← Product Code [ILY](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/ILY) · K142256

# Applied BioPhotonics Phototherapy System ABPT1003 (K142256)

_Applied Biophotonics , Ltd. · ILY · Jan 7, 2015 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/ILY/K142256

## Device Facts

- **Applicant:** Applied Biophotonics , Ltd.
- **Product Code:** [ILY](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/ILY.md)
- **Decision Date:** Jan 7, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5500
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

Applied BioPhotonics® Phototherapy System ABPT1003 is intended to provide phototherapeutic light to the body. The device emits: Red and infrared light to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis, sprains, strains, and muscle spasm, as well as back, neck and shoulder pain; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.

## Device Story

System uses polychromatic energy from super-luminous LEDs (630nm red, 850nm NIR) to provide photobiostimulation/topical heating; non-invasive application to skin surface. Composed of ABLM1002 LightMachine control unit, LightOS software, and flexible polymeric ABLP103 LightPads. Operated by healthcare professionals in clinical settings. Operator selects treatment protocols via control panel—either manual (wavelength, duration, pulse frequency) or preset. Device provides independent control/monitoring for two LightPad sets. Output is thermal energy elevating tissue temperature; intended to relieve pain/stiffness and increase local blood circulation. Benefits include non-invasive pain management and muscle relaxation.

## Clinical Evidence

No clinical data. Safety and effectiveness supported by bench testing, including electromagnetic compatibility, electrical safety, and software verification/validation per IEC 60601-1, IEC 60601-1-2, IEC 62471, and IEC 62304.

## Technological Characteristics

System uses pulsed LEDs (630nm red, 850nm NIR). Components: ABLM1002 LightMachine, flexible polymeric ABLP103 LightPads. Power: 120-230VAC universal input. Features: integral hardware safety timer, mechanical off switch, inline fuse. Software: revision-controlled, upgradeable. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 62471 (photobiological safety), IEC 62304 (software lifecycle).

## Regulatory Identification

An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

## Predicate Devices

- IllumiMed ([K060792](/device/K060792.md))
- LIGHTWAVE Professional Deluxe ([K082586](/device/K082586.md))
- LumiWave 1X4 Infrared Therapy Device ([K051816](/device/K051816.md))
- HealthLight MicroController, MiniPro, ProNeuroLight and Pro Unit ([K101894](/device/K101894.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, with three faces overlapping to represent the department's focus on people and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 7, 2015

Applied BioPhotonics, Ltd. % Ms. Roshana Ahmed OptumInsight (Canada) Incorporated 4 Innovation Drive Dundas, Ontario L9H 7P3 CANADA

Re: K142256

Trade/Device Name: Applied BioPhotonics® Phototherapy System ABPT1003 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp, Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: ILY Dated: December 17, 2014 Received: December 18, 2014

Dear Ms. Ahmed:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{1}------------------------------------------------

#### Page 2 - Ms. Roshana Ahmed

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# David Krause -S

for

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K142256

Device Name

Applied BioPhotonics® Phototherapy System ABPT1003

#### Indications for Use (Describe)

Applied BioPhotonics® Phototherapy System ABPT1003 is intended to provide phototherapeutic light to the body. The device emits:

- Red and infrared light to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis, sprains, and muscle spasm, as well as back, neck and shoulder pain; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| Preparation Date:             | December 16, 2014                                                                            |
|-------------------------------|----------------------------------------------------------------------------------------------|
| Submitter:                    | Applied BioPhotonics Ltd.                                                                    |
| Address:                      | 15th Floor, Far East Consortium Building 121 De<br>Voeux Road, Central, Hong Kong, China SAR |
| Phone:                        | +886-3-621-8538                                                                              |
| Fax:                          | +886-3-668-3089                                                                              |
| Contact:                      | Eric Hsieh<br>(erichsieh@appliedbiophotonics.com)                                            |
| Identification of the Device: |                                                                                              |
| Proprietary/<br>Trade name:   | Applied BioPhotonics® Phototherapy System<br>ABPT1003                                        |
| Classification Name:          | Lamp, Infrared, Therapeutic Heating<br>Powered Laser Surgical Instrument                     |
| Device Classification:        | II                                                                                           |
| Regulation Number:            | 890.5500<br>878.4810                                                                         |
| Panel:                        | Physical Medicine<br>General & Plastic Surgery                                               |
| Product Code:                 | ILY<br>GEX                                                                                   |

## Predicate Devices

Substantial equivalence is claimed to the following devices as related to intended use, design, and technological characteristics:

- IllumiMed™, PhotoActif, LLC, K060792 ●
- LIGHTWAVE Professional Deluxe, LIGHTWAVE Technologies LLC, K082586 .
- LumiWave 1X4 Infrared Therapy Device, BioCare Systems, Inc., K051816
- HealthLight™ MicroController, MiniPro, ProNeuroLight and Pro Unit, . BioRemedi Therapeutic Systems, Inc., K101894

# Description of the Device

The Applied BioPhotonics® Phototherapy System ABPT1003 is a light therapy device using specific wavelengths of polychromatic energy produced by super-luminous light emitting diodes (LEDs) to treat a variety of skin and body conditions.

{4}------------------------------------------------

Applied BioPhotonics Ltd. Traditional 510(k)

The device can be described as a Class II Low Level Light treatment process employing the application of light and heat, which penetrates the skin surface to the underlying tissues, triggering normal cellular functions. This technology is commonly referred to as photobiostimulation, light emitting diode therapy (LEDT), LLLT, or LED's. The application of LED's to tissue is non-invasive.

Using patent pending and proprietary techniques and computational processes, the Applied BioPhotonics® Phototherapy System ABPT1003 is composed of:

- Proprietary hardware (i.e. ABLM1002 LightMachine) ●
- LightOS software (running in the LightMachine)
- . LightPad sets (ABLP103) comprising flexible, aseptic, polymeric LED pads
- . Cables (for connecting LightPads and LightMachines)

The IR and red spectrum emitted from LEDs in the ABLP103 LightPad sets is intended to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis, sprains, strains, and muscle spasm, as well as back, neck and shoulder pain; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.

The LightMachine control panel displays to the operator what treatment is being performed and provides the user with a choice of driving one or two LightPad sets with any chosen treatment protocol.

The software component of the device controls the user's selection of specific treatment protocols on the LightMachine's control panel. The treatment may be set manually by entering LED-operating conditions (i.e. wavelength, duration, and pulse frequency) or by selecting preset treatment protocols. The software is revision-controlled and can be upgraded by the manufacturer.

This is a prescription use only device. The labeling, instructions and user operations are designed for health care professionals.

## Intended Use/Indications for Use

Applied BioPhotonics® Phototherapy System ABPT1003 is intended to provide phototherapeutic light to the body. The device emits:

- Red and infrared light to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis, sprains, strains, and muscle spasm, as well as back, neck and shoulder pain; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.

{5}------------------------------------------------

# Technological Characteristics

The Applied BioPhotonics® Phototherapy System ABPT1003 employs pulsed LED of the same wavelengths as the predicate devices for the same indicated uses, yet provides the user with a greater degree of independent control of each set of LightPads, as well as greater flexibility in setting, sequencing, and monitoring the LED pulse conditions (wavelength, duration, pulse frequency) before and during treatments.

Applied BioPhotonics® Phototherapy System ABPT1003 affords added safety over the predicate devices with an integral hardware safety timer, AC socket with inline fuse and mechanical off switch, universal input power supply (capable of safely operating at 120VAC or 240VAC inputs).

Based on the comparison table below, the proposed device is substantially equivalent to the predicate devices in intended use, safety and performance claims.

-------------------This space intentionally left blank--------------------------------------

{6}------------------------------------------------

| Item                     | Proposed device                                                                                                                                                                          | Predicate device                                                                                         | Predicate device                                           | Predicate device                          | Predicate device                                                                                            |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|------------------------------------------------------------|-------------------------------------------|-------------------------------------------------------------------------------------------------------------|
|                          | Applied BioPhotonics®<br>Phototherapy System<br>ABPT1003                                                                                                                                 | illumiMed™                                                                                               | LIGHTWAVE Professional<br>Deluxe                           | LumiWave 1X4 Infrared<br>Therapy Device   | HealthLight™<br>MicroController,<br>MiniPro,<br>ProNeuroLight and<br>Pro Unit                               |
| 510(k) Number            | Interchangeable Pads                                                                                                                                                                     | K060792                                                                                                  | K082586                                                    | K051816                                   | K101894                                                                                                     |
| LED                      | Interchangeable Pads                                                                                                                                                                     | Interchangeable Pads &<br>Wands                                                                          | Interchangeable Panels                                     | LED Strap (Faceplate)                     | Interchangeable<br>Panels                                                                                   |
| Power                    | 120-230VAC, switched,<br>fused                                                                                                                                                           | 120VAC, switched                                                                                         | 120VAC                                                     | 120VAC                                    | 120VAC                                                                                                      |
| Software                 | Upgradeable, Revision<br>Controlled                                                                                                                                                      | Upgradable                                                                                               | Fixed                                                      | Fixed                                     | Fixed                                                                                                       |
| User Control             | LCD Panel, Select<br>Treatment (Program),<br>Preset or Manual LED<br>Settings, Automatic LED<br>Sequencing Capable,<br>Independent Control and<br>Monitoring of Each Set<br>of LightPads | Control Panel, Select<br>Protocol (Program), Preset<br>LED Settings, Automatic<br>LED Sequencing Capable | LCD & keyboard, Select<br>Function, Preset LED<br>Settings | High / Low Select, Preset<br>LED Settings | Control Panel, Select<br>Protocol (Program),<br>Preset LED Settings,<br>Automatic LED<br>Sequencing Capable |
| LED Pulse<br>Frequencies | Multiple                                                                                                                                                                                 | Multiple                                                                                                 | Multiple                                                   | Single                                    | Multiple                                                                                                    |
| LED<br>Wavelengths       | 630nm red<br>850nm NIR                                                                                                                                                                   | 430nm blue<br>660nm red<br>940nm NIR                                                                     | N/S blue<br>630nm red<br>830nm NIR                         | 880nm NIR                                 | 430nm blue<br>630nm red<br>880nm NIR                                                                        |
| Safety Timer             | Yes                                                                                                                                                                                      | No                                                                                                       | No                                                         | No                                        | Yes                                                                                                         |
| Pad Surface              | Flexible Pads - Non-<br>Porous                                                                                                                                                           | Flexible Pads – Porous                                                                                   | Rigid Pad – Non-Porous<br>Flexible Pads - Porous           | Rigid Pad – Non-Porous                    | Flexible Pads - Very<br>Porous                                                                              |

{7}------------------------------------------------

## Non-Clinical Testing

A series of studies were performed to assess the safety and effectiveness of the Applied BioPhotonics® Phototherapy System ABPT1003 including:

- Electromagnetic compatibility tests
- Electrical safety tests ●
- . Software verification and validation

Testing was done in compliance with standards including, but not limited to, the following:

- . IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2 Medical electrical equipment -- Part 1-2: General requirements . for basic safety and essential performance - Collateral Standard: Electromagnetic compatibility - Requirements and tests
- . IEC 62471 Photobiological Safety of Lamps and Lamp Systems
- . IEC 62304 Medical device software – Software life cycle processes

All laboratory tests and measurements were performed by an independent test lab certified to make such assessments. All the test results demonstrate that Applied BioPhotonics Phototherapy System ABPT1003 meets the requirements of its predefined acceptance criteria and intended uses.

## Clinical Testing

Clinical test data is not required to support the safety and effectiveness of this device.

#### Conclusion

The Applied BioPhotonics® Phototherapy System ABPT1003 has similar intended use, similar fundamental scientific technology, and similar technological characteristics with the predicate devices. Analysis of performance data including bench and safety testing data demonstrates that the Applied BioPhotonics® Phototherapy System ABPT1003 is as safe and effective as, and is therefore, substantially equivalent to, the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/ILY/K142256](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/ILY/K142256)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
