ARTOCOOL-S WATER CIRCULATING COLD PACK

K964354 · Ormed GmbH · ILO · Aug 29, 1997 · Physical Medicine

Device Facts

Record IDK964354
Device NameARTOCOOL-S WATER CIRCULATING COLD PACK
ApplicantOrmed GmbH
Product CodeILO · Physical Medicine
Decision DateAug 29, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5720
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Localized cold therapy for post-traumatic and post-surgical medical and or surgical conditions.

Device Story

Artrocool®-S is a water-circulating cryotherapy device consisting of an insulated chest, a DC electric pump, and tubing connected to body cuffs. User freezes a water-filled cartridge, places it in the chest, and fills the reservoir with a 4:1 water-alcohol mixture. The pump circulates chilled fluid through the cuff applied to injured or post-surgical body structures. Used in clinical or home settings to provide 6-8 hours of continuous cold therapy for trauma, orthopaedic surgery, sports injuries, and rehabilitation. Healthcare providers or patients use the device to manage pain and swelling; output is the physical application of cold therapy to the patient's skin.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing and comparative performance specifications (cooling time, temperature range, electrical safety, and design features).

Technological Characteristics

System includes insulated chest, DC electric pump, and fluid cuffs. Operates on 6 VDC power (from 115/230 VAC). Recirculating fluid is 4:1 water/alcohol mixture. Cooling provided by frozen water cartridge. Electrical safety compliant with UL-544; ground leakage current <55 micro-amps. Non-sterile; surface disinfection only.

Indications for Use

Indicated for treatment of pain and swelling of acute periarticular processes, pain and swelling following mobilization of shoulder stiffness under anesthesia, pain and swelling postoperatively for bones, joints and soft tissue, and pain and swelling caused by musculoskeletal contusions and athletic injury.

Regulatory Classification

Identification

A water circulating hot or cold pack is a device intended for medical purposes that operates by pumping heated or chilled water through a plastic bag and that provides hot or cold therapy for body surfaces.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 890.9.

Predicate Devices

Submission Summary (Full Text)

{0} Aug 29 1983 K964354 ATTACHMENT III REVISED SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION 6 {1} K964354 # Section 510(k) Premarket Notification ## Summary of Safety and Effectiveness Information ### ORMED, GMBH **Regulatory Authority:** Safe Medical Devices Act of 1990, 21 CFR 807.92 1. **Device Trade Name:** Artrocool®-S Water Circulating Cold Pack 2. **Common Name:** Water Circulating Cold Pack **Classification Name:** Water Circulating Cold Pack 3. **Establishment Name &amp; Registration Number:** Merzhauser Strasse 112 D-79100 Freiberg I. B. Germany 011 49 7 61/45 84 471 (voice) 011 49 7 61/45 84 450 (fax) **Reg. Number:** 2247872 4. **Classification:** § 890.5720 Water circulating hot or cold pack. (a) Identification. A water circulating hot or cold pack is a device intended for medical purposes that operates by pumping heated or chilled water through a plastic bag and that provides hot or cold therapy for body surfaces. (b) Classification. Class II (performance standards). **Product Code:** 89ILO **Device Class:** Class II **Classification Panel:** Physical Medicine 5. **Contact Person:** Mrs. Margit Bayha ORMED, GMBH Merzhauser Strasse 112 D-79100 Freiberg I. B. Germany 011 49 7 61/45 84 471 (voice) 011 49 7 61/45 84 450 (fax) 6. **Special Controls:** Special controls have not been established for this device. 7 {2} 8 7. Device Description: The Artrocool®-S is a typical cold liquid recirculating cryotherapy device. The Artrocool®-S is an insulated ice chest-like box with a closed loop liquid recirculating pump and tubing. The tubing connects to a variety of cuffs designed to be placed on or around injured or post surgical body structures, limbs or joints. The device operates by the use of a water filled cartridge and an insulated chest. The water filled bottle is frozen solid prior to first use. When the bottle is frozen, it is placed within the insulated chamber of the Artrocool. A 4:1 mixture of water and alcohol is poured into the fluid reservoir (balancing container) until full. The pump is circulated with the connecting tubing and cuff attached. Once both are filled and all air bubbles have been purged from the system the unit is ready for service. 8. Substantially Equivalent Device(s): The Artrocool®-S is substantially equivalent to the Theracool™ Water Circulating Cold Pack, the Polar Care™ Water Circulating Cold Pack, the Dr. Kool™ Water Circulating Cold Pack and the Cryo Cuff™ Water Circulating Cold Pack. 9. Comparison to Predicate Device(s): With the exception of the Cryo Cuff™ which is a gravity flow device, the Artrocool®-S and the remaining cold therapy units utilize a small DC electric pump to circulate and recirculate a chilled fluid through an assortment of body contact pads or cuffs. All of the powered units including the Artrocool®-S are UL-544 compliant. The water circulating electric pump has a ground current leakage value of less than 55 micro-amps. None of the units depend on direct refrigeration of the circulating liquid. The Dr. Kool™ and the Artrocool®-S utilize a container of frozen water or chemical gel to cool the isolated recirculating liquid. The rest of the units circulate melt water throughout the system. The intended uses of all of the units are essentially the same and include conditions like trauma, orthopaedic surgery, sports medicine injuries and rehabilitation, rheumatology, general, plastic and reconstructive surgery, chirosurgery and neurology. All the devices are suitable for any traumatic, medical or surgical conditions benefiting from cold therapy. The duration of therapeutic cold conditions for all of the units is variable depending on issues like pad/cuff size, ambient temperature, patient temperature, etc. Time duration of useful cold therapy for the comparison units is from 1 ½ to 10 hours. The Artrocool®-S provides cooling for 6 to 8 hours of continuous use. 10. Packaging: Standard, paper fiber industry typical bulk shipper packaging is utilized. The packaging selected for use is sufficient to identify, protect and transport the device safely. 11. Sterilization/Re-sterilization: May not be sterilized or re-sterilized. Surface disinfection is possible using commercially available non-solvent based disinfectant products. Cleaning and sanitizing instructions are supplied with each Artrocool®-S unit. {3} 9 # 12. Conclusion: Based on the materials, intended uses, design, and effectiveness, the **Arrocool®-S** Water Circulating Cold Pack is equivalent to the referenced legally marketed comparison water circulating cold packs. The feature comparison chart below graphically demonstrates this equivalence. # 13. Feature Comparison Table: | FEATURE | Arrocool® | Theracool™ | Polar Care™ | Dr. Kool™ | Cryo Cuff™ | SE? | | --- | --- | --- | --- | --- | --- | --- | | Intended Use | Trauma, orthopaedics, sports medicine, rheumatology, general, plastic and reconstructive surgery, chirosurgery and neurology | Same | Same | Same | Same | Yes | | Cooling Time | Approx. 8 hours at 43°F | Up to 9½ hours | Approx. 8 hours | Up to 10 hours | Approx. 6-8 hours | Yes | | Cold Source | Water - Ice | Water - Ice | Water - Ice | Frozen Gel | Ice & Water | Yes | | Operating Temperature Range | 39°F to 59°F depending on ambient air temp | Unstated | 32°F - 70°F depending on ambient air temp | Minimum temp. 38°F | Unstated | Yes | | Power Supply | 6 VDC from 115/230 VAC 50/60 Hz | 12 VDC from 110 VAC 60Hz | 12 VDC from 110 VAC 60Hz | 12 VDC from 110 VAC 60Hz | Gravity flow only | Yes | | Ground Leakage Current | <55 micro-amps | <55 micro-amps | <55 micro-amps | <55 micro-amps | <55 micro-amps | Yes | | Recirculating Fluid | 4:1 mixture of water and alcohol - 250 ml. | Water | Water - | Water | Water | Yes | | Disposable cuff covers | Yes | Yes | Yes | Yes | Yes | Yes | | Cuff | 5 styles/sizes | 1 multi-use universal | 4 styles, 4 sizes | 3 styles, 1 size ea. | 4 styles, 2 sizes ea. | Yes | | Performance Standards | UL-544 | UL-544 | UL-544 | UL-544 | Non-electric | Yes | | Weight | Approx. 12.75 lb. filled | Approx. 8 to 12 lb. filled | Approx. 24 lb. w/ cooler | Approx. 8.9 lb. filled | Approx. 10 lb. filled | Yes | {4} DEPARTMENT OF HEALTH &amp; HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. David W. Schlerf Buckman Company, Inc. Representing Ormed, GmBH 1000 Burnett Avenue, Suite 450 Concord, California 94520 Re: K964354 Artrocool®-S Water Circulating Cold Pack K964799 Artrotherm™ Cryotherapy and Thermotherapy Regulatory Class: II Product Code: ILO Dated: July 27, 1997 Received: July 29, 1997 Dear Mr. Schlerf: We have reviewed your Section 510(k) notifications of intent to market the devices referenced above and we have determined these devices are substantially equivalent (for the indications for use stated in the enclosures) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your devices are classified (see above) into either class II (Special Controls) or class III (Premarket Approval), they may be subject to such additional controls. Existing major regulations affecting your devices can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your devices in the Federal Register. Please note: this response to your premarket notification submission does {5} Page 2 - Mr. David W. Schlerf not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in a classification for your devices and thus, permits your devices to proceed to the market. If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-0.jpeg](img-0.jpeg) Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosures {6} Page 1 of 1 510(k) Number: K964354 Device Name: ARTROCOOL™ WATER CIRCULATING COLD PACK ## Intended Use: 1. Localized cold therapy for post-traumatic and post-surgical medical and or surgical conditions. ## Indications For Use: 1. Treatment of pain and swelling of acute periarticular processes. 2. Treatment of pain and swelling following mobilization of shoulder stiffness under anesthesia. 3. Treatment of pain and swelling postoperatively for bones, joints and soft tissue. 4. Treatment of pain and swelling caused by musculoskeletal contusions and athletic injury. PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY Concurrence of CDRH, Office of Device Evaluation (ODE) ![img-1.jpeg](img-1.jpeg) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use ☑ (Optional format 1-2-96) 5
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...