K203552 · Rapid Reboot Recovery Products, LLC · IRP · Dec 22, 2020 · Physical Medicine
Device Facts
Record ID
K203552
Device Name
Rapid Reboot
Applicant
Rapid Reboot Recovery Products, LLC
Product Code
IRP · Physical Medicine
Decision Date
Dec 22, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Rapid Reboot REGEN+, REGEN, or GENESIS Compression Therapy Systems are indicated for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health. The Rapid Reboot REGEN, or GENESIS Compression Therapy Systems simulate kneading and stroking of tissues by using an inflatable garment.
Device Story
Rapid Reboot REGEN+, REGEN, and GENESIS are powered inflatable tube massagers for home or clinical use. Devices consist of an air pump, hoses, and four-chambered inflatable sleeves. System operates by sequentially inflating and deflating air chambers to provide distal-to-proximal compression massage. Microprocessor-controlled software manages timing, pressure, and inflation cycles. Users interact via a 6-inch HD LED touchscreen or a Bluetooth-connected smartphone app to adjust intensity (0-200 mmHg), session duration, and massage mode. Powered by external AC supply or integrated rechargeable lithium-ion battery. Output is mechanical compression of limbs; intended to provide temporary relief of minor muscle aches and pains and increase local circulation. Benefits include portability and customizable therapy settings.
Clinical Evidence
No clinical data. Substantial equivalence is supported by benchtop performance testing, including usability studies, pressure/airflow testing, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), wireless coexistence (ANSI C63.27-2017), and biocompatibility (cytotoxicity, irritation, sensitization).
Technological Characteristics
Powered inflatable tube massager; 4-chambered garment; 12VDC power via IEC 60601-1 supply or integrated Li-ion battery; 6" HD LED touchscreen interface; BLE connectivity; ABS/carbon fiber housing. Software-controlled sequential gradient/peristaltic inflation. Pressure range 0-200 mmHg. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and ANSI C63.27-2017.
Indications for Use
Indicated for healthy individuals seeking temporary relief of minor muscle aches and pains and temporary increase in circulation to treated areas via simulated kneading and stroking of tissues using an inflatable garment.
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 22, 2020
Rapid Reboot Recovery Products, LLC % Prithul Bom Regulatory Technology Services, LLC 1000 Westgate Drive. Suite 510k Saint Paul, Minnesota 55114
Re: K203552
Trade/Device Name: Rapid Reboot Regulation Number: 21 CFR 890.5650 Regulation Name: Powered inflatable tube massager Regulatory Class: Class II Product Code: IRP Dated: December 2, 2020 Received: December 4, 2020
Dear Prithul Bom:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Heather Dean, PhD Assistant Director. Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K203552
Device Name Rapid Reboot
Indications for Use (Describe)
The Rapid Reboot REGEN+, REGEN, or GENESIS Compression Therapy Systems are indicated for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health. The Rapid Reboot REGEN, or GENESIS Compression Therapy Systems simulate kneading and stroking of tissues by using an inflatable garment.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------|
| <div> <input type="checkbox"/> </div> | <div> <input type="checkbox"/> </div> |
| | Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:
l. SUBMITTER Rapid Reboot Recovery Products, LLC 1396 W 200 S, Building 2, Unit A Lindon, Utah 84042 Tel: +1.801.704.5271 ext. 405
Contact Person: David Johnson, CEO Date Prepared: October 7, 2020
11. DEVICE
Name of Device: Rapid Reboot REGEN+ Rapid Reboot REGEN Rapid Reboot GENESIS Classification Name: Powered Inflatable Tube Massager Regulation: 21 CFR §890.5650 Regulatory Class: Class II Product Classification Code: IRP
#### III. PREDICATE AND REFERENCE DEVICE
| Predicate Manufacturer: | Rapid Reboot Recovery Products, LLC |
|-------------------------|-----------------------------------------|
| Predicate Trade Name: | Rapid Reboot Compression Therapy System |
| Predicate 510(k): | K182668 |
| Reference Manufacturer: | NormaTec Industries, LP |
| Reference Trade Name: | NormaTec Pulse 2.0 and Pulse Pro 2.0 |
| Reference 510(k): | K183169 |
#### DEVICE DESCRIPTION IV.
The Rapid Reboot GENESIS, REGEN, and REGEN+ model systems are powered inflatable tube massagers (Product Code IRP). They are indicated for the temporary relief of minor muscle aches and pains and for the temporary increase in circulation to the treated areas in people who are in good health. They simulate kneading and stroking of tissues by using an inflatable garment. The air pump is connected to the dedicated sleeves via a series of hoses, and each sleeve has four (4) compression chambers. The compression massage direction is from limb end to body center (distal to proximal). By inflating the air chambers sequentially and then deflating as one cycle, the pressure can be adjusted to avoid any discomfort to the user. The sleeve works under the action of sensors and microprocessors. Software controls the timing and pressure reflected by the sensor, cycling airflow into and out of the sleeves to compress the body in a controlled and specific manner. Each unit also has a user interface that allows users the ability to control several aspects of the massage: 1.e., intensity (pressure), session duration, and mode. The devices are powered by an external IEC 60601-1 compliant power supply and can also be powered by an internal IEC 62133-compliant lithium-ion battery.
The user interface on the GENESIS, REGEN, and REGEN+ models is a 6" HD LED touchscreen with capacitive sensors identical to most smartphones. The user interface provides for:
{4}------------------------------------------------
- -Starting and stopping the massage treatment;
- Adjusting the time, intensity (pressure), and type of distal-to-proximal sequence (Mode). -
While the main functions, indications for use, and parameters of settings are the same on the three models, there are differences in non-crucial features and resolution of settings that are intended for marketing differentiation and user preferences: e.g., the REGEN and REGEN+ models offer pressure resolution of 5 mmHg from 0 to 200 mmHg for a total of 40 potential pressure settings, while the GENESIS only offers 7 pressure settings. Similarly, the REGEN and REGEN+ allow the user to set the session duration, or time, to the minute between 1 and 179 minutes (2 hrs and 59 minutes), the GENESIS only allows the user to set a session duration of 10, 20, 30, 40, 50, or 60 minutes.
In addition to the user interface on these respective devices, these proposed models have Bluetooth Low Energy (BLE) capability that allows the use of a Rapid Reboot app to control a device. The app mimics the device interface graphics and buttons, allowing the user to use a compatible Android or iOS powered smartphone or tablet to control the devices core functions just as if they were using the interface on the device. The app functionality is limited to mimicking the device interface and does not provide additional settings that alter the therapy provided by the device. When paired with a REGEN or REGEN+, the app does offer additional, non-critical and non-function related for marketing differentiation and user preferences: e.g., informational features in the menu, programs (i.e., saved settings) that can be saved and quickly applied to future sessions, and the ability to access logs showing usage.
#### V. INDICATIONS FOR USE
The Rapid Reboot REGEN+, REGEN, or GENESIS Compression Therapy Systems are indicated for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health. The Rapid Reboot REGEN, or GENESIS Compression Therapy Systems simulate kneading and stroking of tissues by using an inflatable garment.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVCE
The following characteristics were compared between the subject device and the predicate device in order to demonstrate substantial equivalence:
- Indications for Use The predicate and subject device have identical indications for use. ● Both are indicated for temporary relief of minor aches and pains and for temporary increase in circulation to treated areas in people who are in good health. They both simulate kneading and stroking of tissues by using an inflatable garment.
- . Materials - The predicate and subject device are made of similar patient-contacting materials. The exact materials are different between the subject device and predicate device but biocompatibility testing including cytotoxicity, sensitization, and irritation testing has been completed to demonstrate that the new material is as safe as the predicate device.
- Design – The predicate and subject device have the same fundamental design of an external automated air pump, wearable garments, and software controls. The subject device introduces a new software user interface that has been validated to perform as well as the predicate device.
- Energy Source - The predicate device uses a direct wall outlet plug to power the system. The subject device is powered by 12VDC via an IEC 60601-1 compliant power supply (100-240VAC input) Integrated rechargeable battery. Electrical Safety and EMC testing was conducted to ensure that the change in energy source is as safe and effective as the predicate.
{5}------------------------------------------------
- . Other Design Features – The proposed models include a new housing design from the predicate to make them smaller and more portable as well as accommodate the integrated battery and 6" LCD touchscreen.
- Performance Testing The predicate and subject device involved the following testing: .
- Human Factors Engineering Usability Testing Report o
- Minimum, Interval, and Maximum Air Pressure Test o
- Maximum Electric Current Value Test o
- o Maximum Airflow Valve Test
- o Noise Level Test
- 0
- Button and Display Test O
- o
| | Predicate Device<br>(K182668) | Reference Devices<br>(K183169) | Proposed New<br>Devices | Comment |
|---------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model Name<br>510(k) Number | Rapid Reboot<br>Compression Therapy<br>System | NormaTec Pulse 2.0<br>and Pulse Pro 2.0 | Rapid Reboot<br>REGEN+, REGEN,<br>and GENESIS<br>510(k): TBD | N/A |
| Manufacturer | Rapid Reboot<br>Recovery Products,<br>LLC | NormaTec Industries,<br>LP | Rapid Reboot<br>Recovery Products,<br>LLC | N/A |
| Prescriptive | No, OTC | No, OTC | Same (OTC) | N/A |
| Indications for Use | The Rapid Reboot<br>Compression Therapy<br>System is indicated<br>for the temporary<br>relief of minor muscle<br>aches and pains and<br>for the temporary<br>increase in circulation<br>to the treated areas in<br>people who are in<br>good health. It<br>simulates kneading<br>and stroking of tissues<br>by using an inflatable<br>garment. | The NormaTec Pulse<br>2.0 and Pulse Pro 2.0<br>are air pressure<br>massagers intended to<br>temporarily relieve<br>minor muscles aches<br>and/or pains, and to<br>temporarily increase<br>circulation to the<br>treated areas | Same as predicate | N/A |
| Intended Use<br>Environment | Clinics, hospital,<br>athlete training, and<br>home environments | Clinics, hospital,<br>athlete training, and<br>home environments | Same | N/A |
| Power Sources | IEC 60601-1<br>compliant integrated<br>power panel with<br>110V direct wall<br>outlet plug | 15 VDC via an IEC<br>60601-1 compliant<br>power supply (100-<br>240VAC input)<br>Integrated<br>rechargeable battery | 12VDC via an IEC<br>60601-1 compliant<br>power supply (100-<br>240VAC input)<br>Integrated<br>rechargeable battery | N/A |
| Software/Firmware<br>Micro-processor<br>Control | PCB | Microprocessor | Microprocessor | N/A |
| Technology | Compressor and valve<br>system that<br>sequentially inflates<br>cells of Attachment. | Compressor and valve<br>system that<br>sequentially inflates<br>cells of Attachment.<br>Bluetooth<br>communication ability | Same as reference | Proposed<br>devices<br>incorporate an<br>FCC certified<br>BLE module<br>that allow using<br>the Rapid<br>Reboot app as<br>an additional<br>interface |
| Compliance with<br>voluntary<br>standards | ES 60601-1, IEC<br>60601-1-2, IEC<br>60601-1-11 | ES 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>ANSI C63.27-2017 | Same as reference | The proposed<br>devices comply<br>with one<br>additional<br>standard: ANSI<br>C63.27-2017 |
| Device Pressure<br>Range | 0-200 mmHg | 0-110 mmHg | Same as predicate | N/A |
| Intensity Control<br>per Chamber | No | Pulse 2.0: Yes<br>(limited range)<br>Pulse Pro 2.0: Yes | GENESIS: Yes<br>(limited range)<br>REGEN: Yes<br>REGEN+: Yes | N/A |
| Treatment Time | 10, 20, 30 minute<br>time settings | Stays on until the user<br>turns it off or can be<br>set up to turn off in a<br>range of 10 minutes to<br>continuous | GENESIS: 10-60<br>minutes by 10 minute<br>increments<br>REGEN: 1-179<br>minutes by 1 minute<br>increments<br>REGEN+: 1-179<br>minutes by 1 minute<br>increments | N/A |
| Inflation/Deflation<br>Cycle Type | Sequential gradient,<br>peristaltic | Sequential gradient,<br>peristaltic and pulsing | Same as reference | N/A |
| Attachment<br>Contact Surface<br>Material | Interior: 70 denier<br>with thermoplastic<br>polyurethane<br>laminate/extrusion<br>Exterior: 210 denier<br>with thermoplastic<br>polyurethane<br>laminate/extrusion<br>and rip-stop weave | 200 denier nylon with<br>a polyurethane<br>laminate/extrusion | Same as predicate | N/A |
| Number of<br>Inflatable<br>Attachment<br>Chambers | 4 | 5 or less | Same as predicate | N/A |
| Weight | 5.2 lbs | 3.6 lbs (incl. battery) | GENESIS: 4.0 lbs<br>(incl. battery)<br>REGEN: 4.4 lbs (incl.<br>battery)<br>REGEN+: 4.8 lbs<br>(incl. battery) | Differences in<br>weight reflect<br>number of<br>battery packs.<br>GENESIS has<br>one integrated<br>battery pack,<br>REGEN has<br>two, and<br>REGEN+ has<br>three. |
| Dimensions<br>(W x H x D) | 6.5" x 5" x 10" | 4.4" x 3.8" x 8.1" | 5" x 4.5" x 9.5" | The proposed<br>models include<br>a new housing<br>design from the<br>predicate to<br>make them<br>smaller and<br>more portable,<br>as well as<br>accommodate<br>modification<br>such as an<br>integrated<br>battery and 6"<br>HD LCD<br>touchscreen |
| Housing Materials<br>and Constructions | Molded ABS<br>enclosure | Molded ABS<br>enclosure | GENESIS: Same<br>REGEN: Same<br>REGEN+: Molded<br>ABS and carbon fiber<br>enclosure | The GENESIS<br>and<br>REGEN control<br>unit housings<br>are comprised<br>of ABS molded<br>parts. The<br>REGEN+<br>control unit<br>housing is<br>comprised of a<br>combination of<br>molded ABS<br>and carbon fiber<br>parts for<br>marketing<br>differentiation. |
| Interface | Segment LED<br>capacitive touch<br>display | Pulse 2.0: 4.3" color<br>LCD screen with<br>smart switches<br>Pulse Pro 2.0: 4.3"<br>color TFT Screen<br>with capacitive sensor | GENESIS, REGEN,<br>& REGEN+: 6" color<br>HD LCD touchscreen<br>(1440p X 720p) | N/A |
| Patient Contact | Non-conductive<br>attachments | Non-conductive<br>attachments | Same | N/A |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## Sterilization & Shelf-life Testing
The device is shelf-stable and not sterile, therefore sterilization and shelf-life testing was not necessary to demonstrate the safety or performance of the device.
### Biocompatibility Testing
Cytotoxicity, Irritation, and Sensitization testing was conducted to demonstrate the biocompatibility of the patient-contacting materials of the Rapid Reboot devices.
### Electrical safety and electromagnetic compatibility (EMC)
The subject device passed all electrical safety and EMC tests. Electrical Safety Testing was conducted in accordance with IEC 60601-1 Medical Electrical Equipment - General Requirements for Basic Safety and Essential Performance and ANSI IEEE C63.27-2017 American National Standard for Evaluation of Wireless Coexistence. EMC Testing was conducted in accordance with IEC 60601-1-2 General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests.
# Software Verification and Validation Testing
Software Verification and Validation Testing was conducted in accordance with IEC 62304 Medical Device Software - Software Life Cycle Processes.
# Mechanical and acoustic Testing
Mechanical Stress Testing was conducted as described above in the performance testing.
# Animal Study
Animal performance testing was not required to demonstrate safety and effectiveness of the device.
### Clinical Studies
Clinical testing was not required to demonstrate the safety and effectiveness of the Rapid Reboot devices. Instead, substantial equivalence is based upon benchtop performance testing.
### Usabilitv Studies
A Usability Study was conducted to ensure that intended users are able to properly use the device as indicated to fulfil its intended use.
# VIII. CONCLUSIONS
Rapid Reboot REGEN+, REGEN, and Genesis are substantially equivalent to the legally marketed Rapid Reboot Compression Therapy System (primary predicate) for Indication for Use, and technological and performance characteristics. Based on the Safety and Effectiveness test reports it is at least as safe and effective as the predicate device and technologically comparable. It does not raise any new safety and/or effectiveness concerns. Hence, it is clear that Rapid Reboot is substantially equivalent to that of the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.