K201285 · InMode , Ltd. · IPF · Mar 5, 2021 · Physical Medicine
Device Facts
Record ID
K201285
Device Name
InMode Evolve System with the Tone Applicator
Applicant
InMode , Ltd.
Product Code
IPF · Physical Medicine
Decision Date
Mar 5, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EVOLVE System with the Tone Applicator is an electro-muscle and transcutaneous nerve stimulation device for the treatment of different body areas. The EVOLVE System with Tone Applicator is designed to operate in two modes - EMS and TENS. In EMS mode it is used for: Relaxation of muscle spasms; Prevention or retardation of disuse atrophy; Increasing local blood circulation; Muscle re-education; Maintaining or increasing range of motion; Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis. And in TENS mode is intended for: Symptomatic relief and management of chronic, intractable pain; Post-surgical acute pain; Post-trauma acute pain.
Device Story
Computerized electro-muscle and transcutaneous nerve stimulation device; console with LCD touch screen, power supply, and controller; connects to up to four Tone Applicators via cables. Operates in EMS or TENS modes; delivers symmetrical biphasic rectangular electrical pulses to body areas. Used in clinical settings by healthcare providers. Output intensity, pulse width, and frequency are adjustable via the user interface. Benefits include muscle re-education, pain relief, and prevention of venous thrombosis post-surgery. Safety features include automatic overload/no-load trips, patient override control, and automatic shut-off.
Clinical Evidence
Bench testing only. No clinical data provided. Testing validated output waveforms, basic unit characteristics, and compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-10 standards. Biocompatibility confirmed per ISO 10993-1.
Technological Characteristics
Powered muscle/nerve stimulator; 100-240 VAC input; 48V AC/DC power supply. Symmetrical biphasic rectangular waveform; 3-200 Hz frequency; 20-400 μs pulse width; up to 54 Vpeak intensity. Housing: PC Makrolon 2458. Connectivity: Wired applicators. Standards: ANSI AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 60601-1-6.
Indications for Use
Indicated for patients requiring muscle stimulation (spasm relaxation, atrophy prevention, circulation, re-education, range of motion, post-surgical calf stimulation) or pain management (chronic intractable, post-surgical, or post-trauma acute pain).
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 5, 2021
InMode Ltd. % Amit Goren Regulatory Manager A. Stein - Regulatory Affairs Consulting Ltd. 20 Hata'as Str., Suite 102 Kfar Saba, 4442520 Israel
Re: K201285
Trade/Device Name: Evolve System with the Tone Applicator Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF, GZJ Dated: May 10, 2020 Received: May 13, 2020
Dear Amit Goren:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jitendra Virani Acting Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K201285
#### Device Name
EVOLVE System with the Tone Applicator
#### Indications for Use (Describe)
The EVOLVE System with the Tone Applicator is an electro-muscle and transcutaneous nerve stimulation device for the treatment of different body areas.
The EVOLVE System with Tone Applicator is designed to operate in two modes - EMS and TENS.
In EMS mode it is used for:
- Relaxation of muscle spasms
- · Prevention or retardation of disuse atrophy
- Increasing local blood circulation
- · Muscle re-education
- · Maintaining or increasing range of motion
- · Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
#### And in TENS mode is intended for
- · Symptomatic relief and management of chronic, intractable pain
- · Post-surgical acute pain
- Post-trauma acute pain
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(K) SUMMARY
### THE EVOLVE SYSTEM WITH THE TONE APPLICATOR
# 510(k) Number K201285
#### Applicant Name: Company Name: InMode Ltd. Tabor Building, Shaar Yokneam Address: Yokneam 20692 Israel Tel: +972-4-9097470 Fax: +972-4-9097471 E-mail: amit(@asteinrac.com Contact Person: Official Correspondent: Amit Goren A. Stein - Regulatory Affairs Consulting Ltd. Company Name: Address: 20 Hata'as Str., Suite 102 Kfar Saba 4442520 Israel Tel: + 972-9-7670002 Fax: +972-9-7668534 E-mail: amit(@asteinrac.com Date Prepared: February 11, 2021 Trade Name: EVOLVE System with the Tone Applicator Powered muscle stimulator, Transcutaneous electrical Classification Name: nerve stimulator for pain relief Regulation No .: 890.5850, 882.5890 Product Codes: IPF, GZJ Classification: Class II Medical Device
### Predicate Device:
The EVOLVE System with the Tone Applicator is substantially equivalent to the following main and reference predicate devices.
| Manufacturer | Device | 510(k) No. |
|----------------------------------------|-------------|------------|
| Main Predicate | | |
| InMode System with the Tone Applicator | InMode Ltd. | K192249 |
| Reference Predicate | | |
| The EVOLVE System | InMode Ltd. | K183450 |
{4}------------------------------------------------
# Device Description:
The EVOLVE System in combination with Tone Applicator (manufactured by InMode Ltd.), is a computerized device intended to employ EMS (Electrical Muscle Stimulation) and TENS (Transcutaneous Electrical Nerve Stimulation) technologies for various medical applications.
The EVOLVE System with the Tone Applicator consists of an AC/DC power supply unit, controller and user interface including an LCD touch screen. The Tone Applicator is connected to the console via a cable, each of the subject device applicator units consist of a designated cable and of a connection port to be directly connected to one of the four connectors positioned on the rear side of the console. Up to four Tone Applicator units can be connected to the console simultaneously.
The delivery of the electrical energy is controlled by a Start/Stop button positioned on the LCD screen.
The System supports the following components:
- LCD display touch screen .
- Audio loudspeaker ●
- . 48V AC/DC power supply
- Controller
- . Fans
The System operates while connected to the Tone Applicator.
Following are The EVOLVE System with the Tone Applicator specifications:
| Main Line Frequency (nominal): | 50-60 Hz |
|--------------------------------|----------------------------------------------------------|
| Input Voltage (nominal): | 100-240 VAC |
| Input Current (rms) | 4A |
| Dimension: | |
| Console | 46cm W x 46cm D x 100cm H [18.2" W x<br>18.2" D x 44" H] |
| [W x H x D] | |
| Applicator | Tone Applicator |
| [L x D] | 12cm L x 10cm D [4.7'' L x 4'' D] |
| Weight Console: | 33 Kg (73 lbs.) |
{5}------------------------------------------------
| Tone Applicator Weight: | 0.22 Kg [0.5 lbs.] |
|----------------------------------------|---------------------------------------------------|
| Platform modules<br>AC/DC power supply | Converts AC input voltage (100-240Vac) to<br>6Vdc |
| Waveform | Symmetrical Biphasic |
| Shape | Rectangular |
| Intensity (output Voltage) | Up to 50 intensity level (=54 Vpeak) |
| Pulse Width<br>Tone Applicator | 20 to 400 $ μ$ S |
| Frequency<br>Tone Applicator | 3 to 200 Hz |
# Intended Use/Indication for Use:
The EVOLVE System with the Tone Applicator is an electro-muscle and transcutaneous nerve stimulation device for the treatment of different body areas.
The EVOLVE System with Tone Applicator is designed to operate in two modes – EMS and TENS.
In EMS mode it is used for:
- . Relaxation of muscle spasms
- . Prevention or retardation of disuse atrophy
- . Increasing local blood circulation
- . Muscle re-education
- . Maintaining or increasing range of motion
- Immediate postsurgical stimulation of calf muscles to prevent venous . thrombosis
And in TENS mode is intended for:
- Symptomatic relief and management of chronic, intractable pain
- Post-surgical acute pain
- Post-trauma acute pain
{6}------------------------------------------------
### Performance Standards:
The EVOLVE System with the Tone Applicator has been tested and complies with the following FDA recognized consensus standards:
[Rec. Number 19-4] ANSI AAMI ES 60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
[Rec. Number 19-8] IEC 60601-1-2 Edition 4.0 2014-02 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances - Requirements and tests
[Rec. Number 17-16] IEC 60601-2-10 Edition 2.1 2016-04 Medical Electrical Equipment - Part 2-10: Particular Requirements for The Basic Safety and Essential Performance of Nerve and Muscle Stimulators
[Rec. Number 5-89] IEC 60601-1-6 Edition 3.1 2013-10 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance -Collateral standard: Usability
### Sterilization/Disinfection/Cleaning:
The cleaning and reprocessing instructions set forth in the device user manual for all of the reprocessed device Applicator components. The handle should be thoroughly cleaned by the user by using 70% alcohol absorbed pad for at least 30 sec. The user should carefully examine the applicator components prior to their assembly and usage for complete drying and for any visible damage.
The device user manual and device labels, provide users with the proper symbolling and instructions/warnings for safe and effective device use and for a safe discard of the single use components upon their usage.
{7}------------------------------------------------
All device materials in contact with the patient are biocompatible.
# Non-Clinical (Bench) Performance Data:
Bench testing was conducted to demonstrate that the EVOLVE System with the Tone Applicator performs as expected under anticipated conditions of use and to verify that the device performance meets the device design requirements. The device was tested for validation of output waveform, basic unit characteristics, and output specifications.
The bench testing results demonstrated that the device performs as expected under anticipated conditions of use.
## Pre-Clinical (Animal) Performance Data:
Non-Applicable.
# Clinical Performance Data:
Non-Applicable.
{8}------------------------------------------------
# Substantial Equivalence:
The below table summarizes the main comparison aspects between the EVOLVE System with the Tone Applicator and the proposed predicate device.
| Characteristic | Subject Device | Predicate |
|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) file No. | K201285 | K192249 |
| Device Name | EVOLVE System with<br>the Tone Applicator | InMode System with the<br>Tone Applicator |
| Manufacturer | InMode MD Ltd. | InMode MD Ltd. |
| Class, Product Code. | Class II<br>IPF GZJ | Class II<br>IPF, GZJ |
| Design: | The EVOLVE System with<br>Tone Applicator consists of<br>an AC/DC power supply<br>unit, controller and user<br>interface including an LCD<br>touch screen. The delivery<br>of the electrical energy is<br>controlled by a Start/Stop<br>button positioned on the<br>front panel.<br><br>The System supports the<br>following components:<br>● LCD display touch<br>screen<br>● Audio loudspeaker<br>● 48V AC/DC power<br>supply<br>● Controller<br>● The System operates<br>while connected to the<br>Tone Applicator. | The InMode System with<br>Tone Applicator consists of<br>an AC/DC power supply<br>unit, controller and user<br>interface including an LCD<br>touch screen. The delivery of<br>the electrical energy is<br>controlled by a Start/Stop<br>button positioned on the front<br>panel.<br><br>The System supports the<br>following components:<br>● LCD display touch<br>screen<br>● Audio loudspeaker<br>● 48V AC/DC power<br>supply<br>● Controller<br>● The System operates while<br>connected to the Tone<br>Applicator. |
| Mechanism of Action | Muscle contraction by<br>electrical pulsing. | Muscle contraction by<br>electrical pulsing. |
| Components<br>Console | The EVOLVE System<br>consists of the following<br>components:<br>● Console, including a<br>power supply unit,<br>controller and user<br>interface including an<br>LCD touch screen. | The InMode System consists<br>of the following components:<br>● Console, including a<br>power supply unit,<br>controller and user<br>interface including an<br>LCD touch screen. |
| Characteristic | Subject Device | Predicate |
| 510(k) file No. | K201285 | K192249 |
| Device Name | EVOLVE System with<br>the Tone Applicator | InMode System with the<br>Tone Applicator |
| Manufacturer | InMode MD Ltd. | InMode MD Ltd. |
| | • Tone applicator with up<br>to 4 units connected to<br>the console via 4<br>designated cables and 4<br>designated connection<br>ports. | • Tone applicator with up to<br>2 units connected to the<br>console via a cable with<br>splitter and a single<br>connection port. |
| Dimension<br>Console<br>[W x H x D] | 46cm W x 46cm D x<br>100cm H [18.2" W x<br>18.2" D x 44" H] | 35cm W x 35cm D x 100cm<br>H [18.2" W x 18.2" D x<br>40" H] |
| Applicator<br>[L x D] | Tone Applicator<br>12cm L x 10cm D<br>[4.7" L x 4" D] | Tone Applicator<br>12cm L x 10cm D<br>[4.7" L x 4" D] |
| Weight Console | 33.0 Kg [73 lbs.] | 20.0 Kg [44 lbs.] |
| Weight applicator | Tone: 0.22 Kg [0.5 lbs.] | Tone: 0.22 Kg [0.5 lbs.] |
| Performance<br>Specifications:<br>Components<br>Console | Main Line Frequency<br>(nominal)<br>50-60Hz<br>Input Voltage (nominal)<br>100-240VAC<br>Input Current (rms)<br>4A | Main Line Frequency<br>(nominal)<br>50-60Hz<br>Input Voltage (nominal)<br>100-240VAC<br>Input Current (rms)<br>2A |
| Method of Line Current<br>Isolation | Independent transformer<br>isolated | Independent transformer<br>isolated |
| Electrical Type | Type BF | Type BF |
| Patient Leakage Current<br>- Normal Condition (μΑ) | <100uA patient leakage | <100uA patient leakage |
| Patient Leakage Current<br>- Single Fault Condition<br>(μΑ) | <300uA line leakage | <300uA line leakage |
| Number of Output Modes | 2 | 2 |
| Number of Output<br>Channels | 2 | 2 |
| Synchronous or<br>Alternating | See Output Specifications<br>Below | See Output Specifications<br>Below |
| Method of Channel<br>Isolation | Through transformers and<br>isolators | Through transformers and<br>isolators |
| Characteristic | Subject Device | Predicate |
| 510(k) file No. | K201285 | K192249 |
| Device Name | EVOLVE System with<br>the Tone Applicator | InMode System with the<br>Tone Applicator |
| Manufacturer | InMode MD Ltd. | InMode MD Ltd. |
| Regulated Current or<br>Regulated Voltage<br>(output signals only) | Regulated voltage on all<br>channels<br>With current limit | Regulated voltage on all<br>channels<br>With current limit |
| Software/Firmware/Micr<br>oprocessor Control | Yes | Yes |
| Automatic Overload Trip | Yes | Yes |
| Automatic No-Load Trip | Yes | Yes |
| Automatic Shut Off | Yes, On/off switch | Yes, On/off switch |
| Patient Override Control | Yes | Yes |
| Indicator Display | Yes | Yes |
| On/Off Status | Yes | Yes |
| Battery | No battery | No battery |
| Voltage/Current level | Yes, voltage levels | Yes, voltage levels |
| Timer Range (Minutes) | 0-60 [minutes] | 0-60 [minutes] |
| Compliance with 21 CFR<br>890.5850 (IPF) | Yes | Yes |
| Compliance with 21 CFR<br>882.5890 (GZJ) | Yes | Yes |
| Applicator Name | Tone Applicator | Tone Applicator |
| EMS Output Mode | | |
| Output Specifications:<br>Waveform | Symmetrical Biphasic<br>Waveform | Symmetrical Biphasic<br>Waveform |
| Pulse Shape | Rectangular | Rectangular |
| Maximum Output Voltage<br>(± 10%) | 56V @500Ω | 56V @500Ω |
| | 56V @2 kΩ | 56V @2 kΩ |
| | 56V @10kΩ | 56V @10kΩ |
| Maximum Output Current<br>(± 10%) | 92.86 mA @ 500 Ω | 98.46 mA @ 500 Ω |
| | 26.7 mA @ 2 kΩ | 27.3 mA @ 2 kΩ |
| | 5.4 mA @ 10 kΩ | 5.4 mA @ 10 kΩ |
| Pulse Width (μsec.) - The<br>output active positive pulse<br>width | 20 to 400 [μs] | 20 to 400 [μs] |
| Frequency (Hz) | 3 to 200 [Hz] | 3 to 200 [Hz] |
| Net Charge @ 500 ohms<br>[μC/pulse] | 0 [μC] @ 500Ω | 0 [μC] @ 500Ω |
| Characteristic | Subject Device | Predicate |
| 510(k) file No. | K201285 | K192249 |
| Device Name | EVOLVE System with<br>the Tone Applicator…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.