← Product Code [JFB](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/JFB) · K102281

# THERMAL MASSAGE BED, MODEL 2D-GX (K102281)

_Inarex Corporation · JFB · Dec 22, 2010 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/JFB/K102281

## Device Facts

- **Applicant:** Inarex Corporation
- **Product Code:** [JFB](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/JFB.md)
- **Decision Date:** Dec 22, 2010
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 890.5880
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

Thermal Massage Bed (2D-GX) is a device intended to provide patients with muscle relaxation therapy by delivering heat and soothing massage. Additionally, the infrared lamps provide topical heating for: -Temporary relief of minor muscle and joint pain, and stiffness -Temporary relief of minor joint pain associated with arthritis -Temporary increase in local circulation where applied -Relaxation of muscles

## Device Story

Thermal massage bed provides muscle relaxation therapy via heat and mechanical massage. Device features upper mattress with motor-driven germanium ceramic rollers (horizontal/vertical movement) and lower mattress with heating panel. User operates device via remote control with FND screen/LED indicators; selects auto or custom modes (1-4). Sensors (photo, Hall effect) manage roller positioning and safety limits. System includes auxiliary handheld units for direct body heating. Device used in home or clinic settings by patients. Output consists of thermal energy and mechanical massage; healthcare providers/users monitor temperature/time settings to achieve therapeutic effects (40-45°C). Benefits include pain relief and muscle relaxation.

## Clinical Evidence

Bench testing and limited clinical performance study. Study evaluated skin surface temperature elevation in 5 subjects using 30°C and 65°C settings. Results demonstrated the device elevates skin temperature to the therapeutic range (40-45°C) at the 65°C setting. 30°C setting provides comfort warming only. Safety verified via compliance with IEC 60601-1, IEC 60601-2-38, and IEC 60601-1-2.

## Technological Characteristics

Thermal massage bed; 2,000mm x 700mm x 550mm; 75kg. Germanium ceramic rollers; heating film/panel. Energy source: 120V/230V, 50/60Hz, max 230VA. Sensing: Photo sensor (vertical), Hall effect sensor (horizontal). Temperature range: 30°C-65°C. MCU: Atmega32. Standards: IEC 60601-1, IEC 60601-2-38, IEC 60601-1-2.

## Regulatory Identification

A multi-function physical therapy table is a device intended for medical purposes that consists of a motorized table equipped to provide patients with heat, traction, and muscle relaxation therapy.

## Predicate Devices

- Thermal massage bed (2D-LX) ([K080554](/device/K080554.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) Summary

[as required by 807.92(c)]

# A. 510K Number

K 10228 1

# B. 510(K) Preparer

Company name: Business Association Co., Ltd.

Address: Migun Techno world II C-dong 524, #533-1 Yongsan-dong Yuseong-gu, Daejeon, KOREA

### C. 510(K) Owner

INAREX CORPORATION Name of contact person: Heui-Jong Baek Address: 737-5 Ancheong-Dong Gwangsan-Gu, Gwangju, Korea Telephone Number: +82-62-952-1051 Fax Number: +82-62-952-1054

### D. Regulatory Information

1. Classification Name: Table, physical therapy, multi function

- 2. Common/Usual Name: Thermal massage bed
- 3. Proprietary Name: Thermal massage bed (2D-GX)
- 4. Classification /Product Code: Class II / JFB (21 CFR 890.5880)

#### E. Identification of the legally marketed device

- 1. Classification Name: Table, Physical therapy, Multi function
- 2. Common/Usual Name: Thermal massage bed
- 3. Proprietary Name: Thermal massage bed (2D-LX)
- 4. Classification /Product Code: Class II / JFB (21 CFR 890.5880)
- 5. 510(K) Number: K080554

#### F. Intended Use

Thermal Massage Bed (2D-GX) is a device intended to provide patients with muscle relaxation therapy by delivering heat and soothing massage. Additionally, the lamps provide topical heating. -temporary relief of minor muscle and joint pain, and stiffness -temporary relief of minor joint pain associated with arthritis

-temporary increase in local circulation where applied

-Relaxation of muscles

#### G. Device Description

This device is intended to provide patient with muscle relaxation therapy by delivering heat and soothing massage. This device has auto mode, and four custom modes to massage the back and apply heat to calves and thighs. For the upper mattress, there are germanium

DEC 2 2 2010

{1}------------------------------------------------

ceramic roller controlled left and right by motor. Germanium ceramic roller also can be controlled up and down. The users can choose the auto mode or custom mode by using the remote control and the remote control has the FND screen which shows the temperature and LED which indicates operation condition.

| Feature                          | Specification       |                                              |
|----------------------------------|---------------------|----------------------------------------------|
| Upper and Lower<br>body mattress | Temperature setting | 30°C/35°C/40°C/45°C/50°C/55°C/60°C/65°C/OFF. |
|                                  | Vertical control    | Operated by using up/down photo sensor       |
|                                  | Horizontal Control  | Operated by using Hall effect sensor         |
| AUX8                             | Temperature setting | 30°C/35°C/40°C/45°C/50°C/55°C/60°C/65°C/OFF. |
| Time setting                     |                     | 15/30//45/60/15 (Minutes)                    |
| Mode                             |                     | Auto mode<br>Custom mode1,2,3,4              |
| Weight                           |                     | 75Kg                                         |
| Dimensions                       |                     | 2,000mm(L)x 700mm(w) x 550mm(H)              |

# H. Technological Characteristics:

# I. Substantial Equivalence Comparison

This device, INAREX CORPORATION, thermal massage bed (2D-GX) and predicate device are substantially equivalent and have the same intended use. Technological and performance difference do not raise any new questions of safety or effectiveness. Comparison analysis, including comparison tables, of the subject device versus the predicate device is provided as below:

| Model Name                | 2D-GX                                                                                                                                                                                                          | 2D-LX                                                                                                                                                                                                                                                                    |                                                                                      |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| 510(K) Number             | None yet                                                                                                                                                                                                       | K080554                                                                                                                                                                                                                                                                  |                                                                                      |
| Classification            | Table, physical therapy, multi<br>function (Class II),<br>21 CFR 890.5880                                                                                                                                      | Same                                                                                                                                                                                                                                                                     |                                                                                      |
| Intended Use              | Thermal Massage Bed (2D-GX) is<br>a device intended to provide<br>patients with muscle relaxation<br>therapy by delivering heat and<br>soothing massage. Additionally, the<br>lamps provide topical heating.   | Same                                                                                                                                                                                                                                                                     |                                                                                      |
| Indications               | Thermal Massage Bed (2D-GX) is<br>a device intended to provide<br>patients with muscle relaxation<br>therapy by delivering heat and<br>soothing massage.<br>Additionally, the lamps provide<br>topical heating | Same                                                                                                                                                                                                                                                                     |                                                                                      |
|                           |                                                                                                                                                                                                                | -temporary relief of minor muscle<br>and joint pain, and stiffness                                                                                                                                                                                                       |                                                                                      |
|                           |                                                                                                                                                                                                                | -temporary relief of minor joint<br>pain associated with arthritis                                                                                                                                                                                                       |                                                                                      |
|                           |                                                                                                                                                                                                                | -temporary increase in local<br>circulation where applied                                                                                                                                                                                                                |                                                                                      |
|                           |                                                                                                                                                                                                                | -Relaxation of muscles                                                                                                                                                                                                                                                   |                                                                                      |
| Standards                 |                                                                                                                                                                                                                | IEC60601-1<br>IEC60601-1-2<br>IEC60601-2-38.                                                                                                                                                                                                                             | Same                                                                                 |
|                           | Rated supply volt                                                                                                                                                                                              | 120V~/230V~                                                                                                                                                                                                                                                              | Same                                                                                 |
|                           | Supply Frequency                                                                                                                                                                                               | 50/60Hz                                                                                                                                                                                                                                                                  | Same                                                                                 |
|                           | Power Consumption                                                                                                                                                                                              | Max.230VA                                                                                                                                                                                                                                                                | 200W                                                                                 |
|                           | Dimensions                                                                                                                                                                                                     | 2,000mm(L)x700mm(W)x550mm(H)                                                                                                                                                                                                                                             | 2,000mm(L)x700mm(W)x700mm(H)                                                         |
|                           | Weight                                                                                                                                                                                                         | 75Kg                                                                                                                                                                                                                                                                     | 88Kg                                                                                 |
|                           | Safe Working Load                                                                                                                                                                                              | Max.170Kg                                                                                                                                                                                                                                                                | Same                                                                                 |
|                           | Use temperature                                                                                                                                                                                                | 30°C~65 °C                                                                                                                                                                                                                                                               | Same                                                                                 |
| Upper<br>body<br>mattress | General<br>Description                                                                                                                                                                                         | Programmed to apply thermal<br>massage to upper body by<br>operating germanium roller.                                                                                                                                                                                   | Programmed to apply<br>thermal massage to upper<br>body by operating jade<br>roller. |
|                           | Up/Down                                                                                                                                                                                                        | Rack Gear type                                                                                                                                                                                                                                                           | Cam Follower type                                                                    |
|                           | Up/Down<br>limmit                                                                                                                                                                                              | -Photo Sensor applied<br>-LED for checking the working<br>condition                                                                                                                                                                                                      | Mechanical limmit switch                                                             |
|                           | LAMP<br>Left &<br>right<br>transmission<br>limmit                                                                                                                                                              | Same<br>-Hall effect sensor applied<br>-LED for checking the working<br>condition                                                                                                                                                                                        | 12v 10w<br>Mechanical limmit switch                                                  |
|                           | Massage<br>roller                                                                                                                                                                                              | Germanium ceramic                                                                                                                                                                                                                                                        | Jade                                                                                 |
| Lower<br>body<br>mattress | General<br>Description                                                                                                                                                                                         | Apply heat to the lower body on the<br>lower body mattress.<br>-For the component of lower<br>mattress, there are heating film<br>protection , heating panel, heating<br>film protection, non woven fabric,<br>Germanium ceramic and net<br>material ( written in order) | Give thermal massage<br>therapy by operating jade<br>roller.                         |
|                           | Heating<br>panel                                                                                                                                                                                               | -Two heating film protection<br>protect heating panel back and forth                                                                                                                                                                                                     | Not Applicable                                                                       |
| Control panel             |                                                                                                                                                                                                                | It contains terminal for the power<br>supply, auxiliary hand-held unit,<br>remote control, and connecting<br>cables.                                                                                                                                                     | Same                                                                                 |
| Auxilia                   | General<br>Description                                                                                                                                                                                         |                                                                                                                                                                                                                                                                          |                                                                                      |
| ry<br>handheld unit       |                                                                                                                                                                                                                | Can apply heat directly to the body<br>by one set of germanium hand-held unit                                                                                                                                                                                            | Can apply heat directly to<br>the body by two sets of jade<br>hand-held unit.        |
|                           | 2 Aux                                                                                                                                                                                                          | Eliminated                                                                                                                                                                                                                                                               | Applied                                                                              |
|                           | 8 Aux                                                                                                                                                                                                          | The caps are made of Germanium<br>ceramic                                                                                                                                                                                                                                | Applied<br>The caps are made of Jade                                                 |
|                           | Outer Cover                                                                                                                                                                                                    | Removable and hand washable for<br>easy cleaning                                                                                                                                                                                                                         | Same                                                                                 |
| Remote<br>controller      | General<br>Description                                                                                                                                                                                         | A user friendly remote control that<br>controls all functions automatically<br>or manually.                                                                                                                                                                              | Same                                                                                 |
|                           | Buttons                                                                                                                                                                                                        | Rubber pad type (similar to TV<br>remote control)                                                                                                                                                                                                                        | Sticker pad type                                                                     |
|                           | Temp./Time                                                                                                                                                                                                     | Display area is wider than 2D-LX                                                                                                                                                                                                                                         |                                                                                      |
|                           | Repeating<br>term                                                                                                                                                                                              | Repeating term is divided into four<br>terms.                                                                                                                                                                                                                            | None                                                                                 |
|                           | Height<br>control                                                                                                                                                                                              | (L1, L2, L3) 3 Steps, height is<br>measured from the head side (upper<br>side).                                                                                                                                                                                          | (L1, L2, L3)3 Steps, height<br>is measured from the lower<br>side.                   |
|                           | Software                                                                                                                                                                                                       | MCU Atmega32                                                                                                                                                                                                                                                             | MCU Atmega128                                                                        |
|                           |                                                                                                                                                                                                                | -Due to change of MCU Atmega32 to MCU Atmega128, program<br>size2D-LX (75KB) has been changed into 2D-GX (95KB). In                                                                                                                                                      |                                                                                      |
|                           |                                                                                                                                                                                                                | addition, due to the change of I/O from the chip itself, firmware is                                                                                                                                                                                                     |                                                                                      |
|                           |                                                                                                                                                                                                                | changed and sources of each part are distinguishable and those                                                                                                                                                                                                           |                                                                                      |
|                           |                                                                                                                                                                                                                | make the device modifiable.                                                                                                                                                                                                                                              |                                                                                      |
|                           | Feature                                                                                                                                                                                                        | Thermal method is modified by using Germanium ceramic and<br>heat on the lower part.                                                                                                                                                                                     |                                                                                      |
| Risk<br>analysis          | Potential<br>new<br>hazard<br>introduced by the<br>modification                                                                                                                                                | -Heat transfer rate of germanium ceramic is higher than that of jade<br>and this makes temperature goes higher than 2D-LX.                                                                                                                                               |                                                                                      |
|                           |                                                                                                                                                                                                                | -There is no control system for overload of motor                                                                                                                                                                                                                        |                                                                                      |
|                           | Causes of<br>the<br>hazards                                                                                                                                                                                    | -Temperature of upper germanium roller is higher than that of jade<br>roller.                                                                                                                                                                                            |                                                                                      |
|                           |                                                                                                                                                                                                                | -If there is no bimetal and control system, temperature goes up to<br>120°C                                                                                                                                                                                              |                                                                                      |
|                           |                                                                                                                                                                                                                | -Surface temperature of 8 Auxiliary handheld unit is higher than<br>2D-LX.                                                                                                                                                                                               |                                                                                      |
|                           |                                                                                                                                                                                                                | -If the upper body left &right limmit sensor(hall sensor) is bad, it is<br>hard to control(Activation is checkable by naked eyes)                                                                                                                                        |                                                                                      |
|                           |                                                                                                                                                                                                                | -There is no other control system for motor load. (Even if there is<br>large load while motor is activating, output keeps coming out.)                                                                                                                                   |                                                                                      |
|                           | Potential<br>effects on<br>the user                                                                                                                                                                            | There is possibility of burning by germanium ceramic roller or<br>cam.                                                                                                                                                                                                   |                                                                                      |
|                           | The<br>severity                                                                                                                                                                                                | 4                                                                                                                                                                                                                                                                        |                                                                                      |

{2}------------------------------------------------

·

·

.

{3}------------------------------------------------

1

:

·

{4}------------------------------------------------

| level                                               |                                                                                                                                                          |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| The<br>actions<br>taken to<br>mitigate<br>the risks | -Temperature is controlled by temperature sensor                                                                                                         |
|                                                     | -If the temperature control system is bad, power turns OFF by<br>bimetal.                                                                                |
|                                                     | -Temperature control of lower heat is designed to make lower<br>temperature than actual temperature. (Checking temperature<br>variation)                 |
|                                                     | -If the Hall effect sensor and Photo sensor are not connected, error<br>occurs and system stops.                                                         |
|                                                     | -System turns OFF when the remote control and 2D-GX is<br>separated.                                                                                     |
|                                                     | -Motor should go to head side, and up/down position should be<br>down side when the device is off. Device does not turn off if it is<br>not initialized. |

# J. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence is as follows:

The Model 2D-GX thermal massage bed has been tested to and meets the requirements of the following standards.

IEC60601-1 :88+A1:91+A2:95 Medical electrical equipment-Part1:General requirements for safety

IEC 60601-2-38:1996+A1:1999Medical electrical equipment Part2: Particular requirements for the safety of electrically operated hospital beds

IEC60601-1-2:2007 Medical electrical equipment-Part1-2: General requirements for basic safety and essential performance-Collateral standard: Electromagnetic compatibility-Requirements and tests.

## K. Discussion of Clinical Tests Performed:

The performance test was performed in order to show the safety of the device and to show the device can make the therapeutic range (40-45℃) from the users' skin surface temperature. By testing 30℃ and 65℃ (minimum and maximum of the device temperature setting) for 5 people, the device is able to elevate skin surface temperature from normal baseline temperature to the therapeutic range (40~~45℃) when the device activated by 65℃ setting. And for 30℃ setting, device has not been shown to provide a therapeutic effect and is indicated for comfort warming only.

## L. Conclusion:

The non-clinical testing and clinical study results demonstrate that the 2D-GX, thermal massage bed is safe, accurate. It also demonstrates that the 2D-GX, thermal massage bed is substantially equivalent to the predicate device, 2D-LX, currently sold on the U.S. market.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three lines representing its wings, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MID 20993-0002

Inarex Corporation % Ms. Eileen Yang Business Association Co., Ltd. Migun Techno World II C-dong 524 #533-1 Yongsan-dong Yuseong-gu Daejeon, KOREA

DEC 2 2 2010

Re: K102281

Trade/Device Name: Thermal Massage Bed, Model No. 2D-GX Regulation Number: 21 CFR 890.5880 Regulation Name: Multi-function physical therapy table Regulatory Class: Class II Product Code: JFB Received: November 30, 2010

Dear Ms. Yang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{6}------------------------------------------------

# Page 2 - Ms. Eileen Yang

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

AC is. nh
for.

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

# Indications for Use

DEC 2 2 2010

K102281 510(k) Number (if known):

Device Name: Thermal Massage Bed Model No. 2D-GX

Indications for Use: Thermal Massage Bed (2D-GX) is a device intended to provide patients with muscle relaxation therapy by delivering heat and soothing massage. Additionally, the infrared lamps provide topical heating for:

-Temporary relief of minor muscle and joint pain, and stiffness

-Temporary relief of minor joint pain associated with arthritis

-Temporary increase in local circulation where applied

-Relaxation of muscles

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use x (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign Off

(Division Sign-Off Division of Surgical, Orthopedic, and Restorative Devices

Page 1 of 1

510(k) Number K102281

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/JFB/K102281](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/JFB/K102281)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
