← Product Code [ILY](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/ILY) · K982636

# PATIENT WARMER- MODEL PW810 (K982636)

_Fisher & Paykel Healthcare, Ltd. · ILY · Oct 26, 1998 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/ILY/K982636

## Device Facts

- **Applicant:** Fisher & Paykel Healthcare, Ltd.
- **Product Code:** [ILY](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/ILY.md)
- **Decision Date:** Oct 26, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5500
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The Fisher & Paykel Healthcare PW810 is an Infrared Lamp, as per 21 CFR §890.5500. It emits energy at infrared frequencies to provide topical heating, and is used to provide thermal support for patients who have sustained or are susceptible to excessive or undesirable heat loss. This may include use in pre- and post-operative environments, operating rooms, intensive care units, burns units and emergency departments. Situations requiring frequent intervention or unobstructed access to a patient, and those where injury precludes surface contact, may indicate a radiant warming source should be used.

## Device Story

PW810 Patient Warmer provides radiant heat via infrared heating element housed in parabolic reflector. Device includes controller unit, height-adjustable stainless steel pole, and five-arm base with casters. Input: skin temperature via double-thermistor sensor probe. Operation: controller automatically adjusts heater power based on metabolic/environmental conditions in Patient mode; maintains 25% power in Standby mode. Output: radiant heat; visual/audible alarms for temperature, equipment fault, power failure. Used in ORs, ICUs, burns units, emergency departments by clinicians. Benefits: maintains patient body temperature without requiring extensive surface contact; allows unobstructed access for frequent clinical intervention.

## Clinical Evidence

Clinical verification studies conducted to assess warming performance. Results demonstrated the device's ability to progressively warm patients with reduced body temperature and maintain set temperatures accurately for stable conditions.

## Technological Characteristics

Infrared lamp (21 CFR §890.5500). Components: rod infrared heating element, parabolic reflector, double-thermistor skin sensor. Construction: stainless steel tubing, five-arm base. Standards: IEC 60601-1, IEC 60601-1-2, UL 2601-1, IEC 60601-2-21 (infant radiant warmer performance requirements).

## Regulatory Identification

An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

## Predicate Devices

- Aragona Medical, Thermal Ceiling OPN XII UL ([K871179](/device/K871179.md))
- Aragona Medical, Mobile Thermal Ceiling MTC XI UL ([K871178](/device/K871178.md))
- Fisher & Paykel Healthcare, IW910 Infant Warmer ([K970432](/device/K970432.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Fisher & Paykel Healthcare. The words "Fisher & Paykel" are on the top line, with a line underneath. The word "HEALTHCARE" is on the second line in a larger font.

**OCT 26 1998**

K982636

Fisher & Paykel Electronics Limited
25 Carbine Road, Panmure, Auckland, New Zealand
P.O. Box 14-348, Panmure, Auckland, New Zealand
Tel: +64-9-574-0100 Fax: +64-9-574-0158

24 July, 1998

# 510(k) Summary of Safety and Effectiveness Information

Model No. / Name: PW810 Patient Warmer Classification Name: Infrared Lamp - 89 ILY Physical Medicine Devices, 21 CFR §890.5500 (Class II) Aragona Medical, Thermal Ceiling OPN XII UL, K871179 Predicate Devices: Aragona Medical, Mobile Thermal Ceiling MTC XI UL, K871178 Fisher & Paykel Healthcare, IW910 Infant Warmer, K970432

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR \$807.92.

## (a)(1) - (a)(3) (refer to information above and concluding this summary)

## (a)(4) Description of the Device

The PW810 Patient Warmer consists of a heater assembly and controller unit, with mounting pole and base sections.

The heater assembly includes a single rod infrared heating element housed inside a parabolic reflector. An observation lamp is mounted at the back of the heater unit. The heater can be rotated to either side of the warmer, away from the patient. A metal grill on the underside of the heater prevents contact with the element.

The controller unit supports the heater assembly, and contains the device electronics, control panel and labels. The front panel contains the control buttons, temperature displays, main power switch and temperature sensor socket. Controls include operating mode and temperature selection, and lamp operation. LED displays include indicators for operating mode, alarms, lamp, and heater power, with 3-digit temperature displays. The main label is located on the controller rear panel. and the power socket is mounted in the underside.

A single pole mounting system supports the controller unit, with two sections of stainless steel tubing providing a height adjustment option. A height indication scale is etched into the pole sections to enable optimum heater-to-patient distance to be set. The lower pole section is mounted into a stabilizer weight attached to a five-arm base unit, with casters which include foot-operated brake levers.

Accessories for the patient warmer include the skin temperature sensor probe and an accessory hook for IV bags or other small items.

{1}------------------------------------------------

## 510(k) Summary continued - Fisher & Paykel PW810 Patient Warmer

In Patient mode, the PW810 provides stable control of skin temperature by automatically adjusting the heater power to compensate for varving metabolic and environmental conditions. In Standby mode, power is maintained at a constant level of 25% ready for use.

A double-thermistor sensor probe measures the patient's skin temperature, and audible and visual alarms alert the user to high or low temperature situations, equipment fault, power failure and periodic reminders to reassess the patient's clinical condition. Independent safety features are included to control maximum output and avoid thermal injury to the patient.

#### (a)(5) Statement of the Intended Use

The Fisher & Paykel Healthcare PW810 is an Infrared Lamp, as per 21 CFR §890.5500. It emits energy at infrared frequencies to provide topical heating, and is used to provide thermal support for patients who have sustained or are susceptible to excessive or undesirable heat loss.

This may include use in pre- and post-operative environments, operating rooms, intensive care units, burns units and emergency departments. Situations requiring frequent intervention or unobstructed access to a patient, and those where injury precludes surface contact, may indicate a radiant warming source should be used.

#### (a)(6) Technological Characteristics Summary

The technological characteristics of the PW810 Patient Warmer are equivalent to those of the IW910 predicate device. There are no changes to the design, materials, energy source, performance or other aspects of the device which affect substantial equivalence.

## (b)(1) Discussion of the Non-Clinical Tests

Testing for the PW810 has been carried out in the areas of mechanical, electrical and thermal safety, environmental conditions and electromagnetic compatibility. functional verification, temperature control, and irradiance distribution patterns. The device meets the requirements of the IEC 60601-1 and IEC 60601-1-2 standards, and the deviations relevant to the USA in UL 2601-1.

The device also meets applicable performance and safety requirements from the standard for Infant Radiant Warmers, IEC 60601-2-21, including temperature distribution and variance, accuracy of temperature control, and maximum irradiance levels for overall IR and near IR spectrum regions.

## (b)(2) Discussion of the Clinical Tests

Clinical verification studies demonstrated the ability of the warmers to progressively warm up patients with reduced body temperature, and the ability to control the set temperature accurately for a stable situation.

{2}------------------------------------------------

510(k) Summary continued - Fisher & Paykel PW810 Patient Warmer

#### (b)(3) Conclusions Demonstrating Safety, Effectiveness and Performance

The testing carried out for the PW810 Patient Warmer indicates that it meets design and performance functional requirements. Clinical verification studies demonstrate the successful use of the warmer and its ability to provide effective patient warming. The proposed device meets the requirements of international and US medical electrical equipment standards for safety, and key performance and safety requirements from the particular standard for infant radiant warmers which are applicable to general patient warming.

signed:

te: 24 July 1998

Chris Mander Fisher & Paykel Healthcare Ltd

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the edge. In the center of the seal is an abstract image of an eagle.

OCT 26 1998

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Chris Mander Requlatory Affairs Engineer Fisher & Paykel Healthcare Limited 25 Carbine Road, Panmure P.O. Box 14-348, Panmure Auckland, New Zealand

K982636 Re: Patient Warmer - Model PW810 Requlatory Class: II Product Code: ILY Dated: July 24, 1998 Received: July 29, 1998

Dear Mr. Mander:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labelinq, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ಕ್ಷ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2 - Mr. Chris Mander

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D.
Director

Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Fisher & Paykel Healthcare. The words "Fisher & Paykel" are on the top line, with a line underneath. The word "HEALTHCARE" is on the second line in all caps.

Fisher & Paykel Electronics Limited
25 Carbine Road, Panmure, Auckland, New Zealand
P.O. Box 14-348, Panmure, Auckland, New Zealand
Tel: +64-9-574-0100 Fax: +64-9-574-0158

24 July, 1998

# Fisher & Paykel PW810 Patient Warmer

[510(k)] Number:

## PREMARKET NOTIFICATION 510(k) INDICATIONS FOR USE STATEMENT

The Fisher & Paykel Healthcare PW810 Patient Warmer is an Infrared Lamp (as per 89 ILY, 21 CFR §890.5500) containing an infrared heating element to be positioned above an operating table or similar patient bed, in order to maintain a patient's body temperature by means of controlled radiant heat.

The PW810 is intended to provide topical heating to patients in clinical situations where external thermal support is required or desirable. This may include during or following surgical procedures, hypothermia recovery, and severe burns patients.

Situations which necessitate unobstructed access to a patient may indicate the need for a radiant heat source, in preference to thermal regulation systems which rely on extensive surface contact with a patient. Minimal direct contact by a warming system may also be required for particular injuries to patients.

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General Restorativ
510(k) Number

**Prescription Use**
**(Per 21 CFR 801.109)**

Prescription Use (Per 21 CFR §801.109)

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/ILY/K982636](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/ILY/K982636)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
