THERAPY CONCEPTS WATER CIRCULATING TREATMENT SYSTEM
Device Facts
| Record ID | K961621 |
|---|---|
| Device Name | THERAPY CONCEPTS WATER CIRCULATING TREATMENT SYSTEM |
| Applicant | Therapy Concepts, Inc. |
| Product Code | ILO · Physical Medicine |
| Decision Date | Oct 31, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 890.5720 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The Therapy Concepts Water Circulating Treatment System will be used to pump chilled or room temperature water through a plastic pad that is placed on a mattress underneath the user and in so doing, will provide cold temperature therapy intended to temporarily assist in the localized reduction of minor swelling and pain. Pre-chilled water will be provided by adding ice cubes and water to the reservoir. The cooling rate is adjusted by adjusting the rate of flow of the water through the plastic pad.
Device Story
Water circulating treatment system; consists of 1.5L water reservoir, low voltage DC pump, wall-mounted 7.5V DC power supply, plastic/foam pad with internal channels, and nylon carrying case. User adds ice and water to reservoir; pump circulates fluid through pad placed on mattress under user. Cooling rate controlled by adjusting water flow rate. Used for localized reduction of minor swelling and pain.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
System components: 1.5L reservoir, DC pump, 7.5V DC power supply, plastic/foam pad with internal circulation channels. Energy source: 7.5V DC. Form factor: modular system with carrying case.
Indications for Use
Indicated for patients requiring cold temperature therapy for the temporary localized reduction of minor swelling and pain.
Regulatory Classification
Identification
A water circulating hot or cold pack is a device intended for medical purposes that operates by pumping heated or chilled water through a plastic bag and that provides hot or cold therapy for body surfaces.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 890.9.
Predicate Devices
- HOT/ICE Cold Therapy Cooler System (catalog number 42000)
Related Devices
- K955057 — MODEL 1100 COLD THERAPY DEVICE · Smith and Nephew Donjoy, Inc. · Mar 8, 1996
- K970533 — MODEL 1100 COLD THERAPY DEVICE · Smith & Nephew, Inc. · Apr 15, 1997
- K962537 — INCARE HOT/ICE LITE MACHINE · Hollister, Inc. · Nov 27, 1996
- K964799 — ARTROTHERM CRYOTHERAPY AND THERNOTHERAPY · Ormed GmbH · Aug 29, 1997
- K070402 — MODIFICATION TO POLAR CARE 500 · Breg, Inc. · Mar 29, 2007