iHelmet Laser Comb

K202631 · Slinph Technologies Co., Ltd. · OAP · Dec 8, 2020 · Physical Medicine

Device Facts

Record IDK202631
Device NameiHelmet Laser Comb
ApplicantSlinph Technologies Co., Ltd.
Product CodeOAP · Physical Medicine
Decision DateDec 8, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The iHelmet Laser Comb is indicated to promote hair growth in males with androgenic alopecia who have Norwood-Hamilton classifications of Ha-V or females with androgenic alopecia who have Ludwig-Savin Classifications of I-4, II-1, II-2, or frontal and both with Fitzpatrick Skin Phototypes I-IV.

Device Story

Handheld, comb-shaped low-level laser therapy (LLLT) device; emits 650nm laser light to scalp; comb teeth part hair to ensure light delivery. Powered by rechargeable lithium battery; one-button operation; audible timer automatically terminates treatment. Used in home setting by patient. Output is therapeutic laser light; intended to stimulate hair growth. Clinical benefit derived from LLLT mechanism; device safety managed via automatic shut-off and compliance with laser safety standards.

Clinical Evidence

Bench testing only. Device evaluated for electrical safety (IEC 60601-1, IEC 60601-1-11), electromagnetic compatibility (IEC 60601-1-2), laser safety (IEC 60825-1), and battery safety (IEC 62133-2). Biocompatibility testing performed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and sensitization). No clinical trial data presented.

Technological Characteristics

Handheld comb-shaped LLLT device; 12-16 visible red light-emitting diodes (650nm±10nm); power <5mW. Powered by rechargeable lithium battery. Materials identical to K162782. Standards: IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, IEC 60825-1, IEC 62133-2, ISO 10993-5, ISO 10993-10.

Indications for Use

Indicated for males with androgenic alopecia (Norwood-Hamilton IIa-V) and females with androgenic alopecia (Ludwig-Savin I-4, II-1, II-2, or frontal) with Fitzpatrick Skin Phototypes I-IV to promote hair growth.

Regulatory Classification

Identification

An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.

Special Controls

*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a circular seal with an abstract design of human figures. To the right of the seal, there is a blue square with the letters 'FDA' in white. Next to the blue square, the words 'U.S. FOOD & DRUG ADMINISTRATION' are written in a bold, sans-serif font. The words 'FOOD & DRUG' are on the top line, and the word 'ADMINISTRATION' is on the bottom line. December 8, 2020 Slinph Technologies Co., Ltd % Cassie Lee Manager Guangzhou GLOMED Biological Technology Co., Ltd. 2231, Building 1, Rui Feng Center, Kaichung Road, Huangpu District, Guangzhou, Guangdong 51006 China Re: K202631 Trade/Device Name: iHelmet Laser Comb Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: September 6, 2020 Received: September 11, 2020 Dear Cassie Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Purva Pandya Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K202631 Device Name iHelmet Laser Comb (Model: SC8 Dual, SC12, SC16Pro) Indications for Use (Describe) The iHelmet Laser Comb is indicated to promote hair growth in males with androgenic alopecia who have Norwood-Hamilton classifications of Ha-V or females with androgenic alopecia who have Ludwig-Savin Classifications of I-4, II-1, II-2, or frontal and both with Fitzpatrick Skin Phototypes I-IV. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary (K202631) This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 890.5500. #### 1. Submitter Information - � Sponsor Name: SLINPH TECHNOLOGIES CO., LTD - Address: Room 211, Building B,1970 Cultural and Creative Industrial Park, Minzhi Street, � Longhua, Shenzhen - Phone: +86-0755-83461353 � - � Fax: +86-0755-83461353 - Contact Person (including title): Jian Zou (General Manager) � - E-mail: Kevinzou@slinph.com � #### Application Correspondent: 2. - � Contact Person: Ms. Cassie Lee - Guangzhou GLOMED Biological Technology Co., Ltd. � - � Address: 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District, Guangzhou, Guangdong, China - � Tel: +86 20 8266 2446 - Email: regulatory@glomed-info.com � #### Subject Device Information 3. | Type of 510(k): | Traditional | |----------------------|---------------------------| | Trade Name: | iHelmet Laser Comb | | Classification Name: | Laser, comb, hair | | Review Panel: | General & Plastic Surgery | | Product Code: | OAP | | Regulation Number: | 890.5500 | | Regulation Class: | 2 | #### Predicate Device Information 4. {4}------------------------------------------------ 510(k) number K202631 Sponsor: SLINPH TECHNOLOGIES CO., LTD Subject iHelmet Laser Comb, Model: SC8 Dual, SC12, SC16Pro Device: File No.: 510(k) Summary Sponsor: OMM IMPORTS INC DBA ZERO GRAVITY Classification Name: Infrared Lamp Trade Name: Laser Hair Therapy 510(K) Number: K183329 Review Panel: General & Plastic Surgery Product Code: OAP Regulation Number: 890.5500 Regulation Class: 2 #### 5. Device Description The iHelmet Laser Comb is a comb-shaped low level laser therapy (LLLT) device that emits laser light with the intention to promote hair growth. The device provides distributed laser to the scalp at 650+/-10nm while the comb teeth simultaneously part the user's hair to ensure the laser light reaches the user's scalp. The Laser Hair Therapy is designed as handheld product, and it consists of main unit, the charging dock and a power cable, as well as it is powered by the built-in rechargeable lithium battery. The device is one-button operated, and has an audible timer that automatically turns the lasers off after the treatment is completed. #### 6. Intended Use The iHelmet Laser Comb is indicated to promote hair growth in males with androgenic alopecia who have Norwood-Hamilton classifications of Ila-V or females with androgenic alopecia who have Ludwig-Savin Classifications of 1-4, II-1, II-2, or frontal and both with Fitzpatrick Skin Phototypes I-IV. #### 7. Test Summary The iHelmet Laser Comb (Model: SC8 Dual, SC12, SC16Pro) has been evaluated the safety and performance by lab bench testing according to the following standards: | Standards No. | Standard Title | Version | Date | Detail | |----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|------------|--------------| | IEC 60601-1 | Medical Electrical Equipment -<br>Part 1: General Requirements<br>for Safety | 2005+A1:2012 | 07/09/2014 | Attachment 9 | | IEC 60601-1-11 | Medical Electrical Equipment -<br>Part 1-11: General<br>Requirements for Basic Safety<br>and Essential Performance -<br>Collateral Standard:<br>Requirements for Medical<br>Electrical Equipment and | Edition 2.0,<br>2015-01 | 06/27/2016 | Attachment 9 | {5}------------------------------------------------ 510(k) number K202631 Sponsor: SLINPH TECHNOLOGIES CO., LTD Subject iHelmet Laser Comb, Model: SC8 Dual, SC12, SC16Pro Device: File No.: 510(k) Summary | | Medical Electrical Systems<br>Used | | | | |---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|------------|--------------| | IEC 60601-1-2 | Medical electrical equipment -<br>Part 1-2: General requirements<br>for basic safety and essential<br>performance - Collateral<br>standard: Electromagnetic<br>compatibility - Requirements<br>and tests | Edition 4.0<br>2014-02 | 09/17/2018 | Attachment 9 | | IEC 60825-1 | Safety of laser products - Part<br>1: Equipment classification,<br>and requirements [Including:<br>Technical Corrigendum 1<br>(2008), Interpretation Sheet 1<br>(2007), Interpretation Sheet 2<br>(2007)] | Edition 2.0,<br>2007-03 | 07/09/2014 | Attachment 9 | | IEC 62133-2 | Secondary cells and batteries<br>containing alkaline or other<br>non-acid electrolytes - Safety<br>requirements for portable<br>sealed secondary cells, and for<br>batteries made from them, for<br>use in portable applications -<br>Part 2: Lithium systems | Edition 1.0<br>2017-02 | 12/23/2019 | Attachment 9 | #### Comparison to Predicate Device 8. Compared with the predicate device, the subject device has similar in the design principle, the intended use, the indications for use, functions and applicable standards. The differences between the subject device and the predicate device do not raise any new questions of safety or effectiveness. | Elements of<br>Comparison | Subject Device | Predicate Device | Remark | |---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------| | Sponsor | SLINPH TECHNOLOGIES CO., LTD | OMM IMPORTS INC DBA ZERO<br>GRAVITY | -- | | Device name | iHelmet Laser Comb (Model: SC8 Dual,<br>SC12, SC16Pro) | Laser Hair Therapy | -- | | K Number | Pending | K183329 | -- | | Product Code | OAP | OAP | SE | | Classification | Class II | Class II | SE | | Location for use | OTC | OTC | SE | | Intended Use | The iHelmet Laser Comb is indicated to<br>promote hair growth in males with<br>androgenic alopecia who have<br>Norwood-Hamilton classifications of IIa-<br>V or females with androgenic alopecia<br>who have Ludwig-Savin Classifications<br>of I-4, II-1, II-2, or frontal and both with | Laser Hair Therapy is indicated to<br>promote hair growth in males with<br>androgenic alopecia who have<br>IIa-V or females with androgenic<br>alopecia who have Ludwig-Savin<br>Classifications of I-4, II-1, II-2, or | SE | {6}------------------------------------------------ 510(k) number K202631 Sponsor: SLINPH TECHNOLOGIES CO., LTD Subject iHelmet Laser Comb, Model: SC8 Dual, SC12, SC16Pro Device: File No.: 510(k) Summary | Elements of<br>Comparison | Subject Device | Predicate Device | Remark | |-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|--------------| | Fitzpatrick Skin Phototypes I-IV. | | frontal and both with Fitzpatrick<br>Skin Phototypes I-IV. | | | Type of Laser | Visible red light-emitting diodes | Visible red light-emitting diodes | SE | | Light Class | Class I | Class 3R | SE<br>Note 1 | | Amount of Light-<br>emitting diodes | For model SC8 Dual: 16 pcs<br>For model SC12: 12 pcs<br>For model SC16 Pro: 16 pcs | 12 pcs | SE<br>Note 2 | | Power | <5mW | <5mW | SE | | Wavelength | For model SC8 Dual: 650nm±10nm,<br>624 nm<br>For model SC12: 650nm±10nm<br>For model SC16Pro: 650nm±10nm | 655±5nm | SE<br>Note 2 | | Treatment time | For SC8 Dual: 11 minutes per<br>treatment, 3 times per week (every<br>other day)<br>For SC12: 8 minutes per treatment, 3<br>times per week (every other day)<br>For SC16Por: 6 minutes per treatment,<br>3 times per week (every other day) | 8 minutes per treatment, 3 times<br>per week (every other day) | SE<br>Note 2 | | Applicable<br>people | Norwood- Hamilton Ila~V (males);<br>Ludwig-Savin I-4, II-1, II-2, or frontal<br>(females) | Norwood- Hamilton Ila~V (males);<br>Ludwig-Savin I-4, II-1, II-2, or<br>frontal (females) | SE | | Applicable skin | Fitzpatrick Skin Phototypes I- IV | Fitzpatrick Skin Phototypes I- IV | SE | | Appearance<br>design | Comb | Comb | SE | | Safety | Complied with: IEC 60601-1-2; IEC<br>60601-1; IEC 60601-1- 11; IEC 60825-<br>1 | Complied with: IEC 60601-1-2; IEC<br>60601-1; IEC 60601-1- 11; IEC<br>60825-1 | SE | | Biocompatibility | Complied with: ISO 10993-5; ISO<br>10993-10 | Complied with: ISO 10993-5; ISO<br>10993-10 | SE<br>Note 3 | ### Comparison in Detail(s): ### Note 1: Although the "Light Class" of the subject device is different from the predicate device, they all meet the IEC 60825-1 requirements. So the differences will not raise any safety or effectiveness issue. ### Note 2: Although the "Amount of Light-emitting diodes", "Wavelength" and "Treatment time" of the subject device are slightly different from the predicate device's, they are very similar to the predicate {7}------------------------------------------------ | 510(k) number | K202631 | |-----------------|----------------------------------------------------| | Sponsor: | SLINPH TECHNOLOGIES CO., LTD | | Subject Device: | iHelmet Laser Comb, Model: SC8 Dual, SC12, SC16Pro | | File No.: | 510(k) Summary | device and comply with the regulation requirements, so the differences will not raise any safety or effectiveness any issue. ### Note 3: The iHelmet Laser Comb is contact directly with the scalp and skin of human body and the contact duration is less than 24 hours. All materials of the iHelmet Laser Comb (Model: SC8 Dual, SC12, SC16Pro) is identical to the materials of the iHelmet Hair Growth System (K162782), the formulation and processing of them also identical and there are no other chemicals have been added to the iHelmet Laser Comb (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents, etc.). So, they are complied with biocompatibility standards ISO 10993-5 (Cytotoxicity) and ISO 10993-10 (Irritation and Sensitization). - ISO 10993-5, Biological Evaluation of Medical Devices Part 5: Tests for In Vitro Cytotoxicity; A - ISO 10993-10, Biological Evaluation of Medical Devices Part 10: Tests for Irritation and Skin A Sensitization. ### Finial Conclusion: The subject device "iHelmet Laser Comb (Model: SC8 Dual, SC12, SC16Pro)" has all features of the predicate device. The few differences do not affect the safety and effectiveness of the subject device. Thus, the subject device is substantially equivalent to the predicate device K18329. #### 9. Summary Prepared Date 5 December 2020
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...