90 mW power. Operated by the user, it delivers low-level light radiation to specific anatomic areas to provide topical heating. This heat application aims to relieve minor muscle and joint pain, arthritis, and muscle spasms, while promoting muscle relaxation and increasing local blood circulation. It is intended for over-the-counter use.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and intended use compared to the predicate device.
Technological Characteristics
Hand-held infrared lamp system; utilizes laser, infrared, and visible red light-emitting diodes. Wavelengths: 600-700 nm and 860-960 nm. Power output: 60-90 mW. Components: desktop control unit, hand-held emitter, 110V power adapter. Operates as a non-invasive heating device per 21 CFR 890.5500.
Indications for Use
Indicated for temporary relief of minor muscle and joint pain, arthritis, muscle spasm, stiffness, and to promote muscle relaxation and increase local blood circulation via topical heating.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
K061614 — TERRAQUANT MQ2000 V. 5 AND HANDYRX MQ2007 LASERS · Escada International, Inc. · Oct 6, 2006
K032379 — QUASAR, MODEL MD21 · Quantum Skincare, Inc. · Nov 20, 2003
K080102 — TQ SOLO AND SOLO PRO AND LS50 ACCESSORY AND MR4 MULTI RADIANCE THERAPY SYSTEM · Escada International, Inc. · Aug 20, 2008
K101716 — HVR PAIN RELIEF DEVICE · H.V.R., Inc. · Jan 11, 2011
K080689 — SPECTRA THERAPY SYSTEM A1000 LASER DEVICE · Premier Dynamics, Inc. · Jul 9, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
1 of 2
# JUL 2 5 2005 510(k) Summary for TerraQuant MQ2000 Laser Therapy Device
#### 1. SPONSOR
Escada International, Inc. 27100 Richmond Road Solon, Ohio 44139
Contact Person: Max Kanarsky Telephone: 440-542 0762
Date Prepared: March 10, 2005
#### 2. DEVICE NAME
Proprietary Name: TerraQuant MQ2000 Laser Therapy Device Common/Usual Name: heating lamp Classification Name: Infrared lamp
#### 3. PREDICATE DEVICES
Quantum WARP 10 Light Delivery System -- K032229
#### 4. DEVICE DESCRIPTION
The TerraQuant MQ2000 is a non-invasive pain therapy system, which utilizes heating lamps consisting of laser, infrared, and visible red light emitting diodes (LED). It combines the clinically accepted therapeutic treatment of numerous predicate light therapy systems currently in commercial distribution into one complete, compact system.
The TerraQuant MQ2000 system consists of a desktop control unit, a hand-held emitter from which the laser and other radiances are released, and a 110 V power adaptor. The laser and light therapy releases radiation with wavelengths that fall within the range as defined in 21 CFR 890.5500 for an infrared lamp.
. :
{1}------------------------------------------------
KO4/3055
2 of 2
#### 5. INTENDED USE
The TerraQuant MQ2000 Laser Therapy Device is a non-invasive infrared lamp intended to provide topical heating. The Terraquant MQ2000 in indicated for temporary relief of minor muscle and joint pain, arthritis and muscle spasm, relieving stiffness, promoting relaxation of muscle tissue, and to temporarily increase local blood circulation where heat is indicated.
#### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE 6.
The TerraQuant MQ2000 device and the predicate device are substantially equivalent in intended use in that they are both heating lamps intended for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm, relieving stiffness, promoting relaxation of muscle tissue, and to temporarily increase local blood circulation where heat is indicated. The proposed device and predicate devices are intended to provide relief to various areas of the body depending on the site of pain.
The proposed TerraQuant MQ2000 Laser Therapy System and the predicate devices are also substantially equivalent in technological characteristics in that they all consist of hand-held, heating infrared lamps that deliver low level laser to various anatomic areas.
{2}------------------------------------------------
### Name of Manufacturer: Escada International, Inc. Laser Model Name and Number: Terraquant MQ2000
Laser Type: (Circle all that apply)
Alexandrite. Argon, CO2, Copper-Vapor, Diode, Dye, Nd:YAG, Erbium, Hol: YAG, Krypton, Ruby, KTP/532, Excimer, HENE, Accessory, Other
### Indications in this application:
The TerraQuant MQ2000 is a non-invasive infrared lamp intended to provide topical heating. The TerraQuant MQ2000 is indicated for temporary relief of minor muscle and joint pain, arthritis and muscle spasm, relieving stiffness, promoting relaxation of muscle tissue, and to temporarily increase local blood circulation where heat is indicated.
### FDA Document Control Number: K043055
FDA Product Code: 79GEX
Reviewer Computer Initials: CYH
Date of Clearance Letter: 07/18/05
Basis of Approval: (Circle all that apply)
Predicate Device (PD), Clinical Data (CD), Animal Data (AD), Specifications (SPECS), Bench Test Data (BTD), Historical Information (HI), Other
Description of Laser: Operation Modes: (Circle all that apply) CW, Pulsed, Q-Switched, Mode Locked, Contact, Free Beam, Other
Wavelength in Nanometers: 900, 860-960, 600-700
Power/Energy Range (Watts/Joules): 60-90 mW
Pulse Width:
Repitition Rate:
Delivery System: hand held emitter Comments:
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized caduceus, which is a staff with two snakes entwined around it, often associated with medicine and healthcare. The seal is in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 25 2005
Escada International, Inc. c/o Ms. Mary McNamara-Cullinane, RAC Medical Device Consultants, Inc. . . 49 Plain Street North Attleboro, Massachusetts 02760
\$
Re: K043055
Trade/Device Name: TerraQuant MQ2000 Laser Therapy Device Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: II Product Code: ILY Dated: July 5, 2005 Received: July 6, 2005
Dear Ms. McNamara-Cullinane:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
## Page 2- Mary McNamara-Cullinane, RAC
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter with and by John finding of substantial equivalence of your device to a legally predicated predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
E.A.
Miriam C. Provost, Ph.D.
Miriam C. Provost, Ph.D Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
.
Enclosure
{5}------------------------------------------------
# Indications for Use
KO43055
510(k) Number (if known):
Device Name: TerraQuant MQ2000 Laser Therapy Device
.. .
Indications for Use:
The TerraQuant MQ2000 is a non-invasive infrared lamp intended to provide topical heating. The Terraquant MQ2000 is indicated for temporary relief of minor muscle and joint pain, arthritis and muscle spasm, relieving stiffness, promoting relaxation of muscle tissue, and to temporarily increase local blood circulation where heat is indicated.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use X (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Eula A
Jon Sign-Off) vision of General, Restorative rd Neurological Devices
March 10, 2005
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.