← Product Code [IRP](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IRP) · K210281

# Fist Assist Devices Model FA-1 (K210281)

_Fist Assist Devices, LLC · IRP · Jun 17, 2021 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IRP/K210281

## Device Facts

- **Applicant:** Fist Assist Devices, LLC
- **Product Code:** [IRP](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IRP.md)
- **Decision Date:** Jun 17, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5650
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The Fist Assist Model FA-1 is an arm air pressure massager intended to temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas.

## Device Story

Wearable, battery-operated, intermittent pneumatic compression device for arm use. Consists of control module with miniature pump and elastic cloth wrap with internal air bladder. User operates via single push-button switch. Device performs pre-programmed pressure cycles: inflates to 60 mmHg (held 20s), deflates to ≤10 mmHg (held 55s). Cycle repeats until manual power-off or automatic shut-off after 1 hour. Used over clothing in home or clinical settings. Provides mechanical massage to increase circulation and relieve muscle aches. Benefits patient through non-invasive, portable muscle therapy.

## Clinical Evidence

Bench testing only. Performance data includes ISTA 3A-2018 transportation testing, AAMI/ANSI ES 60601-1 electrical safety testing, IEC 60601-1-2 EMC evaluation, and mechanical functional testing. Software validation performed per FDA guidance. No clinical data provided.

## Technological Characteristics

Wearable intermittent pneumatic compression device. Materials: Molded ABS enclosure (94VB), elastic cloth wrap. Powered by (2) AA batteries. Sensing/actuation: Electronically controlled miniature air pump and valve system. Dimensions: 0.35 lbs. Connectivity: None. Sterilization: Non-sterile. Software: Pre-programmed fixed cycle logic.

## Regulatory Identification

A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.

## Predicate Devices

- NormaTec Pulse 2.0 and Pulse Pro 2.0 ([K183169](/device/K183169.md))
- Fig LLC, PowerPlay PPRT-01 ([K122154](/device/K122154.md))
- ManaMed Inc., PlasmaFlight ([K200351](/device/K200351.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 17, 2021

Fist Assist Devices, LLC Tej Singh CEO. Founder 25599 Fernhill Road Los Altos, California 94024

Re: K210281

Trade/Device Name: Fist Assist Devices Model FA-1 Regulation Number: 21 CFR 890.5650 Regulation Name: Powered Inflatable Tube Massager Regulatory Class: Class II Product Code: IRP Dated: March 19, 2021 Received: March 22, 2021

Dear Tej Singh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K210281

Device Name Fist Assist Devices Model FA-1

Indications for Use (Describe)

The Fist Assist Model FA-1 is an arm massager intended to temporarily relieve minor muscle aches and to temporarily increase circulation to the treated areas.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

[as required by 21 CFR 807.92(c)]

### Fist Assist Devices, Model FA-1

### 510(k) K210281

| DATE PREPARED:         | May 25, 2021                                                                                                                            |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT              | Fist Assist Devices, LLC.<br>25599 Fernhill Dr.<br>Los Altos Hills, CA 94024                                                            |
| CONTACT                | Tej Singh, MD, MBA<br>CEO and Founder<br>Fist Assist Devices, LLC<br>Email: fistassistdevices@gmail.com                                 |
| TRADE NAME:            | Fist Assist Devices Model FA-1                                                                                                          |
| DEVICE CLASSIFICATION: | Class II per 21 CFR 890.5650                                                                                                            |
| CLASSIFICATION NAME:   | Powered Inflatable Tube Massager                                                                                                        |
| PRODUCT CODE           | IRP                                                                                                                                     |
| PREDICATE DEVICES:     | 1. NormaTec Pulse 2.0 and Pulse Pro 2.0 (K183169)<br>2. Fig LLC, PowerPlay PPRT-01 (K122154)<br>3. ManaMed Inc., PlasmaFlight (K200351) |

#### INDICATION FOR USE:

The Fist Assist Model FA-1 is an arm air pressure massager intended to temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas.

### DEVICE DESCRIPTION:

The Fist Assist Model FA-1 is a wearable, non-sterile, battery operated, intermittent pneumatic compression device for use on the arm over clothing. It is composed of two major components that are permanently attached to one another:

- 1. the control module that contains the electronics with miniature pump and
- 2. the wrap made of elastic cloth that holds the control module and internal air bladder and uses hook tape for attachment to any part of the wrap exterior.

The control module and wrap are integral and permanently attached. It is powered by (2) AA batteries.

{4}------------------------------------------------

Fist Assist Devices, LLC Model FA-1

The Fist Assist Model FA-1 has a single operation mode. When powered on by depressing the single push button switch, it goes through pressure cycles wherein the air bladder inflates to 60 mmHg and is held at that pressure for 20 seconds; the air bladder is then deflated and a minimal pressure of 10 mmHg (or less) and is held for 55 seconds until the next inflation to 60 mmHg begins again. The cycle repeats continuously until the power is turned off using the single push button switch. The device will turn off automatically after 1 hour of continuous use. When wearing this device, the device should only be used over full clothing and never on direct skin.

#### COMPARISON WITH PREDICATE DEVICE:

The indications for use of the Model FA-1 and its primary predicate device is identical. Additionally, there are specific design features of the Model FA-1 device that are similar to the predicate devices. Table 5-1 below provides a summary of the comparison of the Model FA-1 to its predicates. Further information on predicate comparison is located in Section 12, Substantial Equivalence Discussion.

{5}------------------------------------------------

### Fist Assist Devices, LLC Model FA-1

|                                                          | Subject Device:<br>Fist Assist<br>Devices, LLC<br>Model FA-1                                                                                                                                                                | Primary Predicate<br>Device:<br>NormaTec Industries, LP<br>Pulse 2.0 and Pulse Pro<br>2.0                                                                                                                       | Predicate Device:<br>Fig, LLC<br>PowerPlay PPRT-01                                                                                                                                                                                                                                                                                                                                                    | Predicate Device<br>ManaMed, Inc.<br>PlasmaFlight                                                                                                                                                                                                                                                                                                                                         |
|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                                             | Fist Assist<br>Devices, LLC                                                                                                                                                                                                 | NormaTec Industries, LP                                                                                                                                                                                         | Fig, LLC                                                                                                                                                                                                                                                                                                                                                                                              | ManaMed, Inc                                                                                                                                                                                                                                                                                                                                                                              |
| 510k Number                                              | K210281                                                                                                                                                                                                                     | K183169                                                                                                                                                                                                         | K122154                                                                                                                                                                                                                                                                                                                                                                                               | K200351                                                                                                                                                                                                                                                                                                                                                                                   |
| Indications<br>for Use                                   | The Fist Assist,<br>Model FA-1 is an<br>arm air pressure<br>massager intended<br>to temporarily<br>relieve minor<br>muscle aches<br>and/or pains, and<br>to temporarily<br>increase<br>circulation to the<br>treated areas. | The NormaTec Pulse and<br>Pulse Pro are air pressure<br>massagers intended to<br>temporarily relieve minor<br>muscle aches and/or pains,<br>and to temporarily increase<br>circulation to the treated<br>areas. | The PowerPlay model PPRT-<br>01 is intended to be an over-<br>the-counter portable inflatable<br>tube massage system which<br>simulates kneading and<br>stroking of tissue with the<br>hands and by use of inflatable<br>pressure wraps. This device<br>can be used to: temporarily<br>increase blood circulation in<br>the treated areas; and<br>temporary relief of minor<br>muscle aches and pains | The PlasmaFlight is intended<br>to be an over-the-counter<br>portable inflatable tube<br>massage system which<br>simulates kneading and<br>stroking of tissue with the<br>hands and by use of inflatable<br>pressure wraps. This device<br>can be used to: temporarily<br>increase blood circulation in<br>the treated areas; and<br>temporary relief of minor<br>muscle aches and pains. |
| Technology                                               | A single bladder<br>wrap actuated by<br>an electronically<br>controlled small<br>internal air pump<br>and valve system.                                                                                                     | Compressor and valve<br>system that sequentially<br>inflates cells of appliance.<br>Bluetooth communication<br>ability.                                                                                         | Single bladder compression<br>wraps actuated by and<br>electronically controlled air<br>pump unit and solenoid valves                                                                                                                                                                                                                                                                                 | Airbladder encased in a<br>covering controlled by a<br>micropump controlled by<br>microprocessor                                                                                                                                                                                                                                                                                          |
|                                                          | Subject Device:<br>Fist Assist<br>Devices, LLC<br>Model FA-1                                                                                                                                                                | Primary Predicate<br>Device:<br>NormaTec Industries, LP<br>Pulse 2.0 and Pulse Pro<br>2.0                                                                                                                       | Predicate Device:<br>Fig, LLC<br>PowerPlay PPRT-01                                                                                                                                                                                                                                                                                                                                                    | Predicate Device<br>ManaMed, Inc.<br>PlasmaFlight                                                                                                                                                                                                                                                                                                                                         |
| Compliance<br>with<br>EMC/IEC<br>Voluntary<br>standards? | Yes, ES 60601-1,<br>IEC 60601-1-2,<br>IEC 60601-1-11                                                                                                                                                                        | Yes, ES 60601-1,<br>IEC 60601-1-2, IEC<br>60601-1-11                                                                                                                                                            | Information Not Available                                                                                                                                                                                                                                                                                                                                                                             | Compliant to Voluntary<br>Standards                                                                                                                                                                                                                                                                                                                                                       |
| Device<br>Pressure<br>Range                              | 10-60mmHg                                                                                                                                                                                                                   | 0-110 mmHg                                                                                                                                                                                                      | 30-70 mmHg                                                                                                                                                                                                                                                                                                                                                                                            | 0-55 mmHg                                                                                                                                                                                                                                                                                                                                                                                 |
| Weight                                                   | 0.35 pounds                                                                                                                                                                                                                 | 3.6 pounds                                                                                                                                                                                                      | 1.1 pounds                                                                                                                                                                                                                                                                                                                                                                                            | .51 pounds                                                                                                                                                                                                                                                                                                                                                                                |
| Housing<br>Materials                                     | Molded ABS<br>enclosure (94VB)                                                                                                                                                                                              | Molded ABS enclosure<br>(94V0)                                                                                                                                                                                  | Plastic Housing                                                                                                                                                                                                                                                                                                                                                                                       | Plastic Case                                                                                                                                                                                                                                                                                                                                                                              |
| Sterilization<br>Method                                  | Non-Sterile                                                                                                                                                                                                                 | Non-Sterile                                                                                                                                                                                                     | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                           | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                               |
| Max.<br>Inflation<br>Pressure                            | 60 mmHg,<br>preprogrammed                                                                                                                                                                                                   | 110mmHg, set by user                                                                                                                                                                                            | 70mmHg, set by user                                                                                                                                                                                                                                                                                                                                                                                   | 55 mmHg                                                                                                                                                                                                                                                                                                                                                                                   |
| Deflation<br>Hold Time                                   | 55 seconds                                                                                                                                                                                                                  | Default 30 sec, but can be<br>manually adjusted between<br>15 to 90 sec.                                                                                                                                        | Approximately 30 seconds for<br>each port used,<br>preprogrammed                                                                                                                                                                                                                                                                                                                                      | Information Not Available                                                                                                                                                                                                                                                                                                                                                                 |
| Prescription<br>Device                                   | No (OTC)                                                                                                                                                                                                                    | No (OTC)                                                                                                                                                                                                        | No (OTC)                                                                                                                                                                                                                                                                                                                                                                                              | No (OTC)                                                                                                                                                                                                                                                                                                                                                                                  |
|                                                          | Subject Device:<br>Fist Assist<br>Devices, LLC<br>Model FA-1                                                                                                                                                                | Primary Predicate Device:<br>NormaTec Industries, LP<br>Pulse 2.0 and Pulse Pro 2.0                                                                                                                             | Predicate Device:<br>Fig, LLC<br>PowerPlay PPRT-01                                                                                                                                                                                                                                                                                                                                                    | Predicate Device<br>ManaMed, Inc.<br>PlasmaFlight                                                                                                                                                                                                                                                                                                                                         |
| Pre-<br>Programmed<br>Treatment<br>time                  | 60-minute<br>preprogrammed<br>treatment. Cycle<br>can be stopped by<br>user.                                                                                                                                                | 10 minutes to 2h55 minute<br>treatment time set by user.<br>Cycle can be stopped by<br>user.                                                                                                                    | 20 minutes preprogrammed<br>per port, up to 60-minute total<br>treatment if all 3 ports are<br>used. Cycle can be stopped by<br>user.                                                                                                                                                                                                                                                                 | 60-minute preprogrammed<br>treatment.                                                                                                                                                                                                                                                                                                                                                     |

## Table 5-1 Comparison Summary of the Model FA-1 to Predicates

{6}------------------------------------------------

## Fist Assist Devices, LLC Model FA-1

{7}------------------------------------------------

## Fist Assist Devices, LLC Model FA-1

{8}------------------------------------------------

Fist Assist Devices, LLC Model FA-1

### NON-CLINICAL TESTING / PERFORMANCE DATA:

Non-clinical testing of the Model FA-1 was performed following applicable standards.

### TRANSPORTATION TESTING

Packaging testing of the Model FA-1 was performed to ISTA 3A-2018, General Simulation Performance Test Procedure for Parcel Delivery System 70kg (or 150lbs.) or less using the small packaging process sequence of testing. The results of test concluded that all units tested met the requirements of standard.

### BIOCOMPATIBILITY

The biocompatibility of the materials has not been verified by the FDA and contact of the cuffs/accessories to direct skin may lead to skin irritation, skin sensitization and/or cytotoxicity.

### SOFTWARE VALIDATION

Software validation was performed using the "General Principle of Software Validations: Final Guidance for Industry and FDA Staff, dated January 11, 2002". All tested variables were found to meet acceptance criteria established.

### ELECTRICAL TESTING

Electrical Testing was performed using the test standards AAMI/ANSI ES 60601-1:2005/(R)2012 And A1:2012: Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance. Testing concluded that the Device met all test requirements for the electrical safety.

### EMC EVALUATION

Electromagnetic Compatibility (EMC) testing of the Device was performed using the test standard IEC 60601-1-2:2014: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests. Testing concluded that the Device met all test requirements for EMC.

### BENCH TESTING

Mechanical testing was performed to confirm the functionality of key process parameters were performed. All parameters tested were found to meet acceptance criteria established.

### CONCLUSION:

The Model FA-1 is indicated for the identical intended use and similar technological characteristics such as components, design, materials, sterilization method, shelf life and operating principles as the predicate devices. Performance data demonstrates that the device functions as intended. Therefore, the Model FA-1 is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IRP/K210281](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IRP/K210281)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
