HOVEROUND HOVERGLIDE MODEL HOVERGLIDE MK-2

K093946 · Hoveround Corp. · ILK · Mar 22, 2010 · Physical Medicine

Device Facts

Record IDK093946
Device NameHOVEROUND HOVERGLIDE MODEL HOVERGLIDE MK-2
ApplicantHoveround Corp.
Product CodeILK · Physical Medicine
Decision DateMar 22, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5150
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The intended use is to assist transfer of an individual up or down a straight, residential, indoor staircase while in a seated position.

Device Story

Hoveround HoverGlide Stairlift is a powered patient transport device designed for residential use. It assists individuals in navigating straight, indoor staircases while seated. The device operates via an electric motor to move a seat along a rail system installed on the staircase. It is intended for over-the-counter use by individuals with mobility limitations. The device provides a mechanical means of vertical transport, reducing the physical effort required for stair navigation and enhancing patient safety and independence within the home environment.

Clinical Evidence

Bench testing only.

Technological Characteristics

Powered patient transport device; electric motor-driven seat on a rail system; designed for straight, indoor residential staircases; Class II device; 21 CFR 890.5150.

Indications for Use

Indicated for individuals requiring assistance to transfer up or down straight, residential, indoor staircases while in a seated position.

Regulatory Classification

Identification

Powered patient stairway chair lifts: A powered patient stairway chair lift is a motorized lift equipped with a seat and permanently mounted in one location that is intended for use in mitigating mobility impairment caused by injury or other disease by moving a person up and down a stairway. All other powered patient transport: A powered patient transport is a motorized device intended for use in mitigating mobility impairment caused by injury or other disease by moving a person from one location or level to another, such as up and down flights of stairs (e.g., attendant-operated portable stair-climbing chairs). This generic type of device does not include motorized three-wheeled vehicles or wheelchairs.

Special Controls

(i) Appropriate analysis and nonclinical testing (such as that outlined in the currently FDA-recognized edition of American Society of Mechanical Engineers (ASME) A18.1 “Safety Standard for Platform Lifts and Stairway Chair Lifts”) must demonstrate that the safety controls are adequate to prevent a free fall of the chair in the event of a device failure; (ii) Appropriate analysis and nonclinical testing must demonstrate the ability of the device, including armrests, to withstand the rated load with an appropriate factor of safety; (iii) Appropriate restraints must be provided to prevent the user from falling from the device (such as that outlined in the currently FDA-recognized edition of ASME A18.1 “Safety Standard for Platform Lifts and Stairway Chair Lifts”); (iv) Appropriate analysis and nonclinical testing (such as that outlined in the currently FDA-recognized editions of AAMI/ANSI/IEC 60601-1-2, “Medical Electrical Equipment—Part 1-2: General Requirements for Safety—Collateral Standard: Electromagnetic Compatibility—Requirements and Tests,” and ASME A18.1 “Safety Standard for Platform Lifts and Stairway Chair Lifts”) must validate electromagnetic compatibility and electrical safety; and (v) Appropriate analysis and nonclinical testing must demonstrate the resistance of the device upholstery to ignition.

Submission Summary (Full Text)

{0}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and the USA. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Hoveround Corporation % Mr. Tony DiGiovanni 2151 Whitfield Industrial Parkway Sarasota, Florida 34243 Re: K093946 Trade/Device Name: Hoveround HoverGlide Stairlift Regulation Number: 21 CFR 890.5150 Regulation Name: Powered patient transport Regulatory Class: II Product Code: ILK Dated: December 21, 2009 Received: December 23, 2009 MAR 2 2 2010 Dear Mr. DiGiovanni: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Page 2 – Mr. Tony DiGiovanni If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. sincerely yours, Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: Hoveround HoverGlide stairlift Indications for Use: The intended use is to assist transfer of an individual up or down a straight, residential, indoor staircase while in a seated position. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use X (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) KATHLEEN M. QUINN FOR M. MELKERSON (Division Sign-Off) Division of Surgical. Orinopedic, and Restorative Devices 510(k) Number K093946
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...