← Product Code [PFW](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/PFW) · K233210

# sam CS Long Duration Ultrasound Device (K233210)

_ZetrOZ Systems, LLC · PFW · Oct 25, 2023 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/PFW/K233210

## Device Facts

- **Applicant:** ZetrOZ Systems, LLC
- **Product Code:** [PFW](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/PFW.md)
- **Decision Date:** Oct 25, 2023
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 890.5300
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The sam CS Long Duration Ultrasound Device is intended for home use to apply ultrasonic energy to generate deep heat within body tissues and to increase local circulation.

## Device Story

Device consists of rechargeable power controller, timer, and ultrasound-generating applicators; uses single-use disposable ultrasonic coupling patches. Operates at 3 MHz frequency; delivers 0.65 W (single applicator) or 1.3 W (dual applicator) continuous wave ultrasound. Intended for home use by patients. Device generates deep heat in tissues and increases local circulation over 1-hour treatment duration. Firmware manages timing and power button activation; no control function. Healthcare provider prescribes device; patient self-administers treatment. Benefits include therapeutic deep heat application for circulation improvement.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing and adherence to recognized consensus standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 10993).

## Technological Characteristics

Ultrasonic diathermy device; piezoelectric generation; 3 MHz frequency; 0.65W/1.3W output. Materials: Lead Zirconate-Titanate crystal, TPX lens, ABS plastic applicator. Dimensions: 3.4x3.4x1.4 cm controller. Power: 120/240 VAC or lithium battery. Non-sterile. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 10993. Firmware-controlled timing and power activation.

## Regulatory Identification

An ultrasonic diathermy for use in applying therapeutic deep heat for selected medical conditions is a device that applies to specific areas of the body ultrasonic energy at a frequency beyond 20 kilohertz and that is intended to generate deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies.

## Predicate Devices

- sam X1 Long Duration Ultrasound Device ([K211513](/device/K211513.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

October 25, 2023

ZetrOZ Systems, LLC Sabrina Lewis OAR Director 56 Quarry Road Trumbull, Connecticut 06611

Re: K233210

Trade/Device Name: sam CS Long Duration Ultrasound Device Regulation Number: 21 CFR 890.5300 Regulation Name: Ultrasonic Diathermy Regulatory Class: Class II Product Code: PFW Dated: September 28, 2023 Received: September 28, 2023

Dear Sabrina Lewis:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Lauren E. Woodard -S

for Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Rehabilitation Devices

{2}------------------------------------------------

OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known) K233210

Device Name sam CS Long Duration Ultrasound Device

Indications for Use (Describe)

The sam CS Long Duration Ultrasound Device is intended for home use to apply ultrasonic energy to generate deep heat within body tissues and to increase local circulation.

| Type of Use (Select one or both, as applicable)                                   |                                                                                  |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span>     ☑ Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>     ☐ Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(K) SUMMARY - K233210

| Device Trade Name: | sam CS Long Duration Ultrasound Device                                                |
|--------------------|---------------------------------------------------------------------------------------|
| Manufacturer:      | ZetrOZ Systems, LLC<br>56 Quarry Road<br>Trumbull, CT 06611                           |
| Contact:           | Ms. Sabrina Lewis<br>QAR Director<br>Phone: 888-202-9831<br>Email: sabrina@zetroz.com |
| Date Prepared:     | October 24, 2023                                                                      |
| Classification:    | 21 CFR 890.5300; Ultrasonic diathermy.                                                |
| Class:             | II                                                                                    |
| Product Code:      | PFW                                                                                   |
| Predicate Devices: | sam X1 Long Duration Ultrasound Device (K211513)                                      |

#### Indications for Use:

The sam CS Long Duration Ultrasound Device is intended for home use to apply ultrasonic energy to generate deep heat within body tissues and to increase local circulation.

### Device Description:

The sam CS Long Duration Ultrasound Device consists of:

- Rechargeable Power Controller and Timer ●
- Ultrasound Generating Applicators

Single use disposable accessories

- Ultrasonic Coupling Patch .
The Power Controller can be used to power the Applicator to generate ultrasonic energy at one frequency (3 MHz) and one power setting (0.65 W) per Applicator. In single Applicator mode, the patient can receive 0.65 W at 3 MHz. In dual Applicator mode, the patient can receive 1.3 W at 3 MHz. The Applicators are applied to the skin with onetime use Ultrasonic Coupling Patches.

The sam CS Long Duration Ultrasound Device contains firmware that controls the timing display and inputs from the power button to activate/deactivate the device. There is no control function of this firmware.

{5}------------------------------------------------

The system is intended for prescription home use to apply ultrasonic energy for a long duration (1 hour) to generate deep heat within body tissues and to increase local circulation.

| Table 1: Substantial Equivalence Summary                                                                                                                                                                              |                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                     |  |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|--|
| Information                                                                                                                                                                                                           | ZetrOZ sam CS Long<br>Duration Ultrasound<br>Device<br>(Subject)                                                                                                                                 | ZetrOZ sam X1 Long<br>Duration Ultrasound<br>Device<br>(K211513)                                                                                                                                                                                                                                                                                                      | Comparison<br>(to sam X1)                                                                                                           |  |
| Classification Name                                                                                                                                                                                                   | Ultrasonic Diathermy<br>Device                                                                                                                                                                   | Ultrasonic Diathermy<br>Device                                                                                                                                                                                                                                                                                                                                        | Identical                                                                                                                           |  |
| Service Type                                                                                                                                                                                                          | Physical Medicine                                                                                                                                                                                | Physical Medicine                                                                                                                                                                                                                                                                                                                                                     | Identical                                                                                                                           |  |
| Classification                                                                                                                                                                                                        | 21 CFR 890.5300                                                                                                                                                                                  | 21 CFR 890.5300                                                                                                                                                                                                                                                                                                                                                       | Identical                                                                                                                           |  |
| Class                                                                                                                                                                                                                 | II                                                                                                                                                                                               | II                                                                                                                                                                                                                                                                                                                                                                    | Identical                                                                                                                           |  |
| Indications                                                                                                                                                                                                           | The sam CS Long<br>Duration Ultrasound<br>Device is intended for<br>home use to apply<br>ultrasonic energy to<br>generate deep heat within<br>body tissues and to<br>increase local circulation. | The sam X1 Long<br>Duration Ultrasound<br>Device is intended for<br>home use to apply<br>ultrasonic energy to<br>generate deep heat within<br>body tissues for the<br>treatment of selected<br>medical conditions such as<br>the relief of pain, the relief<br>of muscle spasms, the<br>treatment of joint<br>contractures, and the local<br>increase in circulation. | Similar. The<br>indication statement is<br>a subset of the<br>previous indication for<br>use. The intended use<br>remains the same. |  |
| Manufacturer                                                                                                                                                                                                          | ZetrOZ                                                                                                                                                                                           | ZetrOZ                                                                                                                                                                                                                                                                                                                                                                | Identical                                                                                                                           |  |
| Console/Generator<br>Dimensions (L x W x H<br>cm)                                                                                                                                                                     | 3.4 cm x 3.4 cm x 1.4 cm                                                                                                                                                                         | 3.4 cm x 3.4 cm x 1.4 cm                                                                                                                                                                                                                                                                                                                                              | Identical                                                                                                                           |  |
| Treatment Head<br>Dimensions (L x W x H<br>cm)                                                                                                                                                                        | 3.81 cm L x 3.30 cm W x<br>1.14 cm H                                                                                                                                                             | 3.81 cm L x 3.30 cm W x<br>1.14 cm H                                                                                                                                                                                                                                                                                                                                  | Identical                                                                                                                           |  |
| Console/Generator Weight<br>(kg)                                                                                                                                                                                      | 0.02 kg                                                                                                                                                                                          | 0.02 kg                                                                                                                                                                                                                                                                                                                                                               | Identical                                                                                                                           |  |
| Treatment Head Weight<br>(kg)                                                                                                                                                                                         | 0.01 kg                                                                                                                                                                                          | 0.01 kg                                                                                                                                                                                                                                                                                                                                                               | Identical                                                                                                                           |  |
| Power Supply                                                                                                                                                                                                          | 120/240 VAC with 5V DC<br>Input Power Jack and<br>Lithium Battery Powered                                                                                                                        | 120/240 VAC with 5V DC<br>Input Power Jack and<br>Lithium Battery Powered                                                                                                                                                                                                                                                                                             | Identical                                                                                                                           |  |
| Leakage Current                                                                                                                                                                                                       | 0.3 mA                                                                                                                                                                                           | 0.3 mA                                                                                                                                                                                                                                                                                                                                                                | Identical                                                                                                                           |  |
| Crystal Material                                                                                                                                                                                                      | Lead Zirconate-Titanate                                                                                                                                                                          | Lead Zirconate-Titanate                                                                                                                                                                                                                                                                                                                                               | Identical                                                                                                                           |  |
| Technology of ultrasound<br>generation (e.g.,<br>piezoelectric,<br>magnetoconstructive)                                                                                                                               | Piezoelectric                                                                                                                                                                                    | Piezoelectric                                                                                                                                                                                                                                                                                                                                                         | Identical                                                                                                                           |  |
| Treatment Mode(s)                                                                                                                                                                                                     | Two discrete settings of<br>power at same Frequency                                                                                                                                              | Two discrete settings of<br>power at same Frequency                                                                                                                                                                                                                                                                                                                   | Identical                                                                                                                           |  |
| Beam Type (collimated or<br>divergent)                                                                                                                                                                                | Divergent                                                                                                                                                                                        | Divergent                                                                                                                                                                                                                                                                                                                                                             | Identical                                                                                                                           |  |
| Transducer Diameter (cm)                                                                                                                                                                                              | 5 cm                                                                                                                                                                                             | 5 cm                                                                                                                                                                                                                                                                                                                                                                  | Identical                                                                                                                           |  |
| Information                                                                                                                                                                                                           | ZetrOZ sam CS Long<br>Duration Ultrasound<br>Device<br>(Subject)                                                                                                                                 | ZetrOZ sam X1 Long<br>Duration Ultrasound<br>Device<br>(K211513)                                                                                                                                                                                                                                                                                                      | Comparison<br>(to sam X1)                                                                                                           |  |
| Acoustic Working<br>Frequency and Accuracy<br>(MHz)                                                                                                                                                                   | 3MHz ± 20%                                                                                                                                                                                       | 3MHz ± 20%                                                                                                                                                                                                                                                                                                                                                            | Identical                                                                                                                           |  |
| Effective Radiating Area<br>and Accuracy (cm2)                                                                                                                                                                        | One: 6 cm2<br>Two:12 cm2<br>± 20%                                                                                                                                                                | One: 6 cm2<br>Two:12 cm2<br>± 20%                                                                                                                                                                                                                                                                                                                                     | Identical                                                                                                                           |  |
| Beam Nonuniformity Ratio<br>and Accuracy                                                                                                                                                                              | BNR: <5:1 ± 20%                                                                                                                                                                                  | BNR: <5:1 ± 20%                                                                                                                                                                                                                                                                                                                                                       | Identical                                                                                                                           |  |
| Output Mode: (Continuous<br>Wave/Amplitude –<br>Modulated Wave)                                                                                                                                                       | Continuous Wave - 100%<br>duty cycle                                                                                                                                                             | Continuous Wave - 100%<br>duty cycle                                                                                                                                                                                                                                                                                                                                  | Identical                                                                                                                           |  |
| Maximum Timer Setting<br>and Accuracy                                                                                                                                                                                 | 1 Hour +/- 1 minute                                                                                                                                                                              | 1 Hour +/- 1 minute                                                                                                                                                                                                                                                                                                                                                   | Identical                                                                                                                           |  |
| Beam Maximum Intensity<br>and Accuracy (W/cm2)                                                                                                                                                                        | 0.132 W/cm2 ± 20%                                                                                                                                                                                | 0.132 W/cm2 ± 20%                                                                                                                                                                                                                                                                                                                                                     | Identical                                                                                                                           |  |
| Maximum Value of the<br>Output Power (Rated<br>Output Power) and<br>Accuracy (W)                                                                                                                                      | Single Applicator:<br>0.65W ± 20%<br><br>Dual Applicator:<br>1.3W ± 20%                                                                                                                          | Single Applicator:<br>0.65W ± 20%<br><br>Dual Applicator:<br>1.3W ± 20%                                                                                                                                                                                                                                                                                               | Identical                                                                                                                           |  |
| Maximum Value of the<br>Effective Intensity and<br>Accuracy (Not to exceed 3<br>W/cm2 *)                                                                                                                              | 0.264 W/cm2 ± 20%                                                                                                                                                                                | 0.264 W/cm2 ± 20%                                                                                                                                                                                                                                                                                                                                                     | Identical                                                                                                                           |  |
| For Amplitude Modulated<br>Waves                                                                                                                                                                                      | Not Amplitude Modulated                                                                                                                                                                          | Not Amplitude Modulated                                                                                                                                                                                                                                                                                                                                               | Identical                                                                                                                           |  |
| Peak Temperature Rise vs.<br>Time and Tissue Depth to<br>Maximum Treatment Time<br>(for fixed Treatment Head<br>Placement) (deg C)                                                                                    | 7°C at 1 cm<br>4°C at 3 cm<br>2°C at 5 cm<br>Treatment time: 1 hour                                                                                                                              | 7°C at 1 cm<br>4°C at 3 cm<br>2°C at 5 cm<br>Treatment time: 1 hour                                                                                                                                                                                                                                                                                                   | Identical                                                                                                                           |  |
| Maximum Patient Contact<br>Surface Temperature of<br>Treatment Head under<br>Simulated or Actual Use<br>Conditions for all<br>Operating Conditions<br>(Continually operated for<br>maximum treatment time)<br>(deg C) | 44 °C                                                                                                                                                                                            | 44 °C                                                                                                                                                                                                                                                                                                                                                                 | Identical                                                                                                                           |  |
| Therapeutically Applied                                                                                                                                                                                               | Ultrasound Coupling Patch                                                                                                                                                                        | Ultrasound Coupling Patch                                                                                                                                                                                                                                                                                                                                             | Identical                                                                                                                           |  |
| Applicator Type                                                                                                                                                                                                       | Up to two circular<br>Applicators with 3 MHz<br>output                                                                                                                                           | Up to two circular<br>Applicators with 3 MHz<br>output                                                                                                                                                                                                                                                                                                                | Identical                                                                                                                           |  |
| Applicator Type<br>Applicator Emitting<br>Surface Areas (cm2)                                                                                                                                                         | Up to two circular<br>Applicators<br>One Applicator : 5 cm2<br>Two Applicators :10 cm2                                                                                                           | Up to two circular<br>Applicators<br>One Applicator : 5 cm2<br>Two Applicators : 10 cm2                                                                                                                                                                                                                                                                               | Identical                                                                                                                           |  |
| Coupling Bandage                                                                                                                                                                                                      | ABS Plastic with<br>integrated coupling<br>medium                                                                                                                                                | ABS Plastic with<br>integrated coupling<br>medium                                                                                                                                                                                                                                                                                                                     | Identical                                                                                                                           |  |
|                                                                                                                                                                                                                       | ZetrOZ sam CS Long<br>Duration Ultrasound<br>Device                                                                                                                                              | ZetrOZ sam X1 Long<br>Duration Ultrasound<br>Device                                                                                                                                                                                                                                                                                                                   | Comparison                                                                                                                          |  |
| Information                                                                                                                                                                                                           | (Subject)                                                                                                                                                                                        | (K211513)                                                                                                                                                                                                                                                                                                                                                             | (to sam X1)                                                                                                                         |  |
| Applicator Lens Material                                                                                                                                                                                              | TPX                                                                                                                                                                                              | TPX                                                                                                                                                                                                                                                                                                                                                                   | Identical                                                                                                                           |  |
| Environmental - Operating<br>Temperature Range                                                                                                                                                                        | 0°C to +50°C (32°F to<br>+122°F)                                                                                                                                                                 | 0°C to +50°C (32°F to<br>+122°F)                                                                                                                                                                                                                                                                                                                                      | Identical                                                                                                                           |  |
| Performance Standards                                                                                                                                                                                                 | 21 CFR 1050.10                                                                                                                                                                                   | 21 CFR 1050.10                                                                                                                                                                                                                                                                                                                                                        | Identical                                                                                                                           |  |
| Sterility                                                                                                                                                                                                             | Non Sterile                                                                                                                                                                                      | Non Sterile                                                                                                                                                                                                                                                                                                                                                           | Identical                                                                                                                           |  |
| Designed to meet Electrical<br>Safety Standards                                                                                                                                                                       | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11                                                                                                                                                   | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11                                                                                                                                                                                                                                                                                                                        | Identical                                                                                                                           |  |
| Biocompatibility                                                                                                                                                                                                      | Yes                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                                                                   | Identical                                                                                                                           |  |
| Mechanical safety                                                                                                                                                                                                     | Yes                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                                                                   | Identical                                                                                                                           |  |
| Radiation safety (if not<br>radioactive state as such)                                                                                                                                                                | Not Radioactive                                                                                                                                                                                  | Not Radioactive                                                                                                                                                                                                                                                                                                                                                       | Identical                                                                                                                           |  |
| Software/Firmware                                                                                                                                                                                                     | Yes - For measurement of<br>battery level                                                                                                                                                        | Yes - For measurement of<br>battery level                                                                                                                                                                                                                                                                                                                             | Identical                                                                                                                           |  |
| Output Channels                                                                                                                                                                                                       | Two Independent Power<br>Channels                                                                                                                                                                | Two Independent Power<br>Channels                                                                                                                                                                                                                                                                                                                                     | Identical                                                                                                                           |  |

{6}------------------------------------------------

{7}------------------------------------------------

### Substantial Equivalence:

The sam CS Long Duration Ultrasound Device is substantially equivalent to the predicate device with respect to intended use, design, function, and performance. The purpose of this special 510(k) is to adjust the indications for use statement. The output and strength of the signal are identical to the predicate.

### Non-Clinical Performance Data:

The sam CS Long Duration Ultrasound Device is in accordance with these recognized consensus standards:

IEC 60601-1 Ed. 3.1 2012 - Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

IEC 60601-1-2 Ed. 4.0 2014-02 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests

IEC 60601-1-11 Ed. 2.0 2015-01 Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment ISO 10993 Biological evaluation of medical devices - Part 1-Ed. 5-2018: Evaluation and testing within a risk management process; Part 5 Ed. 3-2009: Tests for in vitro cytotoxicity; Part 10 Ed. 3-2010: Tests for irritation and skin sensitization

## Clinical Performance Data:

Clinical testing was not necessary to support substantial equivalence of the sam CS device.

### Conclusion:

The sam CS Long Duration Ultrasound Device possesses the same intended use and technological characteristics as the predicate device. Therefore, the sam CS Long Duration Ultrasound Device is substantially equivalent to the legally marketed predicate.

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/PFW/K233210](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/PFW/K233210)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
