← Product Code [IPF](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IPF) · K160299

# geko(TM) Plus R-2 Neuromuscular Stimulator (K160299)

_Firstkind Limited · IPF · Apr 8, 2016 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IPF/K160299

## Device Facts

- **Applicant:** Firstkind Limited
- **Product Code:** [IPF](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IPF.md)
- **Decision Date:** Apr 8, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 890.5850
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The geko(TM) Plus R-2 is intended for: - Increasing local blood circulation; and - Immediate post-surgical stimulation of the calf muscles to prevent venous thrombosis.

## Device Story

Disposable, battery-powered neuromuscular stimulator; integrated unit with constant current pulse generator, silver electrode, and hydrogel coating. Applied to posterior knee to stimulate common peroneal nerve; induces calf muscle contraction via motor neuron activation to increase blood flow. Operated by patient or clinician; single-use, 30-hour duration. User controls intensity via two buttons; device delivers asymmetric biphasic waveform at 1 Hz. Output mimics walking cadence/energy. Healthcare providers use device to prevent post-surgical venous thrombosis; patient benefits from improved circulation and reduced thrombosis risk.

## Clinical Evidence

Bench testing only. Verification included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and nerve/muscle stimulator safety (IEC 60601-2-10). Hardware/firmware validation confirmed output waveform characteristics via oscilloscope tracings at 500Ω, 2kΩ, and 10kΩ. Biocompatibility established per ISO 10993-1.

## Technological Characteristics

Disposable, over-molded plastic casing; CR2032 lithium coin cell. Silver electrode with hydrogel coating. Constant current, asymmetric biphasic waveform; 1 Hz frequency. 8 discrete intensity settings (pulse widths 50-560 μs). Microprocessor-controlled with automatic overload/no-load trip and shut-off. Complies with IEC 60601-1, IEC 60601-2-10, and IEC 60601-1-2.

## Regulatory Identification

A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

## Predicate Devices

- geko™ T-2 Neuromuscular Stimulator ([K152677](/device/K152677.md))

## Reference Devices

- geko™ T-1 ([K133638](/device/K133638.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 8, 2016

Firstkind Limited % Sheila Hemeon-Heyer President Hever Regulatory Solutions LLC 125 Cherry Lane Amherst, Massachusetts 01002

Re: K160299

Trade/Device Name: Geko™ Plus R-2 Neuromuscular Stimulator Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF Dated: March 10. 2016 Received: March 11, 2016

Dear Sheila Hemeon-Heyer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Carlos L. Pena -S

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K160299

Device Name

geko(TM) Plus R-2 Neuromuscular Stimulator

Indications for Use (Describe)

The geko(TM) Plus R-2 is intended for:

- Increasing local blood circulation; and
- Immediate post-surgical stimulation of the calf muscles to prevent venous thrombosis.

| Type of Use (Select one or both, as applicable)                                                               |                                                                                                              |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span>                      Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span>                      Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summarv

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

| A. Submitter:              | Firstkind Limited<br>Hawk House<br>Peregrine Business Park<br>High Wycombe, UK<br>HP13 7DL           |
|----------------------------|------------------------------------------------------------------------------------------------------|
| Contact:<br>Tel:<br>Email: | Rachel Fallon, Chief Technology Officer<br>+44 (0) 1494 572045<br>rachel.fallon@firstkindmedical.com |

- Date Prepared: B. April 8, 2016
#### C. Device Name and Classification Information:

| Trade Name:          | gekoTM Plus R-2 Neuromuscular Stimulator |
|----------------------|------------------------------------------|
| Common/usual Name:   | Powered Muscle Stimulator                |
| Classification Name: | Stimulator, Muscle, Powered              |
| Product Code, CFR:   | IPF, 21 CFR 890.5850                     |
| Panel code:          | 89                                       |
| Class:               | II                                       |

- D. Predicate Device: K152677, geko™ T-2 Neuromuscular Stimulator
#### ய் Device Description:

The geko™ Plus R-2 Neuromuscular Stimulator (geko™ Plus R-2) is a disposable, fully integrated unit composed of a constant current pulse generator with embedded software and battery enclosed in an over-molded plastic casinq, and a silver electrode with a hydrogel coating which provides a means of attachment of the device and electrical contact with the patient. Two buttons are used to control the On/Off function and increase or decrease the intensity level of the device output, which is achieved through changes in the delivered pulse width. The geko™ Plus R-2 is applied so that the electrodes lie over the common peroneal nerve behind the knee. Stimulation of the common peroneal nerve causes contraction of the calf muscles through the direct activation of the motor neurons, resulting in increased blood flow.

{4}------------------------------------------------

The stimulus intensity varies with the pulse width, which can be set to one of eight levels for the geko™ Pus R-2. The asymmetric biphasic waveform results in a net charge of zero to the patient during each pulse cycle. The pulse rate is fixed at a frequency of 1 Hz and is used to isometrically stimulate the leg and foot muscles with a cadence and energy similar to that of walking.

Electrical contact is made with the patient through a hydrogel layer applied during manufacture to the integrated electrode. The patient contacting materials have been previously tested per the requirements of ISO 10993-1 and shown to be biocompatible for prolonged (up to 30 days) contact with intact skin. There are no separate electrode leads or electrodes.

### F. Indications for Use:

The geko™ Plus R-2 is intended for:

- Increasing local blood circulation, and ●
- Immediate post-surgical stimulation of the calf muscles to prevent venous ● thrombosis.

### G. Contraindications

Powered muscle stimulators should not be used on patients with cardiac demand pacemakers.

Powered muscle stimulators should not be used on patients with recently diagnosed deep vein thrombosis.

{5}------------------------------------------------

## H. Substantial Equivalence

| Parameter                                                                                                                             | PredicateTM gekoTM T-2                                                                                                                           | ProposedTM gekoTM Plus R-2                                                                                        |
|---------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Intended Use and<br>Indications for Use                                                                                               | • Increasing local blood<br>circulation<br>• Immediate post-surgical<br>stimulation of the calf<br>muscles to prevent venous<br>thrombosis       | Same                                                                                                              |
| Clinical application                                                                                                                  | Prescription use only for use in<br>a clinical or home use setting.<br>Single patient use for up to 30<br>hours<br>Disposable (cannot be reused) | Same                                                                                                              |
| Anatomical Sites                                                                                                                      | The electrodes are applied to<br>the posterior aspect of the knee<br>only for stimulation of the<br>peroneal nerve.                              | Same                                                                                                              |
| Shelf Life                                                                                                                            | 24 months                                                                                                                                        | Same                                                                                                              |
| Power source                                                                                                                          | One CR2032 primary lithium<br>coin cell. Not replaceable by<br>user                                                                              | Same                                                                                                              |
| -Method of Line<br>Current Isolation                                                                                                  | N/A                                                                                                                                              | N/A                                                                                                               |
| -Patient Leakage<br>Current<br>-Normal Condition                                                                                      | < 20μΑ                                                                                                                                           | Same                                                                                                              |
| -Single Fault<br>Condition                                                                                                            | < 20μΑ                                                                                                                                           | Same                                                                                                              |
| # output modes                                                                                                                        | Single mode with 7 discrete<br>stimulation settings<br>corresponding to the 7 pulse<br>widths.                                                   | Single mode with 8 discrete<br>stimulation settings<br>corresponding to the 8 pulse<br>widths.                    |
| # output channels                                                                                                                     | Single channel                                                                                                                                   | Same                                                                                                              |
| -Synchronous or<br>alternating                                                                                                        | N/A (single channel)                                                                                                                             | Same                                                                                                              |
| -Method of channel<br>isolation                                                                                                       | Capacitor                                                                                                                                        | Same                                                                                                              |
| Parameter                                                                                                                             | Predicate<br>geko™ T-2                                                                                                                           | Proposed<br>geko™ Plus R-2                                                                                        |
| Regulated current or<br>regulated voltage                                                                                             | Current regulated                                                                                                                                | Same                                                                                                              |
| Microprocessor<br>controlled?                                                                                                         | Yes                                                                                                                                              | Same                                                                                                              |
| Automatic overload<br>trip                                                                                                            | Yes                                                                                                                                              | Same                                                                                                              |
| Automatic no-load<br>trip                                                                                                             | Yes                                                                                                                                              | Same                                                                                                              |
| Automatic shut-off                                                                                                                    | Yes                                                                                                                                              | Same                                                                                                              |
| Patient over-ride<br>control                                                                                                          | Yes                                                                                                                                              | Same                                                                                                              |
| Indicator display                                                                                                                     |                                                                                                                                                  |                                                                                                                   |
| - On/Off status                                                                                                                       | Yes                                                                                                                                              | Same                                                                                                              |
| - Low battery                                                                                                                         | Yes (automatic off)                                                                                                                              | Same                                                                                                              |
| -Voltage / current<br>level                                                                                                           | N/A (fixed constant current)                                                                                                                     | Same                                                                                                              |
| -Charge level (pulse<br>width)                                                                                                        | Yes, number of LED flashes indicate stimulation level                                                                                            | Same                                                                                                              |
| Timer range in<br>minutes                                                                                                             | 1800 minutes maximum<br>(device is disabled after 30<br>hours battery run time)                                                                  | Same                                                                                                              |
| Compliance with<br>voluntary standards                                                                                                | Yes<br>IEC 60601-1:2005, 3rd ed<br>IEC 60601-2-10:2012, 2nd ed<br>EN 60601-1-2:2007, 3rd ed<br>ISO 10993-1                                       | Same                                                                                                              |
| Compliance with 21<br>CFR 898                                                                                                         | N/A (electrodes are integral<br>with the device, there are no<br>separate leads)                                                                 | Same                                                                                                              |
| Weight                                                                                                                                | 10 g                                                                                                                                             | Same                                                                                                              |
| Dimensions                                                                                                                            | 7.8" x 1.2" x 0.4"                                                                                                                               | Same                                                                                                              |
| Housing material and<br>construction                                                                                                  | Plastic injection molding                                                                                                                        | Same                                                                                                              |
| Waveform                                                                                                                              | Biphasic (asymmetrical<br>biphasic with zero net DC)<br>Rectangular, with charge<br>balancing second phase                                       | Same                                                                                                              |
| Parameter                                                                                                                             | Predicate<br>geko™ T-2                                                                                                                           | Proposed<br>geko™ Plus R-2                                                                                        |
| Maximum output<br>voltage                                                                                                             | 13.5 V @ 500 Ω<br>54 V @ 2000 Ω<br>255 V @ 10,000 Ω                                                                                              | 27 V @ 500 Ω<br>108 V @ 2000 Ω<br>255 V @ 10,000 Ω                                                                |
| Maximum output<br>current                                                                                                             | 27 mA @ 500 Ω<br>27 mA @2000 Ω<br>25.5 mA @ 10,000 Ω                                                                                             | 54 mA @ 500 Ω<br>54 mA @2000 Ω<br>25.5 mA @ 10,000 Ω                                                              |
| Pulse width                                                                                                                           | 50, 70, 100, 140, 200, 280, 400 μs                                                                                                               | 50, 70, 100, 140, 200, 280,<br>400, 560 μs                                                                        |
| Frequency                                                                                                                             | 1 Hz                                                                                                                                             | Same                                                                                                              |
| For interferential<br>modes only:<br>-beat Frequency (Hz)                                                                             | N/A                                                                                                                                              | Same                                                                                                              |
| Multiphasic<br>waveforms<br>-Symmetrical phases                                                                                       | No                                                                                                                                               | Same                                                                                                              |
| -Phase duration                                                                                                                       | 50-400 µs for positive phase,<br>second (negative) phase is an<br>exponential decay with a 0.1 s<br>time constant                                | 50-560 µs for positive phase,<br>second (negative) phase is<br>an exponential decay with a<br>0.1 s time constant |
| Net charge                                                                                                                            | 0 µC at 500 Ω                                                                                                                                    | Same                                                                                                              |
| -How achieved                                                                                                                         | Capacitor coupling                                                                                                                               | Same                                                                                                              |
| Maximum phase<br>charge*                                                                                                              | 10.8 µC at 500 Ω                                                                                                                                 | 30.2 µC at 500 Ω                                                                                                  |
| Maximum current<br>density*                                                                                                           | 5.5 mA/cm²                                                                                                                                       | 11.0 mA/cm²                                                                                                       |
| Average (r.m.s.)<br>current density*                                                                                                  | 0.11 mA/cm²                                                                                                                                      | 0.26 mA/cm²                                                                                                       |
| Maximum power<br>density (averaged<br>over the duty cycle)*                                                                           | 0.000000012 W/cm²                                                                                                                                | 0.000000093 W/cm²                                                                                                 |
| Burst mode<br>a) Pulses per burst<br>b) Bursts per second<br>c) Burst duration<br>(seconds)<br>d) Duty Cycle [Line<br>(b) x Line (c)] | N/A (single pulse, no pulse<br>train or burst)                                                                                                   | Same                                                                                                              |
| Parameter                                                                                                                             | Predicate gekoTM T-2                                                                                                                             | Proposed gekoTM Plus R-2                                                                                          |
| ON Time (seconds)<br>OFF Time (seconds)                                                                                               | N/A Stimulation is delivered at<br>1 Hz, with single pulses of<br>50μs to 400μs                                                                  | N/A Stimulation is delivered<br>at 1 Hz, with single pulses of<br>50μs to 560μs                                   |
| Electrodes                                                                                                                            | Hydrogel applied to silver<br>electrode. Biocompatibility for<br>the hydrogel has been<br>established.                                           | Same                                                                                                              |
| Cables/ connectors                                                                                                                    | Integrated device: no separate<br>cables                                                                                                         | Same                                                                                                              |
| Patient-contact                                                                                                                       | Contact is made through<br>integrated self-adhesive<br>electrodes.                                                                               | Same                                                                                                              |

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

*Values are calculated using the device design specifications (i.e., nominal values).

#### Discussion of differences

The primary change from the predicate geko™ T-2 to the new geko™ Plus R-2 is that the new device model provides options for stronger stimulation to enable treatment for patients who do not achieve sufficient stimulation with the geko T-2. The stronger stimulation is achieved primarily by the higher constant current (54 mA vs 27 mA) along with a higher maximum pulse width setting (560 us vs 400 us). There are no changes to the indications for use, fundamental scientific principles, or operation of the device.

The higher current and pulse width settings do not raise new questions of safety or effectiveness. The maximum voltage is the same for both the new and predicate devices (255 V). The maximum current of the new device, although twice that of the predicate, is still low for a powered muscle stimulator, delivering an average (r.m.s) current density of 0.26 mA/cm2, well below the 2 mA/cm² maximum limit as set by IEC 60601-2-10:2012 (Medical electrical equipment: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators) and a maximum power density (averaged over the duty cycle) of 0.000000093 W/cm², well below the maximum limit of 0.25 W/cm² stated in FDA's Guidance Document for Powered Muscle Stimulator 510(k)s. The longer pulse width of 560 us for the new device is the same as the longest pulse width for the original geko™ T-1 device cleared under K133638, which served as the predicate device for the geko™ T-2.

{9}------------------------------------------------

#### -Design Validation Activities

The device changes described in this Special 510(k) were implemented under the company's design change procedures. A risk assessment of the changes resulted in the following verification and validation activities:

Electrical Safety and Electromagnetic Compatibility Testing – The geko™ Plus R-2 has been certified to comply with the applicable clauses of the following standards:

- IEC 60601-1: Medical Electrical Equipment Part 1: General Requirements . for Safety, 3rd edition, 2005
- IEC 60601-2-10: Medical electrical equipment Part 2-10: Particular . requirements for the safety of nerve and muscle stimulators, 2nd edition,
- IEC 60601-1-2: Medical electrical equipment Part 1-2: General . requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests, 30 edition, 2007.

#### Hardware/Firmware Testing

As with the predicate device, the geko™ Plus R-2 hardware and firmware were tested together in order to verify the correct functioning of the device. Testing included the following:

- . Verification of output waveform characteristics via oscilloscope output tracings at 500Ω, 2kΩ and 10kΩ
- Validation of all geko™ Plus R-2 hardware and firmware functionality ●

All test results demonstrated that the geko ™ Pus R-2 meets the predefined device technical and functional requirements specifications.

#### I. Conclusions

The information and testing presented in this 510(k) demonstrated that the geko™ Plus R-2 performs as designed and intended and is substantially equivalent to the predicate device, the geko™ T-2, for increasing local circulation and immediate post-surgical prevention of venous thrombosis.

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IPF/K160299](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IPF/K160299)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
