K222818 · Stryker Medical · ILK · Jan 24, 2023 · Physical Medicine
Device Facts
Record ID
K222818
Device Name
Xpedition ™ Powered Stair Chair (625700000000)
Applicant
Stryker Medical
Product Code
ILK · Physical Medicine
Decision Date
Jan 24, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5150
Device Class
Class 2
Indications for Use
Xpedition transports a patient with a mobility-limiting medical condition or injury, who is physically able to maintain a seated position while restrained, up or down a set of stairs. Xpedition is intended for use in residential and commercial environments, including pre-hospital environments, emergency, and non-emergency applications. All operators, including healthcare professionals such as emergency medical service personnel and medical first responders, must be trained by a qualified trainer before product use.
Device Story
Stryker Xpedition is a powered stair chair for patient transport; supports patients ≥50 lbs (23 kg) up to 500 lbs (227 kg). Input: operator control via buttons on head-end/top handle. Operation: rechargeable battery powers motorized belt/track system for stair traversal; manual backup available. Output: powered ascent/descent of stairs; LED visual feedback for battery/drive status. Used in residential, commercial, and pre-hospital settings by trained EMS/first responders. Operators maneuver chair over terrain; use handles for control. Benefits: facilitates safe patient transport over stairs; reduces physical strain on operators compared to manual carry. Includes patient restraint system, fold-out footrest, and wheel locks.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including mechanical/functional safety (BS ISO 7176-28), climatic testing (ISO 7176-9), power/control systems (ISO 7176-14), static/fatigue strength (ISO 7176-8), brake effectiveness (ISO 7176-3), biocompatibility (ISO 10993 series), EMC/electrical safety (IEC 60601 series), and usability testing.
Technological Characteristics
Powered stair chair; motorized belt/track drive system. Materials: biocompatible per ISO 10993. Power: rechargeable battery. Dimensions: 38" H x 20.375" W; weight 57.4 lbs. Connectivity: LED visual feedback networks. Software: moderate level of concern. Standards: BS ISO 7176-28, ISO 7176-9, ISO 7176-14, ISO 7176-8, ISO 7176-3, ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-12, IEC 60601-1-6, IEC TR 60601-4-2, AIM 7351731.
Indications for Use
Indicated for patients with mobility-limiting medical conditions or injuries who can maintain a seated position while restrained. Intended for use in residential, commercial, and pre-hospital environments by trained operators.
Regulatory Classification
Identification
Powered patient stairway chair lifts: A powered patient stairway chair lift is a motorized lift equipped with a seat and permanently mounted in one location that is intended for use in mitigating mobility impairment caused by injury or other disease by moving a person up and down a stairway. All other powered patient transport: A powered patient transport is a motorized device intended for use in mitigating mobility impairment caused by injury or other disease by moving a person from one location or level to another, such as up and down flights of stairs (e.g., attendant-operated portable stair-climbing chairs). This generic type of device does not include motorized three-wheeled vehicles or wheelchairs.
Special Controls
(i) Appropriate analysis and nonclinical testing (such as that outlined in the currently FDA-recognized edition of American Society of Mechanical Engineers (ASME) A18.1 “Safety Standard for Platform Lifts and Stairway Chair Lifts”) must demonstrate that the safety controls are adequate to prevent a free fall of the chair in the event of a device failure;
(ii) Appropriate analysis and nonclinical testing must demonstrate the ability of the device, including armrests, to withstand the rated load with an appropriate factor of safety;
(iii) Appropriate restraints must be provided to prevent the user from falling from the device (such as that outlined in the currently FDA-recognized edition of ASME A18.1 “Safety Standard for Platform Lifts and Stairway Chair Lifts”);
(iv) Appropriate analysis and nonclinical testing (such as that outlined in the currently FDA-recognized editions of AAMI/ANSI/IEC 60601-1-2, “Medical Electrical Equipment—Part 1-2: General Requirements for Safety—Collateral Standard: Electromagnetic Compatibility—Requirements and Tests,” and ASME A18.1 “Safety Standard for Platform Lifts and Stairway Chair Lifts”) must validate electromagnetic compatibility and electrical safety; and
(v) Appropriate analysis and nonclinical testing must demonstrate the resistance of the device upholstery to ignition.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a recognizable symbol of the agency's role in regulating and overseeing food and drug products in the United States.
January 24, 2023
Stryker Medical Rachel Stallworth Staff Regulatory Affairs Specialist 3800 E. Centre Ave. Portage, Michigan 49002
Re: K222818
Trade/Device Name: Xpedition TM Powered Stair Chair (625700000000) Regulation Number: 21 CFR 890.5150 Regulation Name: Powered Patient Transport Regulatory Class: Class II Product Code: ILK Dated: December 23, 2022 Received: December 23, 2022
Dear Ms. Stallworth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Tushar Bansal -S
for Heather Dean, PhD
Acting Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K222818
Device Name Xpedition TM Powered Stair (625700000000)
Indications for Use (Describe)
Xpedition transports a patient with a mobility-limiting medical condition or injury, who is physically able to maintain a seated position while restrained, up or down a set of stairs. Xpedition is intended for use in residential and commercial environments, including pre-hospital environments, emergency, and non-emergency applications. All operators, including healthcare professionals such as emergency medical service personnel and medical first responders, must be trained by a qualified trainer before product use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### I. SUBMITTER
Stryker Medical 3800 E. Centre Ave Portage, MI 49002
Phone: (269) 213-6493 Email: rachel.stallworth(@stryker.com
Contact: Rachel Stallworth Date Prepared: December 11, 2022
#### II. DEVICE
#### K22818
Name of Device: Xpedition ™ Powered Stair (625700000000) Common or Usual Name: Powered patient transport Classification Name: Transport, Patient, Powered (21 CFR 890.5150) Regulatory Class: II Product Code: ILK
#### III. PREDICATE DEVICE
C-Max +, K130864 This predicate has not been subject to a design-related recall.
#### IV. DEVICE DESCRIPTION
The Stryker Xpedition™ Powered Stair Chair is a seated patient transport device with handles and a powered belt and track system. The chair is designed to support and transport a maximum weight of 500 1b (227kg) up and down stairs. The chair is intended for patients who weigh 50 lb (23 kg) or more and can remain seated in the chair while secured by patient restraint straps. Handles at the head and foot ends allow operators to control the chair during powered or manual ascent or descent on stairs. Operators can push and maneuver the chair over various types of terrain expected in commercial and residential environments, as well as lift patients over obstacles. The chair has a removable patient containment system with attachment points for the chest and waist to secure a patient during transport. A fold-out footrest can be deployed for secure feet placement. A removable, rechargeable battery powers electrical functions including the motorized drive system for traversing stairs,
{4}------------------------------------------------
speed selection, direction, ground lighting activation, battery capacity feedback, drive activation buttons, and LED visual feedback networks. User interfaces at the back of the chair and the top handle allow for drive system control. The chair has several mechanical activations including wheel locks to prevent unintended motion on ground, a latch to fold or unfold the chair, a track deployment mechanism to deploy the stair driving track system, and top and bottom handle length adjustment activations. Options include the footrest and head end flip-up carry handles.
#### V. INDICATIONS FOR USE
Xpedition™ transports a patient with a mobility-limiting medical condition or injury, who is physically able to maintain a seated position while restrained, up or down a set of stairs. Xpedition™ is intended for use in residential and commercial environments including pre-hospital and hospital environments, emergency, and non-emergency applications. All operators, including healthcare professionals such as emergency medical service personnel and medical first responders, must be trained by a qualified trainer before product use.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Xpedition™ Powered Stair Chair and C-Max + (plus) have the same general operating principle, frame construction, and type of power source. The specific areas where the subject device and predicate device have similar or different technological characteristics are described in more detail in the sections below.
## Primary Climbing Method (Different)
The subject device can traverse stairways using two belts driven on a continuous loop. The predicate device uses an electro-mechanical piston mechanism to climb and descend stairs.
## Manual Backup Option (Different)
In the event of power loss, the subject device can be used manually. The predicate device cannot be used manually.
## Stairway Configuration (Similar)
The stairway configuration that the subject device and predicate can traverse in power mode is identical. The subject device can also be used on winding stairs if operators use the lift handles to manually carry the device on winding stairs.
## Maximum Load (Similar)
The subject device can support a maximum load of 500 pounds, compared to the predicate device maximum load of 661 pounds.
{5}------------------------------------------------
## Steps on a full charge (Different)
The subject device covers a minimum of 180 steps on a single charge, whereas the predicate device covers approximately 300 steps on a single charge.
## User Interface (Similar)
The head end operator of the subject device presses buttons at the head end of the chair to ascend and descend stairs at their chosen speed. The predicate device uses a combination of an up/down switch, on/off switch, and a speed control button to ascend and descend stairs.
#### Speed (Different)
The subject device speed ranges from 51-71 steps per minute. The predicate device speed ranges from 3-8 steps per minute.
## Height, Width, and Weight (Different)
The subject device has a height of 38 inches, width of 20.375 inches, and weight of 57.4 lbs. The predicate device has a height of 58.6-69.8 inches, width of 20.748 inches, and weight of 174.8 lbs.
#### Number of Operators Required (Different)
The subject device requires a minimum of two operators, whereas the predicate device instructs one operator and an optional second operator.
#### Storage Configuration (Similar)
The subject device is collapsible, by pulling a latch at the back of the device that folds and unfolds the chair. The predicate device can be disassembled into three parts.
## Climbing Brake (Similar)
The subject device has an electric motor brake that will stop motion on stairs when the operator releases the "Go" button(s) on the user interface. The predicate device has an automatic safety brake.
#### Wheel Lock (Different)
The subject device has a manual foot-activated wheel lock, and the predicate device does not.
## Accessories/Components/Options
The subject device has a patient containment system, head end flip-up carry handles, grooved tracks, a footrest, and a charging system. The predicate device has a head rest, seat belts, different footrests, a power charger for a vehicle, and a magnetic charging system.
{6}------------------------------------------------
#### VII. PERFORMANCE DATA
The following performance data was provided in support of the substantial equivalence determination.
Non-clinical tests were conducted to demonstrate the performance (safety and effectiveness) of the Xpedition™ Powered Stair Chair and demonstrate substantial equivalence to the predicate. The following performance testing for Xpedition™ was conducted:
- . To demonstrate the device's mechanical and functional safety, we performed testing according to BS ISO 7176-28:2012 Wheelchairs- Part 28: Requirements and test methods for stair-climbing devices.
- To demonstrate safe functionality under different climates, we performed ● tests according to ISO 7176-9:2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairs, Sections 8.2 - 8.5 and Section 8.8.
- . To demonstrate general safety of the device, we evaluated testing according to ISO 7176-14:2022 Wheelchairs – Part 14: Power and control systems for electrically powered wheelchairs and scooters - Requirements and test methods.
- . To demonstrate the device's static and fatigue strength, we evaluated testing according to ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strengths.
- . To demonstrate the device's brake fatigue strength, we performed testing according to ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes.
- Biocompatibility testing was performed to the following standards: ●
- ISO-10993-10:2010 Biological evaluation of medical devices o Part 10: Tests for irritation and skin sensitization
- ISO 10993-5 Third edition 2009-06-01 Biological evaluation of O medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO-10993-23:2021 Biological evaluation of medical devices o Part 23: Tests for irritation
- To demonstrate the device's EMC and electrical safety, we performed testing . according to the following testing methods:
- o ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R) 2012 and A2:2010/(R)2012 - Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
{7}------------------------------------------------
- IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: o General requirements for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-1-12 Edition 1.0 2014-06- Medical electrical equipment o -Part 1-12: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems intended for use in the emergency medical services environment
- IEC 60601-1-6 Edition 3.1 2013-10 Medical electrical equipment o Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- IEC TR 60601-4-2 Edition 1.0 2016-05 Medical electrical O equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
- o AIM 7351731, Rev. 2.00, 2017-02-23 Medical Electrical Equipment and System Electromagnetic Immunity Test for Exposure to Radio Frequency Identification Readers
- To ensure the device can be cleaned and disinfected, testing was performed according to the following standards:
- o AAMI TIR12:2020 and AAMI TIR30:2011
- Additional simulated use testing to demonstrate mechanical and electrical ● performance
- Life cycle testing (including powered stair climb performance) O
- Static and dynamic overload testing O
- Power-wash exposure O
- Shipping verification O
- Manual stair climb performance O
- Continuous stair climb performance O
- Software verification and validation testing were conducted, and ● documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern.
{8}------------------------------------------------
- . Usability testing was conducted to ensure the users of the device could operate the device safely and effectively.
The following additional standards were applied to the design and development process of Xpedition™:
- ISO 14971, 3rd Edition:2019 Medical devices Application of risk ● management to medical devices
- IEC 62304-1 Edition 1.1 2015-06 Consolidated Version - Medical device software - Software lifecycle processes
- . IEC 62366-1 Edition 1.0:2015 - Medical devices - Part 1: Application of usability engineering to medical devices
Clinical Studies were not required to demonstrate substantial equivalence to the predicate device.
The above software, mechanical, electrical, electromagnetic compatibility, biocompatibility, functionality, and usability testing demonstrate that the Stryker Xpedition™ Powered Stair Chair meets its functional, performance, safety and efficacy specifications and requirements and is as safe and effective as the predicate device. Based on this testing, the Stryker Xpedition™ Powered Stair Chair was demonstrated to be substantially equivalent to the predicate device
#### VIII. CONCLUSIONS
The subject device, Xpedition™ Powered Stair Chair is substantially equivalent to the legally marketed predicate device, C-Max +, as the differences in the technological characteristics do not raise new questions of safety and effectiveness. Non-clinical performance data demonstrates that the subject device is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.