K243488 · Termosalud · ISA · Feb 6, 2025 · Physical Medicine
Device Facts
Record ID
K243488
Device Name
VMAT PRO
Applicant
Termosalud
Product Code
ISA · Physical Medicine
Decision Date
Feb 6, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5660
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The VMAP PRO is intended for relief of minor muscle aches and pains, temporary increase in local blood circulation and activation of connective tissue.
Device Story
VMAT PRO is a pneumatic pressure pulse device for therapeutic massage. It uses compressed air to drive a ballistic projectile within a handpiece, which impacts a transmitter head in contact with the patient's skin. The device features four interchangeable transmitter heads (15mm, 20mm, 35mm) to accommodate various body areas. An FDA-cleared ultrasound gel is required to ensure proper skin contact and transmission. Operated by a clinician, the device delivers continuous vibrations at fixed frequencies (1-21 Hz) and adjustable pressures (0.5-5.0 bar). The mechanical pulses are intended to provide relief from minor muscle aches and pains, increase local blood circulation, and activate connective tissue. The device is used in clinical settings; it does not involve complex software-based analysis or AI/ML algorithms.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and acoustic field characterization (IEC 63045). Biocompatibility was verified per ISO 10993-1:2018.
Indicated for relief of minor muscle aches and pains, temporary increase in local blood circulation, and activation of connective tissue in patients requiring therapeutic massage.
Regulatory Classification
Identification
A therapeutic massager is an electrically powered device intended for medical purposes, such as to relieve minor muscle aches and pains.
Aquasonic 100 Ultrasound Transmission Gel (K802146)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 6, 2025
TermoSalud % Aubrey Thompson Regulatory Consultant Hoy and Associates Regulatory Consultants 1830 Bonnie Way Sacramento, California 95825
Re: K243488
Trade/Device Name: Vmat Pro Regulation Number: 21 CFR 890.5660 Regulation Name: Therapeutic Massager Regulatory Class: Class I Product Code: ISA Dated: October 31, 2024 Received: November 12, 2024
Dear Aubrey Thompson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{3}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Heather L. Dean -S
for Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
Submission Number (if known)
K243488
Device Name
VMAT PRO
Indications for Use (Describe)
The VMAP PRO is intended for relief of minor muscle aches and pains, temporary increase in local blood circulation and activation of connective tissue.
Type of Use (Select one or both, as applicable)
> | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## 510(K) Summary VMAT PRO K243488
| Applicant | TermoSalud |
|--------------------------------------|----------------------------------------------------------------------------|
| Address | Ataulfo Friera Tarfe, 8 -33211 Gijón, Spain |
| Applicant Contact | Cristina Cifuentes Pantoja |
| Correspondent Contact | Aubrey Thompson, Regulatory Consultant<br>AubreyThompson@hoyregulatory.com |
| Preparation Date | February 6, 2025 |
| Device Trade Name | VMAT PRO (K243488) |
| Device Classification Name | Massager, Therapeutic, Electric |
| Common Name | Therapeutic electric massager |
| Regulation Number | 21 CFR 890.5660 |
| Product Code | ISA |
| Regulatory Class | I |
| Legally Marketed Predicate<br>Device | D-Actor 200 Vibration Massage System (K173692) |
### Device Description:
The VMAT PRO is a pressure pulse device that creates pneumatically generated pressure pulses caused by compressed air delivering ballistic projectile through a handpiece coming into contact with the skin. VMAT PRO is equipped with one handpieces that includes four transmitters [VAT120 (diameter of 20mm), VAT220 (diameter of 20mm), VAT215 (diameter of 15mm), and VAT 135 (diameter 35mm)] that allow radial treatments, ergonomically designed to allow the user to work in various areas of the body. To facilitate the movement of the transmitter on the skin, and guarantee a correct contact, an ultrasound gel FDA cleared should be used (i.e. Aquasonic 100 Ultrasound Transmission Gel K802146).
### Indications for use:
The VMAT PRO is intended for relief of minor muscle aches and pains, temporary increase in local blood circulation, and activation of connective tissue.
{6}------------------------------------------------
## 510(K) Summary VMAT PRO K243488
## Substantial Equivalence—Technological Characteristics:
| Subject Device | Predicate Device | Comparison |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| The VMAT PRO is intended for relief of<br>minor muscle aches and pains, temporary<br>increase in local blood circulation, and<br>activation of connective tissue | The D-ACTOR® 200 Vibration Massage System is intended for: Relief of minor<br>muscle aches and pains, temporary<br>increase in local blood circulation,<br>activation of connective tissue | Identical. |
{7}------------------------------------------------
| Technical Comparison | VMAT PRO | D-ACTOR® 200 Vibration<br>Massage System<br>(K173692) | Comparison |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Modes of action | HP-AT pressure pulse handpiece | Radial pressure waves, or<br>extracorporeal pulse<br>activation respectively | Same |
| Mechanism of action | Pneumatically generated<br>vibrations | Pneumatically generated<br>vibrations | Same |
| Type of acoustic wave<br>generation | Pneumatic/ballistic | Pneumatic/ballistic | Same |
| Pulse repeat rate (1/s) | 1-21 Hz | 1-21 Hz | Same |
| Maximum and<br>Minimum intensity<br>setting | 0.5-5.0 bar | 1-5 bar | Very similar- the<br>pressure on the VMAT<br>PRO goes slightly<br>lower than the<br>predicate device, but<br>it does not impact the<br>safety of the device.<br>Treatment pressure is<br>adjusted for patient<br>comfort, so a slightly<br>lower pressure can be<br>more comfortable for<br>patients. |
| Type of application | Continuous vibration at a fixed<br>frequency | Continuous vibration at a<br>fixed frequency | Same |
| Driving power | 1-5 bar | 1-5 bar | Same |
| Projectile Mass (g) | 3.1 g | 3.0 g | Same |
| Pulse repeat rate | 1-21Hz | 1-21Hz | Same |
| Number of pulses | Variable | variable | Same |
| | | | |
| Max and Min<br>displacement of<br>applicator heads | VAT220:<br>Max: 1.230 mm<br>Min: 0.350 mm<br><br>VAT215:<br>Max: 0.320 mm<br>Min: 0.120 mm<br><br>VAT120:<br>Max: 1.300 mm<br>Min: 0.700 mm<br><br>VAT 135:<br>Max: 0.760 mm<br>Min: 0.300 mm | 0.6 – 2.0 mm | Different. There are<br>different ways of<br>measuring this<br>variable and that not<br>all of them are equally<br>accurate. In our case,<br>it has been measured<br>in a laboratory, using<br>ultrasound.<br><br>The difference in<br>values is minimal and<br>does not pose a risk to<br>the safety of the<br>equipment and the<br>user. |
| Pressure | 0.5-5.0 bar (7.25 psi - 72.5psi) | 1-5 bar | Very similar- the<br>pressure on the VMAT<br>PRO goes slightly<br>lower than the<br>predicate device, but<br>it does not impact the<br>safety of the device.<br>Treatment pressure is<br>adjusted for patient<br>comfort, so a slightly<br>lower pressure can be<br>more comfortable for<br>patients. |
| Beam Pressure<br>Maximum (BPM) | According to IEC 63045 (at 5<br>bar)<br>VAT 220: 45.8 mJ<br>VAT 215: 34 mJ | According to modified<br>protocols from IEC 61846<br>N/A | The measurements<br>taken are different |
{8}------------------------------------------------
## 510(K) Summary VMAT PRO
K243488
{9}------------------------------------------------
## 510(K) Summary VMAT PRO K243488 VAT 120: 24 mJ VAT 135: 34 mJ
{10}------------------------------------------------
## 510(K) Summary VMAT PRO
K243488
| Total Derived focal<br>acoustic pulse energy | N/A | 5bar/6.5mJ<br>3bar/2.4mJ | because the<br>measurement for the<br>acoustic pulse was<br>conducted per IEC<br>63045 standards,<br>which requires Beam<br>Pressure Maximum<br>measurements.<br><br>The energy values of<br>the VMAT PRO are<br>well within the usual<br>range for pressure<br>pulse devices. the<br>difference between<br>the values of our<br>VMAT PRO and the<br>values of D-Actor 200<br>is due to the terms<br>used in the standard.<br><br>See further<br>explanation of the<br>difference in<br>specifications below. |
|----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Positive derived<br>acoustic pulse<br>energy | According to IEC 63045<br>(measured for VAT220 at 5bar)<br><br>VAT 220: 39.9 mJ<br><br>VAT 215: 23 mJ<br><br>VAT 120: 17 mJ<br><br>VAT 135: 32 mJ | | |
| Positive peak pressure | VAT220:<br>5 bar: 8.1 MPa<br>3 bar: 6.7 MPa<br><br>VAT215: | Values of ultrasonic pulse:<br>5bar/18.5MPa<br>3bar/13.4MPa | The difference in these<br>values is due to<br>difference in testing<br>methodology. See<br>discussion below. |
| | | | |
| | 5 bar: 11.3 MPa<br>3 bar: 8.1 MPa<br>VAT120:<br>5 bar: 6.3 MPa<br>3 bar: 4.9 MPa<br>VAT 135:<br>5 bar: 3.8 MPa<br>3 bar: 2.8 MPa | | |
| Negative peak<br>pressure | VAT220:<br>5 bar: 6.5 MPa<br>3 bar: 5.1 MPa | Values of ultrasonic pulse:<br>5bar: 6.8MPa<br>3bar: 5.0MPa | Same as VAT220. The<br>510K summary of the<br>predicate device does<br>not state which<br>transmitter was tested<br>for these values. |
| Derived pulse-<br>intensity integral<br>(Energy Flux Density) | VAT220:<br>5 bar: 0.267 mJ/mm2<br>3 bar: 0.152 mJ/mm2<br><br>VAT215:<br>5 bar: 0.224 mJ/mm2<br>3 bar: 0.135 mJ/mm2 | Values of ultrasonic pulse<br>according to modified<br>protocols from IEC 61846:<br>5bar: .284 mJ/mm2<br>3bar: .176 mJ/mm2 | Different. While the<br>values are not<br>identical, they are very<br>close in to one another<br>for the VAT220 and<br>the remaining<br>transmitters all have |
| | VAT120:<br>5 bar: 0.252 mJ/mm2<br>3 bar: 0.187 mJ/mm2<br><br>VAT 135:<br>5 bar: 0.207 mJ/mm2<br>3 bar: 0.129 mJ/mm2 | lower values than they<br>predicate. | |
| Maximum penetration<br>depth | VAT 220: 33.1 mm<br>VAT 215: 23.1 mm<br>VAT 120: 32 mm<br>VAT 135: 35 mm | Different. The<br>differences are within<br>1 mm of the<br>predicates, which does<br>not impact the safety<br>or efficacy. | |
| Working mode | Continuous | Same | |
| Rise time (measured<br>at 5bar (10%-90%)<br>(μs) | VAT220<br>Ultrasonic pulse: 4.97 μs<br><br>VAT215 | Similar. These rise<br>times are all within the<br>standard range for<br>pressure pulse devices. | |
| | Ultrasonic pulse: 2.41 µs<br>VAT 120:<br>Ultrasonic pulse: 4.28 µs | | See further discussion<br>below. |
| | VAT 135:<br>Ultrasonic pulse: 4.16 µs | | |
| Compressional pulse<br>duration (measured at<br>5 bar in µs ) | VAT220<br>Ultrasonic pulse: 6.6 µs<br>VAT215<br>Ultrasonic pulse: 2.93 μs<br>VAT120:<br>Ultrasonic pulse: 4.78 µs<br>VAT 135:<br>Ultrasonic pulse: 6.96 us | Ultrasonic pulse: 5 µs | Similar. These pulse<br>durations are all within<br>the standard range for<br>pressure pulse devices.<br>See further discussion<br>below. |
| Power Supply | 230 VAC 50 Hz<br>115 VAC 60 Hz (optional with<br>reference M93AF) | 500 VA | |
| Maximum operating<br>Temperature | 18-30C | 10-40c | The VMAT operating<br>temperature is within<br>the range of the<br>predicate's |
| Treatment heads | 4: 20mm, 20mm, 35mm, 15mm | 4: 6mmOD, 15mmOD,<br>20mmOD, 35mmOD | Similar sizes |
| Patient Contacting<br>Materials | Titanium, Stainless Steel,<br>Polyoxymethylene | Steel | Similar for transmitters,<br>different for<br>transmitter casing. |
{11}------------------------------------------------
{12}------------------------------------------------
## 510(K) Summary VMAT PRO
K243488
{13}------------------------------------------------
{14}------------------------------------------------
#### Performance Testing
Verification and validation activities were successfully completed and establish that the VMAT PRO control unit performs as intended. Testing included the following:
IEC 60601-1:2005 (Third Edition) + A1:2012 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2:2014 Medical electrical equipment, Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility Requirements and tests
IEC 63045:2020 Ultrasonics -Non-focusing short pressure pulse sources including ballistic pressure pulse sources - Characteristics of fields.
Software verification and validation testing was conducted, and documentation provided in accordance with FDA's Guidance or the Content of Premarket Submissions for Software Contained in Medical Devices.
#### Biocompatibility
All tissue contacting materials were tested for biocompatibility according to ISO 10993-1:2018 and found to be biocompatible.
Clinical Evidence – N/A. No clinical studies were conducted as part of this submission.
### Substantial Equivalence Discussion
Though the measurements in the substantial equivalence table above some differences between the VMAT PRO and the predicate device, the differences are due to different measuring methods. A key parameter to compare is the derived pulse-intensity integral (energy flux density, measured in mJ/mm²). This parameter demonstrates efficacy because the therapeutic effectiveness of pressure pulses depends on whether the energy of the pressure pulse is distributed over a large area or focused on a locally confined treatment zone (focal zone). The measure of energy concentration is obtained by calculating the energy per area (derived pulse-intensity integral) and is obtained after the pulse has stabilized, not at the first impulse or shot.
### Test Methods
The performance testing for the predicate device was conducted using modified protocols from the IEC 61846 standard. Testing for the VMAT Pro was conducted to the standard IEC 63045. This resulted in several differences in the measurements taken.
{15}------------------------------------------------
## 510(K) Summary VMAT PRO K243488
Devices such as the VMAT PRO do not have a focus, so the term derived focal acoustic pulse energy is not applicable in this case. The term is also not used in IEC 63045, the focus has been replaced by the beam pressure maximum (BPM).
The derived pulse-intensity integral values of the VMAT PRO are within the usual range for pneumatic pressure pulse devices. Therefore, the difference between the values of our VMAT PRO and the values of D-Actor 200 is due to the terms used in the regulations.
### Conclusion
The VMAT PRO and D-Actor 200 are identical in their indications for use and similar in their overall design. The key specification to compare, the derived pulse-intensity integral, is nearly identical between the devices. Differences that exist are due to either differences in the method of pressure wave generation, or differences in the test methods used. The differences do not impact the safety or efficacy of the VMAT PRO device and the key comparisons show that the VMAT PRO is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.