← Product Code [OAP](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/OAP) · K210169

# Hair Growth Device (K210169)

_Dongguan Lescolton Intelligent Electrical Appliance Co., Ltd. · OAP · Apr 16, 2021 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/OAP/K210169

## Device Facts

- **Applicant:** Dongguan Lescolton Intelligent Electrical Appliance Co., Ltd.
- **Product Code:** [OAP](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/OAP.md)
- **Decision Date:** Apr 16, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5500
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The Hair Growth Device is indicated to promote hair growth in males with androgenetic alopecia who have Hamilton-Norwood Classifications of Ila-V and females with androgenetic alopecia who have Ludwig-Savin Classifications of I-II and Fitzpatrick Classification of Skin Phototypes I to IV.

## Device Story

Hair Growth Device LS-D601 is an OTC helmet-based system for home use. It utilizes 26 laser diodes (650-660nm) and 30 red light LEDs (640-660nm) to deliver low-level light therapy to the scalp. The device features an integrated sensor system that automatically pauses therapy if the user's head moves outside the radiation zone and resumes when the correct position is re-established. An audible beep pattern signals the end of the 25-minute treatment cycle. The device is powered by an external adapter. It is intended to promote hair growth in patients with androgenetic alopecia. Users operate the device independently at home; the system provides feedback on treatment completion via audio cues.

## Clinical Evidence

Bench testing only. Includes biocompatibility (ISO 10993-1, -5, -10), electrical safety (ANSI AAMI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), and laser safety (IEC 60825-1). A usability study with 15 participants confirmed that users could understand labeling and operate the device safely and effectively.

## Technological Characteristics

Helmet-based device with 26 laser diodes (650-660nm, <5mW) and 30 red light LEDs (640-660nm). Powered by external adapter. Biocompatible materials per ISO 10993. Complies with IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-57, and IEC 60825-1. Standalone operation.

## Regulatory Identification

An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

## Predicate Devices

- LG Pra.L Derma LD Scalp Care ([K200464](/device/K200464.md))
- iRestore Hair Growth System ([K151662](/device/K151662.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 16, 2021

Dongguan Lescolton Intelligent Electrical Appliance Co., Ltd. % Jinghua Zhou Regulation Control Manager Guangzhou Junyi Information Technology Co., Ltd. Room 304, Building A. No. 62 Nanyun 2nd Road, Science Town Huangpu District, Guangzhou City, Guangdong 510663 China

Re: K210169

Trade/Device Name: Hair Growth Device Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: January 22, 2021 Received: January 22, 2021

Dear Jinghua Zhou:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Purva U. Pandya -S

Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known)

K210169

Device Name Hair Growth Device

Indications for Use (Describe)

The Hair Growth Device is indicated to promote hair growth in males with androgenetic alopecia who have Hamilton-Norwood Classifications of Ila-V and females with androgenetic alopecia who have Ludwig-Savin Classifications of I-II and Fitzpatrick Classification of Skin Phototypes I to IV.

Type of Use (Select one or both, as applicable)

_ Prescription Use (Part 21 CFR 801 Subpart D)

× Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

### Section 5 - 510(k) Summary

Date of Summary Preparation: March 31, 2021

### 1. Submitter's Identifications

Submitter's Name: Dongguan Lescolton Intelligent Electrical Appliance Co., Ltd. Address: 401, Building 1, # 3, Xinxing 3rd Road, jiaoshe, Dongkeng Town, Dongguan City, Guangdong Province, China ZIP Code: 523586 Contact Person: Liu Jianwen Contact Title: General Manager Contact E-mail Address: lescolton2012@outlook.com Telephone: +86-755-33141373

### 2. Correspondent's Identifications

Correspondent's Name: Guangzhou Junyi Information Technology Co., Ltd. Address: Room 304, Building A, No. 62 Nanyun 2nd Road, Science Town, Huangpu District, Guangzhou City, Guangdong, 510663, China ZIP Code: 510663 Contact Person: Jinghua Zhou Contact Title: Regulation Control Manager Contact E-mail Address: admanzhou@126.com Telephone: +86-20-82329549 Fax: +86-20-82329549

### 3. Name of the Device

Trade Name: Hair Growth Device Model: LS-D601 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Common Name: Laser, Comb, Hair Regulatory Class: Class II Product Code: OAP Review Panel: General & Plastic Surgery

### 4. The Predicate Devices

Primary Predicate: K200464 LG Pra.L Derma LD Scalp Care Secondary Predicate: K151662 iRestore Hair Growth System

### 5. Device Description

{4}------------------------------------------------

### K210169

### Dongguan Lescolton Intelligent Electrical Appliance Co., Ltd.

The Hair Growth Device LS-D601 is composed of 26 3R class laser diodes (wavelength: 650-660nm, power<5mW) and 30 red light diodes (wavelength: 640nm-660mm) configured within an outer helmet and protective inner liner.

The use of diode lasers and non-laser LEDs provides for a full coverage of the upper 1/3 of the head; i.e., the area commonly covered with stylized hair. The helmet system will automatically pause to therapy if the subject's head is moved outside of the zone of radiation and will resume therapy when the correct head position is re-established. At the end of the therapy cycle, the system signals that therapy is complete and ready to be powered down, by emitting an audible beep pattern.

#### 6. Intended Use of Device

The Hair Growth Device is indicated to promote hair growth in males with androgenetic alopecia who have Hamilton-Norwood Classifications of Ila-V and females with androgenetic alopecia who have Ludwig-Savin Classifications of I-II and Fitzpatrick Classification of Skin Phototypes I to IV.

{5}------------------------------------------------

### 7. Comparison with Predicate Device

### Table 1 Technical Characteristic in Comparison to Predicate Devices

|                         | Proposed Device                                                                                                                                                                                                                                                                                                                          | Primary Predicate                                                                                                                                                                                                                                                                                                                                  | Secondary Predicate                                                                                                                                                                                                                                                                                                            |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k number             | ------                                                                                                                                                                                                                                                                                                                                   | K200464                                                                                                                                                                                                                                                                                                                                            | K151662                                                                                                                                                                                                                                                                                                                        |
| Proprietary<br>name     | Hair Growth Device                                                                                                                                                                                                                                                                                                                       | LG Pra.L Derma LD<br>Scalp Care                                                                                                                                                                                                                                                                                                                    | iRestore Hair Growth<br>System                                                                                                                                                                                                                                                                                                 |
| Model                   | LS-D601                                                                                                                                                                                                                                                                                                                                  | HGN1                                                                                                                                                                                                                                                                                                                                               | ID-500                                                                                                                                                                                                                                                                                                                         |
| Manufacturer            | Dongguan Lescolton<br>Intelligent Electrical<br>Appliance Co., Ltd.                                                                                                                                                                                                                                                                      | LG Electronics, Inc.                                                                                                                                                                                                                                                                                                                               | Freedom Laser Therapy,<br>Inc.                                                                                                                                                                                                                                                                                                 |
| Regulation<br>number    | 890.5500                                                                                                                                                                                                                                                                                                                                 | 890.5500                                                                                                                                                                                                                                                                                                                                           | 890.5500                                                                                                                                                                                                                                                                                                                       |
| Regulation<br>name      | Infrared Lamp                                                                                                                                                                                                                                                                                                                            | Infrared Lamp                                                                                                                                                                                                                                                                                                                                      | Infrared Lamp                                                                                                                                                                                                                                                                                                                  |
| Regulatory<br>Class     | Class II                                                                                                                                                                                                                                                                                                                                 | Class II                                                                                                                                                                                                                                                                                                                                           | Class II                                                                                                                                                                                                                                                                                                                       |
| Product code            | OAP                                                                                                                                                                                                                                                                                                                                      | OAP                                                                                                                                                                                                                                                                                                                                                | OAP                                                                                                                                                                                                                                                                                                                            |
| Common<br>name          | Laser, Comb, Hair                                                                                                                                                                                                                                                                                                                        | Lamp, non-heating, for<br>promotion of hair growth                                                                                                                                                                                                                                                                                                 | Lamp, non-heating, for<br>promotion of hair growth                                                                                                                                                                                                                                                                             |
| Review panel            | General & Plastic<br>Surgery                                                                                                                                                                                                                                                                                                             | General & Plastic<br>Surgery                                                                                                                                                                                                                                                                                                                       | General & Plastic<br>Surgery                                                                                                                                                                                                                                                                                                   |
| Indications<br>for use  | The Hair Growth Device<br>is indicated to promote<br>hair growth in males with<br>androgenetic alopecia<br>who have<br>Hamilton-Norwood<br>Classifications of IIa-V<br>and females with<br>androgenetic alopecia<br>who have Ludwig-Savin<br>Classifications of I-II and<br>Fitzpatrick Classification<br>of Skin Phototypes I to<br>IV. | The LG Pra.L Derma LD<br>Scalp Care is indicated to<br>promote hair growth in<br>males with androgenetic<br>alopecia who have<br>Hamilton-Norwood<br>Classifications of IIa-V<br>and females with<br>androgenetic alopecia<br>who have Ludwig-Savin<br>Classifications of I-II and<br>Fitzpatrick Classification<br>of Skin Phototypes I to<br>IV. | The iRestore Hair<br>Growth System is<br>indicated to promote hair<br>growth in females with<br>androgenetic alopecia<br>who have Ludwig-Savin<br>Classifications of I - II,<br>males who have<br>Norwood-Hamilton<br>Classifications of IIa - V<br>and for both, Fitzpatrick<br>Classification of Skin<br>Phototypes I to IV. |
| Intended<br>user        | Females & Males                                                                                                                                                                                                                                                                                                                          | Both sex                                                                                                                                                                                                                                                                                                                                           | Females & Males                                                                                                                                                                                                                                                                                                                |
| Type of use             | OTC                                                                                                                                                                                                                                                                                                                                      | OTC                                                                                                                                                                                                                                                                                                                                                | OTC                                                                                                                                                                                                                                                                                                                            |
| Mode of<br>operation    | Low-level laser diodes<br>and light emitting diodes                                                                                                                                                                                                                                                                                      | Low-level laser diodes<br>and light emitting diodes                                                                                                                                                                                                                                                                                                | LLLT Device Type                                                                                                                                                                                                                                                                                                               |
| Wavelength1             | Laser: 650-660nm<br>Red light<br>640-660nm                                                                                                                                                                                                                                                                                               | LD: 650~667nm<br>LED: 645nm~665nm                                                                                                                                                                                                                                                                                                                  | —                                                                                                                                                                                                                                                                                                                              |
| No. of light<br>source2 | Laser diodes: 26<br>LED diodes: 30                                                                                                                                                                                                                                                                                                       | 250                                                                                                                                                                                                                                                                                                                                                | 21 red visible light diode<br>lasers<br>30 red light<br>super-luminescent diodes                                                                                                                                                                                                                                               |
| Power output            | <5mW per light output<br>point                                                                                                                                                                                                                                                                                                           | 5mW per light output<br>point                                                                                                                                                                                                                                                                                                                      | —                                                                                                                                                                                                                                                                                                                              |
| Treatment<br>duration   | 25 minutes every other<br>day for 16 weeks                                                                                                                                                                                                                                                                                               | —                                                                                                                                                                                                                                                                                                                                                  | 25 minutes every other<br>day for 16 weeks                                                                                                                                                                                                                                                                                     |
| Way to get<br>power     | Power adapter                                                                                                                                                                                                                                                                                                                            | Rechargeable battery                                                                                                                                                                                                                                                                                                                               | —                                                                                                                                                                                                                                                                                                                              |
| Bluetooth3              | No                                                                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                                                                | —                                                                                                                                                                                                                                                                                                                              |

{6}------------------------------------------------

Note 1: The wavelengths are all within the range of red light wavelengths. The proposed device emits wavelengths that are similar to predicate device. Differences between the devices do not raise new questions regarding the safety and effectiveness of proposed device and the performance testing done demonstrates that the proposed device can be used safety and effectively.

Note 2: Number of light source of proposed device is less than that of primary predicate K200464, close to secondary predicate K151662. Differences between the devices do not raise new questions regarding the safety and effectiveness of proposed device and the performance testing done demonstrates that the proposed device can be used safety and effectively.

Note 3: The proposed device does not have bluetooth capability. This difference between the devices does not raise new questions regarding the safety and effectiveness of proposed device.

{7}------------------------------------------------

### 8. Non-Clinical Tests Performed:

The following non-clinical testing was provided in this 510(k):

Biocompatibility Testing - The skin contacting materials of the device were subjected to biocompatibility testing per ISO 10993-1:2018, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" for devices in contact with intact skin, including in vitro cytotoxicity, skin sensitization, and skin irritation. All tests passed. The tests conform with the applicable requirements of the following standards:

ISO10993-5:2009, Biological evaluation of medical devices-Part 5: tests for in vitro cytotoxicity.

ISO10993-10:2010, Biological evaluation of medical devices-Part 10: tests for irritation and skin sensitization.

Electrical Safety and Electromagnetic Compatibility Testing - The Hair Growth Device LS-D601 has been tested and conformed with the applicable requirements of the following standards for medical devices used in the home environment:

• ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 Medical electrical equipment-Part 1: General requirements for basic safety and essential performance

· IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests

· IEC 60601-1-11:2015 Medical electrical equipment-Part 1-11: General requirements for basic safety and essential performance - Collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

· IEC 60601-2-57:2011 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests

· IEC 60825-1:2007 Safety of laser products - Part 1: Equipment classification, and requirements

Software Verification and Validation - Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.

### 9. Usability Study:

Testing of 15 participants was conducted showing that the participants were able to understand the user manual and box labeling and were able to safely and effectively use the device.

### 10. Conclusion:

The proposed device uses similar or identical technology as the predicate devices and has the same intended use. Based upon the overall performance characteristics for Hair Growth Device LS-D601, Dongguan Lescolton Intelligent Electrical Appliance Co., Ltd. believes that the technological characteristics of Hair Growth Device LS-D601 does not raise new types of questions regarding its safety and efficacy for its intended use compared with the predicate devices. On the basis of the information provided in this 510(k), Dongguan Lescolton Intelligent Electrical Appliance Co., Ltd. believes that the Hair Growth Device LS-D601 is substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/OAP/K210169](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/OAP/K210169)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
