← Product Code [IRP](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IRP) · K243177

# SACS PRO (K243177)

_Elastimed , Ltd. · IRP · Dec 23, 2024 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IRP/K243177

## Device Facts

- **Applicant:** Elastimed , Ltd.
- **Product Code:** [IRP](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IRP.md)
- **Decision Date:** Dec 23, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5650
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The ElastiMed's SACS PRO device is a portable and lightweight device intended to provide compression in both sustained and intermittent settings and by doing so, aids in the stimulation of blood flow in the legs. The SACS PRO device is intended to temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas. The device can be used in Clinics, hospital, athlete training, and home environments.

## Device Story

Portable, battery-powered compression device; aids blood flow in lower limbs. Imbeds three Electro Active Polymer (EAP) straps; straps stretch/contract via electric charge; sequential upward contractions mimic muscle action. Microprocessor-controlled; user interface includes LED, buzzer, speaker, and micro-USB charging port. Used in clinics, hospitals, athlete training, or home settings; operated by patient or clinician. Provides sustained and intermittent compression; output is mechanical squeezing of limb. Benefits include temporary relief of minor muscle aches/pains and increased circulation. Device is non-sterile, single-patient use.

## Clinical Evidence

Bench testing only. No clinical data. Validation included electrical safety (ANSI AAMI ES60601-1), EMC (ANSI AAMI IEC 60601-1-2), software validation (IEC 62304), packaging/distribution simulation, mechanical stress testing, and functional durability testing. Performance characteristics were verified against predefined requirements.

## Technological Characteristics

Portable, battery-powered (3 AA 1.2V NiMH) compression wrap. Materials: soft, non-latex fabrics. Actuation: three Electro Active Polymer (EAP) straps. Control: microprocessor-based with LED/buzzer interface. Connectivity: micro-USB for charging. Dimensions: <1kg. Sterilization: non-sterile. Software: validated per IEC 62304. Standards: ANSI AAMI ES60601-1, ANSI AAMI IEC 60601-1-2.

## Regulatory Identification

A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.

## Predicate Devices

- Normatec Go ([K221666](/device/K221666.md))
- SACS ([K232300](/device/K232300.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 23, 2024

ElastiMed Ltd. % John J. Smith Partner Hogan Lovells US LLP Columbia Square 555 13th Street, NW Washington, District of Columbia 20004

Re: K243177 Trade/Device Name: SACS PRO Regulation Number: 21 CFR 890.5650 Regulation Name: Powered Inflatable Tube Massager Regulatory Class: Class II Product Code: IRP Dated: September 30, 2024 Received: September 30, 2024

Dear John J. Smith:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of 

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

{2}------------------------------------------------

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Heather L. Dean -S

Heather Dean, PhD Assistant Director, Acute Injury Device Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K243177

Device Name SACS PRO

Indications for Use (Describe)

The ElastiMed's SACS PRO device is a portable and lightweight device intended to provide compression in both sustained and intermittent settings and by doing so, aids in the stimulation of blood flow in the legs.

The SACS PRO device is intended to temporarily relieve minor muscle aches and to temporarily increase circulation to the treated areas. The device can be used in Clinics, hospital, athlete training, and home environments.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | <span></span>  |
|----------------------------------------------|----------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  | <span>☑</span> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) SUMMARY ElastiMed's SACS PRO Device

## Submitter

ElastiMed Ltd.

HaKidma 7 Yokneam Ilit. 2069236 Israel Tel: +972-74-7036220 Fax: +972-74-7036218 E-mail: omer@elastimed.com

Date Prepared: December 10, 2024

Name of Device: SACS PRO

Common or Usual Name: Compression Device

Classification Name: Powered inflatable tube massager

Regulatory Class: II

Product Code: IRP

C.F.R. Section no .: 21 C.F.R. § 890.5650

#### Predicate Devices:

- -Primary Predicate: NormaTec Industries, LP, Normatec Go (K221666)
- -Secondary Predicate: ElastiMed Ltd., SACS , (K232300)

#### Device Description

The ElastiMed SACS PRO is a lightweight, portable battery powered device, aiding in the stimulation of blood flow in the lower limb through sequential compression.

The SACS PRO device imbeds three Electro Active Polymer (EAP) straps, that sequentially stretch and contract when stimulated by an electric charge. The timely sequential upward contractions of the straps compress the limb, allowing for the repetitive squeezing and relieving actions that mimic normal muscle contractions. The device is supplied with a rechargeable battery which is incased in the fabric casing, and a microprocessor which is imbedded within the control box, located at the front of the one touch operation control box also imbeds a buzzer, LED light indicator, micro speaker, and a micro-USB port (for charging). The device can be used in Clinics, hospital, athlete training, and home environments.

{5}------------------------------------------------

## Intended Use / Indications for Use

The ElastiMed's SACS PRO device is a portable and lightweight device intended to provide compression in both sustained and intermittent settings and by doing so, aids in the stimulation of blood flow in the legs.

The SACS PRO device is intended to temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas. The device can be used in Clinics, hospital, athlete training, and home environments.

## Summary of Technological Characteristics

The SACS PRO device is substantially equivalent to both listed predicate devices in that all devices use a microprocessor to provide intermittent compression to simulate muscle contractions in the return of venous flow.

Like the predicate devices, the SACS PRO device is lightweight, portable, and wraps around the lower limb. Identical to the secondary predicate device, the SACS PRO may be used on one or both legs, the wraps operate separately.

The predicates and subject devices all have the same intended use; specifically, the subject and the primary predicate (Normatec Go) have identical Over the-Counter (OTC) indications for use. Both the secondary predicate device have three discrete compression zones with similar on-body pressure and cycle times (2-3 seconds of compression, with approximately 24 seconds of rest period before the cycle repeats, i.e., 2 cycles per minute), providing sequential compression, which cannot be changed. In all devices, compression is controlled by a microprocessor-controlled system having a user interface, which in addition the system, provides battery and system information (including error notification). All compression systems are encased in soft, non-latex fabrics for patient comfort. The SACS PRO, like the secondary predicate device, is supplied with a rechargeable battery, which can be charged when not in use.

The subject device and the secondary predicate device (SACS) are entirely identical in design and physical form. The sole difference between the subject device and the secondary predicate device is the fact that the secondary predicate device is for prescription use only while the subject device (like the primary predicate device) may be supplied as an OTC device. All systems are provided non-sterile and are intended for single patient use.

{6}------------------------------------------------

|                                            | ElastiMed SACS PRO<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Normatec Go<br>K221666<br>(Primary Predicate Device)                                                                                                                                         | ElastiMed SACS<br>K232300<br>(Secondary Predicate Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use<br>and Indications<br>for Use | The ElastiMed's SACS PRO<br>device is a portable and<br>lightweight device intended to<br>provide compression in both<br>sustained and intermittent<br>settings and by doing so, aids<br>in the stimulation of blood flow<br>in the legs.<br><br>The SACS PRO device is<br>intended to temporarily relieve<br>minor muscle aches and/or<br>pains, and to temporarily<br>increase circulation to the<br>treated areas. The device can<br>be used in Clinics, hospital,<br>athlete training, and home<br>environments. | The Normatec Go is an air<br>pressure massager intended<br>to temporarily relieve minor<br>muscle aches and/or pains,<br>and to temporarily increase<br>circulation to the treated<br>areas. | The ElastiMed's SACS device<br>is a portable and lightweight<br>device intended to provide<br>compression in both<br>sustained and intermittent<br>settings by stimulating blood<br>flow in the legs.<br><br>The SACS device is intended<br>for:<br>- Aid in the prevention of<br>DVT.<br>- Enhance blood circulation.<br>- Diminish post-operative<br>pain and swelling.<br>- Reduce wound healing<br>time.<br>- Aid in the treatment and<br>healing of stasis<br>dermatitis, venous stasis<br>ulcers, arterial and<br>diabetic leg ulcers, chronic<br>venous insufficiency,<br>chronic lymphedema, and<br>reduction of edema in the<br>lower limbs.<br>- As a prophylaxis for DVT<br>by persons expecting to<br>be stationary for long<br>periods of time.<br>- Reduction of edema<br>associated with soft tissue<br>injuries, such as burns,<br>postoperative or post-<br>immobilization edema, or<br>ligament sprains.<br><br>The device can be used in the<br>home or clinical setting. |
| Intended Use<br>Environment                | Clinics, hospital, athlete<br>training, and home<br>environments.                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Clinics, hospital, athlete<br>training, and home<br>environments                                                                                                                             | Home or clinical setting                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Technological<br>Characteristics           | Portable device that wraps<br>around the lower limb and can<br>be used on one or both legs                                                                                                                                                                                                                                                                                                                                                                                                                           | Portable device that wraps<br>around the lower limb and can<br>be used on one or both legs                                                                                                   | Portable device that wraps<br>around the lower limb and can<br>be used on one or both legs                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                            | ElastiMed SACS PRO<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Normatec Go<br>K221666<br>(Primary Predicate Device)                                                                                                                                         | ElastiMed SACS<br>K232300<br>(Secondary Predicate Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Main components                            | Three compression / relaxation enabling straps cased within a fabric casing                                                                                                                                                                                                                                                                                                                                                                                                                                          | Control unit mounted to inflatable segment with compressor and valve system that sequentially inflates cells                                                                                 | Three compression / relaxation enabling straps cased within a fabric casing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Weight                                     | lightweight (<1kg)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 1.2 lbs                                                                                                                                                                                      | lightweight (<1kg)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Power Source                               | 3 AA 1.2V NIMH rechargeable batteries                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Integrated rechargeable battery                                                                                                                                                              | 3 AA 1.2V NIMH rechargeable batteries                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Safety Features                            | LED / Buzzer indicators for battery status and general user interface indications                                                                                                                                                                                                                                                                                                                                                                                                                                    | LED / Buzzer indicators for battery status and general user interface indications                                                                                                            | LED / Buzzer indicators for battery status and general user interface indications                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Software                                   | Validated per applicable standards and guidance documents                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Validated per applicable standards and guidance documents                                                                                                                                    | Validated per applicable standards and guidance documents                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Sterility                                  | Provided non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Provided non-sterile                                                                                                                                                                         | Provided non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Opening and closing mechanism              | Hooks and Loops (Velcro)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Hooks and Loops (Velcro)                                                                                                                                                                     | Hooks and Loops (Velcro)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Operation Mode                             | Sustained and intermittent                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Sequential Gradient, Peristaltic and Pulsing                                                                                                                                                 | Sustained and intermittent                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Pressure                                   | Up to 40mmHg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 40 mmHg to 220 mmHg                                                                                                                                                                          | Up to 40mmHg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Cycle Duration                             | 30 seconds                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Stays on until the user turns it off or can be set up to turn off in a range of 15 minutes to 60 minutes                                                                                     | 30 seconds                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |

{7}------------------------------------------------

# Performance Data

Non-clinical validation, including electrical safety, EMC, software validation, environmental / shipping, life cycle (durability and shelf-life), and performance testing verformance attributes, pressure delivery, geometrical attributes, and system durability, have shown that the SACS PRO device has performance characteristics consistent with the predefined requirements for the devices' intended and indications for use. Since the subject device and the secondary predicate device share identical specifications, i.e., identical device master records, all performance data as provided for the approval of the secondary predicate device are applicable for the subject device.

The following table summarizes the evaluations conducted:

{8}------------------------------------------------

| Testing                                                                 | Summary                                                                                                                                                                                               |
|-------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electrical Safety and<br>EMC                                            | EMC was tested according to ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021].                                                                                                                      |
|                                                                         | Electrical Safety was tested according to ANSI AAMI ES60601-1:2005/(R)2012 &<br>A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021].                                            |
|                                                                         | All pass/fail criteria were based on the device's functions, indications, intended use, and<br>essential performance.                                                                                 |
| Software Validation                                                     | Software validation was conducted per FDA's Content of Premarket Submissions for<br>Device Software Functions (June 14, 2023) and ANSI AAMI IEC 62304:2006/A1:2016<br>[Including Amendment 1 (2016)]. |
| Packaging Design<br>Qualification<br>(Distribution Cycle<br>Simulation) | Device safety and performance was validated after simulated distribution cycle<br>conditioning and 6-month accelerated aging.                                                                         |
| Mechanical Evaluation                                                   | Pristine and aged devices were tested to withstand mechanical force both per<br>specification during simulated use and far exceeding requirement specifications<br>(extreme stress testing).          |
| Functional Evaluation                                                   | System-level verification testing of device performance with usability validation was<br>conducted.                                                                                                   |
| Device Durability                                                       | System-level functional testing after expected use life conditioning was conducted to<br>validate device durability.                                                                                  |

# Conclusions

The ElastiMed SACS PRO device has the same intended and indications for use as the primary predicate device. The results of non-clinical testing (conducted for the identical secondary predicate device) demonstrates that the device met all performance requirements. Therefore, the substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IRP/K243177](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IRP/K243177)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
