K040121 · Dermawave, LLC · IMG · Mar 17, 2004 · Physical Medicine
Device Facts
Record ID
K040121
Device Name
DERMAWAVE 5000 COMBI MAX SYSTEM
Applicant
Dermawave, LLC
Product Code
IMG · Physical Medicine
Decision Date
Mar 17, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5860
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
Stimulation, relaxing, and/or repeatedly contracting muscles by passing electrical currents through electrodes contacting the affected body area in the medical specialties of physical medicine, general and plastic surgery and neurology and/or for use in applying therapeutic deep heat for selected medical conditions such as relief of pain, muscle spasms and joint contractures in the medical specialty of physical medicine. The indications of the DermaWave 5000 Combi Max System also include the treatment and relaxation of muscle spasms, prevention or retardation of disuse atrophy, increasing local blood circulation, muscle re-education, maintaining or increasing range of motion, immediate postsurgical stimulation of calf muscles to prevent venous thrombosis, symptomatic relief of chronic intractable pain, management of pain associated with post-traumatic or post-operative conditions using electrical stimulation and are indicated for the application of therapeutic deep heat for the treatment of selected medical conditions such as relief of pain, muscle spasms and joint contractures, relief of pain muscle spasms and joint contractures that may be associated with adhesive capsulitis, bursitis with slight calcification, myositis, soft tissue injuries, shortened tendons due to past injuries and scar tissues, relief of chronic and sub-chronic pain and joint contractures resulting from capsular tightness and capsular scarring using ultrasound. The DermaWave 5000 Combi Max System is also indicated for use in providing topical heating for temporary increase in local blood circulation, temporary relief of minor muscle and joint aches, pains and stiffness and relaxation of muscles; for muscle spasms, minor pain and stiffness associated with arthritis.
Device Story
System console with software/electronics; control/display panel; delivery accessories (electrodes, ultrasound soundheads, infrared heating accessories). Device delivers electrical stimulation, therapeutic deep heat (ultrasound), and topical heat. Used in physical medicine, surgery, and neurology clinics by healthcare providers. Console manages treatment parameters and timing. Output affects muscle activity, blood circulation, and pain perception. Benefits include pain relief, muscle rehabilitation, and prevention of atrophy/thrombosis. Sophisticated internal controls enhance safety compared to predicate consoles.
Clinical Evidence
Bench testing only. Comparative temperature charts provided to demonstrate infrared heating performance against established clinical guidelines for tissue temperature elevation.
Technological Characteristics
System console with integrated software/control electronics; ultrasound soundheads; infrared heating accessories; cutaneous electrodes. Infrared heating designed to maintain tissue temperature between 40C and 45C. System includes sophisticated control architecture for safety management.
Indications for Use
Indicated for patients requiring muscle stimulation, relaxation, or contraction; treatment of muscle spasms; prevention of disuse atrophy; muscle re-education; range of motion maintenance; venous thrombosis prevention; and pain management (chronic, intractable, post-traumatic, post-operative). Also indicated for therapeutic deep heat and topical heating for conditions including adhesive capsulitis, bursitis, myositis, soft tissue injuries, scar tissue, and arthritis-related stiffness.
Regulatory Classification
Identification
An ultrasound and muscle stimulator for use in applying therapeutic deep heat for selected medical conditions is a device that applies to specific areas of the body ultrasonic energy at a frequency beyond 20 kilohertz and that is intended to generate deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies. The device also passes electrical currents through the body area to stimulate or relax muscles.
Predicate Devices
Chatanooga Group Inc., Vectra 2C & Vectra 4C (K982317)
Chaatanooga Group Inc. Vectra Pro 2 & Vectra Pro 4 (K982324)
Dynatronics Corporation, Dynatron 150 Plus Ultrasound (K935728)
{0}------------------------------------------------
### Attachment 14
K040121
1 OF 3
## 510(k) Summary Statement for the DermaWave 5000 Combi Max System
#### 1. General Information
| Submitter: | DermaWave LLC<br>15693 83rd Lane North<br>Loxahatchee, FL 33470 |
|------------|-----------------------------------------------------------------|
|------------|-----------------------------------------------------------------|
Contact Person:
Alan Bunting 15693 83td Lane North Loxahatchee, FL 33470
Summary Preparation Date: December 4, 2003
#### 2. Names
Device Names: DermaWave 5000 Combi Max System (K040121)
Primary Classification Names:
21 CFR 890.5860, IMG Class 2 Stimulator, Ultrasound
- 21 CFR 882.5300, IMI Class 2 Diathermy, Ultrasonic
21 CFR 882.5890 84GZJ Class 2 Transcutaneous Electrical Nerve Stimulator (TENS)
21 CFR 882.1320 GXY Class 2 Electrode, Cutaneous
21CFR 882. 5890 LIH Interferential Current Therapy
21 CFR 890.5500 ILY Class 2 Infrared Lamp
## 3. Predicate Devices
- Chatanooga Group Inc., Vectra 2C & Vectra 4C (K982317) .
- Chaatanooga Group Inc. Vectra Pro 2 & Vectra Pro 4 (K982324) .
- Dynatronics Corporation, Dynatron 150 Plus Ultrasound (K935728) .
- Excel Tech, XL Tek Ultra VM Ultrasound (K001166) .
- MedX 1000 Console (K020017) .
- MedX LCT 100 Laser Accessory (K032231) .
- MedX/Duolight 500 (K032231) .
- MedX 600 LED Accessory (K020017) .
{1}------------------------------------------------
K040121
20F3
## 4. Product Description
The DermaWave 5000 Combi Max System is comprised of the following main components:
- A system console including software and control electronics. .
- A control and display panel. .
- · Delivery device accessories including patient cables and electrodes, ultrasound cables, soundheads, and infrared heating accessories.
#### 5. Indications for Use
The DermaWave 5000 Combi Max System and the delivery accessories that are used with them are indicated for use in the following:
Stimulation, relaxing, and/or repeatedly contracting muscles by passing electrical currents through electrodes contacting the affected body area in the medical specialties of physical medicine, general and plastic surgery and neurology and/or for use in applying therapeutic deep heat for selected medical conditions such as relief of pain, muscle spasms and joint contractures in the medical specialty of physical medicine.
The indications of the DermaWave 5000 Combi Max System also includes the treatment and relaxation of muscle spasms, prevention or retardation of disuse atrophy, increasing local blood circulation, muscle re-education, maintaining or increasing range of motion, immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, symptomatic relief of chronic intractable pain, management of pain associated with posttraumatic or post-operative conditions using electrical stimulation and are indicated for the application of therapeutic deep heat for the treatment of selected medical conditions such as relief of pain, muscle spasms and joint contractures, relief of pain muscle spasms and joint contractures that may be associated with adhesive capsulitis, bursitis with slight calcification, myositis, soft tissue injuries, shortened tendons due to past injuries and scar tissues, relief of chronic and sub-chronic pain and joint contractures resulting from capsular tightness and capsular scarring using ultrasound.
The DermaWave 5000 Combi Max System is also indicated for use in providing topical heating for temporary increase in local blood circulation, temporary relief of minor muscle and joint aches, pains and stiffness and relaxation of muscles; for muscle spasms, minor pain and stiffness associated with arthritis.
#### 6. Rationale for Substantial Equivalence
The DermaWave 5000 Combi Max System shares the same general indications for use in physical medicine, general and plastic surgery and neurology applications and share the same or similar basic characteristics and features and, therefore, are substantially equivalent to the:
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows handwritten text that appears to be a combination of alphanumeric characters. The text includes 'K040121' and '30F3', with the latter positioned below and slightly to the right of the former. The handwriting is somewhat stylized, and the text is oriented diagonally in the frame.
- Chatanooga Group Inc., Vectra 2C & Vectra 4C (K982317) ♥
- Chaatanooga Group Inc. Vectra Pro 2 & Vectra Pro 4 (K982324) .
- Dynatronics Corporation, Dynatron 150 Plus Ultrasound (K935728) .
- Excel Tech, XL Tek Ultra VM Ultrasound (K001166) .
The DermaWave 5000 Combi Max System utilizes identical accessories as those produced by MedX Health and share identical indications for use in physical medicine. Further, the MedX accessories comprising the MedX LCT 100 (K032231), MedX Duolight 500 (K032231) and the MedX 600 (K020017) are already cleared. The DermaWave 100 Accessory has almost identical heating characteristics as the MedX line of infrared lamp accessories and falls within the recommended guidelines for such devices in raising the temperature of tissue to no more then 45C in continuous operation over a set time period. The parameters used to achieve the comparative data are based on widely accepted clinical data that require heat lamp devices to elevate the temperature to above 40C and below 45C for a minimum 10 minute period. Comparative temperature charts are included in Section 10 of this submission.
The MedX and Duolight cleared accessories are connected to a control console that provides treatment time information to the user. The control console supplied by MedX and named the MedX 1000/1100 Console, is limited in its control features. The DermaWave 5000 Combi Max System employs sophisticated controls and control systems within system architecture to enhance the safety of the device in applications for which it is used.
## 7. Safety and Effectiveness Information
Validation documentation and a comparison of the technical characteristics and features were provided to demonstrate that the DermaWave 5000 Combi Max System is safe and effective, when indicated in specific applications in the medical specialties of physical medicine, general and plastic surgery and neurology.
## 8. Conclusion
The DermaWave 5000 Combi Max System was found to be substantially equivalent to the predicate devices and share similar indications for use and characteristics and functional features and are substantially equivalent to the currently marketed predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus. The logo is black and white.
MAR 1 7 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Dermawave, LLC C/o Mr. Derwyn Reuber Intertek Testing Services NA, Inc. 70 Codman Hill Road Boxborough, Massachusetts (01719 Re: K040121 Trade/Device Name: DermaWave 5000 Combi Max System Regulation Numbers: 21 CFR §890.5860, 21 CFR §890.5300, 21 CFR §882.5890, 21 CFR §882.1320, 21 CFR §890.5500 Regulation Names: Ultrasound and Muscle Stimulator, Ultrasonic diathermy, Transcutaneous electrical nerve stimulator for pain relief, Cutaneous electrode, Infrared lamp Regulatory Class: II Product Codes: IMG, IMI, GZJ, GXY, LIH, ILY Dated: March 3, 2004 Received: March 5, 2004
Dear Mr. Reuber:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Dcrwyn Reuber
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark N. Millican
Celia M. Witten, PhD, MD Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K040121 Device Name: DermaWave :500 Combi Max Indications for Use:
Stimulation, relaxing, and/or repeatedly contracting muscles by passing electrical currents through electrodes contacting the affected body area in the medical specialties of physical medicine, general and plastic surgery and neurology and/or for use in applying therapeutic deep heat for selected medical conditions such as relief of pain, muscle spasms and joint contractures in the medical specialty of physical medicine.
The indications of the DermaWave 5000 Combi Max System also include the treatment and relaxation of muscle spasms, prevention or retardation of disuse atrophy, increasing local blood circulation, muscle re-education, maintaining or increasing range of motion, immediate postsurgical stimulation of calf rnuscles to prevent venous thrombosis, symptomatic relief of chronic intractable pain, management of pain associated with post-traumatic or post-operative conditions using electrical stimulation and are indicated for the application of therapeutic deep heat for the treatment of selected medical conditions such as relief of pain, muscle spasms and joint contractures, relief of pain muscle spasms and joint contractures that may be associated with adhesive capsulitis, bursitis with slight calcification, myositis, soft tissue injuries, shortened tendons due to past injuries and scar tissucs, relief of chronic and sub-chronic pain and joint contractures resulting from capsular tightness and capsular scarring using ultrasound.
The DermaWave 5000 Combi Max System is also indicated for use in providing topical heating for temporary increase in local blood circulation, temporary relief of minor muscle and joint aches, pains and stiffness and relaxation of muscles; for muscle spasms, minor pain and stiffness associated with arthritis.
Mark A Millhum
and Neurologi
**510(k) Number** K090121
Over-The-Counter Use Prescription Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.