MODEL DX MICRO COMPUTER CONTROL

K970094 · Invacare Respiratory Corp. · ITI · May 12, 1997 · Physical Medicine

Device Facts

Record IDK970094
Device NameMODEL DX MICRO COMPUTER CONTROL
ApplicantInvacare Respiratory Corp.
Product CodeITI · Physical Medicine
Decision DateMay 12, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.3860
Device ClassClass 2

Indications for Use

The intended use of the Model DX Micro-Computer Control is to activate and control power wheelchair motion.

Device Story

Electronic microcomputer-based motion control system for power wheelchairs; comprises motor controller power module and joystick interface. User inputs via joystick; microprocessor processes inputs to command wheelchair motors. Enables adjustment/programming of wheelchair performance characteristics to suit user needs. Used by wheelchair users in various environments; provides steering and motion activation. Benefits include customized mobility and control. Housed in pressure die-cast aluminum enclosure.

Clinical Evidence

Bench testing only. Device tested for compliance with ISO 7176:1993(E) 'Wheelchairs - Requirements and Test Methods for the Power and Control Systems of Electric Wheelchairs'. Results indicate compliance with standard requirements.

Technological Characteristics

Microcomputer-based electronic motor controller; pressure die-cast aluminum enclosure. Includes microprocessor-based power module, joystick interface, and programmable performance parameters. Complies with ISO 7176:1993(E).

Indications for Use

Indicated for power wheelchair users requiring electronic motion control and performance parameter adjustment.

Regulatory Classification

Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} MAY 12 1997 970094 510(k) SUMMARY # Invacare Corporation's Model Dx Micro Computer Controls for Power Wheelchairs Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared. Invacare Corporation 899 Cleveland Street Elyria, Ohio 44035 Phone: (216) 329-6000 Facsimile: (216) 366-9724 Contact Person: Edward A. Kroll Director, TQM and Regulatory Affairs Date Prepared: January 10, 1997 ## Name of Device and Name/Address of Sponsor Model Dx Micro-Computer Control for Power Wheelchairs Invacare Corporation 899 Cleveland Street Elyria, Ohio 44035 Phone: (216) 329-6000 Facsimile: (216) 366-9724 ## Common or Usual Name Power Wheelchair ## Classification Name Wheelchair, Powered - 89ITI ## Predicate Devices The Model Dx Micro-Computer Controls are substantially equivalent to Invacare Corporations' Model MCC MKIV Micro Computer Controls (K940972 June 2, 1994), and Invacare's Model INT DL40i Wheelchair control (K950274, August 7, 1995). ## Intended Use The intended use of the Model DX Micro-Computer Control is to activate and control power wheelchair motion. m\kroll510k summary.doc {1} # Technological Characteristics and Substantial Equivalence ## A. Device Description The Model Dx Controller is a series of electronic, microcomputer based, motion control devices for use with power wheelchairs. Their intended function and use is to activate and control power wheelchair motion. Additionally, they provide a method of adjusting, selecting and programming the type of power wheelchair performance characteristics which best suit the specific control needs of the wheelchair user. The Dx product line includes a motor controller power module with microprocessor, and a variety of joystick options and accessories which are used to perform various functions. The power module, which contains the system microprocessor, is housed in a pressure die cast aluminum enclosure. The power module stores preprogrammed operating functions, processes user inputs and issues commands to the wheelchair motors. The joystick options are used to engage wheelchair motion and to steer the chair. The options and accessories are used to program the wheelchair performance parameters and to perform additional functions as needed by the wheelchair user. ## B. Substantial Equivalence Products which are substantially equivalent to the Dx Controls are; Invacare’s Model MCC-MKIV Micro Computer Controls (K940972) and Invacare’s (Dynamic Controls) Model INT DL 40i Micro Computer Controls for Power Wheelchairs (K950724). Each of these products are electronic microcomputer based motor control devices for use with power wheelchairs with the same intended function and use which is to activate and control power wheelchair motion. Additionally, all provide the means for selecting, adjusting and programming the type of wheelchair operation parameters and performance characteristics which best suit the particular needs of the wheelchair user. Finally, all of these devices have been granted marketing clearance by FDA under prior submissions. None of the differences between the Dx Controller and its’ predicate devices alter the intended function and use of the device, nor do they raise any new questions pertaining to safety or effectiveness. ## PERFORMANCE DATA The Dx Controller has been tested for compliance with ISO 7176:1993(E) entitled ISO Standard, Wheelchairs - Requirements and Test Methods for the Power and Control Systems of Electric Wheelchairs." Results indicate that the Dx Controller meets the requirements of this standard. mikro05510k summary.doc
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...