K171092 · International Trade Group, Inc. · NGX · Jul 10, 2017 · Physical Medicine
Device Facts
Record ID
K171092
Device Name
XiniWave II Model ECS322P
Applicant
International Trade Group, Inc.
Product Code
NGX · Physical Medicine
Decision Date
Jul 10, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TENS: The device is intended for the temporary relief of pain associated with sore and aching muscles in the lower back, upper extremities (arms) and lower extremities (legs) due to strain from exercise and/or normal household and work activities. NMS: The device is intended to stimulate healthy muscles in order to improve and facilitate muscle performance.
Device Story
XiniWave II Model ECS322P is a battery-powered, Class II nerve and muscle stimulator for OTC and prescription use. Device inputs include user-selected TENS or NMES programs via keypad controls. It transforms electrical energy from 3x1.5V AAA batteries into controlled pulses delivered to skin electrodes (min 25cm² surface area) placed on the patient's body. The device provides 8 pre-set TENS programs for pain relief and 1 user-selectable NMES program for muscle conditioning. Users operate the device in home or non-clinical environments to manage pain or improve muscle performance. Output parameters (pulse rate 2-100Hz, width 50-450μS) are managed by internal firmware. Healthcare providers or patients use the device to achieve temporary pain relief or muscle conditioning, potentially reducing reliance on other pain management interventions.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence is supported by comparative bench testing of electrical output parameters (current, pulse width, frequency, power density) and risk analysis confirming performance is consistent with the predicate device.
Technological Characteristics
Class II powered muscle/nerve stimulator. Materials: standard plastic cabinet, skin electrodes (industry standard). Energy: 3x1.5V AAA alkaline batteries. Output: TENS (80mA max), NMES (90mA max). Connectivity: hardwired cable/plug to electrodes; no remote RF programming. Software: proprietary firmware, locked. Dimensions: 128x67x25mm. Sterilization: N/A.
Indications for Use
Indicated for temporary relief of pain in lower back, arms, and legs due to exercise or daily activity strain; also indicated for stimulation of healthy muscles to improve/facilitate muscle performance. Intended for OTC and prescription use.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, resembling a family.
July 10, 2017
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
International Trade Group, Inc. Brent Reider President & Secretary of International Trade Group, Inc. 4663 Katie Lane Oxford, Ohio 45056
Re: K171092
Trade/Device Name: XiniWave II Model ECS322P Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX, NUH Dated: February 2, 2017 Received: April 12, 2017
Dear Brent Reider:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Hoffmann -S
for
Carlos L. Peña. PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171092
Device Name XiniWave II Model ECS322P
#### Indications for Use (Describe)
TENS: The device is intended for the temporary relief of pain associated with sore and aching muscles in the lower back, upper extremities (arms) and lower extremities (legs) due to strain from exercise and/or normal household and work activities.
NMS: The device is intended to stimulate healthy muscles in order to improve and facilitate muscle performance.
| Type of Use (Select one or both, as applicable) |
|---------------------------------------------------|
| \[x] Prescription Use (Part 21 CFR 801 Subpart D) |
| \[x] Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| General Information | As Required by 21 CFR 807.92(c) |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant Date: | 26 April 2017 |
| Applicant & Spec. Dev. Name:<br>Applicant & Spec. Dev. Address:<br>Applicant & Spec. Dev. Tele .:<br>Applicant & Spec. Dev. FAX:<br>Applicant & Spec. Dev. Contact:<br>Applicant & Spec. Dev. e-mail: | International Trade Group, Incorporated<br>4663 Katie Lane, Suite "O"<br>Oxford, OH 45056 USA<br>614-568-7000<br>614-568-7001<br>Brent C. Reider, President<br>brent.reider@yarlap.com |
| Device Trade Name and Model: | XiniWave II Model ECS322P |
| 510(k): | K171092 |
| Common/Usual Name: | Stimulator, Nerve, Transcutaneous for Pain Relief<br>Over-the-Counter Combined with<br>Powered Muscle Stimulator, Over-the-Counter. |
| Classification Group: | NGX & NUH |
| Classification Name: | Stimulator, Nerve, Transcutaneous for Pain Relief<br>Over-the-Counter Combined with<br>Powered Muscle Stimulator, Over-the-Counter. |
| Classification regulations: | 21CFR 882.5890 OTC: Stimulator, Nerve,<br>Transcutaneous for Pain Relief |
## Applicant Device Description
The Applicant device, XiniWave II Model ECS322P, is a precision Class II device housed in a sturdy lightweight cabinet. The keypad controls and the circuit for the Applicant device are identical to the Predicate. For a maturing user constituency, the cabinet of the Applicant device is larger than the Predicate with a bigger white paper Liquid Crystal Display (LCD), yet the weight of the device is virtually equivalent. The Applicant device is powered by 3 x 1.5V AAA alkaline batteries. The Applicant device offers the user a choice of eight (8) pre-set Transcutaneous Electro Neuro Stimulation (TENS) programs for the temporary relief of pain (80 mA Max.) and one pre-set Neuro-Muscular Electrical Stimulation (NMES) program for muscle conditioning (90 mA Max.). In the NMES mode of operation the user is able to select work/rest stimulation with pulse rates between 2-100Hz and pulse widths between 50-450uS rate may be selected. The mode of action for the Applicant device in TENS or NMES mode is identical to the mode of action to the Predicate.
{4}------------------------------------------------
The Applicant device is supplied with skin electrodes (electrode surface minimum of 25cm -), to stimulate the patient. The skin electrodes (extant for the industry) are the same ones as supplied with the Predicate and delineated in the User's Manual. The Applicant device control unit connects directly to the skin electrode by cable and plug (extant for the industry).
Sold as a kit, the XiniWave II Model ECS322P kit consists of:
One (1) Battery Powered Stimulator (Control Unit) One (1) set of Four (4) skin electrodes Three (3x1.5V) AAA Batteries One (1) User's Manual One (1) case
Individual components that may be lost (e.g., battery compartment door) or which must be replaced throughout normal usage (e.g., battery and electrode(s) can be re-ordered individually, see User's Manual).
#### Predicate Device:
Per 21 CFR 807.92(a)(3), the cited Predicate device is:
| Predicate Trade Name and Model: XiniWave, Model XW-18 | |
|-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate 519(k): | K100441 |
| Predicate 510(k) Holder: | International Trade Group, Inc.<br>(Applicant of XiniWave II Model ECS322P) |
| Predicate Product Code: | NGX & NUH |
| Predicate Common/Usual Name: | Stimulator, Nerve, Transcutaneous for Pain<br>Relief Over-the-Counter Combined with<br>Powered Muscle Stimulator, Over-the-<br>Counter. |
| Predicate Classification Name: | Stimulator, Nerve, Transcutaneous for Pain<br>Relief Over-the-Counter Combined with<br>Powered Muscle Stimulator, Over-the-<br>Counter. |
| Predicate Classification Reg .: | 21CFR 882.5890 OTC: Stimulator, Nerve,<br>Transcutaneous for Pain Relief<br>21CFR 890.5850 OTC: Powered Muscle<br>Stimulator |
| Predicate Indications for Use: | Unchanged to Applicant Device |
| Predicate Essential Technology: | Unchanged to Applicant Device |
{5}------------------------------------------------
## Indications for Use
The indications for use of the Applicant device, see Section 5.0 table 1 below, are consistent with the uses described under 21CFR 882.5890 OTC: Stimulator, Nerve, Transcutaneous for Pain Relief and 21CFR 890.5850 OTC: Powered Muscle Stimulator.
#### Applicant & Predicate (K100441) Indications for Use (Unchanged) Section 5.0 Table 1
TENS: The device is intended for the temporary relief of pain associated with sore and aching muscles in the lower back, upper extremities (arms) and lower extremities (legs) due to strain from exercise and/or normal household and work activities.
NMS: The device is intended to stimulate healthy muscles in order to improve and facilitate muscle performance.
#### Indications for Use Discussion
The indications for use of the Applicant device and the Predicate are unchanged and consistent with the uses described under 21CFR 882.5890 OTC: Stimulator, Nerve, Transcutaneous for Pain Relief and 21CFR 890.5850 OTC: Powered Muscle Stimulator.
#### Device Modifications
The device modifications listed in Section 5.0, Table 2, do not affect the intended use of the device or alter the fundamental scientific technology of the device.
| Device Modifications | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Section 5.0, Table 2 | | |
| Modification | Predicate Device (K100441) | Applicant Device (TBA) |
| Energy<br>Type: | 1 X 9 Volt Alkaline Battery<br>(standard 800mAh)<br>See Tables 3 & 4 later in this Section and Section 12.0, Tables 3 & 4 for Power Density details. | 3 X 1.5 Volt Alkaline Batteries<br>(standard 1,000mAh) |
| Ergonomics: | Paper White LCD - 35 X 17 mm<br>Weight: 0.09 kg. without battery,<br>0.1 kg. with battery. | Paper White LCD - 45 X 35 mm<br>Weight: 0.12 kg. without battery,<br>0.16 kg. with battery. |
| | Channel Outlet/Connectors on Cabinet Top<br>Electrode Positioning Belt for Lower Back | Channel Outlet/Connectors at Cabinet Base<br>No Electrode Positioning Belt for Lower Back |
| Dimensions: | 108 X 62 X 23 mm. without clip | 128 x 67 x 25 mm. without clip |
| Software+: | TENS: All 7 Programs are Pre-set<br>NMES: All 6 Programs are Pre-set | TENS: All 8 Programs, Pre-set<br>NMES: 1 Program, User Select<br>In NMES Mode user is able to select work/rest stimulation with<br>rates between 2-100Hz & width between 50-450μS |
| | Program Run Times are Pre-set | Program Run Times Pre-Set Incremental to Max.<br>Program Lock Feature to set and record 1 program |
| Manual: | English | English & Spanish |
| + The Applicant and Predicate devices are programmed exactly the same way, by hardwire connection to the Circuit Board using<br>identical secure proprietary software codes. Neither the Applicant nor the Predicate device can be programmed by remote radio<br>frequency signals. Electro Magnetic Certification of the device is presented in Section 17.1 of this Application | | |
## Device Substantial Equivalence and Safety
The device modifications do not alter the fundamental scientific technology of the device under 21CFR 882.5890 OTC: Stimulator, Nerve, Transcutaneous for Pain Relief and 21CFR 890.5850 OTC: Powered Muscle Stimulator. The Applicant device uses
Traditional 510(k): Notification of a New Device: XiniWave II Model ECS322P ProCode: NGX & NUH Document Completion Date: 10 July 2017
Submission: S007 Per 21 CFR 807.92(a)(3)(4)(5)(6)(b(1)(2)(3)(c)(d) Page 3 of 7
{6}------------------------------------------------
## SECTION 5.0 510(k) Summary
Substantially Equivalent programs as the Predicate device (per 21CFR 882.5890 OTC and 21CFR 890.5850 OTC). Program comparisons for the Applicant and Predicate are demonstrated in Section 5.0, Table 3 below:
| Device Substantial Equivalence and Safety of the Applicant and Predicate Programs<br>Section 5.0, Table 3 | | | | | | | | | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|-----------|------------|-------------|---------------|--------------|--------------|-------------|---------------------|----------------------|
| Utility | Program Description | | Rate<br>Hz | Pulse<br>μS | Work<br>Sec. | Rest<br>Sec. | Time<br>Min. | Max. mW/cm² | | |
| | Predicate<br>K100441 | Applicant | | | | | | 80mA | Max. mW/cm²<br>90mA | Ave. mW/cm²<br>45 mA |
| Applicant Device has 8 TENS Programs (3 are HAN/TENS) – The Predicate Device has 7 TENS Programs<br>Electrode Surface Min. of 25cm²<br>Assumes 500 Ohm Load | | | | | | | | | | |
| TENS | Program 1 | | 80 | 200 | n/a | n/a | 60 | 2.0 | n/a | 0.6 |
| HAN | | Program 1 | 2/70 | 260/150 | n/a | n/a | 30 | 1.3 | n/a | 0.4 |
| Predicate & Applicant are similar | | | | | | | | | | |
| TENS | Program 2 | | 50 | 100 | n/a | n/a | 60 | 0.6 | n/a | 0.2 |
| HAN | | Program 2 | 2/15 | 260/180 | n/a | n/a | 30 | 0.3 | n/a | 0.1 |
| Predicate & Applicant are similar | | | | | | | | | | |
| TENS | Program 3 | | 10 | 175 | n/a | n/a | 60 | 0.2 | n/a | 0.1 |
| HAN | | Program 3 | 2/100 | 260/130 | n/a | n/a | 30 | 1.7 | n/a | 0.5 |
| Predicate & Applicant are similar | | | | | | | | | | |
| TENS | Program 4 | | 150 | 200 | n/a | n/a | 60 | 3.8 | n/a | 1.2 |
| | | Program 4 | 2/200 | 50/300 | n/a | n/a | 60 | 7.7 | n/a | 2.4 |
| Predicate & Applicant are different | | | | | | | | | | |
| TENS | Program 5 | | 150 | 175 | n/a | n/a | 60 | 3.4 | n/a | 1.1 |
| | | Program 5 | 150 | 200 | n/a | n/a | 60 | 3.8 | n/a | 1.2 |
| Predicate & Applicant are similar | | | | | | | | | | |
| TENS | Program 6 | | 100/65 | 200/100 | n/a | n/a | 60 | 2.6 | n/a | 0.8 |
| | | Program 6 | 100 | 120 | n/a | n/a | 60 | 1.5 | n/a | 0.8 |
| Predicate & Applicant are similar | | | | | | | | | | |
| TENS | Program 7 | | 65/10 | 200/100 | n/a | n/a | 60 | 1.7 | n/a | 0.5 |
| | | Program 7 | 35 | 200 | n/a | n/a | 60 | 0.9 | n/a | 0.3 |
| Predicate & Applicant are similar | | | | | | | | | | |
| TENS | Program 8 | | 10/15 | 150/250 | n/a | n/a | 55 | 0.5 | n/a | 0.2 |
| Applicant Device has only 1 NMES Program – The Predicate Device has 6 NMES Programs | | | | | | | | | | |
| MNES | Program 9 | | 35 | 250 | 8 | 8 | 15 | n/a | 1.4 | 0.4 |
| | | Program 9 | 2/90 | 50/450 | User Variable | | 15 | n/a | 6.5 | 1.6 |
While the identified program output differences appear evocative, the output of the applicant are within extant mode of clinical effectiveness and devices cleared in 21CFR 882.5890 OTC: Stimulator, Nerve, Transcutaneous for Pain Relief and 21CFR 890.5850 OTC: Powered Muscle Stimulator. Additional information has been included in the manual for the user to better understand and optimally managing the fundamental scientific technology of the device. In this way a lay-person seeking pain management therapy and muscle conditioning can derive maximum benefit from the output potential of the device.
The Applicant and Predicate devices are programmed by hardwire connection to the Circuit Board only. Neither the Applicant nor the Predicate device can be programmed
{7}------------------------------------------------
by remote radio frequency signals. The product firmware is written in a secure (proprietary) code.
The skin electrode in the Applicant device is identical to the Predicate device. The Applicant device control unit connects directly to the skin electrode by cable and plug (design extant for the industry).
Water-resistance is designed into the unit for EU mandated performance requirements per IEC 60601-2-11:2013 only. To the Applicants knowledge, it is not the custom to use any device of this type in wet areas. The Applicant is in complete agreement with FDA guidance not to use the device in wet areas. Hence, the Applicant device will not be promoted as water-resistant nor labelled as such per this 510(k) submission.
The manual of the Applicant and Predicate devices are printed in English. The Applicant device is available in Spanish test is available on request. The languages used in the Applicant device have been selected based on the predominant languages spoken in the United States and Commonwealth of Puerto Rico and then cross-referenced with the cumulative immigration data selecting the three largest languages spoken by immigrant populations to enter theses United States over the past 50 years (1965 to 2015). As such the device labelling complies with or exceeds 21 CFR 801.15(a)(6)(3)(c)(1)(2)& (3).
The Risk Assessment prepared by an independent firm and reviewed by independent commentators, in accordance with the Quality System Regulation (including documentation of design inputs. risk analysis, design output, test procedures, verification and validation procedures, and documentation of formal design reviews) demonstrate that the product modifications are as safe, and effective, and that the Applicant device performs as well as the Predicate device [21 CFR 807.92(b)(3)].
The Applicant device uses the identical pre-set NMES programs as the Predicate device and such, the out-put, in the applicable NMES programs is substantially equivalent as demonstrated in Section 5.0 Table 4 below.
| Device Substantial Equivalence and Safety<br>Applicant and Predicate Out-put by NMES Program<br>Section 5.0 Table 4 | | |
|---------------------------------------------------------------------------------------------------------------------|-----------------------------------|----------------------------------------|
| Basic Unit Characteristics | Applicant Device<br>XiniWave | Applicant Device<br>XiniWave II |
| 510(k) | K100441 | To Be Advised |
| Maximum Output Current pulse<br>peak @ 500 Ohms | 80mA +0/- 8% | TENS 80mA +0/- 8%<br>NMES 90mA +0/- 8% |
| Maximum Output Current pulse<br>peak @ 2K Ohms | 50mA +/-10% | 50mA +/-10% |
| Maximum Output Current pulse<br>peak @ 10K Ohms | 19mA +/-10%<br>And thus shuts off | 19mA +/-10%<br>And thus shuts off |
Traditional 510(k): Notification of a New Device: XiniWave II Model ECS322P ProCode: NGX & NUH Document Completion Date: 10 July 2017
Submission: S007 Per 21 CFR 807.92(a)(3)(4)(5)(6)(b(1)(2)(3)(c)(d) Page 5 of 7
{8}------------------------------------------------
| Pulse Width (μS) | 175 μS – 250 μS,<br>program dependent | 50 μS – 250 μS,<br>program dependent |
|-----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Frequency (Hz) | 12 Hz - 150Hz,<br>program dependent | 2 Hz - 100Hz,<br>program dependent |
| Net Charge @ 500 ohms (μC<br>per pulse)<br>(If zero, state method of<br>achieving zero net charge.) | Zero- positive pulse is equal and<br>opposite to negative pulse.<br>Asymmetrical DC zero<br>(Transformer output) | Zero- positive pulse is equal andSymmetrical DC zero<br>(Transformer output) |
| Maximum (Peak) Phase Charge,<br>(μC) at 500 ohms | $80mA x 250μS = 20 μC$<br>This corresponds to the longest<br>pulse at the highest current. | $90mA x 450 μS = 40 μC$<br>This corresponds to the longest<br>pulse at the highest current. |
| Maximum (peak) Current<br>Density, (mA/cm²) 2 X 2 | $3.2 mA/cm²$<br>Surface = 25 cm² (2X2) | $4.3 mA/cm²$<br>Surface = 25 cm² (2X2) |
| Maximum Power Density,<br>(W/cm²)<br>at 500 ohms | $3.8mW/cm²$<br>At maximum frequency of<br>150Hz pulse width 200µS and<br>current of 80mA. | $7.7 mW/cm²$<br>At maximum frequency of<br>100Hz pulse width 450µS and<br>current of 90mA.<br>TENS Program 04 |
Therefore, since the device modifications do not affect the intended use of the device or alter the fundamental scientific technology of the device concordant with the uses described under 21CFR 882.5890 OTC and for muscle conditioning as described under 21CFR 890.5850 OTC: Powered Muscle Stimulator (i.e., Indications for Use unchanged from the Predicate) and the Quality Assurance paradigm, including Risk, demonstrates that the product modifications are as safe, and effective, and performs as well as or better than the Predicate device. As such the Applicant device and does not raise new questions about safety or effectiveness and demonstrates with reasonable assurance based on established controls that the device is at least as safe and effective as a legally marketed device.
## Conclusion
- . The Applicant device is substantially equivalent to the predicate because it has the same intended therapeutic use as the Predicate described under 21CFR 882.5890 OTC: Stimulator, Nerve, Transcutaneous for Pain Relief and 21CFR 890.5850 OTC: Powered Muscle Stimulator:
- The Applicant device has the similar technological characteristics as the Predicate and that any differences do not adversely impact the safety or effectiveness of the Applicant device;
- The therapeutic labelling of the Applicant device is concordant with the Predicate and
- The information submitted to the FDA for the Applicant device does not raise new questions about safety or effectiveness and demonstrates with reasonable assurance based on established controls that the device is at least as safe and effective as the Predicate, a legally marketed device.
{9}------------------------------------------------
Hence, the XiniWave II Model ECS322P device meets the FDA's definition of Substantial Equivalency under the relationships cited above (21 U.S.C. §§ 360(n), 360c(t)(1) & 360c(i); 21 CFR 807.92(a)(3) and suggest when obtained over-the-counter and used by a layperson without oversight by a healthcare practitioner in a non-clinical environment, the applicant device can be used correctly, safely and in a manner that produces temporary pain relief consistent with the uses described under 21CFR 882.5890 OTC and for muscle conditioning as described under 21CFR 890.5850 OTC: Powered Muscle Stimulator.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.