Gymform Total ABS

K202653 · Well Brain International , Ltd. · NGX · Jan 15, 2021 · Physical Medicine

Device Facts

Record IDK202653
Device NameGymform Total ABS
ApplicantWell Brain International , Ltd.
Product CodeNGX · Physical Medicine
Decision DateJan 15, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5850
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Gymform® TOTAL ABS (Model: WB-245) is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The large belt is intended for use on the muscles in abdomen. The small belt for working the leg and arm muscles, intended for use on the muscles in arms, legs (lower extremities) and thighs areas separately.

Device Story

Gymform Total ABS (WB-245) is a battery-operated powered muscle stimulator (EMS) for home use. Device consists of a control unit, large belt (abdomen), and small belt (arms/legs/thighs) with electrode pads. User operates device via buttons on control unit to select one of 6 stimulation modes and 10 intensity levels. Device delivers gentle electrical current to targeted muscle groups to condition, tone, and strengthen muscles. Default treatment time is 10 minutes. Output parameters are verified against safety standards. Device is intended for OTC use by healthy individuals to improve muscle performance.

Clinical Evidence

No clinical testing performed. Safety and performance established via bench testing, including electrical safety (IEC 60601-1, IEC 60601-2-10), EMC (IEC 60601-1-2), and biocompatibility (ISO 10993-5, ISO 10993-10). Usability study conducted with 24 subjects confirmed lay user ability to self-select, apply, and understand labeling/contraindications.

Technological Characteristics

Powered muscle stimulator (EMS) using microcurrent technology. ABS plastic control unit; 3x AAA battery power. Includes large and small belts with electrode pads. Complies with IEC 60601-1, 60601-1-2, 60601-1-11, and 60601-2-10. Biocompatible materials per ISO 10993. 6 stimulation modes, 10 intensity levels, 10-minute default timer.

Indications for Use

Indicated for healthy individuals to stimulate muscles to improve or facilitate muscle performance. Intended for use on abdominal, arm, leg, and thigh muscles.

Regulatory Classification

Identification

A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. January 15, 2021 Well Brain International Ltd. Jet Li Regulation Manager Guangzhou Kinda Biological Technology Co., Ltd 6F, No.1 TianTai road, Science City, LuoGang District, GuangZhou, Guangdong China Re: K202653 Trade/Device Name: Gymform Total ABS Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: December 14, 2020 Received: December 18, 2020 Dear Jet Li: We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Heather Dean, Ph.D. Assistant Director. Acute Injury Devices Team DHT5B: Division of Neuromodulation and Rehabilitation Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K202653 #### Device Name Gymform® TOTAL ABS, Model: WB-245 #### Indications for Use (Describe) The Gymform® TOTAL ABS (Model: WB-245) is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The large belt is intended for use on the muscles in abdomen. The small belt for working the leg and arm muscles, intended for use on the muscles in arms, legs (lower extremities) and thighs areas separately. Type of Use (Select one or both, as applicable) | ☑ New Construction (Lot/SFR, 201+ Lots) | |-----------------------------------------| |-----------------------------------------| __ Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary K202653 This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR807.92. #### 1. Submitter's Information - 510(k) Sponsor: Well Brain International Ltd. � - Address: Room 1403 Fook Yip Building, 53-57 Kwai Fung Crescent, Kwai Chung, Hong Kong, China � - Phone: (852) 2619-0833 � - Fax: (852) 2429-0960 � - � Contact Person: Victor K Wai - � Email: victor@wellbrain-intl.com #### Subject Device Information 2. | • | Trade Name: | Gymform® TOTAL ABS, Model: WB-245 | |---|----------------------|------------------------------------------------------| | • | Common Name: | Powered muscle stimulator | | • | Classification name: | Stimulator, Muscle, Powered, For muscle conditioning | | • | Review Panel: | Physical Medicine | | • | Product Code: | NGX | | • | Regulation Class: | 2 | | • | Regulation Number: | 890.5850 | #### 3. Application Correspondent #### Application Correspondent: Guangzhou KINDA Biology Technology Co.,Ltd. - Address: 6F, No.1 TianTai road, Science City, LuoGang District, GuangZhouCity,China � - Contact Person: Mr. Jet Li � - � Tile: Regulation Manager - � Tel: +86-18588874857 - � Email:med-jl@foxmail.com {4}------------------------------------------------ #### Predicate and Reference Device Information 4. | 510k<br>Number | K130074<br>(Primary<br>predicate) | K102295<br>(Reference<br>device I) | K092476<br>(Reference<br>device II) | K183674<br>(Reference<br>device III) | |----------------------|------------------------------------|------------------------------------------|-------------------------------------|-------------------------------------------------------------| | Sponsor | Well Brain<br>International<br>Ltd | Leto<br>Enterprises<br>Incorporation | SPORT-<br>ELEC<br>S.A. | Shenzhen<br>Leading<br>Perfection<br>Technology<br>Co., Ltd | | Device<br>Name | ABS - A-<br>Round | X2ABS<br>Dual<br>Channel<br>Fitness Belt | Body<br>Control<br>System '4M' | Electronic<br>Pulse<br>Stimulator,<br>Model S3 | | Product<br>Code | NGX | NGX | NGX | NGX | | Regulation<br>Number | 890.5850 | 890.5850 | 890.5850 | 890.5850 | | Regulation<br>Class | 2 | 2 | 2 | 2 | #### 5. Device Description The Gymform® TOTAL ABS( Model: WB-245) consists of a control unit, large belt, small belt and electrode pads. The control unit enclosed in ABS plastic case. The device is design based on Microcurrent technology (Electro Muscle Stimulation-EMS) which uses a gentle current to condition, tone and strengthen muscles. The Gymform® TOTAL ABS( Model: WB-245) is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The TOTAL ABS may be considered a technique or method for muscle training. It is a battery operated muscle stimulation system specifically designed to training muscles. The large belt is intended for use on the muscles in abdomen. The small belt for working the leg and arm muscles, intended for use on the muscles in arms, legs (lower extremities) and thighs areas separately. The stimulator sends gentle electrical current to targeted muscle group through the electrodes placed on the skin. The parameters of the unit are controlled by the buttons. Its intensity level can be adjustable by user. There are 6 modes for output stimulation, and for each mode, there are 10 levels of output intensity can be choose. Default time is 10minutes for output stimulation. {5}------------------------------------------------ Power is derived from 3 pieces of AAA batteries located in a compartment protected by a removable battery cover for the Fitness Belt. ### 6. Intended Use / Indications for Use The Gymform® TOTAL ABS( Model: WB-245) is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The large belt is intended for use on the muscles in abdomen. The small belt for working the leg and arm muscles, intended for use on the muscles in arms, legs (lower extremities) and thighs areas separately. #### 7. Performance data Summary The Device has been evaluated the safety and performance by lab bench testing according to the following standards: #### Electrical safety and electromagnetic compatibility (EMC) 1. IEC 60601-1-2: 2014: Medical Electrical Equipment - Part 1-2: General requirements for basic safety and essential performance. Collateral Standard: Electromagnetic Compatibility 2. IEC 60601-1:2005: Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance 3. IEC 60601-1-11:2015: Medical Electrical Equipment - Part 1-11: General requirements for basic safety and essential performance. Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment 4. IEC 60601-2-10: 2012: Medical Electrical Equipment - Part 2: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators. #### Biocompatibility testing The biocompatibility evaluation for the device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, " May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of {6}------------------------------------------------ Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process, "as recognized by FDA. The battery of testing included the following tests: - Biological evaluation of medical device Part 10: Tests for Intracutaneos Reactivity test and skin sensitization (ISO 10993-10:2010) | Test | Test summary | Conclusion | |--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------| | Cytotoxicity test | Cytotoxicity Test with MTT Method<br>according to ISO 10993-5 | Non cytotoxic | | Skin sensitization | Skin Sensitization with test method:<br>0.9% Sodium Chloride Injection<br>Extract and Sesame Oil Extract<br>according to ISO 10993-10 | No skin sensitization | | Intracutaneous Reactivity test | Skin Intracutaneous Reactivity with<br>test method: 0.9% Sodium Chloride<br>Injection Extract and Sesame Oil<br>Extract according to ISO 10993-10 | No skin irritation | - Biological evaluation of medical device Part 5: Cytotoxicity test- In vitro method (ISO 10993-5:2009) #### Software Verification and Validation Testing Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern, since a failure of the Software Device result in Minor Injury, either to a patient or to a user of the device. #### Performance verification testing: The waveform test report has also been provided to verify the parameters of the below output current pulse waveform parameter was verified: Output voltage, output current, Output waveform frequency, pulse width, maximum current and power density and so on were tested to demonstrate substantial equivalence to predicate devices. #### Usability study testing: {7}------------------------------------------------ Usability testing (OTC Study) was completed in 24 subjects to evaluate device human factors and label comprehension. These usability study tests demonstrate that the device and its labeling can meet with the requirement: 1) the lay user can self-select themselves as being appropriate users of this device by the external box labeling, 2) the lay user can apply the treatment safely and correctly according to the instructions for use, and 3) the lay user can understand all indications, contraindications, warnings and be able to identify whether they are within any contraindicated group; and be able to understand the user manual. #### 8. Clinical test No clinical testing is need #### 9. Comparison to predicate device and conclusion The technological characteristics, features, specifications, mode of operation, and intended use of Gymform® TOTAL ABS( Model: WB-245) is substantially equivalent to the predicate devices quoted above. Microcurrent technology (Electro Muscle Stimulation-EMS) is the technological principle for both the subject and predicate devices. . The device is design based on Microcurrent (Electro Muscle Stimulation-EMS) which uses a gentle current to condition, tone and strengthen muscles. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness. | Elements of<br>Comparison | Subject Device | Primary<br>predicate | Reference<br>device I | Reference<br>device II | Reference device III | Remark | |--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------| | <b>Sponsor</b> | Well Brain<br>International Ltd | Well Brain<br>International Ltd | Leto<br>Enterprises<br>Incorporation | SPORT-ELEC<br>S.A. | Shenzhen Leading<br>Perfection Technology<br>Co., Ltd | -- | | Device Name<br>and Model | Gymform®<br>TOTAL ABS,<br>(Model: WB-245) | ABS -A- Round | X2ABS Dual<br>Channel<br>Fitness Belt | Body Control<br>System '4M' | Electronic Pulse<br>Stimulator, Model S3 | -- | | 510 (K)<br>Number | Applying | K130074 | K102295 | K092476 | K183674 | -- | | Indications<br>for Use | The Gymform®<br>TOTAL ABS(<br>Model: WB-245) | The<br>GYMFORM6<br>ABS-A-ROUND<br>is intended to | The X2ABS<br>Dual Channel<br>Fitness Belt is<br>intended for | The Body<br>Control -4M' is<br>intended for use<br>by healthy | Mode 1 (PMS):<br>To be used for<br>stimulating healthy | Minor<br>difference | | | is intended to | stimulate | use by healthy | persons to apply | muscles in order to | Note 1 | | | stimulate healthy<br>muscles in order<br>to improve or<br>facilitate muscle<br>performance.. | healthy<br>muscles in<br>order to<br>improve or<br>facilitate<br>muscle<br>performance.<br>The ABS-A-<br>ROUND may<br>be considered a<br>technique or<br>method for | persons to<br>apply trans-<br>coetaneous<br>electrical<br>muscle<br>stimulation<br>(EMS) through<br>skin contact<br>electrodes for<br>the follow ing<br>purposes: | trans-<br>coetaneous<br>electrical muscle<br>stimulation<br>(EMS) through<br>skin contact<br>electrodes for<br>the follow ing<br>purposes.<br>Improvement of<br>muscle tone and<br>firmness, for | improve and<br>facilitate muscle<br>performance.<br>Mode 2 (TENS):<br>To be used for<br>temporary relief of pain<br>associated with sore<br>and aching muscles in<br>the shoulder, waist,<br>back, neck, upper<br>extremities (arm), lower<br>extremities (leg),<br>abdomen and bottom<br>due to strain from<br>exercise or normal<br>household work<br>activities | For<br>predicat<br>e device<br>IV, only<br>compare<br>d with its<br>indicatio<br>n for use<br>in EMS<br>mode. | | | small belt for<br>working the leg<br>and arm<br>muscles,<br>intended for use<br>on the muscles<br>in arms, legs<br>(low er<br>extremities) and<br>thighs areas<br>separately. | muscle training.<br>The 3-area belt<br>is intended for<br>use on the<br>muscles in<br>abdomen, left<br>waist and right<br>w aist<br>alternately.<br>The Mini belt<br>accessory is<br>intended for<br>use on the<br>muscles in<br>arms, legs<br>(low er<br>extremities),<br>thighs and<br>buttocks areas<br>separately.. | of muscle tone<br>of the muscles<br>in the<br>abdomen. | strengthening<br>muscles in arms,<br>abdomen, thighs<br>and buttocks<br>areas. | | | | Channel | 1 | 3 | 2 | 2 | 1 | SE | | Synchronous<br>/Alternating<br>channels | Synchronous | -- | Synchronous | -- | Symmetrical<br>rectangular | SE | | Stimulated<br>muscles | Abdomen, arms,<br>legs and thighs | Arms, Waist,<br>abdomen,<br>legs, thighs<br>and buttocks<br>areas | Abdomen | Arms, abdomen,<br>thighs and<br>buttocks areas | Arm, leg, abdomen,<br>hip, feet | SE | | Number of<br>programs | 6 | 6 | 8 | 4 | 1 | SE | | Number of<br>output<br>intensity level | 10 steps | 99 steps | 28 steps | -- | -- | Minor<br>differenc<br>e<br>Note 2 | | Maximum<br>output<br>voltage | 81.6V @ 500Ω<br>96V @ 2kΩ<br>99.2V @ 10kΩ | 108V @ 500 Ω<br>124V @ 2k Ω<br>126V @ 10k Ω | from 0 to 60V-<br>from 0 to<br>1000 Ohm) | from 0 to 60V-<br>from 0 to 1000<br>Ohm) | 40±20% V @ 500 Ω<br>64±20%V @ 2k Ω<br>76±20% V @ 10k Ω | Minor<br>differenc<br>e<br>Note 3 | | Maximum<br>Output<br>Current | 163mA @ 500Ω<br>48mA @ 2kΩ<br>9.92mA @ 10kΩ | 216mA@ 500<br>Ω<br>62mA@ 2k Ω<br>12.6mA@ 10k<br>Ω | From 0 to 60<br>mA (From 0 to<br>1000 Ohm) | From 0 to 70mA | 80±20%<br>mA@500ohm;<br>32±20% @2kohm;<br>7.6±20% mA<br>@10kohm | Minor<br>differenc<br>e<br>Note 3 | | Frequency<br>range(Hz) | 50,60,70 Hz | 2 Hz, 10 Hz,<br>50 Hz, 90 Hz,<br>120 Hz | 8.5 to 64 Hz | 50 to 70 Hz | Mode 1: 4-35Hz | Minor<br>differenc<br>e<br>Note 3 | | Pulse w idth<br>range(us) | 200 | 100 us / 120 us | 220 | 200 | 200 | Minor<br>differenc<br>e<br>Note 3 | | Contraction<br>and<br>Relaxation<br>time | Adjustable, due<br>to different<br>modes. | Adjustable, due<br>to different<br>modes. | Adjustable,<br>due to<br>different<br>modes. | Adjustable, due<br>to different<br>modes. | -- | SE | | Pow er | 3 x 1.5V AAA<br>batteries | 3 x 1.5V AAA<br>batteries | 2 x 1.5V AAA<br>batteries | 3 x 1.5V AAA<br>batteries | USB rechargeable<br>battery, 3.7V | SE | | Electrode<br>Size (cm²) | 34.5 x 2 pieces;<br>total 5 pairs of<br>electrode pad | 128 | 32 | 128 | -- | Minor<br>differenc<br>e<br>Note 4 | | Maximum<br>Current<br>Density(mA/c<br>m² @ 500Ω) | 0.041 (mA/cm²<br>@ 500Ω) | 0.057 mA/cm² @<br>500 Ω | 0.032<br>(for the<br>smallest size<br>electrode 32.0<br>cm²) | -- | Hips pad: 0.079<br>mA/cm²<br>Arms and legs pad:<br>0.104 mA/cm²<br>Abdomen pad: 0.111 | Minor<br>difference<br>e<br>Note 5 | | Maximum<br>power<br>Density | 28.81μW/cm² | 53.8 μW/cm² @<br>500 Ω | 16. 38 μW/cm² | -- | Hips pad:<br>100μW/cm²;<br><br>Arms and legs pad:<br>200μW/cm²;<br><br>Abdomen pad:<br>200μW/cm²;<br><br>Feet pad:30μW/cm² | Minor<br>difference<br>e<br>Note 5 | | | | | | | Feet pad: 0.016 mA/cm² | | | | | | | | | | | Timer Range | Default time is<br>10minutes. | Default time is<br>19 minutes | Default time is<br>10 minutes,<br>minimum time<br>is 5 minutes | Default time is 4<br>minutes 30 sec | 25 min | Minor<br>difference<br>e<br>Note 6 | | LED display | Indicate the<br>following<br>information:<br>On/off<br>status, Mode<br>selection. | -- | Indicate the<br>following<br>information:<br>Sound on/off,<br>Keylock, Low<br>battery,<br>Channel<br>indication,<br>Intensity level,<br>Mode<br>selection. | -- | On/off status, Low<br>battery | SE | | Environment<br>for operating | Temperature: 0<br>~ 35° C<br>Humidity: <85%<br>RH | -- | Temperature:<br>5 ~ 40° C<br>Humidity: 20 ~<br>65% RH | -- | -- | Minor<br>difference<br>e<br>Note 7 | | Environment<br>for storage | Temperature: 0 ~<br>40° C<br>Humidity: 10 ~<br>90% RH | -- | Temperature:<br>0 ~ 40° C<br>Humidity: 10 ~<br>90% RH | -- | | SE | | | All user directly<br>contacting | | All user<br>directly | | All user directly | | {8}------------------------------------------------ {9}------------------------------------------------ {10}------------------------------------------------ {11}------------------------------------------------ | Biocompatibility | materials are<br>compliance with<br>ISO10993-5 and<br>ISO10993-10<br>requirements. | -- | contacting<br>materials are<br>compliance<br>with<br>ISO10993-5<br>and<br>ISO10993-10<br>requirements. | -- | contacting materials<br>are compliance w ith<br>ISO10993-5 and<br>ISO10993-10<br>requirements. | SE | |----------------------|------------------------------------------------------------------------------------|-------------------------------------------------------|--------------------------------------------------------------------------------------------------------|---------------------------------------------------|------------------------------------------------------------------------------------------------|----| | | | | | | | | | Electrical<br>Safety | Comply w ith IEC<br>60601-1 and IEC<br>60601-2-10 | Comply w ith<br>IEC 60601-1<br>and IEC 60601-<br>2-10 | Comply w ith<br>IEC 60601-1<br>and IEC<br>60601-2-10 | Comply w ith IEC<br>60601-1 and<br>IEC 60601-2-10 | Comply w ith IEC<br>60601-1 and IEC<br>60601-2-10 | SE | | EMC | Comply w ith IEC<br>60601-1-2 | Comply w ith<br>IEC 60601-1-2 | Comply w ith<br>IEC 60601-1-2 | Comply w ith IEC<br>60601-1-2 | Comply w ith IEC<br>60601-1-2 | SE | #### Comparison in Detail(s): #### Note 1: Even there is minor difference on the description words about Indications for use between subject device and predicate device. But actually, the indications for use and treatment parts are same to the primary predicate device I; And the indications for use of other predicate devices also cover the indications for use of subject device. So the differences of the statement of indications for use will not raise any safety or effectiveness issue. #### Note 2: There is difference on Number of output intensity level between subject device and predicate devices, but the key parameter which affect treatment should be output voltage, current, waveform. It's level number do not affect its treatment, So the differences of the level number of output intensity will not raise any safety or effectiveness issue. #### Note 3: For maximum output voltage, Maximum output current: Even the maximum output voltage and current for the subject device are lower than the primary predicate device (K130074). The maximum output voltage and current reflect to the setting value of the highest intensity level of the device. Not each user need the highest intensity level, the user will change the level setting according to their muscle stimulation feeling when operation. And the subject device's max output current value are covered in the output voltage and current range between primary predicate device and predicate device IV(K183674), so we can conclude that the minor difference of maximum output voltage/ output current between subject device and predicate devices would not affect effectiveness for the subject devices. {12}------------------------------------------------ Although the "Frequency range", "Pulse width range" and of subject device are a little different from the primary predicate device I, but the other predicate devices output specification had been cover the specification range of subject device, and they are all compliance with IEC 60601-2-10 requirement and FDA guidance requirement for the EMS. So the minor differences of function specification will not raise any safety or effectiveness issue. #### Note 4: Although the electrode size of subject device are a little different from the predicate devices, the maximum current density and maximum power density are all compliance with IEC 60601-2-10 requirement and FDA quidance requirement for the EMS, and the electrode also had 510K clearance. So the differences of electrode size will not raise any safety or effectiveness issue. #### Note 5: Although maximum current density and maximum power density of subject device are a little different from the predicate devices, they are all compliance with IEC 60601-2-10 requirement and FDA guidance requirement for the EMS. And for the maximum power density difference: For the power density difference, according to the Guidance Document for Powered Muscle Stimulator 510(k)s (page 14), the power density is related to the risk of thermal burning. And the maximum power density meet with the maximum allowed value 0.25 (W/cm²) required in FDA guidance. The effectiveness of micro current mainly depend on the output voltage; but not depend on the power density; And the power density (28.81µW/cm²) of subject device is in the range of predicate device (K130074) and the secondary predicate device (K102295), and for predicate device IV(K183674), its power density of 30 uW/cm² also was cleared in its 510K submission, which value is similar to the data of subject device. Based on aforementioned discussion, we can conclude that the power density between the subject device and the predicate do not raise concerns of effectiveness for the subject device. So the differences of maximum current density and maximum power density will not raise any safety or effectiveness issue. #### Note6: There is minor difference on treatment time range between subject device and predicate devices. But the treatment time can be adjusted by user as they want. So the differences of treatment time raise any safety or effectiveness issue. {13}------------------------------------------------ ### Note7: Even there is some minor difference for operation environment between subject device and predicate device, but the operation environment do not affect the stimulation performance on body, and the device comply with IEC60601-1-11 and IEC60601-1; So the differences of treatment time range will not raise any safety or effectiveness issue. #### Finial Conclusion: The GYMFORM® TOTAL ABS,Model: WB-245 is substantially equivalence to the predicate devices. #### 10. Date of the summary prepared: January 14, 2021
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...