LF900 is intended for use by medical professionals and patients at home, who are under medical supervision, in treating many conditions, such as: Primary lymphedema, Edema following trauma and sport injuries, Postimmobilization edema, venous insufficiencies, Lymphedema.
Device Story
LF900 is an intermittent pneumatic compression system designed to improve blood circulation. The device delivers air to a 4-chamber sleeve, pressurizing chambers sequentially from 1st to 4th. It is intended for use in professional healthcare environments and at home under medical supervision. The device includes leg and arm sleeves, extension zippers, and tubing. It operates on 100-127V power, with adjustable pressure (10-140mmHg) and treatment time (5-90 min). The device is used to treat lymphedema and various edema conditions. Healthcare providers or patients operate the device to provide compression therapy, which aids in fluid management and circulation improvement. The device must be used over clothing/socks, never on direct skin.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including electrical safety and electromagnetic compatibility testing per IEC 60601-1 and IEC 60601-1-2, and in-house product performance testing regarding pressure changes and sleeve compatibility.
Technological Characteristics
Intermittent pneumatic compression system; 4-chamber sleeve design; pressure range 10-140mmHg; treatment time 5-90 min; power source 100-127V, 50/60Hz. Compliance with IEC 60601-1 and IEC 60601-1-2. Firmware-controlled operation.
Indications for Use
Indicated for medical professionals and patients at home under medical supervision for treatment of primary lymphedema, edema following trauma/sports injuries, post-immobilization edema, and venous insufficiencies.
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 19, 2023
Daesung Maref Co., Ltd
Su Hyeon So
Jr. RA Specialist
298-24, Gongdan-ro
Gunpo-Si, Gyeonggido 15809
Korea, South
Re: K231437
Trade/Device Name: LF900
Regulation Number: 21 CFR 890.5650
Regulation Name: Powered Inflatable Tube Massager
Regulatory Class: Class II
Product Code: IRP
Dated: June 2, 2023
Received: June 22, 2023
Dear Su Hyeon So:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K231437 - Su Hyeon So
Page 2
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Julia E. 2023.07.19
Slocomb -S 07:58:39 -04'00'
for Heather Dean, PhD
Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation
and Physical Medicine Devices
OHT5: Office of Neurological
and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{2}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K231437 | |
| Device Name LF900 | |
| Indications for Use (Describe) LF900 is intended for use by medical professionals and patients at home, who are under medical supervision, in treating many conditions, such as: Primary lymphedema, Edema following trauma and sport injuries, Postimmobilization edema, venous insufficiencies, Lymphedema. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{3}
510(k) Summary K231437
1. Data Prepared [21 CFR 807.92(a)(a)]
May 10, 2023
2. Submitter's Information [21 CFR 807.92(a)(1)]
- Name of Manufacturer: DAESUNG MAREF CO., LTD.
- Address: 298-24, Gongdan-ro Gunpo-si, Gyeonggido Republic of Korea
- Contact Name: Su Hyeon, So
- Telephone No.: 82-31-459-7211
- Fax No.: 82-31-459-7215
- Email Address: rndra@dsmaref.com
- Registration No.: 3004116008
3. Trade Name, Regulation Name, Classification [21 CFR 807.92(a)(2)]
| | Trade / Device Name | LF900 |
| --- | --- | --- |
| | Regulation Number | 21 CFR 890.5650 |
| | Regulation Name | Massager, Powered Inflatable Tube |
| | Regulation Class | II |
| | Product Code | IRP |
4. Description of the Device [21 CFR 807.92(a)(4), (a)(5)]
| | Operation Principle | This product is an Intermittent Pneumatic Compression system to improve the blood circulation of the human body. The operating principle is that the air from the device will be delivered to the sleeve with 4 air chambers and the air pressurizes sequentially the chambers from 1st to 4th. |
| --- | --- | --- |
| | Indication For Use | LF900 is intended for use by medical professionals and patients at home, who are under medical supervision, in treating many conditions, such as: Primary lymphedema, Edema following trauma and sport injuries, Postimmobilization edema, Venous insufficiencies, Lymphedema. |
1 / 5
{4}
5. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
## Predicate Device
- 510(k) Number:
K150033
- Applicant:
DAESUNG MAREF CO., LTD.
- Trade / Device Name:
LF400
- Regulation Number:
21 CFR 890.5650
- Regulation Name:
Massager, Powered Inflatable Tube
- Regulation Class:
II
- Product Code:
IRP
- Indication For Use
LF-400 is intended for use by medical professionals and patients at home, who are under medical supervision, in treating many conditions, such as: Primary lymphedema, Edema following trauma and sport injuries, Postimmobilization edema, Venous insufficiencies, Lymphedema.
* Predicate device has not been subject to a design-related recall.
## Reference Device
- 510(k) Number:
K203019
- Applicant:
DAESUNG MAREF CO., LTD.
- Trade / Device Name:
LF900
- Regulation Number:
21 CFR 890.5650
- Regulation Name:
Massager, Powered Inflatable Tube
- Regulation Class:
II
- Product Code:
IRP
- Indication For Use
LF900 is intended for use by medical professionals and patients at home, who are under medical supervision, in treating many conditions, such as: Primary lymphedema, Edema following trauma and sport injuries, Postimmobilization edema, Venous insufficiencies, Lymphedema.
{5}
# 6. Determination of Substantial Equivalence
Summary of technological characteristics of the device compared to the predicate device. [21CFR 807.92(a)(6)]
The pressure of the previously approved LF900 (K203019) has been changed from $180\mathrm{mmHg}$ to $140\mathrm{mmHg}$ , and a sleeve type has been added.
Through the table 1 below, we assert that the modified LF900 (K231437) is substantially equivalent to the LF400 (K150033).
[Table 1. Comparison of Proposed Device to Predicate Device]
| | Proposed Device | Predicate Device | Reference Device |
| --- | --- | --- | --- |
| Model Name | LF900 | LF400 | LF900 |
| 510(k) Number | K231437 | K150033 | K203019 |
| Manufacturer | DAESUNG MAREF CO., LTD. | DAESUNG MAREF CO., LTD. | DAESUNG MAREF CO., LTD. |
| Product Code | IRP | IRP | IRP |
| Device Class | II | II | II |
| Regulation Number | 21 CFR 890.5650 | 21 CFR 890.5650 | 21 CFR 890.5650 |
| Regulation Name | Massager, Powered Inflatable Tube | Massager, Powered Inflatable Tube | Massager, Powered Inflatable Tube |
| Indications For Use | LF900 is intended for use by medical professionals and patients at home, who are under medical supervision, in treating many conditions, such as: Primary lymphedema, Edema following trauma and sport injuries, Postimmobilization edema, Venous insufficiencies, Lymphedema. | LF-400 is intended for use by medical professionals and patients at home, who are under medical supervision, in treating many conditions, such as: Primary lymphedema, Edema following trauma and sport injuries, Postimmobilization edema, Venous insufficiencies, Lymphedema. | LF900 is intended for use by medical professionals and patients at home, who are under medical supervision, in treating many conditions, such as: Primary lymphedema, Edema following trauma and sport injuries, Postimmobilization edema, Venous insufficiencies, Lymphedema. |
| Intended Use environment | Professional healthcare environment & Home environment | Professional healthcare environment & Home environment | Professional healthcare environment & Home environment |
{6}
4 / 5
| | Proposed Device | Predicate Device | Reference Device |
| --- | --- | --- | --- |
| Specifications | Leg Sleeve
Arm Sleeve
Extension zipper
One touch Single tubing
One touch Double tubing | Leg Sleeve
Arm Sleeve
Extension zipper
One touch Single tubing
One touch Double tubing | Leg Sleeve
Extension zipper
One touch Single tubing
One touch Double tubing |
| Specifications | | | |
| Power Source | 100-127V~, 50/60Hz | 100-127V~, 50/60Hz | 100-127V~, 50/60Hz |
| Time | 5-90 min | 5~180min | 5-90 min |
| Pressure | 10-140mmHg | 20~140mmHg | 10-180mmHg |
| Number of chamber | 4 | 4 | 4 |
[Table 2. Little difference with Predicate Device]
| Justification to Support Substantial Equivalence |
| --- |
| The pressure of the LF900 was changed to be the same as that of the LF400 to add the arm sleeves used for the LF400. LF900 and LF400 have the same principle of operation and intended use, and the pressure is the same, so it is judged that the two products are substantial equivalent. |
## Non-Clinical Test Summary
According to the pressure change of LF900, retests and gap tests related to electrical safety and electromagnetic compatibility were performed, and the standards recognized by the FDA were complied with.
The following data were provided in support of the substantial equivalence determination:
{7}
1) Electrical Safety, Electromagnetic Compatibility and Performance
The proposed device (LF900) complied with the following standards for the electrical safety and electromagnetic compatibility requirements of the product after the changed maximum pressure and firmware was changed.
- IEC 60601:2005/A1:2012, Medical Electrical Equipment: Part 1: General Requirements for Basic Safety and Essential Performance
- IEC 60601 60601-1-2:2014, Medical Electrical Equipment - Part 1 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests
[Inhouse test] Product Performance Test report, RND-R-PRR-103-02(2022.06.13)
Product performance test according to pressure change of LF900 and compatibility test with the sleeve to be added.
## Clinical Test Summary
Clinical testing was not required to demonstrate the substantial equivalence of the LF900 to its predicate device
The device should only be used over full clothing and socks (because the cuffs cover the feet), never on direct skin.
The biocompatibility of the materials has not been verified by the FDA and contact of the cuffs/accessories to direct skin may lead to skin irritation, skin sensitization and/or cytotoxicity.
## 7. Conclusion [21 CFR 807.92(b)(3)]
The LF900 have similar intended use and technical characteristics to the predicate device.
Based on that information, we conclude that the differences between the proposed device (LF900) and predicate device (LF400) do not introduce a new intended use and do not raise new issues of safety and effectiveness.
5 / 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.