K191862 · Guangzhou Longest Science & Technology Co., Ltd. · IRP · Oct 29, 2019 · Physical Medicine
Device Facts
Record ID
K191862
Device Name
Compression Therapy Device Model LGT-2200SP
Applicant
Guangzhou Longest Science & Technology Co., Ltd.
Product Code
IRP · Physical Medicine
Decision Date
Oct 29, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Compression Therapy Device LGT-2200SP is intended for the temporary relief of minor muscle aches and pains and for the temporary increase in circulation to the treated areas, it can simulate kneading and stroking of tissues by using an inflatable garment.
Device Story
Powered inflatable tube massager; comprises air compressor, intermittent pneumatic controller, 4-chamber sleeves (Polyether Nylon Fabric), and connecting hoses. Operates via sequential inflation of sleeve chambers to simulate manual kneading/stroking. Powered by AC or internal lithium-ion battery. Used by adults in home or clinical settings. User selects treatment mode (Sequential, Squeeze, Flush, Flow, Release, Wave) and time (1-99 min) via controller. Output is mechanical pressure (30-150 mmHg) applied to limbs/torso. Benefits include temporary muscle ache relief and increased local circulation.
Clinical Evidence
No clinical data. Safety and performance verified via bench testing, including electrical safety (ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-11), battery safety (IEC 62133-2), and biocompatibility (ISO 10993-5, ISO 10993-10).
Indicated for adults in good health for temporary relief of minor muscle aches and pains and temporary increase in circulation to treated areas (arm, leg, waist, hip, thigh, calf, foot).
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square are the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" below it.
October 29, 2019
Guangzhou Longest Science & Technology Co., Ltd. % Jet Li Regulation Manager Guangzhou KEDA Biological Tech Co., Ltd. 6F, No.1 TianTai Road, Science City, LuoGang District Guangzhou, CN
Re: K191862
Trade/Device Name: Compression Therapy Device Model LGT-2200SP Regulation Number: 21 CFR 890.5650 Regulation Name: Powered Inflatable Tube Massager Regulatory Class: Class II Product Code: IRP Dated: September 1, 2019 Received: September 5, 2019
Dear Jet Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Vivek Pinto, Ph.D. Director (Acting) DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K191862
Device Name Compression Therapy Device Model: LGT-2200SP
#### Indications for Use (Describe)
Compression Therapy Device LGT-2200SP is intended for the temporary relief of minor muscle aches and for the temporary increase in circulation to the treated areas, it can simulate kneading and stroking of tissues by using an inflatable garment.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
EF
PSC Publishing Services (301) 443-6740
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 890.5650, and there were no prior submissions for the subject device.
#### 1. Submitter Information
Sponsor: GUANGZHOU LONGEST SCIENCE&TECHNOLOGY CO., LTD.
Address: 5&6f, BuildingB4, No.11, Kaiyuan Avenue, Science City Guangzhou Hi-Tech Industrial
Development Zone, Guangzhou Guangdong, CHINA 510530
Contact Person: Xiaobing Luo
Title: Deputy General Manager
Phone: +86-020-66353999 Fax: +86-020-66353920
E-mail: gzlongest@126.com
Application Correspondent: Jet Li Company: Guangzhou KEDA Biological Technology Co., Ltd E-mail: med-jl @foxmail.com Phone: 86-18588874857 Address: 6F, No.1 TianTairoad, Science City, LuoGang District, GuangZhou City, China
#### 2. Subject Device Information
Type of 510(k) submission: Traditional Common Name: Compression Therapy Device Model: LGT-2200SP Classification Name: Powered inflatable tube massager Review Panel: Physical Medicine Product Code: IRP Regulation Number: 21 CFR 890.5650
{4}------------------------------------------------
Regulation Class: 2
- 3. Predicate and Reference Device Information 1) Primary Predicate Device Sponsor: NormaTec Industries Trade/Device Name: NormaTec Pulse and NormaTec Pulse Pro 510K number: K160608 Regulation Number: 21 CFR 890.5650 2) Secondary Predicate Device Sponsor: Rapid Reboot Recovery Products, LLC Device Name: Rapid Reboot Compression Therapy System 510(k) number: K182668 Regulation Number: 21 CFR 890.5650 3) Reference Device Sponsor: XIAMEN SENYANG CO., LTD. Device Name: Pressure Therapy System, model: Pt 1002 510(k) number: K161907 Regulation Number: 21 CFR 890.5650
#### 4. Device Description
Compression therapy device LGT-2200SP is a powered inflatable tube massager, comprised of an air compressor with intermittent pneumatic controller, 4-chamber sleeves covered with Polyether Nylon Fabric, and a connectable hose for connecting the device to the sleeves. The case of main unit body is made from ABS plastic
The device can be powered by an external power supply, or be powered by an internal lithium ion battery. The sleeves have 4-chambers and different applicable body areas, such as the Arm' Leg/ Waist, hip and thigh/ Calf and foot. The sleeves can be inflating sequentially to apply the circulating pressure on the target body areas which controlled by the main unit.
The device simulates manual kneading and stroking of tissues by sequential inflated sleeves, to temporarily relieve minor muscle aches and/or pains and to temporarily increase circulation to the treated areas. It is to be used by adults who are in good health.
#### 5. Intended Use/Indication for use
{5}------------------------------------------------
Compression Therapy Device LGT-2200SP is intended for the temporary relief of minor muscle aches and pains and for the temporary increase in circulation to the treated areas, it can simulate kneading and stroking of tissues by using an inflatable garment.
#### Test Summary 6.
Compression Therapy Device has been evaluated the safety and performance by lab bench testing according to the following standards:
- മ ANSI/AAMI ES60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety, 2005+A1:2012
- IEC 60601-1-2, Medical electrical equipment Part 1-2: General requirements for basic ಡ safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests, 2014
- IEC 60601-1-11, Medical electrical equipment Part 1-11: General requirements for basic x safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment, Edition 2.0 2015-01
- IEC 62133-2:2017, Secondary cells and batteries containing alkaline or other non-acid 区 electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications- Part 2: Lithium systems
- x ISO 10993-5: 2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10: 2010 Biological evaluation of medical devices Part 10: Tests for irritation and X skin sensitization
- Guidance for the Content of Premarket Submissions for Software Contained in Medical x Devices.
## 7. Comparison to Predicate Device
Compare with predicate device, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between subject device and predicate device do not raise any new questions of safety or effectiveness. See the below form.
{6}------------------------------------------------
| Elements of | | Predicate Device | Predicate Device | Predicate Device | |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|
| comparison | Subject Device | (Primary) | (II) | (III) | Verdict |
| Manufacturer | GUANGZHOU LONGEST<br>SCIENCE&TECHNOLOGY CO.,<br>LTD. | NormaTec Industries | Rapid Reboot Recovery Products,<br>LLC | XIAMEN SENYANG CO., LTD. | -- |
| 510K number | TBD | K160608 | K182668 | K161907 | -- |
| Product Name | Compression Therapy Device<br>LGT-2200SP | NormaTec Pulse and NormaTec<br>Pulse Pro | Rapid Reboot Compression<br>Therapy System | Pressure Therapy System, Pt<br>1002 | -- |
| Classification<br>Name | Pow ered inflatable tube<br>massager | Pow ered inflatable tube massager | Pow ered inflatable tube massager | Pow ered inflatable tube massager | SE |
| Regulation<br>Class | 2 | 2 | 2 | 2 | SE |
| Regulation<br>Number | 21 CFR 890.5650 | 21 CFR 890.5650 | 21 CFR 890.5650 | 21 CFR 890.5650 | SE |
| OTC & Rx | OTC | OTC | OTC | Rx | -- |
| | | Indications for Use | | | |
| Indications for<br>Use | Compression Therapy Device<br>LGT-2200SP is intended for the<br>temporary relief of minor muscle<br>aches and pains and for the<br>temporary increase in circulation<br>to the treated areas, it can<br>simulate kneading and stroking<br>of tissues by using an inflatable<br>garment. | The NormaTec Pulse and Pulse<br>Pro is an air pressure massager<br>intended to temporarily relieve<br>minor muscle aches and/or pains,<br>and to temporarily increase<br>circulation to the treated areas. | The Rapid Reboot Compression<br>Therapy System is intended for<br>the temporary relief of minor<br>muscle aches and pains and for<br>the temporary increase in<br>circulation to the treated areas in<br>people who are in good health.<br>The Rapid Reboot Compression<br>Therapy System simulates<br>kneading and stroking of tissues | The device is indicated for use by<br>medical professionals and patient<br>at home, who are under medical<br>supervision, in treating many<br>conditions, such as: Primary<br>lymphedema, edema follow ing<br>trauma and sport injuries, post<br>immobilization edema, venous<br>insufficiencies, lymphedema. | Minor<br>differen<br>t<br>Note 1 |
| Elements of<br>comparison | Subject Device | Predicate Device<br>(Primary) | Predicate Device<br>(II) | Predicate Device<br>(III) | Verdict |
| | | | by using an inflatable garment. | | |
| | | Device design | | | |
| Power Source | AC100-240V, 50/60Hz<br>Battery 11.1V, 6500mAh meet<br>IEC 62133 | 12 VDC via an IEC 60601-1<br>compliant pow er supply (100-240<br>VAC input)<br>Optional Integrated rechargeable<br>battery | 110 V, 60Hz | 110 V, 60Hz | Minor<br>different<br>t<br>Note 2 |
| Power<br>consumption | 90VA | 14W | 30W | 30W | Minor<br>different<br>t<br>Note 2 |
| Dimensions<br>(W*H*D) | 270*148*129mm | 4"*5" *9" | 10" x 6.5" x 5" | 260*170*130mm | Minor<br>different<br>t<br>Note 2 |
| Elements of<br>comparison | Subject Device | Predicate Device<br>(Primary) | Predicate Device<br>(II) | Predicate Device<br>(III) | Verdict |
| Photo | Image: Subject Device | Image: Predicate Device (Primary) | Image: Predicate Device (II) | Image: Predicate Device (III) | Minor<br>different<br>Note 2 |
| Weight | 2.0 Kg(4.4pounds) | 3.6 pounds | 5.8 pounds | 5.8 pounds | Minor<br>different<br>Note 2 |
| Housing<br>Materials | Molded ABS enclosure | Molded ABS enclosure | Molded ABS enclosure | Molded ABS enclosure | SE |
| Sleeves | 4-chamber Arm Sleeve:<br>97.0*28.7 cm | Leg: | Leg:<br>X-Short: 14" x 41"<br>Short: 14" x 43"<br>Medium: 14" x 45" | Small Leg Cuff:90X30cm<br>Large Leg Cuff:110X30cm | Minor<br>different |
| Elements of<br>comparison | Subject Device | Predicate Device<br>(Primary) | Predicate Device<br>(II) | Predicate Device<br>(III) | Verdict |
| | 4-chamber Leg Sleeve:<br>S: $94.2*35.85$ cm<br>M: $106.2*39.0$ cm<br>L: $116.0*44.8$ cm<br><br>Image: 4-chamber Leg Sleeve | Image: Normatec Arm Sleeve | Long: 14" x 48"<br>X-Long: 14" x 52"<br><br>Image: Rapid Reboot Leg Sleeve | Image: Blue Sleeve | Note 3 |
| | 4-chamber Waist, Hip and Thigh<br>Sleeve: $71.3*71.8$ cm<br><br>Image: 4-chamber Waist, Hip and Thigh Sleeve | Hip:<br><br>Image: Normatec Hip Sleeve | Hip:<br>Regular: 26" x 32"<br>Large: 26" 35"<br><br>Image: Rapid Reboot Hip Sleeve | | |
| | 4-chamber Calf and Foot<br>Sleeve: $61.0*30.1$ cm<br><br>Image: 4-chamber Calf and Foot Sleeve | | Arm:<br>Regular: 18" x 38"<br>Long: 18" x 44"<br><br>Image: Rapid Reboot Arm Sleeve | | |
| Elements of | | Predicate Device | Predicate Device | Predicate Device | |
| comparison | Subject Device | (Primary) | (II) | (III) | Verdict |
| | Image: Subject Device | | Image: Predicate Device (II) | | |
| Number of<br>Chambers | 4-chamber | 5 or less | 4 Chambers | 4 Chambers for each unit | SE |
| Sleeve<br>Materials | Polyether Nylon Fabric | 200 denier nylon with a<br>polyurethane laminate/extrusion | Nylon with a Polyurethane<br>laminate | Thermoplastic Urethane | Minor<br>different<br>Note 3 |
| Mode of<br>Compression | Sequential | Sequential | Sequential | Sequential | SE |
| Device<br>Pressure<br>range | 30-150mmHg | 30-110 mmHg | 0-200 mmHg | 0-250mmHg | Minor<br>different<br>Note 4 |
| Elements of<br>comparison | Subject Device | Predicate Device<br>(Primary) | Predicate Device<br>(II) | Predicate Device<br>(III) | Verdict |
| Treatment<br>Time | 1min-99min, default as 30min,<br>step of 5min | Stays on until the user turns it off<br>or can be set up to turn off in a<br>range of 10mins to continuous | User determines therapy time.<br>Choose from 10, 20, or 30 minute<br>session time, with option to add<br>additional 10 minutes to any<br>therapy time. | 0 - 30mins | Minor<br>different<br>t<br>Note 5 |
| Work Mode | Image: Numbered boxes 1-4<br><br>SEQUENTIAL: ①→②→③→④<br>SQUEEZE: ①②→③④<br>FLUSH: ①→①②→①②③→①<br>②③④<br>FLOW: ④→③→②→①<br>RELEASE: ③④→①②<br>WAVE: SEQUENTIAL+<br>SEQUENTIAL+FLOW | Sequential<br>Image: Diagram of NormaTec Pulse<br>Image: Diagram of Zone pulses | Model A<br>Image: Model A<br>Model B<br>Image: Model B | Model A<br>Image: Model A<br>Model B<br>Image: Model B | Minor<br>different<br>t<br>Note 6 |
| Elements of<br>comparison | Subject Device | Predicate Device<br>(Primary) | Predicate Device<br>(II) | Predicate Device<br>(III) | Verdict |
| Patient contact | Non-conductive appliances | Non-conductive appliances | Image: Predicate Device II | Image: Predicate Device III | SE |
| Software/Firm<br>ware/Micropro<br>cessor Control | Microprocessor | Microprocessor | Microprocessor | Microprocessor | SE |
| Technology | Compressor and valve<br>system w hich sequentially<br>inflates cells of appliance | Compressor and valve system<br>w hich sequentially inflates cells of<br>appliance | Compressor and valve<br>system w hich sequentially<br>inflates cells of appliance | Compressor and valve<br>system w hich sequentially<br>inflates cells of appliance | SE |
| FDA-Recognized Standards | | | | | |
| Electrical<br>safety, EMC | ANSI/AAMI ES60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 62133<br>ISO 10993-5<br>ISO 10993-10 | ES 60601-1<br>IEC 60601-1-2;<br>IEC 60601-1-11 | IEC 60601-1:2014;<br>IEC 60601-1-2:2014;<br>EN ISO 10993-5:2009<br>EN ISO 10993-10:2010 | IEC 60601-1<br>IEC 60601-1-2<br>ISO10993-5<br>ISO10993-10 | Minor<br>different<br>t<br>Note 7 |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
#### Note 1
Although there is a little difference about the cation for use, the meanings are the same. This difference does not affect he safety and effectiveness.
## Note 2
Although the subject device design and specification parameter between the predicate are different, they are both complied with ANSVAAM ES60601 -1. So the minor differences on such parance) do not affect the safety and effectiveness.
### Note 3
Although the device provides four kind sleeves which have minor difference about size and appearance, the sleeve chamber number are the same and the applicable body areas are included in the predical of sleeves is also complied with ISO 1099-5 and ISO 10993-10. Therefore, the difference would not raise adversely impact on safety and effectiveness.
### Note 4
Although the pressure range of subject devices, the range is smaller and contained by the predicate device (1), sothe difference of pressure range would not raise adversely impact on safety and effectiveness.
### Note 5
Although the treatment time range of subject device is to 99mins, which seems to be longer than the predicated with is suitable for daily use. And the treatment time is optionally set by userwith 5mins step, which is more exact to the difference of pressure range would not raise adversely impact on safety and effectiveness.
{14}------------------------------------------------
## Note 6
Although the subject device provides six kinds of work mode, the SEQUENTIAL and FLUSH mode are the same with the other work modes of subject device just have different about the different thambers. The treatment pressure range are the same under different work modes, so the difference of pressure range would not raise adversely impact on safety and effectiveness.
## Note 7
The subject device references more sufficient safety standards than the safety of the safety of the device is verified via tests.
## 7. Summary for clinical test
Clinical performance is not deemed necessary.
### 8. Conclusion
The subject device Compression Therapy Device has all features for intended use. Thus, the subject device is substantially equivalent to the predicate devices.
### 9. Summary Prepared Date
21 Oct. 2019
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.