Browse hierarchy: [Physical Medicine (PM)](/submissions/PM) → [Subpart F — Physical Medicine Therapeutic Devices](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices) → [21 CFR 890.5290](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/890.5290) → IMJ — Diathermy, Shortwave, For Use In Applying Therapeutic Deep Heat

# IMJ · Diathermy, Shortwave, For Use In Applying Therapeutic Deep Heat

_Physical Medicine · 21 CFR 890.5290 · Class 2_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ

## Overview

- **Product Code:** IMJ
- **Device Name:** Diathermy, Shortwave, For Use In Applying Therapeutic Deep Heat
- **Regulation:** [21 CFR 890.5290](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/890.5290)
- **Device Class:** 2
- **Review Panel:** [Physical Medicine](/submissions/PM)
- **3rd-party reviewable:** yes

## Identification

A nonthermal shortwave therapy is a prescription device that applies to the body pulsed electromagnetic energy in the RF bands of 13.56 MHz or 27.12 MHz and that is intended for adjunctive use in the palliative treatment of postoperative pain and edema of soft tissue by means other than the generation of deep heat within body tissues as described in paragraph (a) of this section.

## Classification Rationale

Class II (special controls). The device is classified as class II.

## Special Controls

*Classification: Class II (special controls).* The device is classified as class II. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible.
(ii) Appropriate analysis/testing must demonstrate that the device is electrically safe and electromagnetically compatible in its intended use environment.
(iii) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Non-clinical performance testing must characterize the output waveform of the device and demonstrate that the device meets appropriate output performance specifications. The output characteristics and the methods used to determine these characteristics, including the following, must be determined:
(A) Peak output power;
(B) Pulse width;
(C) Pulse frequency;
(D) Duty cycle;
(E) Characteristics of other types of modulation that may be used;
(F) Average measured output powered into the RF antenna/applicator;
(G) Specific absorption rates in saline gel test load or other appropriate model;
(H) Characterization of the electrical and magnetic fields in saline gel test load or other appropriate model for each RF antenna and prescribed RF antenna orientation/position; and
(I) Characterization of the deposited energy density in saline gel test load or other appropriate model.
(iv) A detailed summary of the clinical testing pertinent to use of the device to demonstrate the effectiveness of the device in its intended use.
(v) Labeling must include the following:
(A) Output characteristics of the device;
(B) Recommended treatment regimes, including duration of use; and
(C) A detailed summary of the clinical testing pertinent to the use of the device and a summary of the adverse events and complications.
(vi) Nonthermal shortwave therapy devices marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with these special controls.

## Recent Cleared Devices (20 of 44)

Showing 20 most recent of 44 cleared devices.

| Record | Device Name | Applicant | Decision Date | Decision |
| --- | --- | --- | --- | --- |
| [K250376](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ/K250376.md) | ViVY | Viatherm Therapeutics, LLC | Nov 5, 2025 | SESE |
| [K243112](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ/K243112.md) | Curapuls 670 | Enraf-Nonius, B.V. | Jun 25, 2025 | SESE |
| [K223620](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ/K223620.md) | Reprieve by RegenesisTM | Regenesis Biomedical, Inc. | May 24, 2023 | SESE |
| [K182363](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ/K182363.md) | BTL-703 | BTL Industries, Inc. | Jan 24, 2019 | SESE |
| [K173300](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ/K173300.md) | ViaTherm BOOST | Viatherm Therapeutics, LLC | May 1, 2018 | SESE |
| [K162240](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ/K162240.md) | ProMedTek Model C1400 Shortwave Diathermy Device | Promedtek, Inc. | Nov 17, 2016 | SESE |
| [K161862](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ/K161862.md) | ThermoPro | Zimmer Medizinsysteme GmbH | Nov 14, 2016 | SESE |
| [K153456](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ/K153456.md) | Megapulse III Shortwave Diathermy | Accelerated Care Plus | Jun 14, 2016 | SESE |
| [K131926](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ/K131926.md) | THERMOPULSE | Ibramed Equipamentos Medicos | Jul 24, 2014 | SESE |
| [K131899](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ/K131899.md) | SW-THERM | Shenzhen Dongdixin Technology Co., Ltd. | Mar 14, 2014 | SESE |
| [K121123](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ/K121123.md) | BIOFUSIONARY BEBE | Rocky Mountain Biosystems, Inc. | Mar 10, 2014 | SESE |
| [K120093](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ/K120093.md) | BTL ELITE | BTL Industries, Inc. | Oct 5, 2012 | SESE |
| [K083433](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ/K083433.md) | INTELECT SWD 100, MODEL 1600; SENIOR SOLUTIONS SHORTWAVE DIATHERMY, MODEL 1601 | Chattanooga Group | Mar 18, 2009 | SESE |
| [K042554](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ/K042554.md) | AUTO THERM 390, MODEL ME 390 | Mettler Electronics Corp. | Oct 8, 2004 | SESE |
| [K030382](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ/K030382.md) | SELITHERM, MODEL C100 | Selicor, Inc. | Apr 1, 2003 | SESE |
| [K022846](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ/K022846.md) | SW-180 SHORTWAVE THERAPY UNIT | Ito Co., Ltd. | Jan 28, 2003 | SESE |
| [K022458](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ/K022458.md) | AUTO*THERM 395, MODEL ME 395 | Mettler Electronics Corp. | Sep 26, 2002 | SESE |
| [K020227](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ/K020227.md) | SELICOR, MODELS A100 & D100 | Selicor, Inc. | Jul 30, 2002 | SESE |
| [K973732](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ/K973732.md) | MEGAPULSE II | Pti | Jan 23, 1998 | SESE |
| [K964083](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ/K964083.md) | ULTRATHERM (908) | Sie-Med, Inc. | Apr 3, 1997 | SESE |

## Top Applicants

- Rich-Mar Corp. — 3 clearances
- BTL Industries, Inc. — 2 clearances
- Ehti Medical Corp. — 2 clearances
- Elmed, Inc. — 2 clearances
- General Theraphysical, Inc. — 2 clearances

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IMJ)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com