← Product Code [JFB](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/JFB) · K240168

# N7-S (K240168)

_Nuga Medical Co., Ltd. · JFB · Sep 12, 2024 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/JFB/K240168

## Device Facts

- **Applicant:** Nuga Medical Co., Ltd.
- **Product Code:** [JFB](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/JFB.md)
- **Decision Date:** Sep 12, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5880
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The intended use of the N7-S is to provide patients with muscle relaxation therapy by delivering heat and soothing massage. Additionally, it provides topical heating for; -Temporary relief of minor muscle and joint pain, and stiffness. -Temporary relief of minor joint pain associated with arthritis. -The temporary increase in local circulation where applied. -Relaxation of muscle. The air massager module of N7-S is indicated for the temporary relief of minor muscle aches and for temporary increase in blood circulation to the treated areas. The air massager module of N7-S stimulates kneading of tissues by using an inflatable garment.

## Device Story

N7-S is a multi-function physical therapy table providing heat therapy and massage; includes internal smart heating module with four elements for cervical, thoracic, and lumbar vertebrae; auxiliary heating units for arms; 9-ball projector; abdominal heating pad; and air massager module. Air massager uses inflatable garment with air compression to knead/stroke tissues. Device operated by patient (OTC) via remote control; provides automated and manual modes. Heat therapy targets muscle relaxation and pain relief; air massage targets circulation and muscle aches. Benefits include temporary pain relief and improved local circulation. Used in home or clinic settings.

## Clinical Evidence

No clinical data; bench testing only. Performance testing confirmed temperature range, overheating protection, sensor function, skin surface temperature, and air massager pressure performance met requirements.

## Technological Characteristics

Multi-function physical therapy table with internal heating elements, auxiliary heating, 9-ball projector, abdominal heat pad, and air massager. Materials: Oxford and Nylon sleeves. Sensing: Thermistor-based temperature control. Energy: 12V heating system. Connectivity: Remote control interface. Standards: ES60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62304.

## Regulatory Identification

A multi-function physical therapy table is a device intended for medical purposes that consists of a motorized table equipped to provide patients with heat, traction, and muscle relaxation therapy.

## Predicate Devices

- Ceragem Automatic Thermal Massager (CGM MB-1701 & CGM MB-1702) ([K220572](/device/K220572.md))
- N5-1 ([K153054](/device/K153054.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text.

September 12, 2024

Nuga Medical Co., Ltd. Jun Lee Official Correspondent 185. Jiraeul-ro. Jijeong-myeon Wonju-si, Gangwon-do 26355 Korea. South

Re: K240168

Trade/Device Name: N7-S Regulation Number: 21 CFR 890.5880 Regulation Name: Multi-Function Physical Therapy Table Regulatory Class: Class II Product Code: JFB, IRP Dated: September 3, 2024 Received: September 3, 2024

Dear Jun Lee:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Lauren E. Woodard -S

for Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K240168

Device Name N7-S

#### Indications for Use (Describe)

The intended use of the N7-S is to provide patients with muscle relaxation therapy by delivering heatand soothing massage. Additionally, it provides topical heating for;

-Temporary relief of minor muscle and joint pain, and stiffness.

-Temporary relief of minor joint pain associated with arthritis.

-The temporary increase in local circulation where applied.

-Relaxation of muscle.

The air massager module of N7-S is indicated for the temporary relief of minor muscle aches and for temporary increase in blood circulation to the treated areas. The air massager module of N7-S stimulates kneading of tissues by using an inflatable garment.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Nuga Best. The logo consists of a pink circular graphic on the left, followed by the text "NUGA BEST" in blue, bold letters. The circular graphic appears to depict two stylized figures or abstract shapes, also in pink.

## 510(k) Summary

### 1. Date Prepared

September 3, 2024

#### 2. Submitter Information & Contact Person

| Name of Manufacturer | NUGA MEDICAL Co., Ltd.                                                     |
|----------------------|----------------------------------------------------------------------------|
| Address              | 185, Jiraeul-ro, Jijeong-myeon, Wonju-si, Gangwon-do,<br>Republic of Korea |
| Contact Name         | Lee Jun Woo/Assistant Researcher                                           |
| Telephone No.        | +82-33-730-0001                                                            |
| Fax No.              | +82-33-730-0008                                                            |
| Email Address        | nuga227@nuga.kr                                                            |

### 3. Trade Name, Common Name, Classification

Common name: Multi-Function Physical Therapy Table Trade name: N7-S

| Classification Description              | 21 CFR Section | Product Code |
|-----------------------------------------|----------------|--------------|
| Table, Physical Therapy, Multi Function | 890.5880       | JFB          |
| Massager, Powered Inflatable Tube       | 890.5650       | IRP          |

As stated in 21 CFR, parts 890.5880, and 890.5650, each of these generic types of devices has been classified as Class II.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for NUGA BEST. The logo consists of a pink circular graphic on the left, followed by the text "NUGA BEST" in blue, bold letters. The graphic appears to depict three stylized figures in a circle, possibly representing people or movement.

#### Identification of Predicate Device 4.

The identified predicate devices within this submission are shown as follow:

| Predicate device A |                                       |
|--------------------|---------------------------------------|
| 510(k) Number      | K220572                               |
| Applicant          | CERAGEM Co, Ltd.                      |
| Regulation Name    | Multi-Function Physical Therapy Table |
| Trade Name         | Ceragem Automatic Thermal Massager    |
| Device Name        | CGM MB-1701 & CGM MB-1702             |
| Predicate device B |                                       |
| 510(k) Number      | K153054                               |
| Applicant          | NUGA MEDICAL Co., Ltd.                |
| Regulation Name    | Multi-Function Physical Therapy Table |
| Trade Name         | N5-1                                  |
| Device Name        | N5-1                                  |

N7-S is an updated version of the predicate device B with an addition of a leg massaging feature. Therefore, it is used as a secondary predicate device.

#### 5. Description of the Device

The N7-S is a multi-function physical therapy device offering muscle relaxation through heat therapy and massage.

It has an internal heating unit (smart module), and this module, equipped with four individual heating elements, delivers targeted heat therapy to the cervical, thoracic, and lumbar vertebrae. An auxiliary heating unit is used for the heating units located on both arms to boost the heating effect. For more focused heat therapy and massage, 9-ball projector, Abdominal heating pad are also available. Additionally, it includes an air massager module that features air compression technology to deliver a relaxing and therapeutic massage.

#### 6. Indications for Use

The intended use of the N7-S is to provide patients with muscle relaxation therapy by delivering heat and soothing massage. Additionally, it provides topical heating for;

- Temporary relief of minor muscle and joint pain, and stiffness.
- Temporary relief of minor joint pain associated with arthritis.
- The temporary increase in local circulation where applied.
- Relaxation of muscle.

The air massager module of N7-S is indicated for the temporary relief of minor muscle aches and pains and for temporary increase in blood circulation to the treated areas. The air massager module of N7-S stimulates kneading and stroking of tissues by using an inflatable garment.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Nuga Best. The logo consists of a pink circular graphic on the left, followed by the text "NUGA BEST" in blue. The word "BEST" is slightly smaller and has a registered trademark symbol next to it.

### 7. Determination of Substantial Equivalence

## 7.1. Multi-Function Physical Therapy Table

| Name                   | Subject device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Predicate device A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate device B                                                                                                                                                                                                                                                                                                                                                                                                                                             | SE<br>decision                     |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|
| Device<br>Name         | N7-S                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | CGM MB-1701 & CGM<br>MB-1702                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | N5-1                                                                                                                                                                                                                                                                                                                                                                                                                                                           | -                                  |
| Product<br>Code        | JFB, IRP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | JFB, IRP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | JFB, ILY, ISA                                                                                                                                                                                                                                                                                                                                                                                                                                                  | -                                  |
| Manufacturer           | NUGA MEDICAL Co.,<br>Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | CERAGEM Co, Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | NUGA MEDICAL Co.,<br>Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                      | -                                  |
| 510(k) No.             | -                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | K220572                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | K153054                                                                                                                                                                                                                                                                                                                                                                                                                                                        | -                                  |
| Indications<br>for use | The intended use of the<br>N7-S is to provide<br>patients with muscle<br>relaxation therapy by<br>delivering heat and<br>soothing massage.<br>Additionally, it provides<br>topical heating for;<br>- Temporary relief of<br>minor muscle and joint<br>pain, and stiffness.<br>- Temporary relief of<br>minor joint pain<br>associated with arthritis.<br>- The temporary increase<br>in local circulation where<br>applied.<br>- Relaxation of muscle.<br>The air massager module<br>of the N7-S is indicated<br>for the temporary relief<br>of minor muscle aches<br>and pains and for<br>temporary increase in<br>blood circulation to the<br>treated areas.<br>The air massager module<br>of the N7-S stimulates<br>kneading and stroking of<br>tissues by using an<br>inflatable garment. | The intended use of the<br>Ceragem Automatic<br>Thermal Massager, Model<br>CGM MB-1701 & CGM<br>MB-1702 is to provide<br>patients with muscle<br>relaxation therapy by<br>delivering heat and<br>soothing massage.<br>Additionally, it provides<br>topical radiant infrared<br>heat for;<br>- Temporary relief of<br>minor muscle and joint<br>pain, and stiffness.<br>- Temporary relief of<br>minor joint pain associated<br>with arthritis.<br>- The temporary increase<br>in local circulation where<br>applied.<br>- Relaxation of muscle.<br>The Air Massager (Only<br>CGM MB-1701) is<br>indicated for the temporary<br>relief of minor muscle<br>aches and pains and for<br>temporary increase in<br>blood circulation to the<br>treated areas. The Air Cell<br>Massager stimulates<br>kneading and stroking of<br>tissues by using an<br>inflatable garment. | The intended use of the<br>N5-1 is to provide<br>patients with muscle<br>relaxation therapy by<br>delivering heat and<br>soothing<br>massage. Additionally,<br>it provides topical<br>heating for;<br>-Temporary relief of<br>minor muscle and joint<br>pain, and stiffness.<br>- Temporary relief of<br>minor joint pain<br>associated with<br>arthritis.<br>- The temporary<br>increase in local<br>circulation where<br>applied.<br>- Relaxation of muscle. | Same with<br>Predicate<br>device A |

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for Nuga Best. The logo consists of a stylized graphic of two figures in pink, followed by the text "NUGA BEST" in blue. The word "BEST" is followed by a trademark symbol.

| Name       | Subject device         | Predicate device A               | Predicate device B     | SE<br>decision                           |
|------------|------------------------|----------------------------------|------------------------|------------------------------------------|
| Components | Main Table             | Main Table                       | Main assembly (Mat)    | Similar<br>with<br>Predicate<br>device A |
|            | Remote Control         | Remote Control                   | Remote controller      |                                          |
|            | 9-ball projector       | 3-Sphere projector               | 5-ball Projector       |                                          |
|            | Abdominal heat pad     | Abdominal Vibration<br>Projector | -                      |                                          |
|            | Power cord             | Power cord                       | -                      |                                          |
|            | Outer fabric           | Outer fabric                     | -                      |                                          |
|            | Headrest               | Head cushion                     | -                      |                                          |
|            | Air Cell Massager      | Air Cell Massager                | -                      |                                          |
|            | -                      | Supporting Mat                   | -                      |                                          |
|            | Projector Cover        | Projector Cover                  | -                      |                                          |
|            | Auxiliary Heating Part | -                                | Auxiliary Heating Part |                                          |

The subject device includes all external components that predicate device A has.

However, the subject device has the following differences:

Abdominal Heat Pad: The Abdominal Heat Pad of the subject device is equivalent to the heating function of the Abdominal Vibration Projector of predicate device A.

The headrest in the subject device corresponds to the head cushion in the predicate device A.

The subject device includes all components that predicate device B has.

Auxiliary Heating Unit: Both the subject device and the predicate device B include an Auxiliary Heating Unit in the Main Table Unit. The two devices use the same heating wire.

The aforementioned minor differences do not affect the safety and effectiveness of the device.

| Remote<br>Control            | Yes           | Yes           | Yes           | Same      |
|------------------------------|---------------|---------------|---------------|-----------|
| Operation<br>Method          | Auto / Manual | Auto / Manual | Auto / Manual | Same      |
| Heating<br>Device<br>Voltage | 12V           | 24V           | N/A           | Different |
| Heating<br>Device<br>Power   | 8W            | 28.8V         | N/A           | Different |

There is a difference in the voltage and power of the heating device between the subject device and the predicate devices. However, while having higher voltage and power than equivalent devices may pose a risk, lower values are safe. Additionally, there is no difference in thermal performance compared to predicate devices, and they are substantially equivalent. Therefore, these differences do not affect safety or efficacy.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for NUGA BEST. The logo consists of a pink circular graphic on the left, followed by the text "NUGA BEST" in blue. The word "BEST" is followed by a superscripted circle with an asterisk inside.

| Name                               | Subject device                                                                                                            | Predicate device A                                                                                                                                                                             | Predicate device B                                           | SE<br>decision                                                                                                                         |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Temperature<br>Range               | - Main Table: 40°C-60°C<br>(104°F-140°F)<br>-Auxiliary: 30°C-60°C<br>(86°F-140°F)<br>-External: 30°C-60°C<br>(86°F-140°F) | -External, Main, Auxiliary:<br>30°C-60°C (86°F-140°F)                                                                                                                                          | -External, Main,<br>Auxiliary:<br>40°C-60°C<br>(104°F-140°F) | Main<br>Table unit<br>same with<br>predicate<br>device B<br>Auxiliary<br>and<br>External<br>unit same<br>with<br>predicate<br>device A |
| Extra<br>Overheating<br>Protection | Yes                                                                                                                       | Yes                                                                                                                                                                                            | Yes                                                          | Same                                                                                                                                   |
| Temperature<br>Sensor              | Thermistor                                                                                                                | Thermistor                                                                                                                                                                                     | Thermistor                                                   | Same                                                                                                                                   |
| Heating part                       | Internal projector(Smart<br>module)                                                                                       | Internal projector                                                                                                                                                                             | Internal projector                                           | Same                                                                                                                                   |
|                                    | Auxiliary Heating Part<br>9-ball projector                                                                                | Internal heating element<br>3-Sphere projector                                                                                                                                                 | Auxiliary Heating Part<br>5-ball Projector                   |                                                                                                                                        |
|                                    | Abdominal heat pad                                                                                                        | Abdominal Vibration<br>Projector                                                                                                                                                               | -                                                            |                                                                                                                                        |
| Mode                               | 10 Automatic Modes<br>10 Intensive Modes<br>Manual Mode<br>Projector Mode<br>Air Massage Mode                             | Mode A<br>16 Modes<br>Intensive Mode<br>Semi-Automatic Mode<br>Semi-Automatic Master<br>Mode<br>Manual Mode<br>Manual Master Mode<br>Abdominal Vibration<br>Projector Mode<br>Air Massage Mode | Manual Mode and Auto<br>Mode                                 | Similar                                                                                                                                |

The foundations for both automatic mode and manual mode are the same.

These differences are not sigmificant changes since they are additional features for user convenience and do not raise any questions about safety or effectiveness.

#### 7.2. The air massager

| Name                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Subject device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Predicate device A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | SE decision |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Device Name                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | N7-S                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | CGM MB-1701 & CGM MB-1702                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | -           |
| Product Code                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | JFB, IRP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | JFB, IRP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | -           |
| Manufacturer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | NUGA MEDICAL Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | CERAGEM Co, Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | -           |
| Name                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Subject device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Predicate device A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | SE decision |
| 510(k) No.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | -                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | K220572                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | -           |
| Output pressure<br>range                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 30~140 mmHg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 48~240 mmHg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Different   |
| Air pressure<br>level                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 8 levels settings:<br>Level 1 : 30 mmHg<br>Level 2 : 60 mmHg<br>Level 3 : 80 mmHg<br>Level 4 : 90 mmHg<br>Level 5 : 110 mmHg<br>Level 6 : 120 mmHg<br>Level 7 : 130 mmHg<br>Level 8 : 140 mmHg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 9 levels settings:<br>Level 1 : 48mmHg<br>Level 2 : 72mmHg<br>Level 3 : 96mmHg<br>Level 4 : 120mmHg<br>Level 5 : 144mmHg<br>Level 6 : 168mmHg<br>Level 7 : 192mmHg<br>Level 8 : 216mmHg<br>Level 9 : 240mmHg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Different   |
| The air pressure levels of the subject device are divided into 8 levels whereas the predicate device's air pressure<br>levels are divided into 9 levels. The minimum and maximum pressure level of the subject is significantly lower<br>than the predicate device, so there is no safety issue concerned and as for the efficacy, it remains valid since its<br>performance has been demonstrated by the performance testing data included in this submission.                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |             |
| Mode types                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Auto/Manual                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Sequential/Peristaltic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Similar     |
| Sequential mode of predicate device can be substituted through the Manual mode for each chamber of the subject<br>device, and the Peristaltic mode can be substituted through the function of evenly applying air pressure to each<br>part through the Auto mode of the subject device, with the cycle starting accordingly. So, these similarities do not<br>affect safety and effectiveness.                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |             |
| Therapy time                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 15 or 30 minutes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 18 minutes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Different   |
| The predicate device applies a maximum pressure of 240 mmHg for a treatment duration of up to 18 minutes,<br>while the subject device employs a maximum pressure of 160 mmHg for a 30-minute treatment. The extended<br>treatment duration does not necessarily impact safety and efficacy. However, despite the lower pressure in the<br>subject device, the longer treatment time requires caution regarding blood pressure elevation. Consequently,<br>precautions for using the Air Massage Mode are provided in User Manual page 6, taking into consideration the<br>differences from the predicate device. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |             |
| Deflation time                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 4~8s                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 1~10s                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Different   |
| Name                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Subject device…

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/JFB/K240168](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/JFB/K240168)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
